510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

Chiron Diagnostics Automated Chemiluminescence System (Asc) Centaur Analyzer

K971418510(k) TraditionalCleared Jul 17, 199792-day reviewClass II21 CFR 866.6010Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Chiron Diagnostics Corp. · MOI System, Test, Immunological, Antigen, Tumor

Cleared on Jul 17, 1997 as substantially equivalent to K902336 (Ciba Corning Diagnostics Corp., 1990) after a 92-day review. Its family tree reaches back 1 generation to 1990. 21 later devices descend from it. It has been recalled 4 times.

Decided
Jul 17, 1997
Review
92 days
Class
Class II
Product code
MOI
Recalls
4
Predicates
1

Its longest line of descent runs back 1 generation to K902336 (1990). 21 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2009: 2 descendants cleared2003: 1 descendant cleared2004: 1 descendant cleared2016: 1 descendant cleared2017: 3 descendants cleared2015: 1 descendant cleared2007: 1 descendant cleared2012: 3 descendants cleared2010: 3 descendants cleared2023: 2 descendants cleared2022: 1 descendant cleared2014: 1 descendant cleared2024: 1 descendant clearedK902336 · Ciba Corning Acs:180 Automated Chemiluminescence · Ciba Corning Diagnostics Corp. · 19901990200020102020
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 2 descendants3 generations forward: 5 descendants4 generations forward: 7 descendants5 generations forward: 4 descendants6 generations forward: 1 descendant K902336 · Ciba Corning Acs:180 Automated Chemiluminescence · Ciba Corning Diagnostics Corp. · 19900+5
  1. Root · 1 generation backK902336 Ciba Corning Acs:180 Automated ChemiluminescenceCiba Corning Diagnostics Corp. · 1990
  2. This deviceK971418 Chiron Diagnostics Automated Chemiluminescence System (Asc) Centaur…Chiron Diagnostics Corp. · 1997
  3. Descendant · cites it directly · 18 of its ownK032525 Bayer Advia Centaur Automated Chemiluminescence AnalyzerBayer Healthcare, LLC · 2003
  4. Descendant · 2 generations on · 17 of its ownK041133 Modification to Bayer Advia Centaur Automated Chemiluminescence…Bayer Healthcare, LLC · 2004 · Recalled
  5. Descendant · 3 generations on · 4 of its ownK151792 Trinidad IM Thyroid Stimulating Hormone (TSH) Assay, Trinidad IM TSH…Siemens Healthcare Diagnostics, Inc. · 2015 · Recalled
  6. Descendant · 3 generations on · 4 of its ownK082638 Advia Centaur System with Streamlab Analytical WorkcellSiemens Healthcare Diagnostics, Inc. · 2009 · Recalled
  7. Descendant · 3 generations on · 4 of its ownK071224 Dimension Vista Vitamin B12 and Folate Flex Reagent Cartridges and…Dade Behring, Inc. · 2007 · Recalled

Showing 5 of 21 descendants. See all descendants

Descendants cleared per year

2003: 12004: 12007: 12009: 22010: 32012: 32014: 12015: 12016: 12017: 32022: 12023: 22024: 1
2003high 3 (2010) · low 0 (2005)2026

Most influential descendants

Showing 5 of 21 · See all
  • K032525Bayer Advia Centaur Automated Chemiluminescence AnalyzerBayer Healthcare, LLC · 2003 · cites it directly18descendants
  • K041133Modification to Bayer Advia Centaur Automated Chemiluminescence AnalyzerBayer Healthcare, LLC · 2004 · 2 generations on17descendants
  • K151792Trinidad IM Thyroid Stimulating Hormone (TSH) Assay, Trinidad IM TSH Calibrators…Siemens Healthcare Diagnostics, Inc. · 2015 · 3 generations on4descendants
  • K082638Advia Centaur System with Streamlab Analytical WorkcellSiemens Healthcare Diagnostics, Inc. · 2009 · 3 generations on4descendants
  • K071224Dimension Vista Vitamin B12 and Folate Flex Reagent Cartridges and Loci 4 CalibratorDade Behring, Inc. · 2007 · 3 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K902336Ciba Corning Acs:180 Automated ChemiluminescenceCiba Corning Diagnostics Corp.1990—
0 (this device)K971418Chiron Diagnostics Automated Chemiluminescence System (Asc) Centaur AnalyzerChiron Diagnostics Corp.1997—
+1K032525Bayer Advia Centaur Automated Chemiluminescence AnalyzerBayer Healthcare, LLC2003—
+2K041133Modification to Bayer Advia Centaur Automated Chemiluminescence AnalyzerBayer Healthcare, LLC20048
+3K151792Trinidad IM Thyroid Stimulating Hormone (TSH) Assay, Trinidad IM TSH…Siemens Healthcare Diagnostics, Inc.201522
+3K082638Advia Centaur System with Streamlab Analytical WorkcellSiemens Healthcare Diagnostics, Inc.20093
+3K071224Dimension Vista Vitamin B12 and Folate Flex Reagent Cartridges and Loci 4…Dade Behring, Inc.20071

