510(k) Report

510(k) · Cleared

Cleared

Filmarray Respiratory Panel (RP)

K123620510(k) TraditionalCleared Feb 11, 201380-day reviewClass II21 CFR 866.3980

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biofire Diagnostics, Inc. · OCC Respiratory Virus Panel Nucleic Acid Assay System

Cleared on Feb 11, 2013 as substantially equivalent to K103175 (Idaho Technology, Inc., 2011) and 2 other predicates after an 80-day review. Its family tree reaches back 5 generations to 2008. 18 later devices descend from it.

Decided
Feb 11, 2013
Review
80 days
Class
Class II
Product code
OCC
Recalls
None
Predicates
3

Its longest line of descent runs back 5 generations to DEN070013 (2008). 6 ancestors in all. 18 later devices descend from it, 4 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2011: 1 other ancestor2015: 2 descendants cleared2019: 2 descendants cleared2014: 2 descendants cleared2017: 5 descendants cleared2016: 2 descendants cleared2025: 1 descendant cleared2018: 2 descendants cleared2023: 1 descendant cleared2026: 1 descendant clearedDEN070013 · ID-Tag Respiratory Viral Panel · Luminex Molecular Diagnostics, Inc. · 2008K081483 · Xtag Respiratory Viral Panel · Luminex Molecular Diagnostics, Inc. · 2008 · recalledK091667 · Xtag Respiratory Viral Panel, Models I019A0110, I019C011, I019D0112, I019E0113, S019-0116 · Luminex Molecular Diagnostics, Inc. · 2009 · recalledK103175 · Filmarray Respiratory Panel (RP) · Idaho Technology, Inc. · 2011K120267 · Filmarray RP Panel · Idaho Technology, Inc. · 201220052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 4 descendants2 generations forward: 6 descendants3 generations forward: 3 descendants4 generations forward: 3 descendants5 generations forward: 2 descendants K120267 · Filmarray RP Panel · Idaho Technology, Inc. · 2012K103175 · Filmarray Respiratory Panel (RP) · Idaho Technology, Inc. · 2011K091667 · Xtag Respiratory Viral Panel, Models I019A0110, I019C011, I019D0112, I019E0113, S019-0116 · Luminex Molecular Diagnostics, Inc. · 2009 · recalledK081483 · Xtag Respiratory Viral Panel · Luminex Molecular Diagnostics, Inc. · 2008 · recalledDEN070013 · ID-Tag Respiratory Viral Panel · Luminex Molecular Diagnostics, Inc. · 2008−50+5
  1. Root · 5 generations backDEN070013 ID-Tag Respiratory Viral PanelLuminex Molecular Diagnostics, Inc. · 2008
  2. 4 generations backK081483 Xtag Respiratory Viral PanelLuminex Molecular Diagnostics, Inc. · 2008 · Recalled
  3. 3 generations backK091667 Xtag Respiratory Viral Panel, Models I019A0110, I019C011, I019D0112…Luminex Molecular Diagnostics, Inc. · 2009 · Recalled
  4. 2 generations backK103175 Filmarray Respiratory Panel (RP)Idaho Technology, Inc. · 2011
  5. 1 generation backK120267 Filmarray RP PanelIdaho Technology, Inc. · 2012
  6. This deviceK123620 Filmarray Respiratory Panel (RP)Biofire Diagnostics, Inc. · 2013
  7. Descendant · cites it directly · 10 of its ownK143080 FilmArray Respiratory Panel (RP) for use with Multi-instrument…Biofire Diagnostics, LLC · 2015
  8. Descendant · 2 generations on · 6 of its ownK160068 FilmArray Respiratory Panel (RP) for use with FilmArray TorchBiofire Diagnostics, LLC · 2016 · Recalled
  9. Descendant · cites it directly · 5 of its ownK133673 Illumigene Pertusis Dna Amplification Assay, Illumigene Pertussis…Meridian Bioscience, Inc. · 2014
  10. Descendant · 3 generations on · 4 of its ownK170604 FilmArray Respiratory Panel 2 (RP2)Biofire Diagnostics, LLC · 2017
  11. Descendant · 2 generations on · 2 of its ownK163626 ARIES Bordetella Assay; ARIES Bordetella Assay Protocol File KitLuminex Corporation · 2017 · Recalled

Showing 5 of 18 descendants. See all descendants

Descendants cleared per year

2014: 22015: 22016: 22017: 52018: 22019: 22023: 12025: 12026: 1
2014high 5 (2017) · low 0 (2020)2026

