510(k) Report

510(k) · Cleared

Cleared

Spectralis Hra + Oct and Variants, Spectralis FA+ Oct, Spectralis Icga +Oct, Spectralis Oct Blue Peak, Spectralis Oct

K121993510(k) TraditionalCleared Oct 25, 2012111-day reviewClass II21 CFR 886.1570

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Heidelberg Engineering GmbH · OBO Tomography, Optical Coherence

Cleared on Oct 25, 2012 as substantially equivalent to K063191 (Heidelberg Engineering, 2006) and 2 other predicates after a 111-day review. Its family tree reaches back 5 generations to 1995. 15 later devices descend from it.

Decided
Oct 25, 2012
Review
111 days
Class
Class II
Product code
OBO
Recalls
None
Predicates
3

Its longest line of descent runs back 5 generations to K941234 (1995). 13 ancestors in all, the oldest K946058 (1995). 15 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2006: 1 other ancestor2003: 2 other ancestors2004: 1 other ancestor2002: 2 other ancestors2001: 1 other ancestor1995: 1 other ancestor2016: 1 descendant cleared2017: 1 descendant cleared2018: 4 descendants cleared2021: 1 descendant cleared2019: 1 descendant cleared2025: 2 descendants cleared2020: 1 descendant cleared2023: 1 descendant cleared2024: 2 descendants cleared2026: 1 descendant clearedK941234 · Non-Mydriatic Retinal Camera · Canon USA, Inc. · 1995K980246 · Non-Mydriatic Retinal Camera, Model CR6-45NM · Canon USA, Inc. · 1998K014274 · Non-Mydriatic Fundus Camera, Model NM-1000 · Nidek, Inc. · 2002K041367 · Topcon Model TRC-NW200 Non-Mydriatic Retinal Camera · Topcon Corp. · 2004K062295 · Iri Integrated Retinal Imager · Medivision Medical Imaging, Ltd. · 20061995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 3 other ancestors3 generations back: 3 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant3 generations forward: 3 descendants4 generations forward: 2 descendants5 generations forward: 2 descendants6 generations forward: 1 descendant7 generations forward: 1 descendant8 generations forward: 2 descendants9 generations forward: 2 descendants K062295 · Iri Integrated Retinal Imager · Medivision Medical Imaging, Ltd. · 2006K041367 · Topcon Model TRC-NW200 Non-Mydriatic Retinal Camera · Topcon Corp. · 2004K014274 · Non-Mydriatic Fundus Camera, Model NM-1000 · Nidek, Inc. · 2002K980246 · Non-Mydriatic Retinal Camera, Model CR6-45NM · Canon USA, Inc. · 1998K941234 · Non-Mydriatic Retinal Camera · Canon USA, Inc. · 1995−50+5
  1. Root · 5 generations backK941234 Non-Mydriatic Retinal CameraCanon USA, Inc. · 1995
  2. 4 generations backK980246 Non-Mydriatic Retinal Camera, Model CR6-45NMCanon USA, Inc. · 1998
  3. 3 generations backK014274 Non-Mydriatic Fundus Camera, Model NM-1000Nidek, Inc. · 2002
  4. 2 generations backK041367 Topcon Model TRC-NW200 Non-Mydriatic Retinal CameraTopcon Corp. · 2004
  5. 1 generation backK062295 Iri Integrated Retinal ImagerMedivision Medical Imaging, Ltd. · 2006
  6. This deviceK121993 Spectralis Hra + Oct and Variants, Spectralis FA+ Oct, Spectralis…Heidelberg Engineering GmbH · 2012
  7. Descendant · cites it directly · 14 of its ownK152205 Spectralis HRA + OCT and variants (e.g.s below), Spectralis FA+OCT…Heidelberg Engineering GmbH · 2016
  8. Descendant · 2 generations on · 13 of its ownK172649 Spectralis HRA + OCT and variants, Spectralis FA+OCT, Spectralis…Heidelberg Engineering GmbH · 2017
  9. Descendant · 3 generations on · 8 of its ownK181594 Spectralis HRA+OCT and variants with OCT Angiography ModuleHeidelberg Engineering GmbH · 2018 · Recalled
  10. Descendant · 4 generations on · 7 of its ownK182569 Spectralis HRA+OCT and variants with High Magnification ModuleHeidelberg Engineering GmbH · 2018
  11. Descendant · 5 generations on · 6 of its ownK192391 Spectralis HRA+OCT and VariantsHeidelberg Engineering GmbH · 2019

