510(k) Report

510(k) · Cleared

Cleared

Quickprofile Single Drug of Abuse Device and Quick Profile Multi-Drugs of Abuse Screen Device

K130213510(k) TraditionalCleared Oct 18, 2013262-day reviewClass II21 CFR 862.3870

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Lumiquick Diagnostics, Inc. · LDJ Enzyme Immunoassay, Cannabinoids

Cleared on Oct 18, 2013 as substantially equivalent to K061718 (Innovacon, Inc., 2006) after a 262-day review. Its family tree reaches back 5 generations to 1995.

Decided
Oct 18, 2013
Review
262 days
Class
Class II
Product code
LDJ
Recalls
None
Predicates
1

Its longest line of descent runs back 5 generations to K945610 (1995). 26 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 5 generations backK945610 Abusign (TM) Amp, Doa-Amp, Accusign (TM) Amp-Doa-Amp, Biosign (TM)…Princeton BioMeditech Corp. · 1995
  2. 4 generations backK011673 Acon Amp One Step Amphetamine Test Strip, Acon Amp One Step…ACON Laboratories, Inc. · 2001
  3. 3 generations backK020313 Acon One Step Multi-Drug Multi-Line Screen Test Card; Acon One Step…Acon Laboratories Co. · 2002
  4. 2 generations backK031759 Acon Spectrum Multi-Drug Multi-Line Drug Screen Test Card and Acon…ACON Laboratories, Inc. · 2003 · Recalled
  5. 1 generation backK061718 Innovacon Spectrum II Test CardInnovacon, Inc. · 2006 · Recalled
  6. This deviceK130213 Quickprofile Single Drug of Abuse Device and Quick Profile…Lumiquick Diagnostics, Inc. · 2013
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K945610Abusign (TM) Amp, Doa-Amp, Accusign (TM) Amp-Doa-Amp, Biosign (TM) Amp, Doa-AmpPrinceton BioMeditech Corp.1995—
−4K011673Acon Amp One Step Amphetamine Test Strip, Acon Amp One Step Amphetamine Test…ACON Laboratories, Inc.2001—
−3K020313Acon One Step Multi-Drug Multi-Line Screen Test Card; Acon One Step Multi-Drug…Acon Laboratories Co.2002—
−2K031759Acon Spectrum Multi-Drug Multi-Line Drug Screen Test Card and Acon Spectrum…ACON Laboratories, Inc.20031
−1K061718Innovacon Spectrum II Test CardInnovacon, Inc.20066
0 (this device)K130213Quickprofile Single Drug of Abuse Device and Quick Profile Multi-Drugs of…Lumiquick Diagnostics, Inc.2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199620
199810
199910
200010
200170
200220
200320
200430
200510
200610
FIG. 1 — Lineage of K130213, 1995–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LDJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LDJ device

  1. #1 K232732 Rapid Marijuana (THC) Test Strip 20, Rapid…score 9.4 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K061718Innovacon Spectrum II Test Card
Summary page 2
…creen Devices are substantially equivalent to the predicate device noted below: Device Name: Innovacon Spectrum 11 5 10(k) numbers for predicate devices are: k06 1718 6-2 LumiQuic…
Innovacon, Inc.20066

Also cleared against K061718: 6 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 18 years and passes through 5 companies; a new generation was cleared every 3.3 years on median.
  • 2 of the 26 ancestors have a recall of any class.
  • 6 other devices also cite its predicate K061718.

How does its category compare?

All LDJ devices

LDJ Enzyme Immunoassay, Cannabinoids: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · LDJ

2in 2024

2008: 22009: 12011: 42012: 42013: 32014: 22015: 32016: 12019: 12023: 12024: 2
2007high 4 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LDJ

48in 2015

2011: 219 days median2012: 193 days median2013: 162 days median2015: 48 days median
2007high 219 (2011) · low 48 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LDJ

1in 2023

2008: 42023: 1
2007high 4 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LDJ

1in 2015

2015: 1
2007high 1 (2015) · low 0 (2007)2025

MAUDE reports naming product code LDJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20082—40
20091—00
20100—00
2011421900
2012419300
2013316200
20142—00
201534801
20161—00
20170—00
20180—00
20191—00
20200—00
20210—00
20220—00
20231—10
20242—00
20250—00
20260—0—

What is it indicated for?

LumiQuick's QuickProfile Single Drugs of Abuse and QuickProfile Multi-Drugs of Abuse Screen Devices are rapid chromatographic immunoassays for the qualitative and simultaneous detection of one to nine of the following drugs in a variety of combinations in human urine. Both devices are available in test strip, dip panel, cassette panel and cup formats. The designed cutoff concentrations and direct calibrator for these drugs are as follows: AMP Amphetamine…

Read the full statement (178 words)

LumiQuick's QuickProfile Single Drugs of Abuse and QuickProfile Multi-Drugs of Abuse Screen Devices are rapid chromatographic immunoassays for the qualitative and simultaneous detection of one to nine of the following drugs in a variety of combinations in human urine. Both devices are available in test strip, dip panel, cassette panel and cup formats. The designed cutoff concentrations and direct calibrator for these drugs are as follows:

AMP Amphetamine 1000 ng/ml

BAR Secobarbital 300 ng/ml

BZO Oxazepam 300 ng/ml

COC Benzoylecgonine 300 ng/ml

MTD Methadone 300 ng/ml

MAMP Methamphetamine 1000 ng/ml

OPI Morphine 300 ng/ml

PCP Phencyclidine 25 ng/ml

THC THC I 1-nor-A9-THC-9-COOH 50 ng/ml

These devices are intended for prescription use only. These assays provide only a preliminary analytical test result. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography / Mass Spectrometry (GC/MS) or Liquid Chromatography / Mass Spectrometry (LC/MS) are the preferred confirmatory method.

Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K130213: Quickprofile Single Drug of Abuse Device and Quick Profile Multi-Drugs of Abuse Screen Device.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K130213

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026