Most influential descendants

Showing 5 of 18 · See all
  • K143080FilmArray Respiratory Panel (RP) for use with Multi-instrument FilmArray SystemBiofire Diagnostics, LLC · 2015 · cites it directly10descendants
  • K160068FilmArray Respiratory Panel (RP) for use with FilmArray TorchBiofire Diagnostics, LLC · 2016 · 2 generations on6descendants
  • K133673Illumigene Pertusis Dna Amplification Assay, Illumigene Pertussis External Control Kit…Meridian Bioscience, Inc. · 2014 · cites it directly5descendants
  • K170604FilmArray Respiratory Panel 2 (RP2)Biofire Diagnostics, LLC · 2017 · 3 generations on4descendants
  • K163626ARIES Bordetella Assay; ARIES Bordetella Assay Protocol File KitLuminex Corporation · 2017 · 2 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5DEN070013ID-Tag Respiratory Viral PanelLuminex Molecular Diagnostics, Inc.2008—
−4K081483Xtag Respiratory Viral PanelLuminex Molecular Diagnostics, Inc.20081
−3K091667Xtag Respiratory Viral Panel, Models I019A0110, I019C011, I019D0112…Luminex Molecular Diagnostics, Inc.20091
−2K103175Filmarray Respiratory Panel (RP)Idaho Technology, Inc.2011—
−1K120267Filmarray RP PanelIdaho Technology, Inc.2012—
0 (this device)K123620Filmarray Respiratory Panel (RP)Biofire Diagnostics, Inc.2013—
+1K143080FilmArray Respiratory Panel (RP) for use with Multi-instrument FilmArray SystemBiofire Diagnostics, LLC2015—
+2K160068FilmArray Respiratory Panel (RP) for use with FilmArray TorchBiofire Diagnostics, LLC20161
+1K133673Illumigene Pertusis Dna Amplification Assay, Illumigene Pertussis External…Meridian Bioscience, Inc.2014—
+3K170604FilmArray Respiratory Panel 2 (RP2)Biofire Diagnostics, LLC2017—
+2K163626ARIES Bordetella Assay; ARIES Bordetella Assay Protocol File KitLuminex Corporation20171

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
201110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201420
201520
201620
201750
201820
201920
202310
202510
202610
FIG. 1 — Lineage of K123620, 2008–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OCC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OCC device

  1. #1 K262747 ID NOW Influenza A & B 2; ID NOW RSV; ID NOW…score 15.2 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K103175Filmarray Respiratory Panel (RP)
Summary page 1
…ion Number: 21 CER 866.3980 Classification Name: Respiratory Viral Panel Multiplex Nucleic Acid Assay Predicate Device: FilmArray Respiratory Panel (RP) (K103 175, KI 10764 and K120267) Intended Use: FilmArray® Respiratory Panel (RP) is a multiplexed nucleic acid test intended for u…
Idaho Technology, Inc.2011—
*K110764Filmarray RP Panel
Summary page 4
…t FilmArray Respiratory Panel FilmArray Respiratory Panel 1K03175, K1 10764 and K120267) Organisms Influenza A, Influenza A subtype HI1, Same Detected Influenza A subtype H-3,…
Idaho Technology, Inc.2011—
*K120267Filmarray RP Panel
Summary page 1
…3980 Classification Name: Respiratory Viral Panel Multiplex Nucleic Acid Assay Predicate Device: FilmArray Respiratory Panel (RP) (K103 175, KI 10764 and K120267) Intended Use: FilmArray® Respiratory Panel (RP) is a multiplexed nucleic acid test intended for use with the Fil…
Idaho Technology, Inc.2012—

Also cleared against K103175: 4 other devices →

What descends from it? (18)

See all 18

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K143080FilmArray Respiratory Panel (RP) for use with Multi-instrument FilmArray SystemBiofire Diagnostics, LLC201510
K133673Illumigene Pertusis Dna Amplification Assay, Illumigene Pertussis External Control Kit…Meridian Bioscience, Inc.20145
K143653Verigene Respiratory Pathogens Flex Nucleic Acid Test (RP Flex)Nanosphere, Inc.20151
K183597QIAstat-Dx Respiratory PanelQIAGEN GmbH20190

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 5 years and passes through 3 companies; a new generation was cleared every 12 months on median.
  • 2 of the 6 ancestors have a recall of any class.
  • Of its descendants, K143080 (Biofire Diagnostics, LLC, 2015) has gone furthest, with 10 descendants of its own.
  • 4 other devices also cite its predicate K103175.

