510(k) Report

510(k) · Cleared

Recalled · 2011

Innovacon Spectrum II Test Card

K061718510(k) SpecialCleared Nov 13, 2006147-day reviewClass II21 CFR 862.3100Recalls 6

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Innovacon, Inc. · DKZ Enzyme Immunoassay, Amphetamine

Cleared on Nov 13, 2006 as substantially equivalent to K031759 (ACON Laboratories, Inc., 2003) and 6 other predicates after a 147-day review. Its family tree reaches back 4 generations to 1995. 80 later devices descend from it. It has been recalled 6 times.

Decided
Nov 13, 2006
Review
147 days
Class
Class II
Product code
DKZ
Recalls
6
Predicates
7

Its longest line of descent runs back 4 generations to K945610 (1995). 25 ancestors in all. 80 later devices descend from it, 7 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2006: 1 other ancestor2005: 1 other ancestor2004: 3 other ancestors2003: 2 other ancestors2002: 2 other ancestors2001: 7 other ancestors2000: 1 other ancestor1999: 1 other ancestor1998: 1 other ancestor1996: 2 other ancestors2014: 4 descendants cleared2016: 9 descendants cleared2013: 2 descendants cleared2015: 9 descendants cleared2011: 1 descendant cleared2012: 1 descendant cleared2018: 9 descendants cleared2020: 4 descendants cleared2023: 7 descendants cleared2019: 6 descendants cleared2017: 2 descendants cleared2025: 7 descendants cleared2021: 3 descendants cleared2024: 9 descendants cleared2022: 3 descendants cleared2026: 4 descendants clearedK945610 · Abusign (TM) Amp, Doa-Amp, Accusign (TM) Amp-Doa-Amp, Biosign (TM) Amp, Doa-Amp · Princeton BioMeditech Corp. · 1995K011673 · Acon Amp One Step Amphetamine Test Strip, Acon Amp One Step Amphetamine Test Device · ACON Laboratories, Inc. · 2001K020313 · Acon One Step Multi-Drug Multi-Line Screen Test Card; Acon One Step Multi-Drug Multi-Line Screen Test Device · Acon Laboratories Co. · 2002K031759 · Acon Spectrum Multi-Drug Multi-Line Drug Screen Test Card and Acon Spectrum Multi-Drug Multi-Line Drug Screen Test Card · ACON Laboratories, Inc. · 2003 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 6 other ancestors2 generations back: 1 other ancestor3 generations back: 9 other ancestors4 generations back: 5 other ancestors 1 generation forward: 7 descendants2 generations forward: 23 descendants3 generations forward: 20 descendants4 generations forward: 16 descendants5 generations forward: 9 descendants6 generations forward: 5 descendants K031759 · Acon Spectrum Multi-Drug Multi-Line Drug Screen Test Card and Acon Spectrum Multi-Drug Multi-Line Drug Screen Test Card · ACON Laboratories, Inc. · 2003 · recalledK020313 · Acon One Step Multi-Drug Multi-Line Screen Test Card; Acon One Step Multi-Drug Multi-Line Screen Test Device · Acon Laboratories Co. · 2002K011673 · Acon Amp One Step Amphetamine Test Strip, Acon Amp One Step Amphetamine Test Device · ACON Laboratories, Inc. · 2001K945610 · Abusign (TM) Amp, Doa-Amp, Accusign (TM) Amp-Doa-Amp, Biosign (TM) Amp, Doa-Amp · Princeton BioMeditech Corp. · 19950+5
  1. Root · 4 generations backK945610 Abusign (TM) Amp, Doa-Amp, Accusign (TM) Amp-Doa-Amp, Biosign (TM)…Princeton BioMeditech Corp. · 1995
  2. 3 generations backK011673 Acon Amp One Step Amphetamine Test Strip, Acon Amp One Step…ACON Laboratories, Inc. · 2001
  3. 2 generations backK020313 Acon One Step Multi-Drug Multi-Line Screen Test Card; Acon One Step…Acon Laboratories Co. · 2002
  4. 1 generation backK031759 Acon Spectrum Multi-Drug Multi-Line Drug Screen Test Card and Acon…ACON Laboratories, Inc. · 2003 · Recalled
  5. This deviceK061718 Innovacon Spectrum II Test CardInnovacon, Inc. · 2006
  6. Descendant · cites it directly · 55 of its ownK122809 Advin Multi-Drug Screen Test Cassette, Dip Card and CupAdvin Biotech · 2013
  7. Descendant · 2 generations on · 16 of its ownK140748 One Step Single/Multi-Drug Test Cup, One Step Single/Multi-Drug Test…Co-Innovation Biotech Co., Ltd. · 2014
  8. Descendant · 2 generations on · 13 of its ownK182701 Wondfo T-Cup Multi-Drug Urine Test CupGuangzhou Wondfo Biotech Co., Ltd. · 2018 · Recalled
  9. Descendant · 2 generations on · 13 of its ownK142800 Rapid Single/Multi-drug Test Cup and Rapid Single/Multi-drug Test…Co-Innovation Biotech Co., Ltd. · 2014
  10. Descendant · 2 generations on · 12 of its ownK181768 AssureTech Panel Dip Tests, AssureTech Quick Cup TestsAssure Tech. (Hangzhou) Co, Ltd. · 2018

