Recalled · Class II · 2025Unicel DXH 800 Coulter Cellular Analysis System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Beckman Coulter, Inc. · GKZ Counter, Differential Cell
Cleared on Mar 22, 2013 after a 373-day review. The summary cites one number, which could not be confirmed as its predicate (see the table below). 1 of the 8 devices in its family, counting itself, has a Class I recall. 7 later devices descend from it. It has been recalled 11 times.
- Decided
- Mar 22, 2013
- Review
- 373 days
- Class
- Class II
- Product code
- GKZ
- Recalls
- 11
- Predicates
- 1
7 later devices descend from it, 1 citing it directly. 1 of the 8 devices in this family, counting this one, has a Class I recall.
- This deviceK120771 Unicel DXH 800 Coulter Cellular Analysis SystemBeckman Coulter, Inc. · 2013
- Descendant · cites it directly · 6 of its ownK140911 Unicel DXH 800 Coulter Cellular Analysis System/ Unicel DXH…Beckman Coulter, Inc. · 2014 · Recalled (Class I)
- Descendant · 2 generations on · 1 of its ownK193124 Unicel DxH 800 Coulter Cellular Analysis SystemBeckman Coulter · 2020 · Recalled
- Descendant · 2 generations on · 1 of its ownK162414 UniCel DxH SlideMaker Stainer Coulter Cellular Analysis SystemBeckman Coulter, Inc. · 2016
- Descendant · 3 generations on · 0 of its ownK240252 UniCel DxH 900 Coulter Cellular Analysis System; UniCel DxH…Beckman Coulter, Inc. · 2024 · Recalled
- Descendant · 2 generations on · 0 of its ownK230878 QScout Lab; QScout RLDAd Astra Diagnostics, Inc. · 2023
Showing 5 of 7 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
- K140911 Unicel DXH 800 Coulter Cellular Analysis System/…cites it directly · 2014
Most influential descendants
Showing 5 of 7 · See all- K140911Unicel DXH 800 Coulter Cellular Analysis System/ Unicel DXH Slidemaker Stainer Coulter…Beckman Coulter, Inc. · 2014 · cites it directly6descendants
- K193124Unicel DxH 800 Coulter Cellular Analysis SystemBeckman Coulter · 2020 · 2 generations on1descendant
- K162414UniCel DxH SlideMaker Stainer Coulter Cellular Analysis SystemBeckman Coulter, Inc. · 2016 · 2 generations on1descendant
- K240252UniCel DxH 900 Coulter Cellular Analysis System; UniCel DxH Slidemaker Stainer II…Beckman Coulter, Inc. · 2024 · 3 generations on0descendants
- K230878QScout Lab; QScout RLDAd Astra Diagnostics, Inc. · 2023 · 2 generations on0descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K120771 | Unicel DXH 800 Coulter Cellular Analysis System | Beckman Coulter, Inc. | 2013 | — |
| +1 | K140911 | Unicel DXH 800 Coulter Cellular Analysis System/ Unicel DXH Slidemaker Stainer… | Beckman Coulter, Inc. | 2014 | Class I |
| +2 | K193124 | Unicel DxH 800 Coulter Cellular Analysis System | Beckman Coulter | 2020 | 1 |
| +2 | K162414 | UniCel DxH SlideMaker Stainer Coulter Cellular Analysis System | Beckman Coulter, Inc. | 2016 | — |
| +3 | K240252 | UniCel DxH 900 Coulter Cellular Analysis System; UniCel DxH Slidemaker Stainer… | Beckman Coulter, Inc. | 2024 | 2 |
| +2 | K230878 | QScout Lab; QScout RLD | Ad Astra Diagnostics, Inc. | 2023 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2014 | 1 | 1 |
| 2015 | 1 | 0 |
| 2016 | 1 | 0 |
| 2019 | 1 | 0 |
| 2020 | 1 | 0 |
| 2023 | 1 | 0 |
| 2024 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new GKZ device
- #1 K250943 Sysmex XR-Series (XR-10) Automated Hematology…score 13.2 · 3 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.
