510(k) Report

510(k) · Cleared

Recalled · Class II · 2025

Illumigene Group B Streptococcus Carrot Broth Claim

K121044510(k) TraditionalCleared May 1, 201225-day reviewClass I21 CFR 866.3740Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Meridian Bioscience, Inc. · NJR Nucleic Acid Amplification Assay System, Group B Streptococcus, Direct Specimen Test

Cleared on May 1, 2012 as substantially equivalent to K112125 (Meridian Bioscience, Inc., 2011) after a 25-day review. Its family tree reaches back 5 generations to 1994. It has been recalled once.

Decided
May 1, 2012
Review
25 days
Class
Class I
Product code
NJR
Recalls
1
Predicates
1

Its longest line of descent runs back 5 generations to K936112 (1994). No later device cites it as a predicate yet.

  1. Root · 5 generations backK936112 Strep B OiaBiostar, Inc. · 1994
  2. 4 generations backK991828 Strep B OiaBiostar, Inc. · 1999
  3. 3 generations backK022504 Idi-Strep B AssayInfectio Diagnostic, Inc. · 2002
  4. 2 generations backK062948 Smart GBS and Smartcycle DX System and Software, Version 1.7BCepheid · 2006
  5. 1 generation backK112125 Illumigene Group B Streptococcus ,External Control KitMeridian Bioscience, Inc. · 2011
  6. This deviceK121044 Illumigene Group B Streptococcus Carrot Broth ClaimMeridian Bioscience, Inc. · 2012
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K936112Strep B OiaBiostar, Inc.1994—
−4K991828Strep B OiaBiostar, Inc.1999—
−3K022504Idi-Strep B AssayInfectio Diagnostic, Inc.2002—
−2K062948Smart GBS and Smartcycle DX System and Software, Version 1.7BCepheid2006—
−1K112125Illumigene Group B Streptococcus ,External Control KitMeridian Bioscience, Inc.2011—
0 (this device)K121044Illumigene Group B Streptococcus Carrot Broth ClaimMeridian Bioscience, Inc.2012—
FIG. 1 — Lineage of K121044, 1994–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NJR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NJR device

  1. #1 K222638 Xpert® Xpress GBS, GeneXpert® Dx System…score 7.0 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K112125Illumigene Group B Streptococcus ,External Control Kit
Summary page 3
…lity Control. J.Substantial Equivalence Information: 1. Predicate device name(s): illumigene Group BStreptococcus (GBS) Assay 2. Predicate 510(k) numbers: K112 125 3. Comparison with predicates: Similarities…
Meridian Bioscience, Inc.2011—

Also cleared against K112125: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 7, 2025Class IIZ-0569-2026Nonconforming Material/ComponentThe cap on the reaction buffer tube may be missing the black rubber gasket resulting in the potential for leakage and reduced buffer volume.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 4 companies; a new generation was cleared every 4.1 years on median.
  • None of the 5 ancestors has a recall.
  • 2 other devices also cite its predicate K112125.

How does its category compare?

All NJR devices

NJR Nucleic Acid Amplification Assay System, Group B Streptococcus, Direct Specimen Test: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · NJR

1in 2023

2010: 12011: 12012: 32013: 12015: 12016: 12017: 22018: 32022: 12023: 1
2007high 3 (2012) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NJR

87in 2018

2012: 162 days median2018: 87 days median
2007high 162 (2012) · low 87 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NJR

1in 2025

2007: 12011: 12012: 22015: 12022: 12024: 22025: 1
2007high 2 (2012) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · NJR

14in 2024

2014: 12017: 12020: 12021: 22022: 32023: 82024: 14
2007high 14 (2024) · low 0 (2007)2025

MAUDE reports naming product code NJR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—10
20080—00
20090—00
20101—00
20111—10
2012316220
20131—00
20140—01
20151—10
20161—00
20172—01
201838700
20190—00
20200—01
20210—02
20221—13
20231—08
20240—214
20250—10
20260—0—

What is it indicated for?

The illumigene Group B Streptococcus (GBS) assay, performed on the illumipro-10, is a qualitative in vitro diagnostic for the detection of Streptococcus agalactiae in enriched cultures obtained from vaginal/rectal swab specimens from antepartum women. Enriched cultures are obtained by 18-24 hour incubation of vaginal/rectal swab specimens in selective broth medium, either Lim Broth, TransVag Broth or Carrot Broth. The illumigene GBS assay utilizes…

Read the full statement (161 words)

The illumigene Group B Streptococcus (GBS) assay, performed on the illumipro-10, is a qualitative in vitro diagnostic for the detection of Streptococcus agalactiae in enriched cultures obtained from vaginal/rectal swab specimens from antepartum women. Enriched cultures are obtained by 18-24 hour incubation of vaginal/rectal swab specimens in selective broth medium, either Lim Broth, TransVag Broth or Carrot Broth.

The illumigene GBS assay utilizes loop-mediated isothermal DNA amplification (LAMP) technology to detect Streptococcus agalactiae by targeting a segment of the Streptococcus agalactiae genome. Results from the illumigene GBS assay can be used as an aid in establishing the GBS colonization status of antepartum women. This assay is not intended to diagnose or monitor treatment for GBS infections.

The illumigene GBS assay does not provide susceptibility results. Culture isolates are needed for performing susceptibility testing as recommended for penicillin-allergic women.

illumigene Group B Streptococcus is intended for use in hospital, reference or state laboratory settings. The device is not intended for point-of-care use.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K121044: Illumigene Group B Streptococcus Carrot Broth Claim.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K121044

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026