510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

Steritec Smart-Read Biological Test Pack with Instant Readout Integrator

K111150510(k) TraditionalCleared Jul 8, 201174-day reviewClass II21 CFR 880.2800Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Steritec Products Mfg Co, Inc. · FRC Indicator, Biological Sterilization Process

Cleared on Jul 8, 2011 after a 74-day review. FDA’s file for it is a 510(k) statement, which names no predicate. It has been recalled 3 times.

Decided
Jul 8, 2011
Review
74 days
Class
Class II
Product code
FRC
Recalls
3
Predicates
0

FDA’s file for it is a 510(k) statement, which names no predicate.

FIG. 1 — Lineage of K111150Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRC device

  1. #1 K250044 Celerity 5 HP Biological Indicator (LCB052)…score 14.5 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 10, 2013Class IIZ-1827-2013Nonconforming Material/ComponentMesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest Steam and Smart-Read Biological Indicators (BI) because the readout time is not in specification with the label claims.
Jun 10, 2013Class IIZ-1826-2013Nonconforming Material/ComponentMesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest Steam and Smart-Read Biological Indicators (BI) because the readout time is not in specification with the label claims.
Jun 10, 2013Class IIZ-1825-2013Nonconforming Material/ComponentMesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest Steam and Smart-Read Biological Indicators (BI) because the readout time is not in specification with the label claims.

No ancestor or descendant has a Class I recall.

How does its category compare?

All FRC devices

FRC Indicator, Biological Sterilization Process: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · FRC

6in 2025

2007: 32008: 12009: 62010: 42011: 62012: 92013: 42014: 132015: 32016: 22017: 102018: 182019: 52020: 102021: 52022: 52023: 62024: 82025: 62026 (partial year): 1
2007high 18 (2018) · low 1 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRC

29in 2025

2007: 83 days median2009: 183 days median2010: 165 days median2011: 118 days median2012: 178 days median2013: 260 days median2014: 174 days median2015: 130 days median2017: 130 days median2018: 97 days median2019: 88 days median2020: 150 days median2021: 107 days median2022: 193 days median2023: 190 days median2024: 105 days median2025: 29 days median
2007high 260 (2013) · low 29 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRC

8so far in 2026, the most since 2007

2009: 12012: 22013: 22022: 42023: 12026 (partial year): 8
2007high 8 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRC

9in 2025

2007: 1002008: 602009: 452010: 572011: 372012: 462013: 392014: 512015: 1822016: 4822017: 1822018: 222019: 302020: 162021: 102022: 92023: 202024: 222025: 9
2007high 482 (2016) · low 9 (2022)2025

MAUDE reports naming product code FRC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20073830100
20081—060
20096183145
20104165057
20116118037
20129178246
20134260239
201413174051
201531300182
20162—0482
2017101300182
20181897022
2019588030
202010150016
20215107010
2022519349
20236190120
20248105022
202562909
20261—8—

What is it indicated for?

SteriTec Smart Read Biological Test Pack with Instant Readout Integrator is designed specifically for testing of 3 or 4 minutes at 132°C (270°F) and 3 minutes at 135°C (275°F) in pre-vacuum steam sterilizers. The test pack consists of a self-contained biological indicator containing Geobacillus stearothermophilus inside a small package of porous and nonporous materials, simulating the biological indicator 16 towel test packs defined by AAMI ST79. When used as directed, the Chemical Indicator Record Card gives visible indication that sterilizing conditions were met.

Indications for use as stated in the FDA 510(k) statement.

Cite this page

510(k) Report. “K111150: Steritec Smart-Read Biological Test Pack with Instant Readout Integrator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K111150

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026