510(k) Report

510(k) · Cleared

Cleared

Fibernet Embolic Protection Device

K111987510(k) SpecialCleared Aug 17, 201135-day reviewClass II21 CFR 870.1250

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Lumen Biomedical, Inc. · NTE Temporary Carotid Catheter For Embolic Capture

Cleared on Aug 17, 2011 as substantially equivalent to K082348 (Lumen Biomedical, Inc., 2008) after a 35-day review. Its family tree reaches back 7 generations to 2001.

Decided
Aug 17, 2011
Review
35 days
Class
Class II
Product code
NTE
Recalls
None
Predicates
1

Its longest line of descent runs back 7 generations to K010579 (2001). 25 ancestors in all, the oldest K003992 (2001). No later device cites it as a predicate yet.

  1. Root · 7 generations backK010579 Stent Support Guide CatheterMedtronic Vascular · 2001 · Recalled
  2. 6 generations backK030201 Modification to Export Aspiration CatheterMedtronic Vascular · 2003
  3. 5 generations backK040869 Export CatheterMedtronic Vascular · 2004
  4. 4 generations backK050139 Modification to: Export CatheterMedtronic Vascular · 2005
  5. 3 generations backK053372 Lbi Catheter SystemLumen Biomedical, Inc. · 2006
  6. 2 generations backK071529 Xtract Catheter SystemLumen Biomedical, Inc. · 2007 · Recalled
  7. 1 generation backK082348 Fibernet Embolic Protection SystemLumen Biomedical, Inc. · 2008 · Recalled
  8. This deviceK111987 Fibernet Embolic Protection DeviceLumen Biomedical, Inc. · 2011
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K010579Stent Support Guide CatheterMedtronic Vascular200127
−6K030201Modification to Export Aspiration CatheterMedtronic Vascular2003—
−5K040869Export CatheterMedtronic Vascular2004—
−4K050139Modification to: Export CatheterMedtronic Vascular2005—
−3K053372Lbi Catheter SystemLumen Biomedical, Inc.2006—
−2K071529Xtract Catheter SystemLumen Biomedical, Inc.20071
−1K082348Fibernet Embolic Protection SystemLumen Biomedical, Inc.20082
0 (this device)K111987Fibernet Embolic Protection DeviceLumen Biomedical, Inc.2011—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200110
200230
200330
200430
200550
200620
200710
FIG. 1 — Lineage of K111987, 2001–2011Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NTE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NTE device

  1. #1 K222694 EmPro EPS (EP4514C-190…score 6.5 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K082348Fibernet Embolic Protection System
Summary page 1
…tem Common Name: Embolic Protection Device Classification Name: Catheter, Percutaneous Predicate(s): FiberNet Embolic Protection System (K082348) Intended Use: The FiberNet® Embolic Protection System is indicated for use as a guidewire…
Lumen Biomedical, Inc.20082

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 10 years and passes through 2 companies; a new generation was cleared every 1.3 years on median.
  • 6 of the 25 ancestors have a recall of any class.

How does its category compare?

All NTE devices

NTE Temporary Carotid Catheter For Embolic Capture: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · NTE

2in 2023

2007: 12008: 32009: 32010: 12011: 42012: 32014: 12015: 32016: 12018: 12019: 12020: 12021: 12022: 12023: 2
2007high 4 (2011) · low 0 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NTE

105in 2015

2008: 95 days median2009: 86 days median2011: 105 days median2012: 30 days median2015: 105 days median
2007high 105 (2011) · low 30 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NTE

3in 2023

2008: 62011: 12021: 1, 1 Class I2023: 32026 (partial year): 2, 2 Class I
2007high 6 (2008) · low 0 (2007)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · NTE

405in 2025

2007: 1192008: 3682009: 3432010: 2942011: 3042012: 2612013: 2582014: 1992015: 1482016: 1482017: 1412018: 1582019: 2332020: 2382021: 3272022: 3262023: 3542024: 3982025: 405
2007high 405 (2025) · low 119 (2007)2025

MAUDE reports naming product code NTE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—0119
20083956368
20093860343
20101—0294
201141051304
20123300261
20130—0258
20141—0199
201531050148
20161—0148
20170—0141
20181—0158
20191—0233
20201—0238
20211—1 (1)327
20221—0326
20232—3354
20240—0398
20250—0405
20260—2 (2)—

What is it indicated for?

The FiberNet® Embolic Protection System is indicated for use as a guidewire and emboli protection system to capture and remove embolic material (thrombus/debris) produced while performing percutaneous transluminal interventional procedures in carotid arteries in high surgical risk patients with reference vessel diameters of 3.5 to 7.0 mm.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K111987: Fibernet Embolic Protection Device.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K111987

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026