510(k) Report

510(k) · Cleared

Cleared

Spectralis HRA + OCT and variants (e.g.s below), Spectralis FA+OCT, Spectralis ICGA+OCT, Spectralis OCT Blue Peak, Spectralis OCT ith Multicolor

K152205510(k) TraditionalCleared May 6, 2016274-day reviewClass II21 CFR 886.1570

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Heidelberg Engineering GmbH · OBO Tomography, Optical Coherence

Cleared on May 6, 2016 as substantially equivalent to K101223 (Heidelberg Engineering, 2010) and 1 other predicate after a 274-day review. Its family tree reaches back 6 generations to 1995. 14 later devices descend from it.

Decided
May 6, 2016
Review
274 days
Class
Class II
Product code
OBO
Recalls
None
Predicates
2

Its longest line of descent runs back 6 generations to K941234 (1995). 15 ancestors in all, the oldest K946058 (1995). 14 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2012: 1 other ancestor2006: 1 other ancestor2003: 2 other ancestors2004: 1 other ancestor2002: 2 other ancestors2001: 1 other ancestor1995: 1 other ancestor2017: 1 descendant cleared2018: 4 descendants cleared2021: 1 descendant cleared2019: 1 descendant cleared2025: 2 descendants cleared2020: 1 descendant cleared2023: 1 descendant cleared2024: 2 descendants cleared2026: 1 descendant clearedReference device: K072259 · Ophthalmoscope F-10 · Nidek Co., Ltd. · 2008Reference device: K111628 · P200T · Optos Plc. · 2011K941234 · Non-Mydriatic Retinal Camera · Canon USA, Inc. · 1995K980246 · Non-Mydriatic Retinal Camera, Model CR6-45NM · Canon USA, Inc. · 1998K014274 · Non-Mydriatic Fundus Camera, Model NM-1000 · Nidek, Inc. · 2002K041367 · Topcon Model TRC-NW200 Non-Mydriatic Retinal Camera · Topcon Corp. · 2004K062295 · Iri Integrated Retinal Imager · Medivision Medical Imaging, Ltd. · 2006K101223 · Spectralis Hra+Oct, Spectralis FA_Oct, Spectralis Icga+ Oct, Spectralis Hra, Spectralis FA · Heidelberg Engineering · 20101995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 2 other ancestors3 generations back: 3 other ancestors4 generations back: 3 other ancestors 1 generation forward: 1 descendant2 generations forward: 3 descendants3 generations forward: 2 descendants4 generations forward: 2 descendants5 generations forward: 1 descendant6 generations forward: 1 descendant7 generations forward: 2 descendants8 generations forward: 2 descendants K101223 · Spectralis Hra+Oct, Spectralis FA_Oct, Spectralis Icga+ Oct, Spectralis Hra, Spectralis FA · Heidelberg Engineering · 2010K062295 · Iri Integrated Retinal Imager · Medivision Medical Imaging, Ltd. · 2006K041367 · Topcon Model TRC-NW200 Non-Mydriatic Retinal Camera · Topcon Corp. · 2004K014274 · Non-Mydriatic Fundus Camera, Model NM-1000 · Nidek, Inc. · 2002K980246 · Non-Mydriatic Retinal Camera, Model CR6-45NM · Canon USA, Inc. · 1998K941234 · Non-Mydriatic Retinal Camera · Canon USA, Inc. · 1995−50+5
  1. Root · 6 generations backK941234 Non-Mydriatic Retinal CameraCanon USA, Inc. · 1995
  2. 5 generations backK980246 Non-Mydriatic Retinal Camera, Model CR6-45NMCanon USA, Inc. · 1998
  3. 4 generations backK014274 Non-Mydriatic Fundus Camera, Model NM-1000Nidek, Inc. · 2002
  4. 3 generations backK041367 Topcon Model TRC-NW200 Non-Mydriatic Retinal CameraTopcon Corp. · 2004
  5. 2 generations backK062295 Iri Integrated Retinal ImagerMedivision Medical Imaging, Ltd. · 2006
  6. 1 generation backK101223 Spectralis Hra+Oct, Spectralis FA_Oct, Spectralis Icga+ Oct…Heidelberg Engineering · 2010
  7. This deviceK152205 Spectralis HRA + OCT and variants (e.g.s below), Spectralis FA+OCT…Heidelberg Engineering GmbH · 2016
  8. Descendant · cites it directly · 13 of its ownK172649 Spectralis HRA + OCT and variants, Spectralis FA+OCT, Spectralis…Heidelberg Engineering GmbH · 2017
  9. Descendant · 2 generations on · 8 of its ownK181594 Spectralis HRA+OCT and variants with OCT Angiography ModuleHeidelberg Engineering GmbH · 2018 · Recalled
  10. Descendant · 3 generations on · 7 of its ownK182569 Spectralis HRA+OCT and variants with High Magnification ModuleHeidelberg Engineering GmbH · 2018
  11. Descendant · 4 generations on · 6 of its ownK192391 Spectralis HRA+OCT and VariantsHeidelberg Engineering GmbH · 2019
  12. Descendant · 5 generations on · 5 of its ownK201252 Spectralis HRA+OCT and variantsHeidelberg Engineering GmbH · 2020

