Flexlab 3.6, Accelerator A3600

K121012510(k) TraditionalCleared Aug 31, 2012150-day reviewClass II21 CFR 862.1600Recalls 33

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Inpeco S.P.A. · CEM Electrode, Ion Specific, Potassium

Cleared on Aug 31, 2012 as substantially equivalent to K093318 (Abbott Laboratories, 2010) after a 150-day review. Its family tree reaches back 6 generations to 1990. It has been recalled 33 times.

Decided
Aug 31, 2012
Review
150 days
Class
Class II
Product code
CEM
Recalls
33
Predicates
1

Its longest line of descent runs back 6 generations to K902336 (1990). 10 ancestors in all, the oldest K872494 (1987). No later device cites it as a predicate yet.

  1. Root · 6 generations backK902336 Ciba Corning Acs:180 Automated ChemiluminescenceCiba Corning Diagnostics Corp. · 1990
  2. 5 generations backK971418 Chiron Diagnostics Automated Chemiluminescence System (Asc) Centaur…Chiron Diagnostics Corp. · 1997 · Recalled
  3. 4 generations backK032525 Bayer Advia Centaur Automated Chemiluminescence AnalyzerBayer Healthcare, LLC · 2003
  4. 3 generations backK041133 Modification to Bayer Advia Centaur Automated Chemiluminescence…Bayer Healthcare, LLC · 2004 · Recalled
  5. 2 generations backK082638 Advia Centaur System with Streamlab Analytical WorkcellSiemens Healthcare Diagnostics, Inc. · 2009 · Recalled
  6. 1 generation backK093318 Accelerator ApsAbbott Laboratories · 2010 · Recalled
  7. This deviceK121012 Flexlab 3.6, Accelerator A3600Inpeco S.P.A. · 2012
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K902336Ciba Corning Acs:180 Automated ChemiluminescenceCiba Corning Diagnostics Corp.1990—
−5K971418Chiron Diagnostics Automated Chemiluminescence System (Asc) Centaur AnalyzerChiron Diagnostics Corp.19974
−4K032525Bayer Advia Centaur Automated Chemiluminescence AnalyzerBayer Healthcare, LLC2003—
−3K041133Modification to Bayer Advia Centaur Automated Chemiluminescence AnalyzerBayer Healthcare, LLC20048
−2K082638Advia Centaur System with Streamlab Analytical WorkcellSiemens Healthcare Diagnostics, Inc.20093
−1K093318Accelerator ApsAbbott Laboratories20102
0 (this device)K121012Flexlab 3.6, Accelerator A3600Inpeco S.P.A.2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
199110
199810
200510
FIG. 1 — Lineage of K121012, 1990–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in CEM? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CEM device

  1. #1 K253491 ISE indirect K for Gen.2; ISE indirect Na for…score 11.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K093318Accelerator Aps
Summary page 2
…e 2 === Identification of Predicate Device: Predicate Instrument or 510(k) Number Product Code Assay Al'S Accelerator K093318 JQP Description: The following is a brief description of the FlexLab 3.6 system. The FlexLab 3.6 Automa…
Abbott Laboratories20102

