510(k) · Cleared
ClearedBone Plus BCP Eagle Eye
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Megagen Implant Co., Ltd. · LYC Bone Grafting Material, Synthetic
Cleared on Aug 21, 2012 as substantially equivalent to K090950 (Megagen Implant Co., Ltd., 2010) after a 25-day review. Its family tree reaches back 6 generations to 1999.
- Decided
- Aug 21, 2012
- Review
- 25 days
- Class
- Class II
- Product code
- LYC
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 6 generations to K990131 (1999). 25 ancestors in all, the oldest P800035 (1981). No later device cites it as a predicate yet.
- Root · 6 generations backK990131 Pro Osteon Implant 500R Resorbable Bone Graft SubstituteInterpore Cross Intl. · 1999
- 5 generations backK020986 Mastergraft Resorbable CeramicMedtronic Sofamor Danek, Inc. · 2002
- 4 generations backK032268 MBCPBiomatlante · 2003
- 3 generations backK043005 Modification to MBCPBiomatlante · 2005
- 2 generations backK051885 MBCPBiomatlante · 2005
- 1 generation backK090950 Bone Plus BCPMegagen Implant Co., Ltd. · 2010
- This deviceK122240 Bone Plus BCP Eagle EyeMegagen Implant Co., Ltd. · 2012
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −6 | K990131 | Pro Osteon Implant 500R Resorbable Bone Graft Substitute | Interpore Cross Intl. | 1999 | — |
| −5 | K020986 | Mastergraft Resorbable Ceramic | Medtronic Sofamor Danek, Inc. | 2002 | — |
| −4 | K032268 | MBCP | Biomatlante | 2003 | — |
| −3 | K043005 | Modification to MBCP | Biomatlante | 2005 | — |
| −2 | K051885 | MBCP | Biomatlante | 2005 | — |
| −1 | K090950 | Bone Plus BCP | Megagen Implant Co., Ltd. | 2010 | — |
| 0 (this device) | K122240 | Bone Plus BCP Eagle Eye | Megagen Implant Co., Ltd. | 2012 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1981 | 1 | 0 |
| 1985 | 1 | 0 |
| 1987 | 1 | 0 |
| 1992 | 1 | 0 |
| 1993 | 1 | 0 |
| 1995 | 4 | 0 |
| 1996 | 1 | 0 |
| 1998 | 2 | 0 |
| 1999 | 1 | 0 |
| 2000 | 3 | 0 |
| 2004 | 2 | 0 |
| 2005 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in LYC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LYC device
- #1 K253317 Cytrans™ Elashieldscore 14.3 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K090950 | Bone Plus BCPSummary page 3…MegaGen Implant Co., Ltd. Predicate Devices The subject device is substantially equivalent to the following predicate devices: .Bone PlusTM BCP (K090950) manufactured by Megagen Implant Co., Ltd. Comparison to Predicate Devices Testing and other comparisons have established tha… | 2010 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 13 years and passes through 4 companies; a new generation was cleared every 2.1 years on median.
- None of the 25 ancestors has a recall.
How does its category compare?
All LYC devicesLYC Bone Grafting Material, Synthetic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.
FDA decisions per year · LYC
3in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LYC
267in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LYC
1in 2018
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · LYC
2in 2025
MAUDE reports naming product code LYC, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 13 | 80 | 0 | 23 |
| 2008 | 8 | 73 | 0 | 25 |
| 2009 | 14 | 100 | 1 | 25 |
| 2010 | 13 | 164 | 2 | 28 |
| 2011 | 9 | 224 | 0 | 21 |
| 2012 | 5 | 120 | 2 | 10 |
| 2013 | 2 | — | 0 | 12 |
| 2014 | 1 | — | 0 | 9 |
| 2015 | 0 | — | 0 | 14 |
| 2016 | 2 | — | 2 | 140 |
| 2017 | 0 | — | 0 | 19 |
| 2018 | 0 | — | 1 | 14 |
| 2019 | 0 | — | 0 | 15 |
| 2020 | 3 | 156 | 0 | 5 |
| 2021 | 0 | — | 0 | 12 |
| 2022 | 2 | — | 0 | 6 |
| 2023 | 0 | — | 0 | 9 |
| 2024 | 0 | — | 0 | 6 |
| 2025 | 3 | 267 | 0 | 2 |
| 2026 | 3 | 266 | 0 | — |
What is it indicated for?
Bone PlusTM BCP Eagle Eye is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
* Augmentation or reconstructive treatment of alveolar ridge
* Filling of periodontal defects
* Filling of defects after root resection, apicocectomy, and cystectomy
* Filling of extraction sockets to enhance preservation of the alveolar ridge
* Elevation of maxillary sinus floor
* Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)
* Filling of peni-implant defects in conjunction with products intended for Guided Bone Regeneration
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K122240 · printed Nov 6, 2026
Cite this page
510(k) Report. “K122240: Bone Plus BCP Eagle Eye.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K122240
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026