510(k) Report

510(k) · Cleared

Recalled · Class II · 2014

Ariatele Telemetry Transmitter (96281)

K121480510(k) SpecialCleared Jun 6, 201219-day reviewClass II21 CFR 870.1025Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Spacelabs Healthcare · MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)

Cleared on Jun 6, 2012 as substantially equivalent to K983996 (Spacelabs Medical, Inc., 1999) after a 19-day review. Its family tree reaches back 4 generations to 1994. It has been recalled 2 times.

Decided
Jun 6, 2012
Review
19 days
Class
Class II
Product code
MHX
Recalls
2
Predicates
1

Its longest line of descent runs back 4 generations to K933143 (1994). 18 ancestors in all, the oldest K781836 (1978). No later device cites it as a predicate yet.

Other ancestors, by year cleared 1996: 1 other ancestor1995: 3 other ancestors1994: 2 other ancestors1986: 1 other ancestor1978: 1 other ancestor1997: 1 other ancestor1990: 2 other ancestors1989: 2 other ancestors1991: 1 other ancestorK933143 · Model Eli 300 Electrocardiograph · Mortara Instrument, Inc. · 1994K962970 · Spacelabs Medical 12-Lead Diagnostic Ecg Analysis Option · Spacelabs Medical, Inc. · 1997K972502 · Spacelabs Medical Integrated Multiparameter Module 90496 · Spacelabs Medical, Inc. · 1998 · recalledK983996 · Spacelabs Medical Ultraview Digital Telemtry System · Spacelabs Medical, Inc. · 19991980199020002010
  1. Root · 4 generations backK933143 Model Eli 300 ElectrocardiographMortara Instrument, Inc. · 1994
  2. 3 generations backK962970 Spacelabs Medical 12-Lead Diagnostic Ecg Analysis OptionSpacelabs Medical, Inc. · 1997
  3. 2 generations backK972502 Spacelabs Medical Integrated Multiparameter Module 90496Spacelabs Medical, Inc. · 1998 · Recalled
  4. 1 generation backK983996 Spacelabs Medical Ultraview Digital Telemtry SystemSpacelabs Medical, Inc. · 1999
  5. This deviceK121480 Ariatele Telemetry Transmitter (96281)Spacelabs Healthcare · 2012
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K933143Model Eli 300 ElectrocardiographMortara Instrument, Inc.1994—
−3K962970Spacelabs Medical 12-Lead Diagnostic Ecg Analysis OptionSpacelabs Medical, Inc.1997—
−2K972502Spacelabs Medical Integrated Multiparameter Module 90496Spacelabs Medical, Inc.19983
−1K983996Spacelabs Medical Ultraview Digital Telemtry SystemSpacelabs Medical, Inc.1999—
0 (this device)K121480Ariatele Telemetry Transmitter (96281)Spacelabs Healthcare2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197810
198610
198920
199020
199110
199420
199530
199610
199710
FIG. 1 — Lineage of K121480, 1994–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MHX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MHX device

  1. #1 K251702 IntelliVue Patient Monitor MX750; IntelliVue…score 23.0 · 40 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K983996Spacelabs Medical Ultraview Digital Telemtry System
Summary page 1
…Predicate Predicate Equivalent Devices: 5 10(k) Number Manufacturer/ Model Spacelabs Healthcare K983996 Spacelabs Medical, Inc. AriaTele Telemetry Spacelabs Medical…
Spacelabs Medical, Inc.1999—

Also cleared against K983996: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 2, 2014Class IIZ-2586-2014Software designThe transmitter display and the Telemetry Central Station display may show a SpO2 value when the sensor is not connected to the patient or in certain sensor failed conditions. This is an expansion of recall Z-0871-2014 because additional…
Dec 19, 2013Class IIZ-0871-2014Software designThe AriaTele Model 96281 with SpO2 monitoring Option C is recalled because the AriaTele (transmitter) display and the Telemetry Central Station display may show a SpO2 value when the sensor is not connected to the patient or in certain…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 3 companies; a new generation was cleared every 2.9 years on median.
  • 2 of the 18 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K983996.

How does its category compare?

All MHX devices

MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · MHX

8in 2025

2007: 152008: 162009: 132010: 202011: 122012: 162013: 212014: 132015: 142016: 132017: 142018: 112019: 72020: 112021: 52022: 152023: 82024: 52025: 82026 (partial year): 14
2007high 21 (2013) · low 5 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MHX

192in 2025

2007: 96 days median2008: 58 days median2009: 57 days median2010: 90 days median2011: 81 days median2012: 88 days median2013: 120 days median2014: 248 days median2015: 203 days median2016: 149 days median2017: 134 days median2018: 203 days median2019: 243 days median2020: 196 days median2021: 195 days median2022: 189 days median2023: 229 days median2024: 181 days median2025: 192 days median2026 (partial year): 205 days median
2007high 248 (2014) · low 57 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MHX

21so far in 2026, the most since 2007

2007: 32008: 22009: 72010: 132011: 162012: 112013: 142014: 62015: 52016: 72017: 62018: 122019: 62022: 62023: 32024: 32025: 152026 (partial year): 21
2007high 21 (2026) · low 0 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MHX

3,761in 2025

2007: 1312008: 2322009: 3822010: 9552011: 9492012: 6762013: 1,9972014: 1,6092015: 1,2842016: 1,1052017: 1,1152018: 2,1572019: 1,4622020: 1,7142021: 2,4232022: 1,4702023: 2,2542024: 2,8142025: 3,761
2007high 3,761 (2025) · low 131 (2007)2025

MAUDE reports naming product code MHX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715963131
200816582232
200913577382
2010209013955
2011128116949
2012168811676
201321120141,997
20141324861,609
20151420351,284
20161314971,105
20171413461,115
201811203122,157
2019724361,462
20201119601,714
2021519502,423
20221518961,470
2023822932,254
2024518132,814
20258192153,761
20261420521—

What is it indicated for?

The Spacelabs Healthcare AriaTele Telemetry Transmritter (96281), when used in conjunction with a Spacelabs Healthcare Ultraview patient monitor and telemetry receiver, provides a means for the continuous monitoring of electrocardiographic signals in order to detect abnormal cardiac rhythms, including life-threatening events such as high and low heart rates, asystole, and ventricular fibrillation. Optionally, on adult patients, additional abnormal…

Read the full statement (117 words)

The Spacelabs Healthcare AriaTele Telemetry Transmritter (96281), when used in conjunction with a Spacelabs Healthcare Ultraview patient monitor and telemetry receiver, provides a means for the continuous monitoring of electrocardiographic signals in order to detect abnormal cardiac rhythms, including life-threatening events such as high and low heart rates, asystole, and ventricular fibrillation. Optionally, on adult patients, additional abnormal cardiac rhythms, such as ventricular runs, tachycardia, and ST segment deviations are detected.

The 96281 also provides a means for both continuous and episodic monitoring of pulse blood oxygen saturation signals in order to detect desaturation caused by abnormal pulmonary! circulatory functions.

The 96281 is intended for use with either adult or neonatal patient populations in a hospital environment.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K121480: Ariatele Telemetry Transmitter (96281).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K121480

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026