K121663510(k) TraditionalCleared Dec 5, 2012183-day reviewClass II21 CFR 892.5050

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Elekta, Ltd. · IYE Accelerator, Linear, Medical

Cleared on Dec 5, 2012 as substantially equivalent to K111332 (Elekta, Ltd., 2012) and 2 other predicates after a 183-day review. Its family tree reaches back 9 generations to 1996. 4 of the 88 devices in its family, counting itself, have a Class I recall. 1 later device descends from it.

Decided
Dec 5, 2012
Review
183 days
Class
Class II
Product code
IYE
Recalls
None
Predicates
3

Its longest line of descent runs back 9 generations to K954874 (1996). 86 ancestors in all, the oldest K851659 (1985). 1 later device descends from it, 1 citing it directly. 4 of the 88 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2008: 2 other ancestors2006: 2 other ancestors2004: 6 other ancestors2003: 6 other ancestors2002: 4 other ancestors2001: 8 other ancestors2000: 6 other ancestors, 1 with a Class I recall11999: 5 other ancestors1998: 5 other ancestors, 2 with a Class I recall21996: 7 other ancestors, 1 with a Class I recall12005: 1 other ancestor1997: 5 other ancestors1994: 5 other ancestors1992: 3 other ancestors1995: 2 other ancestors1993: 2 other ancestors1985: 1 other ancestor1991: 4 other ancestors1987: 1 other ancestor1990: 1 other ancestor1989: 1 other ancestor2014: 1 descendant clearedK954874 · Spectef Protocol · GE Medical Systems · 1996K980914 · Emory Cardiac Tool Box · Emory University · 1998K992450 · Emory Cardiac Toolbox* 2.0 Executing on Nuclear Medicine Computers · Emory University · 1999K014033 · Emory Cardiac Toolbox 2.1 Executing on Nuclear Medicine Computers · Syntermed, Inc. · 2002K040141 · Emory Cardiac Toolbox Version 2.6 · Syntermed, Inc. · 2004K052379 · Mim 3.5 (Circa) · Mimvista Corp. · 2005K060816 · Mim 4.0 (Neuro · Mimvista Corp. · 2006 · recalledK071964 · Mim 4.1 (Seastar) · Mimvista Corp. · 2007 · recalledK111332 · Clarity Obp System · Elekta, Ltd. · 20121990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 10 other ancestors, 3 with a Class I recall33 generations back: 16 other ancestors, 1 with a Class I recall14 generations back: 15 other ancestors5 generations back: 13 other ancestors6 generations back: 12 other ancestors7 generations back: 9 other ancestors 1 generation forward: 1 descendant K111332 · Clarity Obp System · Elekta, Ltd. · 2012K071964 · Mim 4.1 (Seastar) · Mimvista Corp. · 2007 · recalledK060816 · Mim 4.0 (Neuro · Mimvista Corp. · 2006 · recalledK052379 · Mim 3.5 (Circa) · Mimvista Corp. · 2005K040141 · Emory Cardiac Toolbox Version 2.6 · Syntermed, Inc. · 2004K014033 · Emory Cardiac Toolbox 2.1 Executing on Nuclear Medicine Computers · Syntermed, Inc. · 2002K992450 · Emory Cardiac Toolbox* 2.0 Executing on Nuclear Medicine Computers · Emory University · 1999K980914 · Emory Cardiac Tool Box · Emory University · 1998K954874 · Spectef Protocol · GE Medical Systems · 1996−50
  1. Root · 9 generations backK954874 Spectef ProtocolGE Medical Systems · 1996
  2. 5 generations between
  3. 3 generations backK060816 Mim 4.0 (NeuroMimvista Corp. · 2006 · Recalled
  4. 2 generations backK071964 Mim 4.1 (Seastar)Mimvista Corp. · 2007 · Recalled
  5. 1 generation backK111332 Clarity Obp SystemElekta, Ltd. · 2012
  6. This deviceK121663 ClarityElekta, Ltd. · 2012
  7. Descendant · cites it directly · 0 of its ownK141855 ClarityElekta, Ltd. · 2014
Show all 9 generations
  1. Root · 9 generations backK954874 Spectef ProtocolGE Medical Systems · 1996
  2. 8 generations backK980914 Emory Cardiac Tool BoxEmory University · 1998
  3. 7 generations backK992450 Emory Cardiac Toolbox* 2.0 Executing on Nuclear Medicine ComputersEmory University · 1999
  4. 6 generations backK014033 Emory Cardiac Toolbox 2.1 Executing on Nuclear Medicine ComputersSyntermed, Inc. · 2002
  5. 5 generations backK040141 Emory Cardiac Toolbox Version 2.6Syntermed, Inc. · 2004
  6. 4 generations backK052379 Mim 3.5 (Circa)Mimvista Corp. · 2005
  7. 3 generations backK060816 Mim 4.0 (NeuroMimvista Corp. · 2006 · Recalled
  8. 2 generations backK071964 Mim 4.1 (Seastar)Mimvista Corp. · 2007 · Recalled
  9. 1 generation backK111332 Clarity Obp SystemElekta, Ltd. · 2012
  10. This deviceK121663 ClarityElekta, Ltd. · 2012

