510(k) Report

510(k) · Cleared

Cleared

Hitachi Clinical Analyzer S Test Regent Cartridge C-Reactive Protein (CRP)

K113809510(k) TraditionalCleared Aug 17, 2012238-day reviewClass II21 CFR 866.5270

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hitachi Chemical Diagnostics, Inc. · DCN System, Test, C-Reactive Protein

Cleared on Aug 17, 2012 as substantially equivalent to K073277 (Roche Diagnostics Corp., 2008) and 2 other predicates after a 238-day review. Its family tree reaches back 3 generations to 1989.

Decided
Aug 17, 2012
Review
238 days
Class
Class II
Product code
DCN
Recalls
None
Predicates
3

Its longest line of descent runs back 3 generations to K895880 (1989). 10 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2010: 1 other ancestor2009: 1 other ancestor2006: 1 other ancestor1993: 1 other ancestor1996: 1 other ancestor1995: 2 other ancestorsK895880 · Pancreatic a-Amylase Eps · Boehringer Mannheim Corp. · 1989K981897 · Roche Cobas Integra Alkaline Phosphatase Ifcc Liquid Reagent Cassette, Roche Cobas Integra Pancreatic a-Amylase Eps Reae · Roche Diagnostic Systems, Inc. · 1998K073277 · Roche-Creactive Protein (Latex) · Roche Diagnostics Corp. · 2008198519901995200020052010
  1. Root · 3 generations backK895880 Pancreatic a-Amylase EpsBoehringer Mannheim Corp. · 1989
  2. 2 generations backK981897 Roche Cobas Integra Alkaline Phosphatase Ifcc Liquid Reagent…Roche Diagnostic Systems, Inc. · 1998
  3. 1 generation backK073277 Roche-Creactive Protein (Latex)Roche Diagnostics Corp. · 2008
  4. This deviceK113809 Hitachi Clinical Analyzer S Test Regent Cartridge C-Reactive Protein…Hitachi Chemical Diagnostics, Inc. · 2012
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K895880Pancreatic a-Amylase EpsBoehringer Mannheim Corp.1989—
−2K981897Roche Cobas Integra Alkaline Phosphatase Ifcc Liquid Reagent Cassette, Roche…Roche Diagnostic Systems, Inc.1998—
−1K073277Roche-Creactive Protein (Latex)Roche Diagnostics Corp.2008—
0 (this device)K113809Hitachi Clinical Analyzer S Test Regent Cartridge C-Reactive Protein (CRP)Hitachi Chemical Diagnostics, Inc.2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
199520
199610
200610
200910
201010
FIG. 1 — Lineage of K113809, 1989–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DCN? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DCN device

  1. #1 K232624 CardioPhase® hsCRPscore 6.7 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K073277Roche-Creactive Protein (Latex)
Summary page 1
…ystem (CRP) 807.92 (a)(3): Identification of the legally marketed predicate devices Reagent Test: Roche C-Reactive Protein Latex (CRPLX)- k073277 Instrument Platform: Roche Cobas 8000 Modular Series Analyzer - k100853 Alfa Wasser…
Roche Diagnostics Corp.2008—
*K100853Cobas 8000 Modular Series Analyzer
Summary page 1
…agent Test: Roche C-Reactive Protein Latex (CRPLX)- k073277 Instrument Platform: Roche Cobas 8000 Modular Series Analyzer - k100853 Alfa Wasserman Diagnostic Technologies S40 Clinical Analyzer - k091413 807.92 (a)(4): Device Description The Hit…
Roche Diagnostics Corp.201011
*K091413S-Test Cholesterol (Cho), Model RC0009, S-Test HDL Cholesterol (HDL), Model RC0015…
Summary page 1
…Instrument Platform: Roche Cobas 8000 Modular Series Analyzer - k100853 Alfa Wasserman Diagnostic Technologies S40 Clinical Analyzer - k091413 807.92 (a)(4): Device Description The Hitachi Clinical Analyzer is an automatic, bench-top, wet chemistry system intended for u…
Alfa Wassermann, Inc.2009—

Also cleared against K073277: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 4 companies; a new generation was cleared every 8.7 years on median.
  • 5 of the 10 ancestors have a recall of any class.
  • 1 other device also cites its predicate K073277.

How does its category compare?

All DCN devices

DCN System, Test, C-Reactive Protein: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · DCN

2in 2023

2007: 12008: 22009: 32010: 12012: 12017: 12018: 32020: 12023: 2
2007high 3 (2009) · low 0 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DCN

234in 2018

2009: 111 days median2018: 234 days median
2007high 234 (2018) · low 111 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DCN

1in 2020

2008: 12019: 12020: 12026 (partial year): 1
2007high 1 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DCN

7in 2025

2010: 22011: 52012: 22014: 22015: 22016: 12018: 22019: 82021: 12022: 12023: 32024: 52025: 7
2007high 8 (2019) · low 0 (2007)2025

MAUDE reports naming product code DCN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
20082—10
2009311100
20101—02
20110—05
20121—02
20130—00
20140—02
20150—02
20160—01
20171—00
2018323402
20190—18
20201—10
20210—01
20220—01
20232—03
20240—05
20250—07
20260—1—

What is it indicated for?

The Hitachi Clinical Analyzer S TEST reagent cartridge CRP is intended for the quantitative measurement of C-reactive protein in serum, lithium heparin plasma, K3 EDTA plasma, and sodium citrate plasma. The test system is intended for use in clinical laboratories or physician office laboratories. CRP measurements aid in the evaluation of injury to body tissues, and infection and inflammatory disorders. For in vitro diagnostic use only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K113809: Hitachi Clinical Analyzer S Test Regent Cartridge C-Reactive Protein (CRP).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K113809

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026