Unicel DXH 800 Coulter Cellular Analysis System/ Unicel DXH Slidemaker Stainer Coulter Cellular Analysis System

K140911510(k) AbbreviatedCleared Sep 5, 2014149-day reviewClass II21 CFR 864.5220Recalls 17 (3 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Beckman Coulter, Inc. · GKZ Counter, Differential Cell

Cleared on Sep 5, 2014 as substantially equivalent to K120771 (Beckman Coulter, Inc., 2013) after a 149-day review. Its family tree reaches back 1 generation to 2013. 1 of the 8 devices in its family, counting itself, has a Class I recall. 6 later devices descend from it. It has been recalled 17 times, including 3 Class I recalls.

Decided
Sep 5, 2014
Review
149 days
Class
Class II
Product code
GKZ
Recalls
17 (3 Class I)
Predicates
1

Its longest line of descent runs back 1 generation to K120771 (2013). 6 later devices descend from it, 5 citing it directly. 1 of the 8 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2019: 1 descendant cleared2023: 1 descendant cleared2020: 1 descendant cleared2015: 1 descendant cleared2016: 1 descendant cleared2024: 1 descendant clearedK120771 · Unicel DXH 800 Coulter Cellular Analysis System · Beckman Coulter, Inc. · 2013 · recalled2010201520202025
Descendants, by generation 1 generation forward: 5 descendants2 generations forward: 1 descendant K120771 · Unicel DXH 800 Coulter Cellular Analysis System · Beckman Coulter, Inc. · 2013 · recalled0
  1. Root · 1 generation backK120771 Unicel DXH 800 Coulter Cellular Analysis SystemBeckman Coulter, Inc. · 2013 · Recalled
  2. This deviceK140911 Unicel DXH 800 Coulter Cellular Analysis System/ Unicel DXH…Beckman Coulter, Inc. · 2014
  3. Descendant · cites it directly · 1 of its ownK193124 Unicel DxH 800 Coulter Cellular Analysis SystemBeckman Coulter · 2020 · Recalled
  4. Descendant · cites it directly · 1 of its ownK162414 UniCel DxH SlideMaker Stainer Coulter Cellular Analysis SystemBeckman Coulter, Inc. · 2016
  5. Descendant · 2 generations on · 0 of its ownK240252 UniCel DxH 900 Coulter Cellular Analysis System; UniCel DxH…Beckman Coulter, Inc. · 2024 · Recalled
  6. Descendant · cites it directly · 0 of its ownK230878 QScout Lab; QScout RLDAd Astra Diagnostics, Inc. · 2023
  7. Descendant · cites it directly · 0 of its ownK181475 DxH 520 Hematology InstrumentBeckman Coulter · 2019

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2015: 12016: 12019: 12020: 12023: 12024: 1
2015high 1 (2015) · low 0 (2017)2026

Most influential descendants

Showing 5 of 6 · See all
  • K193124Unicel DxH 800 Coulter Cellular Analysis SystemBeckman Coulter · 2020 · cites it directly1descendant
  • K162414UniCel DxH SlideMaker Stainer Coulter Cellular Analysis SystemBeckman Coulter, Inc. · 2016 · cites it directly1descendant
  • K240252UniCel DxH 900 Coulter Cellular Analysis System; UniCel DxH Slidemaker Stainer II…Beckman Coulter, Inc. · 2024 · 2 generations on0descendants
  • K230878QScout Lab; QScout RLDAd Astra Diagnostics, Inc. · 2023 · cites it directly0descendants
  • K181475DxH 520 Hematology InstrumentBeckman Coulter · 2019 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K120771Unicel DXH 800 Coulter Cellular Analysis SystemBeckman Coulter, Inc.201311
0 (this device)K140911Unicel DXH 800 Coulter Cellular Analysis System/ Unicel DXH Slidemaker Stainer…Beckman Coulter, Inc.2014—
+1K193124Unicel DxH 800 Coulter Cellular Analysis SystemBeckman Coulter20201
+1K162414UniCel DxH SlideMaker Stainer Coulter Cellular Analysis SystemBeckman Coulter, Inc.2016—
+2K240252UniCel DxH 900 Coulter Cellular Analysis System; UniCel DxH Slidemaker Stainer…Beckman Coulter, Inc.20242
+1K230878QScout Lab; QScout RLDAd Astra Diagnostics, Inc.2023—
+1K181475DxH 520 Hematology InstrumentBeckman Coulter2019—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
201610
201910
202010
202310
202410
FIG. 1 — Lineage of K140911, 2013–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GKZ device

