Cobas 8000 Modular Series Analyzer

K100853510(k) SpecialCleared Sep 9, 2010167-day reviewClass II21 CFR 866.5040Recalls 11

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Roche Diagnostics Corp. · DCF Albumin, Antigen, Antiserum, Control

Cleared on Sep 9, 2010 as substantially equivalent to K060373 (Roche Diagnostics, 2006) after a 167-day review. Its family tree reaches back 2 generations to 1995. 9 later devices descend from it. It has been recalled 11 times.

Decided
Sep 9, 2010
Review
167 days
Class
Class II
Product code
DCF
Recalls
11
Predicates
1

Its longest line of descent runs back 2 generations to K953239 (1995). 3 ancestors in all. 9 later devices descend from it, 9 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1996: 1 other ancestor2013: 7 descendants cleared2012: 2 descendants clearedK953239 · Boehringer Mannheim/Hitachi 917 Analyzer · Boehringer Mannheim Corp. · 1995 · recalledK060373 · Cobas 6000 Series System · Roche Diagnostics · 2006 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 2 generations back: 1 other ancestor 1 generation forward: 9 descendants K060373 · Cobas 6000 Series System · Roche Diagnostics · 2006 · recalledK953239 · Boehringer Mannheim/Hitachi 917 Analyzer · Boehringer Mannheim Corp. · 1995 · recalled0
  1. Root · 2 generations backK953239 Boehringer Mannheim/Hitachi 917 AnalyzerBoehringer Mannheim Corp. · 1995 · Recalled
  2. 1 generation backK060373 Cobas 6000 Series SystemRoche Diagnostics · 2006 · Recalled
  3. This deviceK100853 Cobas 8000 Modular Series AnalyzerRoche Diagnostics Corp. · 2010
  4. Descendant · cites it directly · 0 of its ownK132462 S Test Reagent Cartridge Bun and S Test Reagent Cartridge CreHitachi Chemical Diagnostics, Inc. · 2013
  5. Descendant · cites it directly · 0 of its ownK130685 Hitachi Clinical Analyzer S Test Reagent Cartridges for LD and AmyHitachi Chemical Diagnostics, Inc. · 2013
  6. Descendant · cites it directly · 0 of its ownK120945 Hitachi Clinical Analyzer S Test Reagent Cartridge Alt, Hitachi…Hitachi Chemical Diagnostics, Inc. · 2013
  7. Descendant · cites it directly · 0 of its ownK131051 Hitachi S Test Reagent Cartridge Total Protein (TP) and Albumin (Alb)Hitachi Chemical Diagnostics, Inc. · 2013
  8. Descendant · cites it directly · 0 of its ownK131217 S Test Reagent Cartridge Total BilirubinHitachi Chemical Diagnostics, Inc. · 2013

