510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

BD Max GBS Assay, BD Max Instrument

K111860510(k) TraditionalCleared Feb 16, 2012231-day reviewClass I21 CFR 866.3740Recalls 7

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Becton, Dickinson & CO · NJR Nucleic Acid Amplification Assay System, Group B Streptococcus, Direct Specimen Test

Cleared on Feb 16, 2012 as substantially equivalent to K090191 (Handylab, Inc., 2010) after a 231-day review. Its family tree reaches back 5 generations to 1994. 2 later devices descend from it. It has been recalled 7 times.

Decided
Feb 16, 2012
Review
231 days
Class
Class I
Product code
NJR
Recalls
7
Predicates
1

Its longest line of descent runs back 5 generations to K936112 (1994). 2 later devices descend from it, 1 citing it directly.

  1. Root · 5 generations backK936112 Strep B OiaBiostar, Inc. · 1994
  2. 4 generations backK991828 Strep B OiaBiostar, Inc. · 1999
  3. 3 generations backK022504 Idi-Strep B AssayInfectio Diagnostic, Inc. · 2002
  4. 2 generations backK062948 Smart GBS and Smartcycle DX System and Software, Version 1.7BCepheid · 2006
  5. 1 generation backK090191 Handylab GBS Assay, Handylab Jaguar Instrument, Handylab Dna…Handylab, Inc. · 2010
  6. This deviceK111860 BD Max GBS Assay, BD Max InstrumentBecton, Dickinson & CO · 2012
  7. Descendant · cites it directly · 1 of its ownK170557 GenePOC GBS LBGenepoc, Inc. · 2017
  8. Descendant · 2 generations on · 0 of its ownK182467 Simplexa GBS Direct, Simplexa GBS Positive Control PackDiasorin Molecular, LLC · 2018

Most influential descendants

Showing 2 of 2 · See all
  • K170557GenePOC GBS LBGenepoc, Inc. · 2017 · cites it directly1descendant
  • K182467Simplexa GBS Direct, Simplexa GBS Positive Control PackDiasorin Molecular, LLC · 2018 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K936112Strep B OiaBiostar, Inc.1994—
−4K991828Strep B OiaBiostar, Inc.1999—
−3K022504Idi-Strep B AssayInfectio Diagnostic, Inc.2002—
−2K062948Smart GBS and Smartcycle DX System and Software, Version 1.7BCepheid2006—
−1K090191Handylab GBS Assay, Handylab Jaguar Instrument, Handylab Dna Extraction Strip…Handylab, Inc.2010—
0 (this device)K111860BD Max GBS Assay, BD Max InstrumentBecton, Dickinson & CO2012—
+1K170557GenePOC GBS LBGenepoc, Inc.2017—
+2K182467Simplexa GBS Direct, Simplexa GBS Positive Control PackDiasorin Molecular, LLC2018—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
201810
FIG. 1 — Lineage of K111860, 1994–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NJR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NJR device

  1. #1 K222638 Xpert® Xpress GBS, GeneXpert® Dx System…score 7.0 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K090191Handylab GBS Assay, Handylab Jaguar Instrument, Handylab Dna Extraction Strip, Handylab…
Summary page 2
…Classification Panel: Microbiology, Clinical Chemistry BD MAX GBS Assay Migration 510(k) Summary Predicate Device(s): K090191 Nucleic Acid Amplification Assay System, Group BStreptococcus, Direct Specimen Test DE…
Handylab, Inc.2010—

Also cleared against K090191: 1 other device →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K170557GenePOC GBS LBGenepoc, Inc.20171

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 23, 2025Class IIZ-0238-2026Software designProduct service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of unauthorized access that may…
Dec 19, 2024Class IIZ-1004-2025Under Investigation by firmPotential for signal drift in specific lots of BD PCR Cartridges.
Oct 31, 2024Class IIZ-0590-2025Software designBD confirmed that product service credentials used by some BD technical support teams to access certain BD products were accessed by an unauthorized actor. Until these product service credentials are updated, there is a risk of…
Jul 29, 2015Class IIZ-1864-2017Software designThe BD MAX System software version 4.44A contains a software anomaly that has the potential to affect some Open System Reagent (OSR) customers utilizing User Defined Protocol (UDP) assays. Customers utilizing only IVD assays are not…
May 31, 2013Class IIIZ-1662-2013Component change controlElectrical power module installed on some in vitro diagnostic instruments does not meet industry standards as labeled.
Dec 11, 2012Class IIZ-0976-2013Process controlCartridges are not sealing properly during PCR testing and may cause false results, indeterminates and/or invalid runs.
Sep 18, 2012Class IIZ-0701-2013Device DesignClinical diagnostic instrument may exhibit minor defects, which could lead to erroneous test results.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 5 companies; a new generation was cleared every 3.5 years on median.
  • None of the 5 ancestors has a recall.
  • Of its descendants, K170557 (Genepoc, Inc., 2017) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K090191.

How does its category compare?

All NJR devices

NJR Nucleic Acid Amplification Assay System, Group B Streptococcus, Direct Specimen Test: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · NJR

1in 2023

2010: 12011: 12012: 32013: 12015: 12016: 12017: 22018: 32022: 12023: 1
2007high 3 (2012) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NJR

87in 2018

2012: 162 days median2018: 87 days median
2007high 162 (2012) · low 87 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NJR

1in 2025

2007: 12011: 12012: 22015: 12022: 12024: 22025: 1
2007high 2 (2012) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · NJR

14in 2024

2014: 12017: 12020: 12021: 22022: 32023: 82024: 14
2007high 14 (2024) · low 0 (2007)2025

MAUDE reports naming product code NJR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—10
20080—00
20090—00
20101—00
20111—10
2012316220
20131—00
20140—01
20151—10
20161—00
20172—01
201838700
20190—00
20200—01
20210—02
20221—13
20231—08
20240—214
20250—10
20260—0—

What is it indicated for?

The EID MAXTM GBS Assay as implemented on the BID MAX TM System is a qualitative in vitro diagnostic test designed to detect Group B Streptococcus (GBS) DNA in Lim Broth cultures, after incubation for greater than or equal to ( !)18 hours, obtained from vaginal-rectal swab specimens from antepartumn pregnant women. The test incorporates automated DNA extraction to isolate the target nucleic acid from the specimen and real-time polymerase chain reaction…

Read the full statement (194 words)

The EID MAXTM GBS Assay as implemented on the BID MAX TM System is a qualitative in vitro diagnostic test designed to detect Group B Streptococcus (GBS) DNA in Lim Broth cultures, after incubation for greater than or equal to ( !)18 hours, obtained from vaginal-rectal swab specimens from antepartumn pregnant women. The test incorporates automated DNA extraction to isolate the target nucleic acid from the specimen and real-time polymerase chain reaction (PCR) to detect a 124 bp region of the cfb gene sequence of the Streptococcus agalactiae chromosome. Results from the BID MAX"' GBS Assay can be used as an aid in determining colonization status inantepartumn women.

The RID MAXIM GES Assay does not provide susceptibility results. Cultured isolates are needed for performing susceptibility testing as recommended for penicillin-allergic women. Subculture to solid media for additional testing when indicated.

The 80D MAX Tm System isintended for invitro diagnostic (IVO) use inperforming FDA cleared or approved nucleic acid testing inclinical laboratories. The ED MAX System iscapable of automated extraction and purification of nucleic acids from multiple specimren types as well as the automated amplification and detection of target nucleic acid sequences by fluorescence-based PCR.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K111860: BD Max GBS Assay, BD Max Instrument.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K111860

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026