510(k) Report

510(k) · Cleared

Cleared

U&U Hydrogen Peroxide Gas Plasma Sterilization Pouch/Roll

K123162510(k) TraditionalCleared Feb 26, 2014505-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

U&U Medical Technology Co, Ltd. · FRG Wrap, Sterilization

Cleared on Feb 26, 2014 as substantially equivalent to K103210 (Advanced Sterilization Products, 2011) after a 505-day review. Its family tree reaches back 2 generations to 1993.

Decided
Feb 26, 2014
Review
505 days
Class
Class II
Product code
FRG
Recalls
None
Predicates
1

Its longest line of descent runs back 2 generations to K921910 (1993). 3 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 2 generations backK921910 Sterrad Chemical IndicatorJohnson & Johnson Medical, Inc. · 1993
  2. 1 generation backK103210 Tyvek Self-Seal Pouches with Sterrad, Tyvek Rolls with Sterrad…Advanced Sterilization Products · 2011
  3. This deviceK123162 U&U Hydrogen Peroxide Gas Plasma Sterilization Pouch/RollU&U Medical Technology Co, Ltd. · 2014
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K921910Sterrad Chemical IndicatorJohnson & Johnson Medical, Inc.1993—
−1K103210Tyvek Self-Seal Pouches with Sterrad, Tyvek Rolls with Sterrad, Tyvek…Advanced Sterilization Products2011—
0 (this device)K123162U&U Hydrogen Peroxide Gas Plasma Sterilization Pouch/RollU&U Medical Technology Co, Ltd.2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510
FIG. 1 — Lineage of K123162, 1993–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRG device

  1. #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K103210Tyvek Self-Seal Pouches with Sterrad, Tyvek Rolls with Sterrad, Tyvek Heat-Seal with…
Summary page 1
…Regulation Number: 21CFR 860.6850 3. Predicate Devices Information: Trade Name: Tyvek Pouch/Roll With STERRAD Chemical Indicator 510(K) Number: K103210 Manufacturer: Advanced Sterilization Products Company. 4. Devices Description: The U&U Hydrogen Peroxide Gas Plas…
Advanced Sterilization Products2011—

Also cleared against K103210: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 3 companies; a new generation was cleared every 17.7 years on median.
  • 1 of the 3 ancestors has a recall of any class.
  • 2 other devices also cite its predicate K103210.

How does its category compare?

All FRG devices

FRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · FRG

9in 2025

2007: 32008: 22009: 42010: 42011: 62012: 102013: 62014: 82015: 122016: 92017: 92018: 52019: 92020: 52021: 42022: 92023: 92024: 112025: 92026 (partial year): 4
2007high 12 (2015) · low 2 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRG

151in 2025

2007: 182 days median2009: 171 days median2010: 213 days median2011: 110 days median2012: 133 days median2013: 146 days median2014: 177 days median2015: 196 days median2016: 131 days median2017: 260 days median2018: 150 days median2019: 258 days median2020: 122 days median2021: 170 days median2022: 217 days median2023: 230 days median2024: 182 days median2025: 151 days median2026 (partial year): 247 days median
2007high 260 (2017) · low 110 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRG

2in 2022

2013: 22017: 292020: 142022: 2
2007high 29 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRG

18in 2025

2007: 12008: 22009: 22011: 122012: 42013: 52014: 132015: 92016: 202017: 72018: 222019: 102020: 212021: 92022: 42023: 132024: 212025: 18
2007high 22 (2018) · low 0 (2010)2025

MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007318201
20082—02
2009417102
2010421300
20116110012
20121013304
2013614625
20148177013
20151219609
20169131020
20179260297
20185150022
20199258010
202051221421
2021417009
2022921724
20239230013
202411182021
20259151018
202642470—

What is it indicated for?

U&U Hydrogen Peroxide Gas Plasma Sterilization Pouch / Roil are intended to be used to enclose medical devices that are to be terminally sterilized in the STERRAD I100NX Hydrogen Peroxide Gas Plasma sterilizer. The recommended Hydrogen Peroxide Gas Plasma cycle parameter: STERRAD IOGONX STANDARD Cycle Sterilization load claim: STERRAD IOONIX Sterilizer Standard cycle: -General medical instruments (metal and nonmetal, including hinged devices)…

Read the full statement (207 words)

U&U Hydrogen Peroxide Gas Plasma Sterilization Pouch / Roil are intended to be used to

enclose medical devices that are to be terminally sterilized in the STERRAD I100NX Hydrogen

Peroxide Gas Plasma sterilizer.

The recommended Hydrogen Peroxide Gas Plasma cycle parameter:

STERRAD IOGONX STANDARD Cycle

Sterilization load claim:

STERRAD IOONIX Sterilizer Standard cycle:

-General medical instruments (metal and nonmetal, including hinged devices)

-Instruments with single-channel stainless steel lumens with an Internal diameter 3mm or larger

and length 150 mm or shorter.

- Polyethylene and Teflon lumen tubing with an Internal diameter of 3 mm or larger and length of

500 mm or shorter.

* These items must be sterilized without any additional load items

- Limit of 20 pieces of tubing per cycle

Load condition: One device per Sterilization Pouch.

Load chamber Two shelves (total weight: 21.4 lbs or 9.7 kg)

Load Pouch: Leave enough malarial beyond the seal for the opener to easily grasp (usually 1-1

'A Inches).

SHELF UFE: It Is recommended that the products are put to their end use withln 2 years of

manufacture. The recommended "best before data and the manufacturing date are stated on

the label. This device was demonstrated to maintain sterility of the contents for 30 days

after Sterilization.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K123162: U&U Hydrogen Peroxide Gas Plasma Sterilization Pouch/Roll.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K123162

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026