Showing 5 of 14 descendants. See all descendants

Descendants cleared per year

2017: 12018: 42019: 12020: 12021: 12023: 12024: 22025: 22026: 1
2017high 4 (2018) · low 0 (2022)2026

Most influential descendants

Showing 5 of 14 · See all
  • K172649Spectralis HRA + OCT and variants, Spectralis FA+OCT, Spectralis ICGA+OCT, Spectralis…Heidelberg Engineering GmbH · 2017 · cites it directly13descendants
  • K181594Spectralis HRA+OCT and variants with OCT Angiography ModuleHeidelberg Engineering GmbH · 2018 · 2 generations on8descendants
  • K182569Spectralis HRA+OCT and variants with High Magnification ModuleHeidelberg Engineering GmbH · 2018 · 3 generations on7descendants
  • K192391Spectralis HRA+OCT and VariantsHeidelberg Engineering GmbH · 2019 · 4 generations on6descendants
  • K201252Spectralis HRA+OCT and variantsHeidelberg Engineering GmbH · 2020 · 5 generations on5descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K941234Non-Mydriatic Retinal CameraCanon USA, Inc.1995—
−5K980246Non-Mydriatic Retinal Camera, Model CR6-45NMCanon USA, Inc.1998—
−4K014274Non-Mydriatic Fundus Camera, Model NM-1000Nidek, Inc.2002—
−3K041367Topcon Model TRC-NW200 Non-Mydriatic Retinal CameraTopcon Corp.2004—
−2K062295Iri Integrated Retinal ImagerMedivision Medical Imaging, Ltd.2006—
−1K101223Spectralis Hra+Oct, Spectralis FA_Oct, Spectralis Icga+ Oct, Spectralis Hra…Heidelberg Engineering2010—
0 (this device)K152205Spectralis HRA + OCT and variants (e.g.s below), Spectralis FA+OCT, Spectralis…Heidelberg Engineering GmbH2016—
+1K172649Spectralis HRA + OCT and variants, Spectralis FA+OCT, Spectralis ICGA+OCT…Heidelberg Engineering GmbH2017—
+2K181594Spectralis HRA+OCT and variants with OCT Angiography ModuleHeidelberg Engineering GmbH20182
+3K182569Spectralis HRA+OCT and variants with High Magnification ModuleHeidelberg Engineering GmbH2018—
+4K192391Spectralis HRA+OCT and VariantsHeidelberg Engineering GmbH2019—
+5K201252Spectralis HRA+OCT and variantsHeidelberg Engineering GmbH2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510
200110
200220
200320
200410
200610
201210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
201840
201910
202010
202110
202310
202420
202520
202610
FIG. 1 — Lineage of K152205, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OBO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OBO device

  1. #1 K250868 SPECTRALIS HRA+OCT and variantsscore 16.2 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K101223Spectralis Hra+Oct, Spectralis FA_Oct, Spectralis Icga+ Oct, Spectralis Hra, Spectralis FA
Summary page 4
…ph Classification Name: Tomography, Optical Coherence (21 CFR 886.1570) Regulatory Class: II Product Code: OBO, MYC III. Predicate Device Heidelberg SPECTRALIS HRA+OCT, K101223, amended by K121993 Reference devices include other medical devices - with wide field objective: Nidek Ophthalmoscope F-10, K0…
Heidelberg Engineering2010—
*K121993Spectralis Hra + Oct and Variants, Spectralis FA+ Oct, Spectralis Icga +Oct, Spectralis…
Summary page 4
…me: Tomography, Optical Coherence (21 CFR 886.1570) Regulatory Class: II Product Code: OBO, MYC III. Predicate Device Heidelberg SPECTRALIS HRA+OCT, K101223, amended by K121993 Reference devices include other medical devices - with wide field objective: Nidek Ophthalmoscope F-10, K072259 - with ultr…
Heidelberg Engineering GmbH2012—
†K072259Ophthalmoscope F-10Nidek Co., Ltd.2008—
†K111628P200TOptos Plc.2011—

