510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

Beacon Tip Torcon NB Advantage Catheter, Slip-Cath Beacon Tip Catheter and Shuttle Select Slip-Cath Catheter, Beacon Tip Centimeter Sizing Catheter, Beacon Tip Cava Vessel Sizing Catheter, Beacon Tip White Vessel Sizing Catheter

K173289510(k) SpecialCleared Nov 15, 201730-day reviewClass II21 CFR 870.1200Recalls 2 (2 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cook Incorporated · DQO Catheter, Intravascular, Diagnostic

Cleared on Nov 15, 2017 as substantially equivalent to K122937 (Cook, Inc., 2012) and 1 other predicate after a 30-day review. Its family tree reaches back 3 generations to 1988. 3 of the 4 devices in its family, counting itself, have a Class I recall. It has been recalled 2 times, including 2 Class I recalls.

Decided
Nov 15, 2017
Review
30 days
Class
Class II
Product code
DQO
Recalls
2 (2 Class I)
Predicates
2

Its longest line of descent runs back 3 generations to K882796 (1988). No later device cites it as a predicate yet. 3 of the 4 devices in this family, counting this one, have a Class I recall.

  1. Root · 3 generations backK882796 Slip-Coat (TM) CathetersCook, Inc. · 1988
  2. 2 generations backK122937 Slip-Cath Beacon Tip Catheter Model SCBR (Product Prefix), Shuttle…Cook, Inc. · 2012 · Recalled (Class I)
  3. 1 generation backK162448 White Sizing Catheter, Aurous Centimeter Sizing Catheter, Cava…Cook Incorporated · 2017 · Recalled (Class I)
  4. This deviceK173289 Beacon Tip Torcon NB Advantage Catheter, Slip-Cath Beacon Tip…Cook Incorporated · 2017

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K882796Slip-Coat (TM) CathetersCook, Inc.1988—
−2K122937Slip-Cath Beacon Tip Catheter Model SCBR (Product Prefix), Shuttle Select…Cook, Inc.2012Class I
−1K162448White Sizing Catheter, Aurous Centimeter Sizing Catheter, Cava Vessel Sizing…Cook Incorporated2017Class I
0 (this device)K173289Beacon Tip Torcon NB Advantage Catheter, Slip-Cath Beacon Tip Catheter and…Cook Incorporated2017—
FIG. 1 — Lineage of K173289, 1988–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DQO device

  1. #1 K254278 Arterial Pressure Monitoring Set/Trayscore 14.4 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K122937Slip-Cath Beacon Tip Catheter Model SCBR (Product Prefix), Shuttle Select Slip-Cath…
Summary page 5
…2 the Shuttle® Select Slip-Cath® Catheter cleared on 14 December 2012 under 510(k) Premarket Notification Number K122937. The subject Beacon Tip Sizing Catheter family (Beacon® Tip Centimeter Sizing Catheter, Beacon® Tip Cava Vessel Sizing Catheter,…
Cook, Inc.2012Class I
*K162448White Sizing Catheter, Aurous Centimeter Sizing Catheter, Cava Vessel Sizing Catheter
Summary page 5
…, the White Sizing Catheter, the Aurous Centimeter Sizing Catheter, and the Cava Vessel Sizing Catheter cleared on 26 May 2017 under 510(k) Premarket Notification Number K162448. Comparison to Predicate Device: The subject Beacon Tip Catheters and Beacon Tip Sizing Catheters are identical to the predicat…
Cook Incorporated2017Class I

Also cleared against K122937: 7 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 2, 2026Class IZ-1936-2026Under Investigation by firmMarker bands may be at an increased risk of cracking/breakage. Potential adverse events that may result include increased procedural time, harms associated with device fragmentation/separation, and vessel injury. In a worst-case scenario…
May 15, 2025Class IZ-1902-2025Process change controlCook Medical identified that catheters supplied in the affected device lots may experience tip separation.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K162448White Sizing Catheter, Aurous Centimeter Sizing Catheter, Cava Vessel Sizing CatheterCook Incorporated · 2017 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K122937Slip-Cath Beacon Tip Catheter Model SCBR (Product Prefix), Shuttle Select Slip-Cath Catheter Model SCBR…Cook, Inc. · 2012 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls

What does this lineage say?

  • The line of descent spans 29 years and passes through 2 companies; a new generation was cleared every 4.4 years on median.
  • 2 of the 3 ancestors have a recall of any class; 2 have a Class I recall.
  • 7 other devices also cite its predicate K122937.

How does its category compare?

All DQO devices

DQO Catheter, Intravascular, Diagnostic: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · DQO

2in 2025

2007: 82008: 32009: 82010: 72011: 62012: 152013: 22014: 32015: 102016: 42017: 102018: 102019: 62020: 32021: 12022: 42023: 42024: 42025: 22026 (partial year): 3
2007high 15 (2012) · low 1 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DQO

148in 2024

2007: 106 days median2008: 80 days median2009: 58 days median2010: 137 days median2011: 74 days median2012: 94 days median2014: 109 days median2015: 209 days median2016: 89 days median2017: 176 days median2018: 234 days median2019: 113 days median2020: 30 days median2022: 181 days median2023: 73 days median2024: 148 days median2026 (partial year): 85 days median
2007high 234 (2018) · low 30 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DQO

1in 2023

2007: 142008: 42009: 72010: 112011: 72012: 22013: 12015: 1, 1 Class I2016: 2, 1 Class I2017: 12018: 22019: 712020: 4, 4 Class I2022: 12023: 1, 1 Class I2026 (partial year): 24, 4 Class I
2007high 71 (2019) · low 0 (2014)2026

Class I (11 since 2007) Class II/III not classified

Adverse event reports per year · DQO

656in 2025

2007: 5642008: 7262009: 9332010: 5562011: 6252012: 5322013: 7532014: 6012015: 6282016: 6142017: 8732018: 7772019: 3862020: 3122021: 3382022: 3792023: 3912024: 5112025: 656
2007high 933 (2009) · low 312 (2020)2025

MAUDE reports naming product code DQO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007810614564
20083804726
20098587933
2010713711556
20116747625
201215942532
20132—1753
201431090601
2015102091 (1)628
20164892 (1)614
2017101761873
2018102342777
2019611371386
20203304 (4)312
20211—0338
202241811379
20234731 (1)391
202441480511
20252—0656
202638524 (4)—

What is it indicated for?

The Beacon Tip Catheters are intended for use in angiographic procedures by physicians trained and experienced in angiographic techniques. Standard techniques for placement of vascular access sheaths, angiographic catheters and wire guides should be employed.

The Beacon Tip Sizing Catheters are intended for use in angiographic procedures by physicians trained and experienced in angiographic techniques. Standard techniques for placement of vascular access sheaths, angiographic catheters and wire guides should be employed. Beacon Tip Sizing Catheters have marker bands that can be used for anatomical measurements.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K173289: Beacon Tip Torcon NB Advantage Catheter, Slip-Cath Beacon Tip Catheter and Shuttle Select Slip-Cath Catheter, Beacon Tip Centimeter Sizing Catheter, Beacon Tip Cava Vessel Sizing Catheter, Beacon Tip White Vessel Sizing Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K173289

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026