| Type | Number | Device | Decided |
|---|---|---|---|
| Kept off the family tree | K081645 | FDA’s K081645 is Agilis NXT Steerable Introducer, Model G408324 Not matched in our reading of the summary (scans can blur digits), so it is kept off the family tree | 2008 |
What descends from it? (7)
1 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K140911 | Unicel DXH 800 Coulter Cellular Analysis System/ Unicel DXH Slidemaker Stainer Coulter… Class I | 2014 | 6 |
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 13, 2025 | Class II | Z-2575-2025 | Component change control | Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts. |
| Apr 24, 2017 | Class II | Z-2158-2017 | Device Design | There is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated with the unusual… |
| Apr 24, 2017 | Class II | Z-2157-2017 | Device Design | There is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated with the unusual… |
| Dec 19, 2016 | Class II | Z-1013-2017 | Under Investigation by firm | Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System may not flag or detect blasts in some blood… |
| Dec 19, 2016 | Class II | Z-1014-2017 | Under Investigation by firm | Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System may not flag or detect blasts in some blood… |
| Dec 1, 2015 | Class II | Z-0433-2016 | Software design | Beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen… |
| Dec 1, 2015 | Class II | Z-0434-2016 | Software design | Beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen… |
| Dec 1, 2015 | Class II | Z-0432-2016 | Software design | Beckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen… |
| Apr 13, 2015 | Class II | Z-1616-2015 | Nonconforming Material/Component | Beckman Coulter has received an increased number of complaints related to Hemoglobin (HGB) on UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems. The use of the DxH Diluent lots listed in the recall notice may eventually result… |
| Nov 8, 2013 | Class II | Z-1828-2014 | Under Investigation by firm | Beckman Coulter is recalling the UniCel DxH800 and UniCel DxH600 Coulter Analysis Systems because the optical degradation may potentially cause a delay in reporting results for nucleated red blood cells (nRBC), automated white blood cell… |
| Sep 24, 2013 | Class II | Z-0147-2014 | Under Investigation by firm | Beckman Coulter is initiating a recall for UniCel DxH 800 Coulter Cellular Analysis System (PN 629029) and UniCel DxH 600 Coulter Cellular Analysis System (PN B23858) because the company's internal investigation indicates that specific… |
The 5 latest of 11 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K140911Unicel DXH 800 Coulter Cellular Analysis System/ Unicel DXH Slidemaker Stainer Coulter Cellular Analysis…Beckman Coulter, Inc. · 2014 · Cites it directly · ◆ in the 2014 descendants bar of FIG. 13 Class I recalls
What does this lineage say?
- Of its descendants, K140911 (Beckman Coulter, Inc., 2014) has gone furthest, with 6 descendants of its own.
How does its category compare?
All GKZ devicesGKZ Counter, Differential Cell: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.
FDA decisions per year · GKZ
4in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · GKZ
108in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · GKZ
1in 2025
Class I (2 since 2007) Class II/III not classified
Adverse event reports per year · GKZ
77in 2025
MAUDE reports naming product code GKZ, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 6 | 117 | 37 | 268 |
| 2008 | 6 | 99 | 5 | 925 |
| 2009 | 6 | 147 | 13 | 559 |
| 2010 | 4 | 290 | 27 | 896 |
| 2011 | 3 | 315 | 28 | 2,905 |
| 2012 | 3 | 408 | 7 | 2,959 |
| 2013 | 1 | — | 6 | 3,070 |
| 2014 | 3 | 165 | 0 | 3,112 |
| 2015 | 3 | 273 | 3 | 1,972 |
| 2016 | 3 | 198 | 3 | 3,338 |
| 2017 | 4 | 280 | 3 | 229 |
| 2018 | 3 | 270 | 2 (2) | 497 |
| 2019 | 5 | 210 | 2 | 908 |
| 2020 | 1 | — | 1 | 342 |
| 2021 | 0 | — | 0 | 391 |
| 2022 | 3 | 638 | 4 | 198 |
| 2023 | 4 | 329 | 1 | 155 |
| 2024 | 3 | 157 | 0 | 115 |
| 2025 | 4 | 108 | 1 | 77 |
| 2026 | 4 | 254 | 0 | — |
What is it indicated for?
The UniCel® DxH 800 Analyzer is a quantitative multi-parameter, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories.
The UniCel® DxH 800 Analyzer identifies and enumerates the parameters indicated below on the following sample types:
* Whole Blood (Venous and Capillary)
o WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV, NE%, NE#, LY%, LY#, MO%, MO#, EO%, EO#, BA%, BA#, NRBC%, NRBC#, RET%, RET#, MRV, IRF
* Pre-Diluted Whole Blood (Venous and Capillary)
o WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV
* Body Fluids (cerebrospinal, serous and synovial)
o TNC and RBC
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K120771 · printed Nov 4, 2026
Cite this page
510(k) Report. “K120771: Unicel DXH 800 Coulter Cellular Analysis System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K120771