Also cleared against K101223: 2 other devices →

What descends from it? (14)

See all 14

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K172649Spectralis HRA + OCT and variants, Spectralis FA+OCT, Spectralis ICGA+OCT, Spectralis…Heidelberg Engineering GmbH201713

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 6 companies; a new generation was cleared every 3.9 years on median.
  • 1 of the 15 ancestors has a recall of any class.
  • Of its descendants, K172649 (Heidelberg Engineering GmbH, 2017) has gone furthest, with 13 descendants of its own.
  • 2 other devices also cite its predicate K101223.

How does its category compare?

All OBO devices

OBO Tomography, Optical Coherence: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · OBO

3in 2025

2007: 22008: 22009: 42010: 22011: 22012: 32013: 22014: 32015: 12016: 42017: 42018: 62019: 52020: 22021: 12022: 22023: 72024: 82025: 32026 (partial year): 2
2007high 8 (2024) · low 1 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OBO

125in 2025

2009: 151 days median2012: 269 days median2014: 152 days median2016: 125 days median2017: 118 days median2018: 87 days median2019: 249 days median2023: 262 days median2024: 157 days median2025: 125 days median
2007high 269 (2012) · low 87 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OBO

1in 2022

2010: 12011: 22012: 22018: 12021: 12022: 12026 (partial year): 1
2007high 2 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OBO

4in 2025

2013: 22016: 12018: 12021: 12023: 32024: 32025: 4
2007high 4 (2025) · low 0 (2007)2025

MAUDE reports naming product code OBO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20082—00
2009415100
20102—10
20112—20
2012326920
20132—02
2014315200
20151—00
2016412501
2017411800
201868711
2019524900
20202—00
20211—11
20222—10
2023726203
2024815703
2025312504
20262—1—

What is it indicated for?

The SPECTRALIS is a non-contact ophthalmic diagnostic imaging device. It is intended for viewing the posterior segment of the eye, including two- and three-dimensional imaging, cross-sectional imaging (SPECTRALIS HRA+OCT and SPECTRALIS OCT), fundus photography, fluorescence imaging (fluorescein angiography, indocyanine green angiography; SPECTRALIS HRA+OCT, SPECTRALIS HRA), autofluorescence imaging (SPECTRALIS HRA+OCT, SPECTRALIS HRA and SPECTRALIS OCT…

Read the full statement (158 words)

The SPECTRALIS is a non-contact ophthalmic diagnostic imaging device. It is intended for viewing the posterior segment of the eye, including two- and three-dimensional imaging, cross-sectional imaging (SPECTRALIS HRA+OCT and SPECTRALIS OCT), fundus photography, fluorescence imaging (fluorescein angiography, indocyanine green angiography; SPECTRALIS HRA+OCT, SPECTRALIS HRA), autofluorescence imaging (SPECTRALIS HRA+OCT, SPECTRALIS HRA and SPECTRALIS OCT BluePeak) and to perform measurements of ocular anatomy and ocular lesions. The device is indicated as an aid in the detection and management of various ocular diseases, including age-related macular degeneration, macular edema, diabetic retinopathy, retinal and choroidal vascular diseases, glaucoma, and for viewing geographic atrophy as well as changes in the eye that result from neurodegenerative diseases. The SPECTRALIS HRA+OCT and SPECTRALIS OCT include normative databases for retinal nerve fiber layer thickness and optic nerve head neuroretinal parameter measurements, which are used to quantitatively compare the retinal nerve fiber layer and neuroretinal rim in the human retina to values found in normal subjects.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K152205: Spectralis HRA + OCT and variants (e.g.s below), Spectralis FA+OCT, Spectralis ICGA+OCT, Spectralis OCT Blue Peak, Spectralis OCT ith Multicolor.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K152205

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026