510(k) · Cleared
ClearedQIAstat-Dx Respiratory Panel
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
QIAGEN GmbH · OCC Respiratory Virus Panel Nucleic Acid Assay System
Cleared on May 18, 2019 as substantially equivalent to K123620 (Biofire Diagnostics, Inc., 2013) after a 148-day review. Its family tree reaches back 6 generations to 2008.
- Decided
- May 18, 2019
- Review
- 148 days
- Class
- Class II
- Product code
- OCC
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 6 generations to DEN070013 (2008). 7 ancestors in all. No later device cites it as a predicate yet.
- Root · 6 generations backDEN070013 ID-Tag Respiratory Viral PanelLuminex Molecular Diagnostics, Inc. · 2008
- 5 generations backK081483 Xtag Respiratory Viral PanelLuminex Molecular Diagnostics, Inc. · 2008 · Recalled
- 4 generations backK091667 Xtag Respiratory Viral Panel, Models I019A0110, I019C011, I019D0112…Luminex Molecular Diagnostics, Inc. · 2009 · Recalled
- 3 generations backK103175 Filmarray Respiratory Panel (RP)Idaho Technology, Inc. · 2011
- 2 generations backK120267 Filmarray RP PanelIdaho Technology, Inc. · 2012
- 1 generation backK123620 Filmarray Respiratory Panel (RP)Biofire Diagnostics, Inc. · 2013
- This deviceK183597 QIAstat-Dx Respiratory PanelQIAGEN GmbH · 2019
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −6 | DEN070013 | ID-Tag Respiratory Viral Panel | Luminex Molecular Diagnostics, Inc. | 2008 | — |
| −5 | K081483 | Xtag Respiratory Viral Panel | Luminex Molecular Diagnostics, Inc. | 2008 | 1 |
| −4 | K091667 | Xtag Respiratory Viral Panel, Models I019A0110, I019C011, I019D0112… | Luminex Molecular Diagnostics, Inc. | 2009 | 1 |
| −3 | K103175 | Filmarray Respiratory Panel (RP) | Idaho Technology, Inc. | 2011 | — |
| −2 | K120267 | Filmarray RP Panel | Idaho Technology, Inc. | 2012 | — |
| −1 | K123620 | Filmarray Respiratory Panel (RP) | Biofire Diagnostics, Inc. | 2013 | — |
| 0 (this device) | K183597 | QIAstat-Dx Respiratory Panel | QIAGEN GmbH | 2019 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2011 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in OCC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new OCC device
- #1 K262747 ID NOW Influenza A & B 2; ID NOW RSV; ID NOW…score 15.2 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K123620 | Filmarray Respiratory Panel (RP)Summary page 5…Predicate Device Manufacturer Product Name 510(k) No. BioFire Diagnostics, Inc. FilmArray® Respiratory Panel (RP) K123620 QIAGEN 510(K) Premarket Notification QIAstat-Dx® Respir… | 2013 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 11 years and passes through 4 companies; a new generation was cleared every 1.2 years on median.
- 2 of the 7 ancestors have a recall of any class.
- 3 other devices also cite its predicate K123620.
How does its category compare?
All OCC devicesOCC Respiratory Virus Panel Nucleic Acid Assay System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2019, this device’s year, is highlighted.
FDA decisions per year · OCC
1in 2024
2026 to Oct 25, outlined.
