510(k) Report

510(k) · Cleared

Cleared

Poly-Chem 90 Ise Module

K112408510(k) TraditionalCleared Nov 4, 201174-day reviewClass II21 CFR 862.1600

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Polymedco, Inc. · CEM Electrode, Ion Specific, Potassium

Cleared on Nov 4, 2011 as substantially equivalent to K024014 (Randox Laboratories, Ltd., 2003) after a 74-day review. Its family tree reaches back 1 generation to 2003.

Decided
Nov 4, 2011
Review
74 days
Class
Class II
Product code
CEM
Recalls
None
Predicates
1

Its longest line of descent runs back 1 generation to K024014 (2003). No later device cites it as a predicate yet.

K024014 · Randox RX Daytona · Randox Laboratories, Ltd. · 2003 · recalled2000200220042006200820102012
  1. Root · 1 generation backK024014 Randox RX DaytonaRandox Laboratories, Ltd. · 2003 · Recalled
  2. This deviceK112408 Poly-Chem 90 Ise ModulePolymedco, Inc. · 2011
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K024014Randox RX DaytonaRandox Laboratories, Ltd.20031
0 (this device)K112408Poly-Chem 90 Ise ModulePolymedco, Inc.2011—
FIG. 1 — Lineage of K112408, 2003–2011Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CEM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CEM device

  1. #1 K253491 ISE indirect K for Gen.2; ISE indirect Na for…score 11.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K024014Randox RX Daytona
Summary page 1
…11 Predicate Device: Randox RX Daytona ISE Module Device common name: Ion selective electrode module; test for sodium, potassium, chloride 5 10(k) number: K024014 Product code: IGS, GEM, CGZ Regulation number: 862.1665, 862.1600, 862.1170 Classification: Class 11 Device…
Randox Laboratories, Ltd.20031

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.

How does its category compare?

All CEM devices

CEM Electrode, Ion Specific, Potassium: FDA decisions, review times, recalls and adverse event reports over 20 years. 2011, this device’s year, is highlighted.

FDA decisions per year · CEM

1in 2025

2007: 12009: 12010: 32011: 32012: 42020: 12023: 22025: 12026 (partial year): 1
2007high 4 (2012) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CEM

180in 2012

2010: 195 days median2011: 74 days median2012: 180 days median
2007high 195 (2010) · low 74 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CEM

4in 2025

2008: 12010: 122011: 62016: 12019: 12020: 32021: 12023: 22025: 4
2007high 12 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CEM

29in 2025

2007: 662008: 2452009: 792010: 1132011: 1262012: 452013: 372014: 42015: 72016: 62017: 322018: 132019: 252020: 272021: 182022: 162023: 302024: 422025: 29
2007high 245 (2008) · low 4 (2014)2025

MAUDE reports naming product code CEM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—066
20080—1245
20091—079
2010319512113
20113746126
20124180045
20130—037
20140—04
20150—07
20160—16
20170—032
20180—013
20190—125
20201—327
20210—118
20220—016
20232—230
20240—042
20251—429
20261—0—

What is it indicated for?

The Poly-Chem 90 ISE Module is for the quantitative in vitro measurement of sodium, potassium, and chloride in human serum on the Poly-Chem 90 clinical chemistry analyzer. Sodium measurements are used in the diagnosis and treatment of aldosteronism (excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension, Addison's disease (caused by…

Read the full statement (125 words)

The Poly-Chem 90 ISE Module is for the quantitative in vitro measurement of sodium,

potassium, and chloride in human serum on the Poly-Chem 90 clinical chemistry analyzer.

Sodium measurements are used in the diagnosis and treatment of aldosteronism

(excessive secretion of the hormone aldosterone), diabetes insipidus (chronic excretion

of large amounts of dilute urine, accompanied by extreme thirst), adrenal hypertension,

Addison's disease (caused by destruction of the adrenal glands), dehydration,

inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte

imbalance.

Potassium measurements are used to monitor electrolyte balance in the diagnosis and

treatment of diseases conditions characterized by low or high blood potassium levels.

Chloride measurements are used in the diagnosis and treatment of electrolyte and

metabolic disorders such as cystic fibrosis and diabetic acidosis.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K112408: Poly-Chem 90 Ise Module.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K112408

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026