510(k) Report

510(k) · Cleared

Cleared

GL Goodlife Series Blood Glucose Monitoring Sysmtem

K113307510(k) TraditionalCleared Jul 27, 2012262-day reviewClass II21 CFR 862.1345

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hmd Biomedical, Inc. · LFR Glucose Dehydrogenase, Glucose

Cleared on Jul 27, 2012 as substantially equivalent to K101509 (Taidoc Technology Corporation, 2011) and 1 other predicate after a 262-day review. Its family tree reaches back 4 generations to 2007. 1 later device descends from it.

Decided
Jul 27, 2012
Review
262 days
Class
Class II
Product code
LFR
Recalls
None
Predicates
2

Its longest line of descent runs back 4 generations to K070472 (2007). 6 ancestors in all, the oldest K023279 (2003). 1 later device descends from it, 1 citing it directly.

  1. Root · 4 generations backK070472 Clever Chek TD-4230 Blood Glucose Monitoring System, RD-4230…Taidoc Technology Corporation · 2007
  2. 3 generations backK090187 Fora G30 Blood Glucose Monitoring System, Model TD-4241Taidoc Technology Corporation · 2009
  3. 2 generations backK094005 Fora G31A/TD-4256A Blood Glucose Monitoring System and Fora…Taidoc Technology Corporation · 2010
  4. 1 generation backK101509 U-Right TD-4279 Blood Glucose Monitoring System, Fora GD40/TD-4272…Taidoc Technology Corporation · 2011
  5. This deviceK113307 GL Goodlife Series Blood Glucose Monitoring SysmtemHmd Biomedical, Inc. · 2012
  6. Descendant · cites it directly · 0 of its ownK140150 Goodlife AC-300 Blood Glucose Monitoring System, Goodlife AC-301…Hmd Biomedical, Inc. · 2015

Most influential descendants

Showing 1 of 1 · See all
  • K140150Goodlife AC-300 Blood Glucose Monitoring System, Goodlife AC-301 Blood Glucose…Hmd Biomedical, Inc. · 2015 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K070472Clever Chek TD-4230 Blood Glucose Monitoring System, RD-4230, Surechek Blood…Taidoc Technology Corporation2007—
−3K090187Fora G30 Blood Glucose Monitoring System, Model TD-4241Taidoc Technology Corporation2009—
−2K094005Fora G31A/TD-4256A Blood Glucose Monitoring System and Fora G31B/TD-4256B…Taidoc Technology Corporation2010—
−1K101509U-Right TD-4279 Blood Glucose Monitoring System, Fora GD40/TD-4272 Blood…Taidoc Technology Corporation2011—
0 (this device)K113307GL Goodlife Series Blood Glucose Monitoring SysmtemHmd Biomedical, Inc.2012—
+1K140150Goodlife AC-300 Blood Glucose Monitoring System, Goodlife AC-301 Blood Glucose…Hmd Biomedical, Inc.2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200320

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
FIG. 1 — Lineage of K113307, 2007–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LFR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K101509U-Right TD-4279 Blood Glucose Monitoring System, Fora GD40/TD-4272 Blood Glucose…
Summary page 2
…Common name: Blood Glucose Test System Submitter: TaiDoc Technology Corporation 510(k) no.: K101509 Product code: NBW, LFR Common name: Glucose Control Solution Submitte…
Taidoc Technology Corporation2011—
*K032985Glucoleader Enhance Self-Monitoring of Blood Glucose System
Summary page 2
…LFR Common name: Glucose Control Solution Submitter: HMD Biomedical, Inc. 510(k) no.: K032985 Product code: JJX 4. Device Description GoodLife AC-300-'305 Blood Glucose Monitoring Syste…
Hmd Biomedical, Inc.2003—

Also cleared against K101509: 3 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K140150Goodlife AC-300 Blood Glucose Monitoring System, Goodlife AC-301 Blood Glucose…Hmd Biomedical, Inc.20150

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 5 years and passes through 2 companies; a new generation was cleared every 1.1 years on median.
  • None of the 6 ancestors has a recall.
  • Of its descendants, K140150 (Hmd Biomedical, Inc., 2015) has gone furthest, with 0 descendants of its own.
  • 3 other devices also cite its predicate K101509.

How does its category compare?

