510(k) Report

510(k) · Cleared

Cleared

Medisign Blood Glucose Monitoring System

K111456510(k) TraditionalCleared Feb 28, 2012279-day reviewClass II21 CFR 862.1345

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Tianjin Empecs Medical Device Co., Ltd. · NBW System, Test, Blood Glucose, Over The Counter

Cleared on Feb 28, 2012 as substantially equivalent to K053529 (Lifescan, Inc., 2006) and 1 other predicate after a 279-day review. Its family tree reaches back 2 generations to 2000. 2 later devices descend from it.

Decided
Feb 28, 2012
Review
279 days
Class
Class II
Product code
NBW
Recalls
None
Predicates
2

Its longest line of descent runs back 2 generations to K000318 (2000). 3 ancestors in all. 2 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2006: 1 other ancestor2014: 1 descendant cleared2017: 1 descendant clearedK000318 · Bionostics Glucose Quality Control Solution for Use with the Fasttake Compact Blood Glucose Monitoring System · Bionostics, Inc. · 2000K022769 · Lifescan Onetouch Ultra Control Solution · Bionostics, Inc. · 2002200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 2 descendants K022769 · Lifescan Onetouch Ultra Control Solution · Bionostics, Inc. · 2002K000318 · Bionostics Glucose Quality Control Solution for Use with the Fasttake Compact Blood Glucose Monitoring System · Bionostics, Inc. · 20000
  1. Root · 2 generations backK000318 Bionostics Glucose Quality Control Solution for Use with the…Bionostics, Inc. · 2000
  2. 1 generation backK022769 Lifescan Onetouch Ultra Control SolutionBionostics, Inc. · 2002
  3. This deviceK111456 Medisign Blood Glucose Monitoring SystemTianjin Empecs Medical Device Co., Ltd. · 2012
  4. Descendant · cites it directly · 0 of its ownK152534 Medisign MM1000 BT Blood Glucose Monitoring System, Medisign MM1100…Tianjin Empecs Medical Device Co., Ltd. · 2017
  5. Descendant · cites it directly · 0 of its ownK133260 Medisign Blood Glucse Monitoring System, Medisign Multi Blood…Tianjin Empecs Medical Device Co., Ltd. · 2014

Most influential descendants

Showing 2 of 2 · See all
  • K152534Medisign MM1000 BT Blood Glucose Monitoring System, Medisign MM1100 BT Blood Glucose…Tianjin Empecs Medical Device Co., Ltd. · 2017 · cites it directly0descendants
  • K133260Medisign Blood Glucse Monitoring System, Medisign Multi Blood Glucose Monitoring SystemTianjin Empecs Medical Device Co., Ltd. · 2014 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K000318Bionostics Glucose Quality Control Solution for Use with the Fasttake Compact…Bionostics, Inc.2000—
−1K022769Lifescan Onetouch Ultra Control SolutionBionostics, Inc.2002—
0 (this device)K111456Medisign Blood Glucose Monitoring SystemTianjin Empecs Medical Device Co., Ltd.2012—
+1K152534Medisign MM1000 BT Blood Glucose Monitoring System, Medisign MM1100 BT Blood…Tianjin Empecs Medical Device Co., Ltd.2017—
+1K133260Medisign Blood Glucse Monitoring System, Medisign Multi Blood Glucose…Tianjin Empecs Medical Device Co., Ltd.2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201410
201710
FIG. 1 — Lineage of K111456, 2000–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NBW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NBW device

  1. #1 K260545 CONTOUR® PLUS BLUE Blood Glucose Monitoring…score 14.6 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K053529Onetouch Ultra 2 Blood Glucose Monitoring System
Summary page 1
…s 1,21 CFR 862.1660, Quality Control Material Produce code: NBW, CGA and lix Predicate Devices: OneTouch Ultra 2 Blood Glucose Monitoring System (K053529) OneTouch Ultra Control Solution (K022769) Device Description: Medisign MM 1000 Blood Glucose Monito…
Lifescan, Inc.2006—
*K022769Lifescan Onetouch Ultra Control Solution
Summary page 1
…NBW, CGA and lix Predicate Devices: OneTouch Ultra 2 Blood Glucose Monitoring System (K053529) OneTouch Ultra Control Solution (K022769) Device Description: Medisign MM 1000 Blood Glucose Monitoring System, Medisign MMI 100 Blood Glucose Monitoring System, and Med…
Bionostics, Inc.2002—

Also cleared against K053529: 5 other devices →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K152534Medisign MM1000 BT Blood Glucose Monitoring System, Medisign MM1100 BT Blood Glucose…Tianjin Empecs Medical Device Co., Ltd.20170
K133260Medisign Blood Glucse Monitoring System, Medisign Multi Blood Glucose Monitoring SystemTianjin Empecs Medical Device Co., Ltd.20140

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 12 years and passes through 2 companies; a new generation was cleared every 9.3 years on median.
  • None of the 3 ancestors has a recall.
  • Of its descendants, K152534 (Tianjin Empecs Medical Device Co., Ltd., 2017) has gone furthest, with 0 descendants of its own.
  • 5 other devices also cite its predicate K053529.