  1. #1 K250943 Sysmex XR-Series (XR-10) Automated Hematology…score 13.2 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K120771Unicel DXH 800 Coulter Cellular Analysis System
Summary page 7
…Number Cleared Code UniCel® DxH 800 Coulter® March K120771 Class II 864.5220 GKZ Cellular Analysis System 22, 2013 Manually prep…
Beckman Coulter, Inc.201311
*—Manually prepared blood films per the manual wedge-pull film technique as described in…
Named without a submission number
—

What descends from it? (6)

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K193124Unicel DxH 800 Coulter Cellular Analysis SystemBeckman Coulter20201
K162414UniCel DxH SlideMaker Stainer Coulter Cellular Analysis SystemBeckman Coulter, Inc.20161
K230878QScout Lab; QScout RLDAd Astra Diagnostics, Inc.20230
K181475DxH 520 Hematology InstrumentBeckman Coulter20190
K141932Aquios CL Flow Cytometer, Aquios Tetra-1 Panel,Aquios Tetra-2+Panel, Aquios Immuno-Trol…Beckman Coulter, Inc.20150

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 7, 2019Class IIZ-0516-2020Under Investigation by firmWhen the UniCel DxH Slidemaker Stainer and UniCel DxH Slidemaker Stainer II dispense probe is slightly bent, there is a potential for cell carryover from the previous sample, and erroneous results could be reported from the slide review.…
Jul 30, 2018Class IZ-1383-2019Software Design ChangeComplaints received of sporadic erroneously elevated platelet results without flags or system messages. Thrombocytopenia may go unrecognized.
Jul 30, 2018Class IZ-1384-2019Software Design ChangeComplaints received of sporadic erroneously elevated platelet results without flags or system messages. Thrombocytopenia may go unrecognized.
Jul 30, 2018Class IZ-1382-2019Software Design ChangeComplaints received of sporadic erroneously elevated platelet results without flags or system messages. Thrombocytopenia may go unrecognized.
Nov 3, 2017Class IIZ-0237-2018Device DesignThe shield for the Single-Tube Station on the DxH instrument may interfere with sample aspiration in the open vial position.
Nov 3, 2017Class IIZ-0238-2018Device DesignThe shield for the Single-Tube Station on the DxH instrument may interfere with sample aspiration in the open vial position.
Nov 3, 2017Class IIZ-0236-2018Device DesignThe shield for the Single-Tube Station on the DxH instrument may interfere with sample aspiration in the open vial position.
Jun 20, 2017Class IIZ-3101-2017Employee errorThe firm has identified that a fitting in some Slidemaker Stainer units may not completely seal after installation.
Apr 24, 2017Class IIZ-2158-2017Device DesignThere is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated with the unusual…
Apr 24, 2017Class IIZ-2157-2017Device DesignThere is a possibility of a data acquisition disruption that may cause an unusual light scatter event pattern for the white blood cell differential that may not have a system flag and/or message. The events associated with the unusual…
Dec 19, 2016Class IIZ-1013-2017Under Investigation by firmBeckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System may not flag or detect blasts in some blood…
Dec 19, 2016Class IIZ-1014-2017Under Investigation by firmBeckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the UniCel DxH 800 and DxH 600 Coulter Cellular Analysis System may not flag or detect blasts in some blood…
Aug 8, 2016Class IIZ-2732-2016Software designThere is a possibility of specimen misidentification for the stand-alone DxH 800 and DxH 600 where the Primary Identifier is configured to Tube Position ID.
Aug 4, 2016Class IIZ-2692-2016Under Investigation by firmBeckman Coulter is recalling the UniCel DxH Slidemaker Stainer Coulter Cellular Analysis System to reduce the potential for flammable liquids coming into contact with electrical components in the Stainer module possibly causing a fire.
Dec 1, 2015Class IIZ-0433-2016Software designBeckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen…
Dec 1, 2015Class IIZ-0434-2016Software designBeckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen…
Dec 1, 2015Class IIZ-0432-2016Software designBeckman Coulter is recalling the DxH 800 Slidemaker Stainer Software version 3.0.2.0 and DxH 600 software version 1.1.1.0 because the software for the DxH systems allows the creation of multiple orders with the same Specimen…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K193124 (Beckman Coulter, 2020) has gone furthest, with 1 descendant of its own.

