510(k) Report

510(k) · Cleared

Recalled · Class II · 2012

Modified Balloon Guide Catheter Model 90253, 90254

K122581510(k) TraditionalCleared Nov 21, 201290-day reviewClass II21 CFR 870.1250Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Concentric Medical, Inc. · DQY Catheter, Percutaneous

Cleared on Nov 21, 2012 as substantially equivalent to K112404 (Concentric Medical, Inc., 2012) after a 90-day review. Its family tree reaches back 4 generations to 2001. 15 later devices descend from it. It has been recalled once.

Decided
Nov 21, 2012
Review
90 days
Class
Class II
Product code
DQY
Recalls
1
Predicates
1

Its longest line of descent runs back 4 generations to K010954 (2001). 15 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2013: 1 descendant cleared2021: 3 descendants cleared2025: 1 descendant cleared2016: 1 descendant cleared2023: 1 descendant cleared2019: 2 descendants cleared2022: 1 descendant cleared2020: 2 descendants cleared2026: 2 descendants cleared2024: 1 descendant clearedK010954 · Concentric Balloon Guide Catheter, Model 90031 · Concentric Medical, Inc. · 2001 · recalledK021899 · Concentric Balloon Guide Catheter, Model 90072 · Concentric Medical, Inc. · 2002 · recalledK102657 · Concentric Balloon Guide Catheter · Concentric Medical, Inc. · 2010K112404 · Concentric Balloon Guide Catheter · Concentric Medical, Inc. · 2012 · recalled200020052010201520202025
  1. Root · 4 generations backK010954 Concentric Balloon Guide Catheter, Model 90031Concentric Medical, Inc. · 2001 · Recalled
  2. 3 generations backK021899 Concentric Balloon Guide Catheter, Model 90072Concentric Medical, Inc. · 2002 · Recalled
  3. 2 generations backK102657 Concentric Balloon Guide CatheterConcentric Medical, Inc. · 2010
  4. 1 generation backK112404 Concentric Balloon Guide CatheterConcentric Medical, Inc. · 2012 · Recalled
  5. This deviceK122581 Modified Balloon Guide Catheter Model 90253, 90254Concentric Medical, Inc. · 2012
  6. Descendant · cites it directly · 13 of its ownK131492 Modified Flowgate Balloon Guide CatheterConcentric Medical, Inc. · 2013
  7. Descendant · 2 generations on · 10 of its ownK153729 8F FlowGate2 Balloon Guide CatheterConcentric Medical · 2016 · Recalled
  8. Descendant · 2 generations on · 1 of its ownK242392 Path BGCCrossroads Neurovascular, Inc. · 2025
  9. Descendant · 3 generations on · 1 of its ownK212340 EMBOGUARD Balloon Guide CatheterNeuravi Limited · 2021
  10. Descendant · 3 generations on · 1 of its ownK193607 BOBBY Balloon Guide CatheterMicroVention, Inc. · 2020 · Recalled

Showing 5 of 15 descendants. See all descendants

Descendants cleared per year

2013: 12016: 12019: 22020: 22021: 32022: 12023: 12024: 12025: 12026: 2
2013high 3 (2021) · low 0 (2014)2026

Most influential descendants

Showing 5 of 15 · See all
  • K131492Modified Flowgate Balloon Guide CatheterConcentric Medical, Inc. · 2013 · cites it directly13descendants
  • K1537298F FlowGate2 Balloon Guide CatheterConcentric Medical · 2016 · 2 generations on10descendants
  • K242392Path BGCCrossroads Neurovascular, Inc. · 2025 · 2 generations on1descendant
  • K212340EMBOGUARD Balloon Guide CatheterNeuravi Limited · 2021 · 3 generations on1descendant
  • K193607BOBBY Balloon Guide CatheterMicroVention, Inc. · 2020 · 3 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K010954Concentric Balloon Guide Catheter, Model 90031Concentric Medical, Inc.20011
−3K021899Concentric Balloon Guide Catheter, Model 90072Concentric Medical, Inc.20021
−2K102657Concentric Balloon Guide CatheterConcentric Medical, Inc.2010—
−1K112404Concentric Balloon Guide CatheterConcentric Medical, Inc.20121
0 (this device)K122581Modified Balloon Guide Catheter Model 90253, 90254Concentric Medical, Inc.2012—
+1K131492Modified Flowgate Balloon Guide CatheterConcentric Medical, Inc.2013—
+2K1537298F FlowGate2 Balloon Guide CatheterConcentric Medical20162
+2K242392Path BGCCrossroads Neurovascular, Inc.2025—
+3K212340EMBOGUARD Balloon Guide CatheterNeuravi Limited2021—
+3K193607BOBBY Balloon Guide CatheterMicroVention, Inc.20201