How does its category compare?

All NBW devices

NBW System, Test, Blood Glucose, Over The Counter: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · NBW

2in 2025

2007: 292008: 372009: 382010: 312011: 302012: 302013: 472014: 242015: 322016: 202017: 252018: 192019: 102020: 72021: 72022: 122023: 52024: 72025: 22026 (partial year): 5
2007high 47 (2013) · low 2 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NBW

243in 2024

2007: 85 days median2008: 128 days median2009: 140 days median2010: 132 days median2011: 328 days median2012: 230 days median2013: 167 days median2014: 202 days median2015: 237 days median2016: 233 days median2017: 244 days median2018: 168 days median2019: 127 days median2020: 121 days median2021: 465 days median2022: 410 days median2023: 193 days median2024: 243 days median2026 (partial year): 190 days median
2007high 465 (2021) · low 85 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NBW

1in 2025

2007: 92008: 22009: 52010: 112011: 12012: 12013: 5, 2 Class I2014: 12016: 12019: 142025: 1
2007high 14 (2019) · low 0 (2015)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · NBW

4,745in 2025

2007: 7,9912008: 7,7042009: 10,9632010: 11,6312011: 22,9092012: 23,4492013: 35,4082014: 38,5242015: 62,1112016: 76,6212017: 37,5382018: 16,9112019: 14,4722020: 10,7032021: 9,1982022: 8,1032023: 7,2562024: 6,2202025: 4,745
2007high 76,621 (2016) · low 4,745 (2025)2025

MAUDE reports naming product code NBW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007298597,991
20083712827,704
200938140510,963
2010311321111,631
201130328122,909
201230230123,449
2013471675 (2)35,408
201424202138,524
201532237062,111
201620233176,621
201725244037,538
201819168016,911
2019101271414,472
20207121010,703
2021746509,198
20221241008,103
2023519307,256
2024724306,220
20252—14,745
202651900—

What is it indicated for?

The system is intended for the quantitative measurement of the concentration of glucose in whole blood drawn from fingertip, palm, and forearm by a single patient (lay user) as an aid in the management of diabetes, is intended for self-testing by persons at home, is for single-patient use only, and should not be shared. It is intended for use the outside of body (in vitro diagnostic use) and not for diagnosis of or screening for diabetes, nor for use on…

Read the full statement (1,483 words)

The system is intended for the quantitative measurement of the concentration of glucose in whole blood drawn from fingertip, palm, and forearm by a single patient (lay user) as an aid in the management of diabetes, is intended for self-testing by persons at home, is for single-patient use only, and should not be shared. It is intended for use the outside of body (in vitro diagnostic use) and not for diagnosis of or screening for diabetes, nor for use on neonates.

The alternative site testing (palm and forearm) in this system can only be used during stead-state blood glucose conditions.

The Medisign- MMIOOO test strip is to be used with Medisign-'MMl000 Blood Glucose Meter, to monitor glucose concentration of capillary whole blood. Medisign" MM 1000 test strips and associated meter are for use in fingertip, forearm, and palm testing. The strips are intended for self-testing by persons at home, are for single-patient use only, and should not be shared. The strips are not for diagnosis of or screening for diabetes nor for neonatal use.

Medisign T m Glucose Control Solutions are for use with Medisign Tm Brand Blood Glucose Meters and MedisignIN Test Strips to check that the meter and test strips are working together properly.

Medisign' Glucose Control Solutions are intended for use by healthcare professionals and people with diabetes mellitus at home, MesisignJN Glucose Control Solutions are for in vitro diagnostic use.

The system is intended for the quantitative measurement of the concentration of glucose in whole blood drawn from fingertip, palm, and forearm by a single patient (lay user) as an aid in the management of diabetes, is intended for self-testing by persons at home, is for single-patient use only, and should not be shared. It is intended for use the outside of body (in vitro diagnostic use) and not for diagnosis of or screening for diabetes, nor for use on neonates.

The alternative site testing (palm and forearm) in this system can only be used during stead-state blood glucose conditions.

The Medisignm MMlIOO test strip is to be used with Medisign- MMlIOO Blood Glucose Meter, to monitor glucose concentration of capillary whole blood. Medisign- MMII0O test strips and associated meter are for use in fingertip, forearm, and palmn testing. The strips are intended for self-testing by persons at home, are for single-patient use only, and should not be shared. The strips are not for diagnosis of or screening for diabetes nor for neonatal use.

Medlisign m ' Glucose Control Solutions are for use with Medisig" Brand Blood Glucose Meters and Medisign Test Strips to check that the meter and test strips are working together properly.

MedisignT m Glucose Control Solutions are intended for use by healthcare professionals and people with diabetes mellitus at home, Mesisign m Glucose Control Solutions are for in vitro diagnostic use.

The system is intended for the quantitative measurement of the concentration of glucose in whole blood drawn from fingertip, palm, and forearm by a single patient (lay user) as an aid in the management of diabetes, is intended for self-testing by persons at home, is for single-patient use only, and should not be shared. It is intended for use the outside of body (in vitro diagnostic use) and not for diagnosis of or screening for diabetes, nor for use on neonates.

