510(k) Report

510(k) · Cleared

Recalled · Class I · 2016

Slip-Cath Beacon Tip Catheter Model SCBR (Product Prefix), Shuttle Select Slip-Cath Catheter Model SCBR (Product Prefix)

K122937510(k) SpecialCleared Dec 14, 201281-day reviewClass II21 CFR 870.1200Recalls 4 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cook, Inc. · DQO Catheter, Intravascular, Diagnostic

Cleared on Dec 14, 2012 as substantially equivalent to K882796 (Cook, Inc., 1988) after an 81-day review. Its family tree reaches back 1 generation to 1988. 4 of the 11 devices in its family, counting itself, have a Class I recall. 9 later devices descend from it. It has been recalled 4 times, including 2 Class I recalls.

Decided
Dec 14, 2012
Review
81 days
Class
Class II
Product code
DQO
Recalls
4 (2 Class I)
Predicates
1

Its longest line of descent runs back 1 generation to K882796 (1988). 9 later devices descend from it, 8 citing it directly. 4 of the 11 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2018: 4 descendants cleared2017: 2 descendants cleared, 2 with a Class I recall22015: 1 descendant cleared, 1 with a Class I recall12019: 1 descendant cleared2016: 1 descendant clearedReference device: K101450 · Cantata Microcatheter · Cook, Inc. · 2010K882796 · Slip-Coat (TM) Catheters · Cook, Inc. · 19881990200020102020
Descendants, by generation 1 generation forward: 8 descendants, 3 with a Class I recall32 generations forward: 1 descendant K882796 · Slip-Coat (TM) Catheters · Cook, Inc. · 19880
  1. Root · 1 generation backK882796 Slip-Coat (TM) CathetersCook, Inc. · 1988
  2. This deviceK122937 Slip-Cath Beacon Tip Catheter Model SCBR (Product Prefix), Shuttle…Cook, Inc. · 2012
  3. Descendant · cites it directly · 1 of its ownK162448 White Sizing Catheter, Aurous Centimeter Sizing Catheter, Cava…Cook Incorporated · 2017 · Recalled (Class I)
  4. Descendant · cites it directly · 1 of its ownK133130 Beacon Tip Torcon NB Advantage Catheter, Torcon NB Advantage CatheterCook, Inc. · 2015 · Recalled (Class I)
  5. Descendant · cites it directly · 0 of its ownK181722 Polyethylene CatheterCook Incorporated · 2019
  6. Descendant · cites it directly · 0 of its ownK180286 Polyethylene CatheterCook Incorporated · 2018
  7. Descendant · cites it directly · 0 of its ownK180054 Royal Flush Plus High-Flow CatheterCook Incorporated · 2018

Showing 5 of 9 descendants. See all descendants

Descendants cleared per year

2015: 12016: 12017: 22018: 42019: 1
2015high 4 (2018) · low 0 (2020)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 9 · See all
  • K162448White Sizing Catheter, Aurous Centimeter Sizing Catheter, Cava Vessel Sizing CatheterCook Incorporated · 2017 · cites it directly1descendant
  • K133130Beacon Tip Torcon NB Advantage Catheter, Torcon NB Advantage CatheterCook, Inc. · 2015 · cites it directly1descendant
  • K181722Polyethylene CatheterCook Incorporated · 2019 · cites it directly0descendants
  • K180286Polyethylene CatheterCook Incorporated · 2018 · cites it directly0descendants
  • K180054Royal Flush Plus High-Flow CatheterCook Incorporated · 2018 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K882796Slip-Coat (TM) CathetersCook, Inc.1988—
0 (this device)K122937Slip-Cath Beacon Tip Catheter Model SCBR (Product Prefix), Shuttle Select…Cook, Inc.2012—
+1K162448White Sizing Catheter, Aurous Centimeter Sizing Catheter, Cava Vessel Sizing…Cook Incorporated2017Class I
+1K133130Beacon Tip Torcon NB Advantage Catheter, Torcon NB Advantage CatheterCook, Inc.2015Class I
+1K181722Polyethylene CatheterCook Incorporated2019—
+1K180286Polyethylene CatheterCook Incorporated2018—
+1K180054Royal Flush Plus High-Flow CatheterCook Incorporated2018—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201511
201610
201722
201840
201910
FIG. 1 — Lineage of K122937, 1988–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQO device

  1. #1 K254278 Arterial Pressure Monitoring Set/Trayscore 14.4 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K882796Slip-Coat (TM) Catheters
Summary page 1
…ices,subject of this submnission, are substantially equivalentto the Slip-CoatTh Catheters, manufactured by Cook Incorporated, which, are; cleared; under 510O(k), number K882796 and the reference device, Cantata Microcatheter, which is cleared under 510()fnumberK101I450. Comparison to Predicate and 'Refer…
Cook, Inc.1988—
†K101450Cantata MicrocatheterCook, Inc.2010—

