510(k) Report

510(k) · Cleared

Cleared

Randox Opiates Assay

K113747510(k) TraditionalCleared Aug 24, 2012248-day reviewClass II21 CFR 862.3650

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Randox Laboratories Limited · DJG Enzyme Immunoassay, Opiates

Cleared on Aug 24, 2012 after a 248-day review. FDA’s file for it is a 510(k) statement, which names no predicate.

Decided
Aug 24, 2012
Review
248 days
Class
Class II
Product code
DJG
Recalls
None
Predicates
0

FDA’s file for it is a 510(k) statement, which names no predicate.

FIG. 1 — Lineage of K113747Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DJG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DJG device

  1. #1 K232659 Dochek® Multi-Drug Urine Test Cup Rx, Dochek®…score 17.6 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

How does its category compare?

All DJG devices

DJG Enzyme Immunoassay, Opiates: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · DJG

3in 2025

2007: 22008: 22009: 22010: 62011: 62012: 52013: 52014: 92015: 122016: 22017: 42018: 52019: 32020: 52021: 32022: 22023: 102024: 52025: 32026 (partial year): 3
2007high 12 (2015) · low 2 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DJG

31in 2025

2010: 187 days median2011: 233 days median2012: 95 days median2013: 60 days median2014: 62 days median2015: 141 days median2017: 101 days median2018: 134 days median2019: 235 days median2020: 30 days median2021: 422 days median2023: 140 days median2024: 150 days median2025: 31 days median2026 (partial year): 157 days median
2007high 422 (2021) · low 30 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DJG

2in 2025

2011: 12019: 12025: 2
2007high 2 (2025) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DJG

1in 2025

2011: 42012: 12013: 42015: 32016: 22017: 12018: 12021: 22022: 12024: 32025: 1
2007high 4 (2011) · low 0 (2007)2025

MAUDE reports naming product code DJG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20082—00
20092—00
2010618700
2011623314
201259501
201356004
201496200
20151214103
20162—02
2017410101
2018513401
2019323510
202053000
2021342202
20222—01
20231014000
2024515003
202533121
202631570—

What is it indicated for?

Randox Opiates Assay: The Randox Laboratories Ltd. Opiates Assay is an invitro diagnostic test for the qualitative and semi-quantitative analysis of Opiates in human urine. The cut off for both the qualitative and semi-quantitative application is 2000ng/mI for morphine to which the assay is calibrated. Qualitative and semi-quantitative results can be utilized in the diagnosis and treatment of Opiate use or overdose. The Randox Opiates Assay has been…

Read the full statement (340 words)

Randox Opiates Assay: The Randox Laboratories Ltd. Opiates Assay is an invitro diagnostic test for the qualitative and semi-quantitative analysis of Opiates in human urine. The cut off for both the qualitative and semi-quantitative application is 2000ng/mI for morphine to which the assay is calibrated. Qualitative and semi-quantitative results can be utilized in the diagnosis and treatment of Opiate use or overdose. The Randox Opiates Assay has been developed for use on the RX series analysers, which includes the RX Daytona and the RX Imola. This in vitro diagnostic device is intended for prescription use only.

The semi-quantitative mode isfor purposes of (1) Enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as GOMS, or (2) Permitting laboratories to establish quality control procedures.

This assay provides only, a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatograph/Mass Spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

Randox Multidrug Calibrator Set: The Randox Multidrug Calibrator Set is for use on human urine samples. They are liquid calibrators containing Methamphetamine, Secobarbital, Methadone, and Morphine. There are 5 levels of calibrator. They have been developed for use in the calibration of Amphetamine, Barbiturates, Methadone and Opiate assays on the RX series analysers, which includes the RX Daytona and RX Imola. This in vitro diagnostic device is intended for prescription use only.

Randox Multidrug Controls, Level 1 & 2: The Randox Multidrug Controls, level 1 and 2 are for use on human urine samples. They are liquid controls containing Methamphetamine, Secobarbital, Methadone, and Morphine. There are 2 levels of controls. They have been developed for use in the quality control of Am phetamine, Barbiturates, Methadone and Opiate assays on the RX series analysers, which includes the RX Daytona and FiX Imola. This in vitro diagnostic device is intended for prescription use only.

Indications for use as stated in the FDA 510(k) statement.

Cite this page

510(k) Report. “K113747: Randox Opiates Assay.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K113747

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)