The alternative site testing (palm and forearm) in this system can only be used during stead-state blood glucose conditions.

The Medisign- MM 1200 test strip is to be used with Medisignm MM 1200 Blood Glucose Meter, to monitor glucose concentration of capillary whole blood. Medisigiw' MM 1200, test strips and associated meter are for use in fingertip, forearm, and palm testing. The strips are intended for self-testing by persons at home, are for single-patient use only, and should not be shared. The strips are not for diagnosis of or screening for diabetes nor for neonatal use.

Medisign m Glucose Control Solutions are for use with Medisign~v Brand Blood Glucose Meters and Medisign TM Test Strips to check that the meter and test strips are working together properly.

Medisign T m Glucose Control Solutions are intended for use by healthcare professionals and people with diabetes mellitus at home, Mesisign~N Glucose Control Solutions are for in vitro diagnostic use.

The system is intended for the quantitative measurement of the concentration of glucose in whole blood drawn from fingertip, palm, and forearm of diabetic patients by healthcare professionals as an aid in the management of diabetes and may be used for testing multiple patients in professional healthcare settings. It is intended for use the outside of body (in vitro diagnostic use) and not for diagnosis of or screening for diabetes, nor for use on neonates.

The alternative site testing (palm and forearm) in this system can only be used during stead-state blood glucose conditions. Only auto-disabling, single use lancing device should be used with this system.

The Medisign-' MMIOOO Multi Blood Glucose Test strip is to be used with Medisignm, MM1000 Multi Blood Glucose Meter, to monitor glucose concentration of capillary whole blood. Medisignm MIOOO00 Multi Test strips and associated meters are for use in fingertip, forearm, and palm testing. The system is intended for use for multiple-patient use by healthcare professionals in healthcare settings. Only auto-disabling, single use lancing devices should be used with this system to prevent transferring disease by blood. The strips are not for diagnosis of or screening for diabetes nor for neonatal use.

Medisign~m Glucose Control Solutions are for use with Medisign~m Brand Blood Glucose Meters and Medisign " Test Strips to check that the meter and test strips are working together properly.

Medisign~m Glucose Control Solutions are intended for use by healthcare professionals and people with diabetes mellitus at home. Mesisign1 N Glucose Control Solutions are for in vitro diagnostic use.

The system is intended for the quantitative measurement of the concentration of glucose in whole blood drawn from fingertip, palm, and forearm of diabetic patients by healthcare professionals as an aid in the management of diabetes and may be used for testing multiple patients in professional healthcare settings. It is intended for use the outside of body (in vitro diagnostic use) and not for diagnosis of or screening for diabetes, nor for use on neonates.

The alternative site testing (palm and forearm) in this system can only be used during stead-state blood glucose conditions. Only auto-disabling, single use lancing device should be used with this system.

The Medisign"- MM 1100 Multi Blood Glucose Test strip is to be used with Medisign-' MMI 100 Multi Blood Glucose Meter, to monitor glucose concentration of capillary whole blood. Medisign-MMII100 Multi Test strips and associated meters are for use in fingertip, forearm, and palm testing. The system is intended for use for multiple-patient use by healthcare professionals in healthcare settings. Only auto-disabling, single use lancing devices should be used with this system to prevent transferring disease by blood. The strips are not for diagnosis of or screening for diabetes nor for neonatal use.

Medisign m Glucose Control Solutions are for use with Medisign' m Brand Blood Glucose Meters and MedisignI m Test Strips to check that the meter and test strips are working together properly.

Medisign m Glucose Control Solutions are intended for use by healthcare professionals and people with diabetes mellitus at home. Mesisign'm Glucose Control Solutions are for in vitro diagnostic use.

The system is intended for the quantitative measurement of the concentration of glucose in whole blood drawn from fingertip, palm, and forearm of diabetic patients by healthcare professionals as an aid in the management of diabetes and may be used for testing multiple patients in professional healthcare settings. It is intended for use the outside of body (in vitro diagnostic use) and not for diagnosis of or screening for diabetes, nor for use on neonates.

The alternative site testing (palm and forearm) in this systemi can only be used during stead-state blood glucose conditions. Only auto-disabling, single use lancing device should be used with this system.

The MedisignM MMv1200 Multi Blood Glucose Test strip is to be used with Medisignm MM 1200 Multi Blood Glucose Meter, to monitor glucose concentration of capillary whole blood. Medisign"' MM1200 Multi Test strips and associated meters are for use in fingertip, forearm, and palm testing. The system is intended for use for multiple-patient use by healthcare professionals in healthcare settings. Only auto-disabling, single use lancing devices should be used with this system to prevent transferring disease by blood. The strips are not for diagnosis of or screening for diabetes nor for neonatal use.

Medisign T m Glucose Control Solutions are for use with Medisignm Brand Blood Glucose Meters and Medisig1Y'm Test Strips to check that the meter and test strips are working together properly.

Medisign m Glucose Control Solutions are intended for use by healthcare professionals and people with diabetes mellitus at home. Mvesisign' Glucose Control Solutions are for in vitro diagnostic use.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K111456: Medisign Blood Glucose Monitoring System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K111456

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026