Syngo.Via

K123375510(k) TraditionalCleared Nov 20, 201219-day reviewClass II21 CFR 892.2050Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens AG · LLZ System, Image Processing, Radiological

Cleared on Nov 20, 2012 after a 19-day review. The summary cites one number, which could not be confirmed as its predicate (see the table below). 1 of the 97 devices in its family, counting itself, has a Class I recall. 96 later devices descend from it. It has been recalled 8 times.

Decided
Nov 20, 2012
Review
19 days
Class
Class II
Product code
LLZ
Recalls
8
Predicates
1

96 later devices descend from it, 3 citing it directly. 1 of the 97 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2013: 2 descendants cleared2014: 3 descendants cleared2015: 4 descendants cleared2018: 6 descendants cleared2017: 6 descendants cleared2020: 10 descendants cleared2016: 2 descendants cleared2025: 12 descendants cleared, 1 with a Class I recall12019: 7 descendants cleared2023: 14 descendants cleared2022: 7 descendants cleared2021: 4 descendants cleared2026: 11 descendants cleared2024: 8 descendants cleared2010201520202025
  1. This deviceK123375 Syngo.ViaSiemens AG · 2012
  2. Descendant · cites it directly · 63 of its ownK123920 Syngo.ViaSiemens AG Medical Solutions · 2013
  3. Descendant · cites it directly · 54 of its ownK130749 Syngo.MR General; Syngo.MR Cardiology; Syngo.MR VascularSiemens Medical Solutions USA, Inc. · 2013
  4. Descendant · 2 generations on · 46 of its ownK150843 syngo®.via (version VB10A)Siemens AG · 2015
  5. Descendant · 2 generations on · 27 of its ownK133401 Syno MR Neurology, Syngo MR Oncology, Syngo. MR Brevis, Syngo MMR…Siemens Medical Solutions USA, Inc. · 2014
  6. Descendant · 3 generations on · 25 of its ownK151353 syngo.MR Neurology, syngo.MR OncologySiemens Medical Solutions USA, Inc. · 2015

Showing 5 of 96 descendants. See all descendants

Descendants cleared per year

2013: 22014: 32015: 42016: 22017: 62018: 62019: 72020: 102021: 42022: 72023: 142024: 82025: 122026: 11
2013high 14 (2023) · low 2 (2013)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 96 · See all
  • K123920Syngo.ViaSiemens AG Medical Solutions · 2013 · cites it directly63descendants
  • K130749Syngo.MR General; Syngo.MR Cardiology; Syngo.MR VascularSiemens Medical Solutions USA, Inc. · 2013 · cites it directly54descendants
  • K150843syngo®.via (version VB10A)Siemens AG · 2015 · 2 generations on46descendants
  • K133401Syno MR Neurology, Syngo MR Oncology, Syngo. MR Brevis, Syngo MMR GeneralSiemens Medical Solutions USA, Inc. · 2014 · 2 generations on27descendants
  • K151353syngo.MR Neurology, syngo.MR OncologySiemens Medical Solutions USA, Inc. · 2015 · 3 generations on25descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K123375Syngo.ViaSiemens AG2012—
+1K123920Syngo.ViaSiemens AG Medical Solutions2013—
+1K130749Syngo.MR General; Syngo.MR Cardiology; Syngo.MR VascularSiemens Medical Solutions USA, Inc.2013—
+2K150843syngo®.via (version VB10A)Siemens AG2015—
+2K133401Syno MR Neurology, Syngo MR Oncology, Syngo. MR Brevis, Syngo MMR GeneralSiemens Medical Solutions USA, Inc.2014—
+3K151353syngo.MR Neurology, syngo.MR OncologySiemens Medical Solutions USA, Inc.2015—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201320
201430
201540
201620
201760
201860
201970
2020100
202140
202270
2023140
202480
2025121
2026110
FIG. 1 — Lineage of K123375, 2012–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
Kept off the family treeK092551FDA’s K092551 is Smith & Nephew Instrument Trays
Our reading of the summary gave K092551, FDA’s number for Smith & Nephew Instrument Trays (Smith & Nephew Inc., Endoscopy Div.). Its FDA name has none of the distinctive words the summary uses for the device it names, “syngo.x”. Kept off the family tree.
Summary page 2
…cific workflows. syngo.via allows to use a variety of advanced applications (clinical applica- tions) designed for syngo.via just as the predicate device .syngo.x,(K09255 19, cleared on August 27, 2009). These applications are medical devices on theft own rights and filed separately. They are no…
Smith & Nephew Inc., Endoscopy Div.2010—

What descends from it? (96)

See all 96

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K123920Syngo.ViaSiemens AG Medical Solutions201363
K130749Syngo.MR General; Syngo.MR Cardiology; Syngo.MR VascularSiemens Medical Solutions USA, Inc.201354
K133521Aze Phoenix SoftwareAze, Ltd.20140

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 27, 2018Class IIZ-1778-2018Software designFunctionality in the report sections "Findings Information" and "Summary of Measured Findings" not functioning properly. The corrected values modified in the report are neither saved, nor printed, nor sent to the information system. The…
Jan 10, 2018Class IIZ-1261-2018Software Manufacturing/Software DeploymentWhen the archiving configuration is changed, data received/created after the upgrade may be flagged as "Not to be archived". This is caused by the automatic function for cleaning up temporary data being disabled by the software upgrade.…
Jan 8, 2018Class IIZ-1303-2018Software designA software functionality in the report sections Findings Information and Summary of Measured Findings is not performing properly. The system will use the original values instead of displaying and saving corrected/modified values. Outdated…
Aug 31, 2017Class IIZ-0808-2018Software Design ChangeA software update was released on August 22, 2017. Performing this software installation and upgrading syngo.via system to the software version VB20A_HF02 will resolve various bugs and inconsistent diagnosis with the old system.
Mar 6, 2017Class IIZ-1689-2017Software designSoftware changes now available to address several issues
Aug 5, 2016Class IIZ-2853-2016Software designSoftware error. Incorrect values for the volume calculation from a freehand VOI at the customer site. In volume calculations of prostate as well as in volume calculations of liver were too high.
Jun 23, 2016Class IIZ-2245-2016Software designIncorrect values for the volume calculation. Software update VB30B via Update Instructions SY018/16/P to resolve software errors.
May 15, 2015Class IIZ-1705-2015Software designmeasurements drawn on the 2nd and subsequent images of the series are not visible on printouts when the series is sent to print.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K250443MAGNETOM Avanto Fit; MAGNETOM Skyra Fit; MAGNETOM Sola Fit; MAGNETOM Viato.MobileSiemens Healthcare GmbH · 2025 · 4 generations on · ◆ in the 2025 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • Of its descendants, K123920 (Siemens AG Medical Solutions, 2013) has gone furthest, with 63 descendants of its own.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2012, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

syngo.via is a software solution intended to be used for viewing, manipulation, communication, and storage of medical images.

It can be used as a stand-alone device or together with a variety of cleared and unmodified syngo based software options.

syngo.via supports interpretation and evaluation of examinations within healthcare institutions; for example, in Radiology, Nuclear Medicine and Cardiology environments.

The system is not intended for the displaying of digital mammography images for diagnosis in the U.S.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K123375: Syngo.Via.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K123375

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026