510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

Imager II Angiographic Catheter

K121694510(k) TraditionalCleared Sep 14, 201299-day reviewClass II21 CFR 870.1200Recalls 4 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Boston Scientific Corp · DQO Catheter, Intravascular, Diagnostic

Cleared on Sep 14, 2012 as substantially equivalent to K011664 (Boston Scientific/Medi-Tech, 2001) and 1 other predicate after a 99-day review. Its family tree reaches back 1 generation to 2001. 2 of the 5 devices in its family, counting itself, have a Class I recall. 2 later devices descend from it. It has been recalled 4 times, including 2 Class I recalls.

Decided
Sep 14, 2012
Review
99 days
Class
Class II
Product code
DQO
Recalls
4 (2 Class I)
Predicates
2

Its longest line of descent runs back 1 generation to K011664 (2001). 2 ancestors in all. 2 later devices descend from it, 1 citing it directly. 2 of the 5 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2002: 1 other ancestor2015: 1 descendant cleared, 1 with a Class I recall12016: 1 descendant clearedK011664 · Imager II Angiographic Catheter · Boston Scientific/Medi-Tech · 2001 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 1 descendant, 1 with a Class I recall12 generations forward: 1 descendant K011664 · Imager II Angiographic Catheter · Boston Scientific/Medi-Tech · 2001 · recalled0
  1. Root · 1 generation backK011664 Imager II Angiographic CatheterBoston Scientific/Medi-Tech · 2001 · Recalled
  2. This deviceK121694 Imager II Angiographic CatheterBoston Scientific Corp · 2012
  3. Descendant · cites it directly · 1 of its ownK133130 Beacon Tip Torcon NB Advantage Catheter, Torcon NB Advantage CatheterCook, Inc. · 2015 · Recalled (Class I)
  4. Descendant · 2 generations on · 0 of its ownK161822 Torcon NB Advantage CathetersCook Incorporated · 2016

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K133130Beacon Tip Torcon NB Advantage Catheter, Torcon NB Advantage CatheterCook, Inc. · 2015 · cites it directly1descendant
  • K161822Torcon NB Advantage CathetersCook Incorporated · 2016 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K011664Imager II Angiographic CatheterBoston Scientific/Medi-Tech20012
0 (this device)K121694Imager II Angiographic CatheterBoston Scientific Corp2012—
+1K133130Beacon Tip Torcon NB Advantage Catheter, Torcon NB Advantage CatheterCook, Inc.2015Class I
+2K161822Torcon NB Advantage CathetersCook Incorporated2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201511
201610
FIG. 1 — Lineage of K121694, 2001–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQO device

  1. #1 K254278 Arterial Pressure Monitoring Set/Trayscore 14.4 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K011664Imager II Angiographic Catheter
Summary page 1
…Diagnostic Intravascular Catheter Classification: 11 Product Code: 000/ 21 CFR 870.1200 Predicate Device: Imager TM 11 Angiographic Catheter (K01 1664, September 07, 2001) and Renegade STO 18 Microcatheter (K023681 November 26, 2002). Device Description: The 5F Selective Imager 1…
Boston Scientific/Medi-Tech20012
*K023681Renegade STC 18 Microcatheter
Summary page 1
…oduct Code: 000/ 21 CFR 870.1200 Predicate Device: Imager TM 11 Angiographic Catheter (K01 1664, September 07, 2001) and Renegade STO 18 Microcatheter (K023681 November 26, 2002). Device Description: The 5F Selective Imager 11Angiographic catheters without sideholes are sterile, single us…
Boston Scientific Corp20021

Also cleared against K011664: 2 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K133130Beacon Tip Torcon NB Advantage Catheter, Torcon NB Advantage Catheter Class ICook, Inc.20151

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 24, 2026Class IZ-3214-2026Process controlTip may fracture or detach, leading to device fragments remaining within the patient or migrating through the vasculature. Patient impact may include prolonged procedure or life-threatening embolism.
Feb 11, 2020Class IZ-1555-2020Environmental controlPotential for tip detachment of Imager II 5F Angiographic Catheters
Dec 15, 2016Class IIZ-2645-2017Employee errorA field report indicated some units within this lot were missing the bottom pouch seal. The compromised seal is completely missing and is obvious to the user.
Dec 13, 2016Class IIZ-1781-2017Employee errorOne lot of Imager II Angiographic Catheters is being recalled due to packaging mislabeling. The product may be labeled as Imager II Contra 2 curve catheters, when in fact the package contains Contra curve catheters.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K133130Beacon Tip Torcon NB Advantage Catheter, Torcon NB Advantage CatheterCook, Inc. · 2015 · Cites it directly · ◆ in the 2015 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • 2 of the 2 ancestors have a recall of any class.
  • Of its descendants, K133130 (Cook, Inc., 2015) has gone furthest, with 1 descendant of its own.
  • 2 other devices also cite its predicate K011664.

How does its category compare?

All DQO devices

DQO Catheter, Intravascular, Diagnostic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · DQO

2in 2025

2007: 82008: 32009: 82010: 72011: 62012: 152013: 22014: 32015: 102016: 42017: 102018: 102019: 62020: 32021: 12022: 42023: 42024: 42025: 22026 (partial year): 3
2007high 15 (2012) · low 1 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQO

148in 2024

2007: 106 days median2008: 80 days median2009: 58 days median2010: 137 days median2011: 74 days median2012: 94 days median2014: 109 days median2015: 209 days median2016: 89 days median2017: 176 days median2018: 234 days median2019: 113 days median2020: 30 days median2022: 181 days median2023: 73 days median2024: 148 days median2026 (partial year): 85 days median
2007high 234 (2018) · low 30 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQO

1in 2023

2007: 142008: 42009: 72010: 112011: 72012: 22013: 12015: 1, 1 Class I2016: 2, 1 Class I2017: 12018: 22019: 712020: 4, 4 Class I2022: 12023: 1, 1 Class I2026 (partial year): 24, 4 Class I
2007high 71 (2019) · low 0 (2014)2026

Class I (11 since 2007) Class II/III not classified

Adverse event reports per year · DQO

656in 2025

2007: 5642008: 7262009: 9332010: 5562011: 6252012: 5322013: 7532014: 6012015: 6282016: 6142017: 8732018: 7772019: 3862020: 3122021: 3382022: 3792023: 3912024: 5112025: 656
2007high 933 (2009) · low 312 (2020)2025

MAUDE reports naming product code DQO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007810614564
20083804726
20098587933
2010713711556
20116747625
201215942532
20132—1753
201431090601
2015102091 (1)628
20164892 (1)614
2017101761873
2018102342777
2019611371386
20203304 (4)312
20211—0338
202241811379
20234731 (1)391
202441480511
20252—0656
202638524 (4)—

What is it indicated for?

The Imager 11Angiographic Catheters are designed to provide a pathway for delivering contrast media to selected sites in the vascular system including the carotid arteries.

Additionally, the 5F Seleactive Imager 11Arigiographic Catheters without sideholes can be used for the controlled and selective delivery of the Boston Scientific lnterlockTh-35 Detachable Coils into the peripheral vasculature.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K121694: Imager II Angiographic Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K121694

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026