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200410
200710
200920
201030
201230
201410
201510
201610
201730
202210
202320
202410
FIG. 1 — Lineage of K971418, 1990–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MOI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MOI device

  1. #1 K240403 Access BR Monitorscore 7.9 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K902336Ciba Corning Acs:180 Automated Chemiluminescence
Summary page 1
…| 3. Predicate Device Information Name: ACS:180® Analyzer Manufacturer: Chiron Diagnostics Corporation (formally Ciba Corning Diagnostics Corp.) 510(k) Number: D.C. #K902336. 4. Device Description The ACS:Centaur system is a stand-alone, continuous operation, immunochemistry analyzer. The system perf…
Ciba Corning Diagnostics Corp.1990—

Also cleared against K902336: 1 other device →

What descends from it? (21)

See all 21

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K032525Bayer Advia Centaur Automated Chemiluminescence AnalyzerBayer Healthcare, LLC200318
K083373Immulite 2000 Free T4Siemens Healthcare Diagnostics, Inc.20091

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 18, 2014Class IIZ-1553-2014Device DesignInstrument Cover Gas Spring failures. Over time, the gas spring may lose its effectiveness and fail to support the cover in its full or partially open position. This may lead to the cover falling.
Jan 12, 2008Not classifiedZ-1119-2008Nonconforming Material/ComponentIncorrect results: False reactive specimens on the ADVIA Centaur HAV IgM assay.
Nov 9, 2006Not classifiedZ-0645-2007OtherFirm received complaints for signal 2 & 4 errors, and shifts in Relative Light Units (RLU). Investigations showed that the On-Board Stability (OBS) of the acid & base reagents may be compromised after 4 days after installation onto the…
Aug 4, 2006Not classifiedZ-1516-06Software designSample/Patient mis-identification (software defect)-the system can associate a test result and sample identification (SID) with an incorrect patient name when the Patient Demographics feature is used and the Patient Identification (PID)…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K032525 (Bayer Healthcare, LLC, 2003) has gone furthest, with 18 descendants of its own.
  • 1 other device also cites its predicate K902336.

How does its category compare?

All MOI devices

MOI System, Test, Immunological, Antigen, Tumor: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MOI

2in 2024

2007: 12011: 12018: 22019: 12020: 22024: 2
2007high 2 (2018) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MOI

Too few MOI 510(k)s in any year since 2007 to give a median.

Recalls per year · MOI

2in 2012

2007: 12012: 2
2007high 2 (2012) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MOI

6in 2025

2007: 22008: 32009: 32010: 222011: 32012: 42013: 32014: 522015: 282016: 52017: 72018: 12019: 52020: 112021: 162022: 162023: 62024: 52025: 6
2007high 52 (2014) · low 1 (2018)2025

MAUDE reports naming product code MOI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—12
20080—03
20090—03
20100—022
20111—03
20120—24
20130—03
20140—052
20150—028
20160—05
20170—07
20182—01
20191—05
20202—011
20210—016
20220—016
20230—06
20242—05
20250—06
20260—0—

What is it indicated for?

The intended use of the ACS:Centaur system is an automated immunoassay analyzer designed to perform in vitro diagnostic immunochemical assays analysis on clinical specimens using the ACS iimmunoassays. The system menu will include endocrine, anemia, reproductive, cardiac, therapeutic drug, and infectious disease assays. All assays are based on chemiluminescent technology.proteins and is an indirect indicator of thyroid status.4

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K971418: Chiron Diagnostics Automated Chemiluminescence System (Asc) Centaur Analyzer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K971418

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026