Median 510(k) review, days · OCC
31in 2019
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · OCC
1in 2024
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · OCC
2in 2025
MAUDE reports naming product code OCC, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 0 |
| 2008 | 4 | 28 | 0 | 0 |
| 2009 | 6 | 24 | 0 | 0 |
| 2010 | 2 | — | 0 | 0 |
| 2011 | 9 | 82 | 0 | 2 |
| 2012 | 6 | 125 | 0 | 1 |
| 2013 | 6 | 79 | 0 | 0 |
| 2014 | 4 | 94 | 1 | 52 |
| 2015 | 8 | 85 | 0 | 1 |
| 2016 | 7 | 92 | 0 | 36 |
| 2017 | 10 | 136 | 0 | 1 |
| 2018 | 4 | 134 | 1 | 1 |
| 2019 | 5 | 31 | 1 | 2 |
| 2020 | 2 | — | 0 | 17 |
| 2021 | 1 | — | 0 | 686 |
| 2022 | 1 | — | 0 | 2 |
| 2023 | 2 | — | 0 | 1 |
| 2024 | 1 | — | 1 | 3 |
| 2025 | 0 | — | 0 | 2 |
| 2026 | 3 | 130 | 0 | — |
What is it indicated for?
The QIAstat-Dx Respiratory Panel is a multiplexed nucleic acid test intended for use with QIAstat-Dx system for the simultaneous qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) eluted in Universal Transport Media (UTM) obtained from individuals suspected of respiratory tract infections. The following organism types and subtypes are identified using the QIAstat-Dx Respiratory…
Read the full statement (463 words)
The QIAstat-Dx Respiratory Panel is a multiplexed nucleic acid test intended for use with QIAstat-Dx system for the simultaneous qualitative detection and identification of multiple respiratory viral and bacterial nucleic acids in nasopharyngeal swabs (NPS) eluted in Universal Transport Media (UTM) obtained from individuals suspected of respiratory tract infections. The following organism types and subtypes are identified using the QIAstat-Dx Respiratory Panel: Adenovirus, Coronavirus 229E, Coronavirus HKU1, Coronavirus NL63, Coronavirus OC43, Human Metapneumovirus A+B, Influenza A, Influenza A H1, Influenza A H3, Influenza A H1N1/pdm09, Influenza B, Parainfluenza Virus 1, Parainfluenza Virus 2, Parainfluenza Virus 3, Parainfluenza Virus 4, Rhinovirus/Enterovirus, Respiratory Syncytial Virus A+B, Bordetella pertussis, Chlamydophila pneumoniae and Mycoplasma pneumoniae.
The detection and identification of specific viral and bacterial nucleic acids from individuals presenting with signs and symptoms of a respiratory infection aids in the diagnosis of respiratory infection if used in conjunction with other clinical and epidemiological information. The results of this test should not be used as the sole basis for diagnosis, treatment or other management decisions. Negative results in the setting of a respiratory illness may be due to infection with pathogens that are not detected by the test or lower respiratory tract infection that is not detected by a nasopharyngeal swab specimen. Positive results do not rule out co-infection with other organisms: the agent(s) detected by the QIAstat-Dx Respiratory Panel may not be the definite cause of disease. Additional laboratory testing (e.g., bacterial and viral culture, immunofluorescence and radiography) may be necessary when evaluating a patient with possible respiratory tract infection.
Due to the small number of positive specimens collected for certain organisms during the prospective clinical study, performance characteristics for Bordetella pertussis and Parainfluenza Virus 1 were established primarily with retrospective clinical specimens. Performance characteristics for Chlamydophila pneumoniae, Parainfluenza Virus 2, Parainfluenza Virus 4, Influenza A subtype H1 and Coronavirus 229E were established primarily using contrived clinical specimens.
Due to the genetic similarity between Human Rhinovirus and Enterovirus, the QIAstat-Dx Respiratory Panel cannot reliably differentiate them. A positive QIAstat-Dx Respiratory Panel Rhinovirus/Enterovirus result should be followed-up using an alternate method (e.g., cell culture or sequence analysis).
Performance characteristics for Influenza A were established when Influenza A H1N1-2009 and A H3 were the predominant Influenza A viruses in circulation. Performance of detecting Influenza A may vary if other Influenza A strains are circulating or a novel Influenza A virus emerges. If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health departments for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K183597 · printed Nov 4, 2026
Cite this page
510(k) Report. “K183597: QIAstat-Dx Respiratory Panel.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K183597
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026