Also cleared against K882796: 2 other devices →

What descends from it? (9)

8 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K162448White Sizing Catheter, Aurous Centimeter Sizing Catheter, Cava Vessel Sizing Catheter Class ICook Incorporated20171
K133130Beacon Tip Torcon NB Advantage Catheter, Torcon NB Advantage Catheter Class ICook, Inc.20151
K181722Polyethylene CatheterCook Incorporated20190
K180286Polyethylene CatheterCook Incorporated20180
K180054Royal Flush Plus High-Flow CatheterCook Incorporated20180
K172635High-Flo Silver Polyethylene CatheterCook Incorporated20180
K171264Royal Flush CatheterCook Incorporated20180
K173289Beacon Tip Torcon NB Advantage Catheter, Slip-Cath Beacon Tip Catheter and Shuttle… Class ICook Incorporated20170

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 15, 2016Class IIZ-2614-2016Unknown/Undetermined by firmIncrease in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or…
Apr 15, 2016Class IIZ-2613-2016Unknown/Undetermined by firmIncrease in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or…
Oct 7, 2015Class IZ-0798-2016Component design/selectionReports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occluding blood flow to end organs may result. Recall expanded…
Jul 2, 2015Class IZ-2319-2015Component design/selectionReports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occluding blood flow to end organs may result.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K162448White Sizing Catheter, Aurous Centimeter Sizing Catheter, Cava Vessel Sizing CatheterCook Incorporated · 2017 · Cites it directly · ◆ in the 2017 descendants bar of FIG. 11 Class I recall
  • K133130Beacon Tip Torcon NB Advantage Catheter, Torcon NB Advantage CatheterCook, Inc. · 2015 · Cites it directly · ◆ in the 2015 descendants bar of FIG. 11 Class I recall
  • K173289Beacon Tip Torcon NB Advantage Catheter, Slip-Cath Beacon Tip Catheter and Shuttle Select Slip-Cath…Cook Incorporated · 2017 · Cites it directly · ◆ in the 2017 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K162448 (Cook Incorporated, 2017) has gone furthest, with 1 descendant of its own.
  • 2 other devices also cite its predicate K882796.

How does its category compare?

All DQO devices

DQO Catheter, Intravascular, Diagnostic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · DQO

2in 2025

2007: 82008: 32009: 82010: 72011: 62012: 152013: 22014: 32015: 102016: 42017: 102018: 102019: 62020: 32021: 12022: 42023: 42024: 42025: 22026 (partial year): 3
2007high 15 (2012) · low 1 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQO

148in 2024

2007: 106 days median2008: 80 days median2009: 58 days median2010: 137 days median2011: 74 days median2012: 94 days median2014: 109 days median2015: 209 days median2016: 89 days median2017: 176 days median2018: 234 days median2019: 113 days median2020: 30 days median2022: 181 days median2023: 73 days median2024: 148 days median2026 (partial year): 85 days median
2007high 234 (2018) · low 30 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQO

1in 2023

2007: 142008: 42009: 72010: 112011: 72012: 22013: 12015: 1, 1 Class I2016: 2, 1 Class I2017: 12018: 22019: 712020: 4, 4 Class I2022: 12023: 1, 1 Class I2026 (partial year): 24, 4 Class I
2007high 71 (2019) · low 0 (2014)2026

Class I (11 since 2007) Class II/III not classified

Adverse event reports per year · DQO

656in 2025

2007: 5642008: 7262009: 9332010: 5562011: 6252012: 5322013: 7532014: 6012015: 6282016: 6142017: 8732018: 7772019: 3862020: 3122021: 3382022: 3792023: 3912024: 5112025: 656
2007high 933 (2009) · low 312 (2020)2025

MAUDE reports naming product code DQO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007810614564
20083804726
20098587933
2010713711556
20116747625
201215942532
20132—1753
201431090601
2015102091 (1)628
20164892 (1)614
2017101761873
2018102342777
2019611371386
20203304 (4)312
20211—0338
202241811379
20234731 (1)391
202441480511
20252—0656
202638524 (4)—

What is it indicated for?

The catheters are intended for use in angiographic procedures by physicians trained and experienced in angiographic techniques.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K122937: Slip-Cath Beacon Tip Catheter Model SCBR (Product Prefix), Shuttle Select Slip-Cath Catheter Model SCBR (Product Prefix).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K122937

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026