510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

Agility

K121328510(k) TraditionalCleared Jun 12, 201240-day reviewClass II21 CFR 892.5050Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Elekta, Ltd. · IYE Accelerator, Linear, Medical

Cleared on Jun 12, 2012 as substantially equivalent to K082122 (Elekta, Ltd., 2008) after a 40-day review. Its family tree reaches back 2 generations to 1996. 37 later devices descend from it. It has been recalled 3 times.

Decided
Jun 12, 2012
Review
40 days
Class
Class II
Product code
IYE
Recalls
3
Predicates
1

Its longest line of descent runs back 2 generations to K963624 (1996). 37 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2013: 1 descendant cleared2018: 3 descendants cleared2014: 2 descendants cleared2015: 2 descendants cleared2019: 1 descendant cleared2017: 4 descendants cleared2016: 1 descendant cleared2021: 4 descendants cleared2022: 2 descendants cleared2026: 4 descendants cleared2025: 2 descendants cleared2020: 2 descendants cleared2023: 7 descendants cleared2024: 2 descendants clearedK963624 · Philips Sli Series Linear Accelerator · Philips Medical Systems, Inc. · 1996 · recalledK082122 · MLCI2 · Elekta, Ltd. · 2008 · recalled1995200020052010201520202025
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 2 descendants3 generations forward: 3 descendants4 generations forward: 5 descendants5 generations forward: 5 descendants6 generations forward: 4 descendants7 generations forward: 5 descendants8 generations forward: 4 descendants9 generations forward: 4 descendants10 generations forward: 3 descendants11 generations forward: 1 descendant K082122 · MLCI2 · Elekta, Ltd. · 2008 · recalledK963624 · Philips Sli Series Linear Accelerator · Philips Medical Systems, Inc. · 1996 · recalled0+5+10
  1. Root · 2 generations backK963624 Philips Sli Series Linear AcceleratorPhilips Medical Systems, Inc. · 1996 · Recalled
  2. 1 generation backK082122 MLCI2Elekta, Ltd. · 2008 · Recalled
  3. This deviceK121328 AgilityElekta, Ltd. · 2012
  4. Descendant · cites it directly · 36 of its ownK123808 AgilityElekta, Ltd. · 2013 · Recalled
  5. Descendant · 2 generations on · 31 of its ownK140528 True Beam, True Beam STX, EdgeVarian Medical Systems, Inc. · 2014 · Recalled
  6. Descendant · 3 generations on · 25 of its ownK143224 TrueBeam-TrueBeam STx-EdgeVarian Medical Systems, Inc. · 2014 · Recalled
  7. Descendant · 4 generations on · 23 of its ownK162472 TrueBeam-TrueBeam STx-EdgeVarian Medical Systems, Inc. · 2017
  8. Descendant · 5 generations on · 15 of its ownK170817 HalcyonVarian Medical Systems, Inc. · 2017

Showing 5 of 37 descendants. See all descendants

Descendants cleared per year

2013: 12014: 22015: 22016: 12017: 42018: 32019: 12020: 22021: 42022: 22023: 72024: 22025: 22026: 4
2013high 7 (2023) · low 1 (2013)2026

Most influential descendants

Showing 5 of 37 · See all
  • K123808AgilityElekta, Ltd. · 2013 · cites it directly36descendants
  • K140528True Beam, True Beam STX, EdgeVarian Medical Systems, Inc. · 2014 · 2 generations on31descendants
  • K143224TrueBeam-TrueBeam STx-EdgeVarian Medical Systems, Inc. · 2014 · 3 generations on25descendants
  • K162472TrueBeam-TrueBeam STx-EdgeVarian Medical Systems, Inc. · 2017 · 4 generations on23descendants
  • K170817HalcyonVarian Medical Systems, Inc. · 2017 · 5 generations on15descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K963624Philips Sli Series Linear AcceleratorPhilips Medical Systems, Inc.19962
−1K082122MLCI2Elekta, Ltd.20081
0 (this device)K121328AgilityElekta, Ltd.2012—
+1K123808AgilityElekta, Ltd.20132
+2K140528True Beam, True Beam STX, EdgeVarian Medical Systems, Inc.20142
+3K143224TrueBeam-TrueBeam STx-EdgeVarian Medical Systems, Inc.20141
+4K162472TrueBeam-TrueBeam STx-EdgeVarian Medical Systems, Inc.2017—
+5K170817HalcyonVarian Medical Systems, Inc.2017—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
201420
201520
201610
201740
201830
201910
202020
202140
202220
202370
202420
202520
202640
FIG. 1 — Lineage of K121328, 1996–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IYE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K082122MLCI2
Summary page 1
…ollimator Classification Name: Medical Linear Accelerator Accessory; 21CFR 892.5050 Product Code: lYE Predicate Device: MLCi2 (082122) Product Description: This Traditional 510(k) describes the addition of the new Agility multileaf collimator beam limit…
Elekta, Ltd.20081

What descends from it? (37)

See all 37

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K123808AgilityElekta, Ltd.201336

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 25, 2013Class IIZ-1084-2014Software designDuring the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value.
Apr 30, 2013Class IIZ-1402-2013Nonconforming Material/ComponentUnexpected movement during beam delivery may not be detected if there are incorrect static tolerances for Linac Geometric Axes loaded in Integrity R3.0.
Jul 27, 2012Class IIZ-2340-2012Error in labelingInformation released with the product Agility did not originally contain information on compatibility with the add-on Dynamic Multileaf Collimator Apex.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 2 companies; a new generation was cleared every 11.7 years on median.
  • 2 of the 2 ancestors have a recall of any class.
  • Of its descendants, K123808 (Elekta, Ltd., 2013) has gone furthest, with 36 descendants of its own.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—

What is it indicated for?

The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to the anatomy to be exposed. The associated Integrity R3.0 software is the interface and control software for the Elekta medical digital linear accelerator and is intended to assist a licensed practitioner in the delivery of radiation to defined target volumes (e.g. lesions, arterio-venous malformations, malignant and benign…

Read the full statement (108 words)

The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to the anatomy to be exposed.

The associated Integrity R3.0 software is the interface and control software for the Elekta medical digital linear accelerator and is intended to assist a licensed practitioner in the delivery of radiation to defined target volumes (e.g. lesions, arterio-venous malformations, malignant and benign tumors), whilst sparing surrounding normal tissue and critical organs from excess radiation. It is intended to be used for single or multiple fractions, delivered as static and/or dynamic beams of radiation, in all areas of the body where such treatment is indicated.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K121328: Agility.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K121328

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026