Showing 5 of 9 descendants. See all descendants

Descendants cleared per year

2012: 22013: 7
2012high 7 (2013) · low 0 (2014)2026

Most influential descendants

Showing 5 of 9 · See all
  • K132462S Test Reagent Cartridge Bun and S Test Reagent Cartridge CreHitachi Chemical Diagnostics, Inc. · 2013 · cites it directly0descendants
  • K130685Hitachi Clinical Analyzer S Test Reagent Cartridges for LD and AmyHitachi Chemical Diagnostics, Inc. · 2013 · cites it directly0descendants
  • K120945Hitachi Clinical Analyzer S Test Reagent Cartridge Alt, Hitachi Clinical Analyzer S Test…Hitachi Chemical Diagnostics, Inc. · 2013 · cites it directly0descendants
  • K131051Hitachi S Test Reagent Cartridge Total Protein (TP) and Albumin (Alb)Hitachi Chemical Diagnostics, Inc. · 2013 · cites it directly0descendants
  • K131217S Test Reagent Cartridge Total BilirubinHitachi Chemical Diagnostics, Inc. · 2013 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K953239Boehringer Mannheim/Hitachi 917 AnalyzerBoehringer Mannheim Corp.199518
−1K060373Cobas 6000 Series SystemRoche Diagnostics200610
0 (this device)K100853Cobas 8000 Modular Series AnalyzerRoche Diagnostics Corp.2010—
+1K132462S Test Reagent Cartridge Bun and S Test Reagent Cartridge CreHitachi Chemical Diagnostics, Inc.2013—
+1K130685Hitachi Clinical Analyzer S Test Reagent Cartridges for LD and AmyHitachi Chemical Diagnostics, Inc.2013—
+1K120945Hitachi Clinical Analyzer S Test Reagent Cartridge Alt, Hitachi Clinical…Hitachi Chemical Diagnostics, Inc.2013—
+1K131051Hitachi S Test Reagent Cartridge Total Protein (TP) and Albumin (Alb)Hitachi Chemical Diagnostics, Inc.2013—
+1K131217S Test Reagent Cartridge Total BilirubinHitachi Chemical Diagnostics, Inc.2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201220
201370
FIG. 1 — Lineage of K100853, 1995–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DCF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K060373Cobas 6000 Series System
Summary page 1
…Product Code: CZP Predicate We claim substantial equivalence to the Roche cobas 6000 Analyzer Series Device Predicate device cleared in K060373. Device The cobas 800%Modular Analyzer Series is a fully automated system for Description clinical chemi…
Roche Diagnostics200610

Also cleared against K060373: 5 other devices →

What descends from it? (9)

9 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K132462S Test Reagent Cartridge Bun and S Test Reagent Cartridge CreHitachi Chemical Diagnostics, Inc.20130
K130685Hitachi Clinical Analyzer S Test Reagent Cartridges for LD and AmyHitachi Chemical Diagnostics, Inc.20130
K120945Hitachi Clinical Analyzer S Test Reagent Cartridge Alt, Hitachi Clinical Analyzer S Test…Hitachi Chemical Diagnostics, Inc.20130
K131051Hitachi S Test Reagent Cartridge Total Protein (TP) and Albumin (Alb)Hitachi Chemical Diagnostics, Inc.20130
K131217S Test Reagent Cartridge Total BilirubinHitachi Chemical Diagnostics, Inc.20130
K130870Hitachi Clinical Analyzer S Test Reagent Cartridge Calcium (CA)Hitachi Chemical Diagnostics, Inc.20130
K130141S Test Reagent Cartridge Alp (Alkaline Phosphatase)Hitachi Chemical Diagnostics, Inc.20130
K113809Hitachi Clinical Analyzer S Test Regent Cartridge C-Reactive Protein (CRP)Hitachi Chemical Diagnostics, Inc.20120
K120369Hitachi Clinical Analyzer S Test Reagent Cartridge Glucose (Glu)Hitachi Chemical Diagnostics, Inc.20120