Also cleared against K093318: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 30, 2025Class IIZ-0596-2026Device DesignThe Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results, this error occurs if the operators open the printer head of mALX to perform…
Oct 30, 2025Class IIZ-0595-2026Device DesignThe Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results. This error occurs if the operators open the printer head of mALX to perform…
Sep 3, 2025Class IIZ-0106-2026No Marketing ApplicationThe FlexLab System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the FDA, so its use may…
Sep 3, 2025Class IIZ-0107-2026No Marketing ApplicationThe FlexLab X System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the FDA, so its use may…
Nov 3, 2023Class IIZ-0560-2024Software Design ChangeFirmware of the Interface Module with Roche Cobas 8000 and Cobas PRO (CO8 IM) may lead to a delayed sample tube processing.
Mar 10, 2023Class IIZ-1561-2023Under Investigation by firmFirmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate…
Mar 10, 2023Class IIZ-1559-2023Under Investigation by firmFirmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate…
Mar 10, 2023Class IIZ-1563-2023Under Investigation by firmFirmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate…
Mar 10, 2023Class IIZ-1560-2023Under Investigation by firmFirmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate…
Mar 10, 2023Class IIZ-1562-2023Under Investigation by firmFirmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate…
Sep 10, 2021Class IIZ-0087-2022Software designThe Firmware (FW) of the Automation System Interface Module to ADVIA Centaur XP/XPT may lead to an incorrect association of test results to sample ID. Improper management of the sample may bring an incorrect patient result.
Aug 7, 2020Class IIZ-2975-2020Software designFirmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results
Aug 7, 2020Class IIZ-2974-2020Software designFirmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results
Aug 7, 2020Class IIZ-2973-2020Software designFirmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results
Jul 14, 2020Class IIZ-0433-2021Software designWhen a Clot Detection Error is generated during the sample aspiration from the Primary Sample Tube the aspirated volume is dispensed into the first empty Secondary Sample Tube. This Secondary Sample Tube is flagged with error 2132 or 1442…
Jul 14, 2020Class IIZ-0434-2021Software designWhen a Clot Detection Error is generated during the sample aspiration from the Primary Sample Tube the aspirated volume is dispensed into the first empty Secondary Sample Tube. This Secondary Sample Tube is flagged with error 2132 or 1442…
Jul 14, 2020Class IIZ-0432-2021Software designWhen a Clot Detection Error is generated during the sample aspiration from the Primary Sample Tube the aspirated volume is dispensed into the first empty Secondary Sample Tube. This Secondary Sample Tube is flagged with error 2132 or 1442…
Apr 7, 2020Class IIZ-2141-2020Software designThe identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation software. This problem can occur only when one of…
Apr 7, 2020Class IIZ-2140-2020Software designThe identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation software. This problem can occur only when one of…
Apr 7, 2020Class IIZ-1889-2020Software design (manufacturing process)The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.
Apr 7, 2020Class IIZ-2003-2020Software designWhen analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples.
Apr 7, 2020Class IIZ-2198-2020Software designPotential for discrepant patient results due to a potential for the primary sample tube to be diluted. In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the…
Apr 7, 2020Class IIZ-1888-2020Software design (manufacturing process)The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.
Apr 7, 2020Class IIZ-2199-2020Software designIn certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3 of sample volume back into the Primary Tube. Evidence from…
Apr 7, 2020Class IIZ-2142-2020Software designThe identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation software. This problem can occur only when one of…
Apr 7, 2020Class IIZ-1887-2020Software design (manufacturing process)The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.
Apr 7, 2020Class IIZ-2004-2020Software designWhen analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples.
Apr 7, 2020Class IIZ-2200-2020Software designIn certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3 of sample volume back into the Primary Tube. Evidence from…
Apr 7, 2020Class IIZ-2114-2020Component design/selectionThe sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms expected by design.
Apr 7, 2020Class IIZ-2115-2020Component design/selectionThe sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms expected by design.
Aug 24, 2018Class IIZ-0361-2019Under Investigation by firmThe firm has identified customer sites where there are modules on which the safety pacemaker label is missing. The information regarding the label included in the Operations Manual has been updated to specify that: 1) the risk of a…
Feb 13, 2018Class IIZ-1798-2018Device DesignModule may freeze without generating user warning. There is a potential risk in delay of sample processing, leading to delayed delivery of test results to patients.
Oct 4, 2016Class IIZ-0587-2017Employee errorInpeco received several notifications from the field about burnt schuko sockets. The customers reported that some smoke was seen coming from the damaged sockets.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 22 years and passes through 6 companies; a new generation was cleared every 4.7 years on median.
  • 6 of the 10 ancestors have a recall of any class.
  • 1 other device also cites its predicate K093318.

How does its category compare?

All CEM devices

CEM Electrode, Ion Specific, Potassium: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · CEM

1in 2025

2007: 12009: 12010: 32011: 32012: 42020: 12023: 22025: 12026 (partial year): 1
2007high 4 (2012) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CEM

180in 2012

2010: 195 days median2011: 74 days median2012: 180 days median
2007high 195 (2010) · low 74 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CEM

4in 2025

2008: 12010: 122011: 62016: 12019: 12020: 32021: 12023: 22025: 4
2007high 12 (2010) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · CEM

29in 2025

2007: 662008: 2452009: 792010: 1132011: 1262012: 452013: 372014: 42015: 72016: 62017: 322018: 132019: 252020: 272021: 182022: 162023: 302024: 422025: 29
2007high 245 (2008) · low 4 (2014)2025

MAUDE reports naming product code CEM, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—066
20080—1245
20091—079
2010319512113
20113746126
20124180045
20130—037
20140—04
20150—07
20160—16
20170—032
20180—013
20190—125
20201—327
20210—118
20220—016
20232—230
20240—042
20251—429
20261—0—

What is it indicated for?

The FlexLab 3.6 Automation is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory.

The system consolidates analytical instruments, such as the ARCHITECT c8000 System into a unified workstation that performs a variety of instrument specific assays such as Sodium, Potassium and Chloride.

Sodium, Potassium and Chloride measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K121012: Flexlab 3.6, Accelerator A3600.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K121012

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026