Showing 5 of 15 descendants. See all descendants

Descendants cleared per year

2016: 12017: 12018: 42019: 12020: 12021: 12023: 12024: 22025: 22026: 1
2016high 4 (2018) · low 0 (2022)2026

Most influential descendants

Showing 5 of 15 · See all
  • K152205Spectralis HRA + OCT and variants (e.g.s below), Spectralis FA+OCT, Spectralis ICGA+OCT…Heidelberg Engineering GmbH · 2016 · cites it directly14descendants
  • K172649Spectralis HRA + OCT and variants, Spectralis FA+OCT, Spectralis ICGA+OCT, Spectralis…Heidelberg Engineering GmbH · 2017 · 2 generations on13descendants
  • K181594Spectralis HRA+OCT and variants with OCT Angiography ModuleHeidelberg Engineering GmbH · 2018 · 3 generations on8descendants
  • K182569Spectralis HRA+OCT and variants with High Magnification ModuleHeidelberg Engineering GmbH · 2018 · 4 generations on7descendants
  • K192391Spectralis HRA+OCT and VariantsHeidelberg Engineering GmbH · 2019 · 5 generations on6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K941234Non-Mydriatic Retinal CameraCanon USA, Inc.1995—
−4K980246Non-Mydriatic Retinal Camera, Model CR6-45NMCanon USA, Inc.1998—
−3K014274Non-Mydriatic Fundus Camera, Model NM-1000Nidek, Inc.2002—
−2K041367Topcon Model TRC-NW200 Non-Mydriatic Retinal CameraTopcon Corp.2004—
−1K062295Iri Integrated Retinal ImagerMedivision Medical Imaging, Ltd.2006—
0 (this device)K121993Spectralis Hra + Oct and Variants, Spectralis FA+ Oct, Spectralis Icga +Oct…Heidelberg Engineering GmbH2012—
+1K152205Spectralis HRA + OCT and variants (e.g.s below), Spectralis FA+OCT, Spectralis…Heidelberg Engineering GmbH2016—
+2K172649Spectralis HRA + OCT and variants, Spectralis FA+OCT, Spectralis ICGA+OCT…Heidelberg Engineering GmbH2017—
+3K181594Spectralis HRA+OCT and variants with OCT Angiography ModuleHeidelberg Engineering GmbH20182
+4K182569Spectralis HRA+OCT and variants with High Magnification ModuleHeidelberg Engineering GmbH2018—
+5K192391Spectralis HRA+OCT and VariantsHeidelberg Engineering GmbH2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510
200110
200220
200320
200410
200610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201610
201710
201840
201910
202010
202110
202310
202420
202520
202610
FIG. 1 — Lineage of K121993, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OBO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OBO device

  1. #1 K250868 SPECTRALIS HRA+OCT and variantsscore 16.2 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K063191Heidelberg Retina Angiograph 2/Optical Coherence Tomograph, Model Hra 2/Oct
Summary page 2
…aucasian subjects. Substantial Equivalence Analysis The Spectralis HRA±OCT has the same intended use and basic technological characteristics as the Heidelberg HRA2/OCT (K063 191). The device has been updated to include new optical layout and components, new electronic components, new mechanical design, and…
Heidelberg Engineering2006—
*K030433Stratusoct with RNFL Normative Database
Summary page 2
…redicate devices incorporating similar technology features to the upgraded Spectralis device include: the STATUS OCT with Retinal Nerve Fiber Layer Normative Database (K(030433) for the similarity in having a normative database, and the IRI Integrated Retinal Imager (K062295) for the auto fluorescence int…
Carl Zeiss Meditec, Inc.20031
*K062295Iri Integrated Retinal Imager
Summary page 2
…de: the STATUS OCT with Retinal Nerve Fiber Layer Normative Database (K(030433) for the similarity in having a normative database, and the IRI Integrated Retinal Imager (K062295) for the auto fluorescence intended use. Performance - Bench Testing The Spectralis HRA+OCT has been tested according to lEG 606…
Medivision Medical Imaging, Ltd.2006—