How does its category compare?

All OCC devices

OCC Respiratory Virus Panel Nucleic Acid Assay System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · OCC

1in 2024

2008: 42009: 62010: 22011: 92012: 62013: 62014: 42015: 82016: 72017: 102018: 42019: 52020: 22021: 12022: 12023: 22024: 12026 (partial year): 3
2007high 10 (2017) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OCC

31in 2019

2008: 28 days median2009: 24 days median2011: 82 days median2012: 125 days median2013: 79 days median2014: 94 days median2015: 85 days median2016: 92 days median2017: 136 days median2018: 134 days median2019: 31 days median2026 (partial year): 130 days median
2007high 136 (2017) · low 24 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OCC

1in 2024

2014: 12018: 12019: 12024: 1
2007high 1 (2014) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OCC

2in 2025

2011: 22012: 12014: 522015: 12016: 362017: 12018: 12019: 22020: 172021: 6862022: 22023: 12024: 32025: 2
2007high 686 (2021) · low 0 (2007)2025

MAUDE reports naming product code OCC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
200842800
200962400
20102—00
201198202
2012612501
201367900
2014494152
201588501
2016792036
20171013601
2018413411
201953112
20202—017
20211—0686
20221—02
20232—01
20241—13
20250—02
202631300—

What is it indicated for?

FilmArray® Respiratory Panel (RP) is a multiplexed nucleic acid test intended for use with the FilmArray Instrument for the simultaneous qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infections. The following organism types and subtypes are identified using the FilmArray RP: Adenovirus, Coronavirus 229E, Coronavirus…

Read the full statement (495 words)

FilmArray® Respiratory Panel (RP) is a multiplexed nucleic acid test intended for use with the FilmArray Instrument for the simultaneous qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) obtained from individuals suspected of respiratory tract infections. The following organism types and subtypes are identified using the FilmArray RP: Adenovirus, Coronavirus 229E, Coronavirus HKU I, Coronavirus NL63, Coronavirus 0C43, Human Metapneumnovirus, In~fluenza A, Influenza A subtype HI, Influenza A subtype H3, Influenza A subtype HI -2009, Influenza B, Parainfluenza Virus 1, Parainfluenza Virus 2, Parainfluenza Virus 3, Parainfluenza Virus 4, Human Rhinovir-us/Enterovirus, Respiratory Syncytial Virus, Bordetellapertussis, Chiamydophila pneumoniae, and Mycoplasia pneutnoniae. The detection and identification of specific viral and bacterial nucleic acids from individuals exhibiting signs and symptoms of a respiratory infection aids in the diagnosis of respiratory infection if used in conjunction with other clinical and epidemiological information. The results of this test should not be used as the sole basis for diagnosis, treatment, or other management decisions. Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by this test or, lower respiratory tract infection that is not detected by a nasopharyngeal swab specimen. Positive results do not rule out co-infection with other organisms: the agent(s) detected by the FilmArray RP may not be the definite cause of disease. Additional laboratory testing (e.g. bacterial and viral culture, iminunofluorescence, and radiography) may be necessary when evaluating a patient with possible respiratory tract infection.

Due to the small number of positive specimens collected for certain organisms during the prospective clinical study, performance characteristics for Bordetellapertussis, Coronavirus 229E, Coronavirus 0C43, Influenza A I, Influenza A H13, Influenza A i-2009, Influenza B, Mycoplasmapneumoniae,Parainfluenza Virus 1, Paraintluenza Virus 2, and Parainfluenza Virus 4 were established primarily with retrospective clinical specimens. Performance characteristics for Chiamydophilapneumoniae were established primarily using contrived clinical specimens.

Due to the genetic similarity between Human Rhinovirus and Enterovimus, the FilmArray RP cannot reliably differentiate them. A positive FilmArray RP Rhinovirus/Enterovirus result should be followed-up using an alternate method (e.g., cell culture or sequence analysis).

The FilmArray RtP assay for Coronavirus 0C43 may cross-react with some isolates of Coronavirus HKU L A dual positive result may be due to cross-reactivity or may indicate a co-infection.

Performance characteristics for Influenza A were established when Influenza A Hl-2009, A HI, and A H3 were the predominant Influenza A viruses in circulation. Performance of detecting Influenza A may vary if other Influenza A strains are circulating or a novel Influenza A virus e .merges,. I~f infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K123620: Filmarray Respiratory Panel (RP).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K123620

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026