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201310
201610
201920
202020
202130
202210
202310
202410
202510
202620
FIG. 1 — Lineage of K122581, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DQY? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQY device

  1. #1 K234115 APRO 55 Catheterscore 17.5 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K112404Concentric Balloon Guide Catheter
Summary page 1
…Manager, Regulatory Affairs Date Prepared: November 16, 2012 Predicate Device: Concentric Balloon Guide Catheter (K112404) Device Description Like the predicate device, the Modified Balloon Guide Catheter is a coaxial-lumen, braid-reinforced, varia…
Concentric Medical, Inc.20121

Also cleared against K112404: 3 other devices →

What descends from it? (15)

See all 15

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K131492Modified Flowgate Balloon Guide CatheterConcentric Medical, Inc.201313
K202916Balloon Guiding CatheterShanghai Heartcare Medical Technology Co., Ltd.20210

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 20, 2012Class IIZ-0655-2013Device DesignFirm received complaints of resistance and possible collapse of the distal tip of the FlowGate Balloon Guide Catheters during use for aspiration.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 1 company; a new generation was cleared every 1.3 years on median.
  • 3 of the 4 ancestors have a recall of any class.
  • Of its descendants, K131492 (Concentric Medical, Inc., 2013) has gone furthest, with 13 descendants of its own.
  • 3 other devices also cite its predicate K112404.

How does its category compare?

All DQY devices

DQY Catheter, Percutaneous: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · DQY

21in 2025

2007: 162008: 312009: 372010: 222011: 362012: 302013: 322014: 192015: 272016: 222017: 262018: 252019: 302020: 202021: 182022: 182023: 182024: 162025: 212026 (partial year): 11
2007high 37 (2009) · low 16 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQY

115in 2025

2007: 85 days median2008: 66 days median2009: 73 days median2010: 74 days median2011: 71 days median2012: 87 days median2013: 92 days median2014: 127 days median2015: 118 days median2016: 117 days median2017: 113 days median2018: 106 days median2019: 111 days median2020: 120 days median2021: 81 days median2022: 74 days median2023: 114 days median2024: 87 days median2025: 115 days median2026 (partial year): 30 days median
2007high 127 (2014) · low 66 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQY

1in 2025

2007: 62008: 122009: 172010: 322011: 62012: 22013: 36, 1 Class I2014: 32015: 32016: 22, 7 Class I2017: 82018: 22019: 12020: 22021: 22023: 172024: 12025: 1, 1 Class I2026 (partial year): 2, 1 Class I
2007high 36 (2013) · low 0 (2022)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · DQY

3,056in 2025

2007: 3132008: 4962009: 8192010: 1,2932011: 1,2112012: 1,1912013: 1,3522014: 1,4972015: 2,3372016: 2,2522017: 2,4892018: 2,8552019: 2,9022020: 2,8892021: 3,0062022: 2,6432023: 3,0492024: 3,1902025: 3,056
2007high 3,190 (2024) · low 313 (2007)2025

MAUDE reports naming product code DQY, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200716856313
2008316612496
2009377317819
20102274321,293
2011367161,211
2012308721,191
2013329236 (1)1,352
20141912731,497
20152711832,337
20162211722 (7)2,252
20172611382,489
20182510622,855
20193011112,902
20202012022,889
2021188123,006
2022187402,643
202318114173,049
2024168713,190
2025211151 (1)3,056
202611302 (1)—

What is it indicated for?

The Concentric Balloon Guide Catheter is indicated for use in facilitating the insertion and guidance of intravascular catheter into a selected blood vessel in the peripheral and neuro vascular systems. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Balloon Guide Catheter is also indicated for use as a conduit for retrieval devices.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K122581: Modified Balloon Guide Catheter Model 90253, 90254.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K122581

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026