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 8, 2020Class IIZ-0754-2021Software designPotential for Changed Configuration Settings on the cobas 8000 modular analyzer series/cobas pro integrated solutions, may cause incorrect results in several affected parameters. In the case of poor sample quality, discrepant results may…
Jul 18, 2019Class IIZ-2477-2019Device DesignQuality issue with high pressure solenoid valves
Mar 15, 2019Class IIZ-1249-2019Device DesignThe firm has confirmed complaints regarding signal drops on the cobas e 801 module after a ProCell II M bottle changeover, which may lead to incorrect medical decisions with respect to diagnostics and patient treatment.
Apr 3, 2018Class IIZ-2136-2018Process designThe syringe plunger on the device may be mounted in a tilted position, which applies mechanical stress to the plunger. In the worst case scenario, the plunger may crack leading to impaired syringe function.
Jul 28, 2017Class IIZ-0675-2018Device DesignA software failure may incorrectly set the system settings to "default" settings, creating a risk of incorrect results.
Jul 18, 2017Class IIZ-0910-2018Device DesignA warning is being added to the cobas 6000 analyzer and the cobas 8000 modular analyzer series Operators Manual to prevent operator injury during maintenance.
Jun 29, 2017Class IIZ-0378-2018Software designRoche has confirmed that a possible sample mismatch issue may occur on the MODULAR ANALYTICS E 170 module, cobas e 601 analyzer, or cobas e 602 analyzer due to a software limitation.
Mar 21, 2017Class IIZ-2049-2017Component change controlDue to a design change in 2015, corrosion on the sample probe electrical contacts can occur and lead to a disturbance of the sample liquid level detection (LLD). When disturbance of the sample liquid level detection (LLD) occurs, the…
Jul 15, 2016Class IIZ-0097-2017Device DesignRoche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
Mar 13, 2014Class IIZ-1449-2014Software Manufacturing/Software DeploymentThe Initial Cassette Volume Check (ICVC) is not activated on the cobas c 502 module of the cobas 8000 modular analyzer series when manually filled cobas c packs are loaded. This issue occurs with tests that use cobas c pack MULTI or empty…
Apr 4, 2013Class IIZ-1368-2013Software designWhen using the cobas e 602 module, Diluent Multiassay may be used for autodilution of certain assays instead of Diluent Universal. This may cause incorrect results under specific circumstances if all of the following conditions are met…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 15 years and passes through 3 companies; a new generation was cleared every 10.5 years on median.
  • 3 of the 3 ancestors have a recall of any class.
  • Of its descendants, K132462 (Hitachi Chemical Diagnostics, Inc., 2013) has gone furthest, with 0 descendants of its own.
  • 5 other devices also cite its predicate K060373.

How does its category compare?

All DCF devices

DCF Albumin, Antigen, Antiserum, Control: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · DCF

1in 2016

2009: 12010: 32011: 12012: 12013: 12014: 12015: 12016: 1
2007high 3 (2010) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DCF

167in 2010

2010: 167 days median
2007high 167 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DCF

1in 2019

2019: 1
2007high 1 (2019) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DCF

12in 2025

2008: 12017: 32018: 22019: 62020: 22021: 32022: 32023: 52024: 92025: 12
2007high 12 (2025) · low 0 (2007)2025

MAUDE reports naming product code DCF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—01
20091—00
2010316700
20111—00
20121—00
20131—00
20141—00
20151—00
20161—00
20170—03
20180—02
20190—16
20200—02
20210—03
20220—03
20230—05
20240—09
20250—012
20260—0—

What is it indicated for?

The cobas 8000 Modular Analyzer Series is a fully automated system for clinical chemistry analysis, intended for the in vitro qualitative and quantitative determination of analytes in body fluids. It is optimized for high throughput workloads in the professional environment using a combination of ion selective electrodes (ISE) and a photometric analysis unit. It is intended for use in conjunction with certain materials to measure a variety of analytes…

Read the full statement (225 words)

The cobas 8000 Modular Analyzer Series is a fully automated system for clinical chemistry analysis, intended for the in vitro qualitative and quantitative determination of analytes in body fluids. It is optimized for high throughput workloads in the professional environment using a combination of ion selective electrodes (ISE) and a photometric analysis unit. It is intended for use in conjunction with certain materials to measure a variety of analytes that may bbe adaptable to the below analyzers depending on the reagents used.

The cobas c701 analyzer is a fully automated, discrete clinical chemistry analyzer intended for the in vitro quantitative! qualitative determination of analytes in body fluids.

The cobas c502 analyzer is a fully automated discrete clinical chemistry analyzer intended for the in vitro quantitative / qualitative determination of analytes in body fluids.

The cobas 8000 ISE module is a fully automated ion- specific analyzer intended for the in vitro potentiometric determination of chloride, potassium and sodium in serum and plasma using ion-selective electrodes. Measurements obtained by this device are used in the diagnosis and treatment of diseases or conditions involving electrolyte imbalance.

IGA-2 is an in vitro test for the quantitative determination of IgA in human serum and plasma on Roche/Hitachi cobas c systems. IgA measurements aid in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K100853: Cobas 8000 Modular Series Analyzer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K100853

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026