Also cleared against K063191: 1 other device →

What descends from it? (15)

See all 15

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K152205Spectralis HRA + OCT and variants (e.g.s below), Spectralis FA+OCT, Spectralis ICGA+OCT…Heidelberg Engineering GmbH201614

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 17 years and passes through 5 companies; a new generation was cleared every 2.9 years on median.
  • 1 of the 13 ancestors has a recall of any class.
  • Of its descendants, K152205 (Heidelberg Engineering GmbH, 2016) has gone furthest, with 14 descendants of its own.
  • 1 other device also cites its predicate K063191.

How does its category compare?

All OBO devices

OBO Tomography, Optical Coherence: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · OBO

3in 2025

2007: 22008: 22009: 42010: 22011: 22012: 32013: 22014: 32015: 12016: 42017: 42018: 62019: 52020: 22021: 12022: 22023: 72024: 82025: 32026 (partial year): 2
2007high 8 (2024) · low 1 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OBO

125in 2025

2009: 151 days median2012: 269 days median2014: 152 days median2016: 125 days median2017: 118 days median2018: 87 days median2019: 249 days median2023: 262 days median2024: 157 days median2025: 125 days median
2007high 269 (2012) · low 87 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OBO

1in 2022

2010: 12011: 22012: 22018: 12021: 12022: 12026 (partial year): 1
2007high 2 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OBO

4in 2025

2013: 22016: 12018: 12021: 12023: 32024: 32025: 4
2007high 4 (2025) · low 0 (2007)2025

MAUDE reports naming product code OBO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20082—00
2009415100
20102—10
20112—20
2012326920
20132—02
2014315200
20151—00
2016412501
2017411800
201868711
2019524900
20202—00
20211—11
20222—10
2023726203
2024815703
2025312504
20262—1—

What is it indicated for?

The Spectralis HRA±OCT is a non-contact ophthalmic diagnostic imaging device. It is intended for viewing the posterior segment of the eye, including two- and three-dimensional imaging, cross-sectional imaging, fundus photography, and fluorescence imaging (fluorescein, indocyanine green and autofluorescence), and to perform measurements of ocular anatomy and ocular lesions. The device is indicated as an aid in the detection and management of various…

Read the full statement (138 words)

The Spectralis HRA±OCT is a non-contact ophthalmic diagnostic imaging device. It is intended for viewing the posterior segment of the eye, including two- and three-dimensional imaging, cross-sectional imaging, fundus photography, and fluorescence imaging (fluorescein, indocyanine green and autofluorescence), and to perform measurements of ocular anatomy and ocular lesions. The device is indicated as an aid in the detection and management of various ocular diseases including: age-related macular degeneration, macular edema, diabetic retinopathy, retinal and choroidal vascular diseases, glaucoma, and for viewing geographic atrophy as well as changes in the eye that result from neurodegenerative diseases. The Spectralis HRA+OCT includes a retinal nerve fiber layer thickness normative database, which is used to quantitatively compare the retinal nerve fiber layer in the human retina to a database of Caucasian normal subjects; the classification result is valid only for Caucasian subjects.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K121993: Spectralis Hra + Oct and Variants, Spectralis FA+ Oct, Spectralis Icga +Oct, Spectralis Oct Blue Peak, Spectralis Oct.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K121993

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026