Class I recalls in this family

Most influential descendants

Showing 1 of 1 · See all
  • K141855ClarityElekta, Ltd. · 2014 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−9K954874Spectef ProtocolGE Medical Systems1996—
−8K980914Emory Cardiac Tool BoxEmory University1998—
−7K992450Emory Cardiac Toolbox* 2.0 Executing on Nuclear Medicine ComputersEmory University1999—
−6K014033Emory Cardiac Toolbox 2.1 Executing on Nuclear Medicine ComputersSyntermed, Inc.2002—
−5K040141Emory Cardiac Toolbox Version 2.6Syntermed, Inc.2004—
−4K052379Mim 3.5 (Circa)Mimvista Corp.2005—
−3K060816Mim 4.0 (NeuroMimvista Corp.20061
−2K071964Mim 4.1 (Seastar)Mimvista Corp.20072
−1K111332Clarity Obp SystemElekta, Ltd.2012—
0 (this device)K121663ClarityElekta, Ltd.2012—
+1K141855ClarityElekta, Ltd.2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198510
198710
198910
199010
199140
199230
199320
199450
199520
199671
199750
199852
199950
200061
200180
200240
200360
200460
200510
200620
200820

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201410
FIG. 1 — Lineage of K121663, 1996–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYE device

  1. #1 K252977 Halcyon, Ethos Radiotherapy System (5.0)score 18.8 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K111332Clarity Obp System
Summary page 1
…l equivalence. Predicate Device Information Clarity® is substantially equivalent to the following legally marketed devices in the United States: o (Clarity® OBP System {K111332, Jan 30, 2012; Product Codes: IYE, IWB, KPQ) — Elekta Ltd. e Calypso® 4D Localization System (K0O60906, Jul 28, 2006 and K080726,…
Elekta, Ltd.2012—
*K060906Calypso 4D Localization System
Summary page 3
…nt positioning during treatment (i.e., intra-fractionally) is substantially equivalent to that previously cleared for marketing with the Calypso® 4D Localization System (K060906, K080726). Both devices provide 3D localization information to assist with patient positioning prior to treatment and monitor the…
Calypso Medical Technologies, Inc.2006—
*K080726Calypso 4D Localization System
Summary page 1
…e United States: o (Clarity® OBP System {K111332, Jan 30, 2012; Product Codes: IYE, IWB, KPQ) — Elekta Ltd. e Calypso® 4D Localization System (K0O60906, Jul 28, 2006 and K080726, May 14, 2008; Product Code IYE} — Calypso® Medical Technologies, Inc. . _intended Use / Indications for Use Clority® is indicat…
Calypso Medical Technologies, Inc.2008—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K141855ClarityElekta, Ltd.20140

Has anything in its family been recalled? (4 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · 2 generations back · ◆ in the 2000 ancestors bar of FIG. 12 Class I recalls
  • K983529Instatrak 3000Visualization Technology, Inc. · 1998 · 2 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall
  • K960330InstatrakVisualization Technology, Inc. · 1996 · 2 generations back · ◆ in the 1996 ancestors bar of FIG. 11 Class I recall
  • K981998Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · 3 generations back · ◆ in the 1998 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 16 years and passes through 5 companies; a new generation was cleared every 1.8 years on median.
  • 17 of the 86 ancestors have a recall of any class; 4 have a Class I recall.
  • Of its descendants, K141855 (Elekta, Ltd., 2014) has gone furthest, with 0 descendants of its own.

How does its category compare?

All IYE devices

IYE Accelerator, Linear, Medical: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · IYE

12in 2025

2007: 232008: 142009: 202010: 232011: 112012: 282013: 182014: 202015: 202016: 212017: 232018: 212019: 192020: 202021: 192022: 122023: 182024: 92025: 122026 (partial year): 14
2007high 28 (2012) · low 9 (2024)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYE

170in 2025

2007: 44 days median2008: 69 days median2009: 63 days median2010: 74 days median2011: 94 days median2012: 106 days median2013: 88 days median2014: 100 days median2015: 108 days median2016: 97 days median2017: 94 days median2018: 58 days median2019: 83 days median2020: 67 days median2021: 86 days median2022: 172 days median2023: 129 days median2024: 239 days median2025: 170 days median2026 (partial year): 120 days median
2007high 239 (2024) · low 44 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYE

1in 2025

2007: 782008: 492009: 222010: 272011: 732012: 202013: 112014: 142015: 82016: 72017: 72018: 72019: 22020: 42021: 52022: 32023: 72024: 12025: 12026 (partial year): 4
2007high 78 (2007) · low 1 (2024)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IYE

101in 2025

2007: 1552008: 1142009: 1742010: 2462011: 2802012: 1282013: 962014: 1242015: 1002016: 972017: 1252018: 852019: 912020: 742021: 682022: 662023: 1282024: 812025: 101
2007high 280 (2011) · low 66 (2022)2025

MAUDE reports naming product code IYE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007234478155
2008146949114
2009206322174
2010237427246
2011119473280
20122810620128
201318881196
20142010014124
2015201088100
20162197797
201723947125
20182158785
20191983291
20202067474
20211986568
202212172366
2023181297128
20249239181
2025121701101
2026141204—

What is it indicated for?

Clarity® is indicated for use in external beam radiation therapy, to provide 3D ultrasound and hybrid imaging of soft tissue anatomy to support radiation therapy simulation and treatment planning, and to guide patient positioning prior to the delivery of treatment.

Clarity® may also be used with an Autoscan Probe for transperineal ultrasound (TPUS) imaging, to continuously monitor the motion of the prostate and to accurately guide patient positioning during the delivery of treatment (i.e., intra-fractionally).

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K121663: Clarity.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K121663

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026