How does its category compare?

All GKZ devices

GKZ Counter, Differential Cell: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · GKZ

4in 2025

2007: 62008: 62009: 62010: 42011: 32012: 32013: 12014: 32015: 32016: 32017: 42018: 32019: 52020: 12022: 32023: 42024: 32025: 42026 (partial year): 4
2007high 6 (2007) · low 0 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GKZ

108in 2025

2007: 117 days median2008: 99 days median2009: 147 days median2010: 290 days median2011: 315 days median2012: 408 days median2014: 165 days median2015: 273 days median2016: 198 days median2017: 280 days median2018: 270 days median2019: 210 days median2022: 638 days median2023: 329 days median2024: 157 days median2025: 108 days median2026 (partial year): 254 days median
2007high 638 (2022) · low 99 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GKZ

1in 2025

2007: 372008: 52009: 132010: 272011: 282012: 72013: 62015: 32016: 32017: 32018: 2, 2 Class I2019: 22020: 12022: 42023: 12025: 1
2007high 37 (2007) · low 0 (2014)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · GKZ

77in 2025

2007: 2682008: 9252009: 5592010: 8962011: 2,9052012: 2,9592013: 3,0702014: 3,1122015: 1,9722016: 3,3382017: 2292018: 4972019: 9082020: 3422021: 3912022: 1982023: 1552024: 1152025: 77
2007high 3,338 (2016) · low 77 (2025)2025

MAUDE reports naming product code GKZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007611737268
20086995925
2009614713559
2010429027896
20113315282,905
2012340872,959
20131—63,070
2014316503,112
2015327331,972
2016319833,338
201742803229
201832702 (2)497
201952102908
20201—1342
20210—0391
202236384198
202343291155
202431570115
20254108177
202642540—

What is it indicated for?

The UniCel® DxH 800 Analyzer is a quantitative multi-parameter, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories. The UniCel® DxH 800 Analyzer identifies and enumerates the parameters indicated below on the following sample types: Whole Blood (venous and capillary) - WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV, NE%, NE#, LY%, LY#, MO%, MO#, EO%, EO#, BA%, BA#, NRBC%…

Read the full statement (145 words)

The UniCel® DxH 800 Analyzer is a quantitative multi-parameter, automated hematology analyzer for in vitro diagnostic use in screening patient populations found in clinical laboratories.

The UniCel® DxH 800 Analyzer identifies and enumerates the parameters indicated below on the following sample types:

Whole Blood (venous and capillary) - WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV, NE%, NE#, LY%, LY#, MO%, MO#, EO%, EO#, BA%, BA#, NRBC%, NRBC#, RET%, RET#, MRV, IRF

Pre-Diluted Whole Blood (venous and capillary) - WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV

Body Fluids (cerebrospinal, serous and synovial) - TNC and RBC

The DxH Slidemaker Stainer is a fully automated slide preparation and staining device that aspirates a whole-blood sample, smears a blood film on a clean microscope slide, and delivers a variety of fixatives, stains, buffers, and rinse solutions to that blood smear.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K140911: Unicel DXH 800 Coulter Cellular Analysis System/ Unicel DXH Slidemaker Stainer Coulter Cellular Analysis System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K140911

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026