All LFR devices

LFR Glucose Dehydrogenase, Glucose: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · LFR

1in 2015

2007: 12009: 12011: 62012: 132013: 32014: 32015: 1
2007high 13 (2012) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LFR

161in 2014

2011: 296 days median2012: 152 days median2013: 212 days median2014: 161 days median
2007high 296 (2011) · low 152 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LFR

1in 2012

2007: 32008: 12009: 42011: 12012: 1
2007high 4 (2009) · low 0 (2010)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LFR

827in 2025

2007: 23,8192008: 6,5842009: 10,8232010: 4,8652011: 3,8022012: 3,3412013: 3,8212014: 3,9602015: 3,3662016: 2,1452017: 1032018: 1082019: 1082020: 1322021: 1642022: 1012023: 1032024: 1712025: 827
2007high 23,819 (2007) · low 101 (2022)2025

MAUDE reports naming product code LFR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—323,819
20080—16,584
20091—410,823
20100—04,865
2011629613,802
20121315213,341
2013321203,821
2014316103,960
20151—03,366
20160—02,145
20170—0103
20180—0108
20190—0108
20200—0132
20210—0164
20220—0101
20230—0103
20240—0171
20250—0827
20260—0—

What is it indicated for?

The GoodLife AC-300 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-300 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients. The GoodLife AC-300 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro…

Read the full statement (2,358 words)

The GoodLife AC-300 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-300 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.

The GoodLife AC-300 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodL-ife AC-300 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.

The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-300 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.

The GoodLife Glucose Control Solutions are for use with the GoodLife AC-300 Blood Glucose Meter and GoodLife AC Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

The GoodLife AC-301 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-301 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.

The GoodLife AC-301 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-301 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.

The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-301 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.

The GoodLife Glucose Control Solutions are for use with the GoodLife AC-301 Blood Glucose Meter and GoodLife AC Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

The GoodLife AC-302 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-302 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.

The GoodLife AC-302 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-302 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.

The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-302 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.

The GoodLife Glucose Control Solutions are for use with the GoodLife AC-302 Blood Glucose Meter and GoodLife AC Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

The GoodLife AC-303 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-303 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.

The GoodLife AC-303 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-303 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.

The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-303 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.

The GoodLife Glucose Control Solutions are for use with the GoodLife AC-303 Blood Glucose Meter and GoodLife AC Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

The GoodLife AC-304 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-304 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.

The GoodLife AC-304 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-304 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.

The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-304 Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.

The GoodLife Glucose Control Solutions are for use with the Good Life AC-304 Blood Glucose Meter and GoodLife AC Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

The GoodLife AC-305 Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood sample drawn from the fingertips only. The GoodLife AC-305 Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.

The GoodLife AC-305 Blood Glucose Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-305 Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use.

The GoodLife AC Blood Glucose Test Strips are for use with the GoodLife AC-30S Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only.

The GoodLife Glucose Control Solutions are for use with the GoodLife AC-SOS Blood Glucose Meter and GoodLife AC Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

The GoodLife AC-300 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-300 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.

The GoodLife AC-300 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The Good~ife AC-300 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.

The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-300 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

The GoodLife Glucose Control Solutions are for use with the GoodL-ife AC-300 Professional Blood Glucose Meter and GoodLife AC Professional Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

The GoodLife AC-301 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-301 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.

The GoodLife AC-301 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-301 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.

The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-301 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

The GoodLife Glucose Control Solutions are for use with the GoodLife AC-301 Professional Blood Glucose Meter and GoodLife AC Professional Blood Glucose Test Strips to check that the meter and teststrips are working together properly and that the test is performing correctly.

The GoodLife AC-302 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-302 Professional Blood Glucose Monitoring System is intended to be Used for testing multiple patients by a health care professional in a clinical setting.

The GoodLife AC-302 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-302 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.

The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-302 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

The GoodLife Glucose Control Solutions are for use with the GoodLife AC-302 Professional Blood Glucose Meter and GoodLife AC Professional Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

The GoodLife AC-303 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-303 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.

The GoodLife AC-303 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-303 Professional Blood Glucose Monitoring System should hot be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.

The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-303 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

The GoodLife Glucose Control Solutions are for use with the GoodLife AC-303 Professional Blood Glucose Meter and GoodLife AC Professional Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

The GoodLife AC-304 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-304 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.

The GoodLife AC-304 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-304 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or fot neonatal use. This system should only be used with single-use, auto-disabling lancing devices.

The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-304 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

The GoodLife Glucose Control Solutions are for use with the GoodLife AC-304 Professional Blood Glucose Meter and GoodLife AC Professional Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

The GoodLife AC-3O5 Professional Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips only. The GoodLife AC-3O5 Professional Blood Glucose Monitoring System is intended to be used for testing multiple patients by a health care professional in a clinical setting.

The GoodLife AC-305 Professional Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by a health care professional in healthcare facilities as an aid to monitor the effectiveness of diabetes control. The GoodLife AC-3O5 Professional Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. This system should only be used with single-use, auto-disabling lancing devices.

The GoodLife AC Professional Blood Glucose Test Strip are for use with the GoodLife AC-305 Professional Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips.

The GoodLife Glucose Control Solutions are for use with the GoodLife AC-305 Professional Blood Glucose Meter and GoodLife AC Professional Blood Glucose Test Strips to check that the meter and test strips are working together properly and that the test is performing correctly.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K113307: GL Goodlife Series Blood Glucose Monitoring Sysmtem.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K113307

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026