Showing 5 of 80 descendants. See all descendants

Descendants cleared per year

2011: 12012: 12013: 22014: 42015: 92016: 92017: 22018: 92019: 62020: 42021: 32022: 32023: 72024: 92025: 72026: 4
2011high 9 (2015) · low 1 (2011)2026

Most influential descendants

Showing 5 of 80 · See all
  • K122809Advin Multi-Drug Screen Test Cassette, Dip Card and CupAdvin Biotech · 2013 · cites it directly55descendants
  • K140748One Step Single/Multi-Drug Test Cup, One Step Single/Multi-Drug Test DipcardCo-Innovation Biotech Co., Ltd. · 2014 · 2 generations on16descendants
  • K182701Wondfo T-Cup Multi-Drug Urine Test CupGuangzhou Wondfo Biotech Co., Ltd. · 2018 · 2 generations on13descendants
  • K142800Rapid Single/Multi-drug Test Cup and Rapid Single/Multi-drug Test DipcardCo-Innovation Biotech Co., Ltd. · 2014 · 2 generations on13descendants
  • K181768AssureTech Panel Dip Tests, AssureTech Quick Cup TestsAssure Tech. (Hangzhou) Co, Ltd. · 2018 · 2 generations on12descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K945610Abusign (TM) Amp, Doa-Amp, Accusign (TM) Amp-Doa-Amp, Biosign (TM) Amp, Doa-AmpPrinceton BioMeditech Corp.1995—
−3K011673Acon Amp One Step Amphetamine Test Strip, Acon Amp One Step Amphetamine Test…ACON Laboratories, Inc.2001—
−2K020313Acon One Step Multi-Drug Multi-Line Screen Test Card; Acon One Step Multi-Drug…Acon Laboratories Co.2002—
−1K031759Acon Spectrum Multi-Drug Multi-Line Drug Screen Test Card and Acon Spectrum…ACON Laboratories, Inc.20031
0 (this device)K061718Innovacon Spectrum II Test CardInnovacon, Inc.2006—
+1K122809Advin Multi-Drug Screen Test Cassette, Dip Card and CupAdvin Biotech2013—
+2K140748One Step Single/Multi-Drug Test Cup, One Step Single/Multi-Drug Test DipcardCo-Innovation Biotech Co., Ltd.2014—
+2K182701Wondfo T-Cup Multi-Drug Urine Test CupGuangzhou Wondfo Biotech Co., Ltd.20181
+2K142800Rapid Single/Multi-drug Test Cup and Rapid Single/Multi-drug Test DipcardCo-Innovation Biotech Co., Ltd.2014—
+2K181768AssureTech Panel Dip Tests, AssureTech Quick Cup TestsAssure Tech. (Hangzhou) Co, Ltd.2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199620
199810
199910
200010
200170
200220
200320
200430
200510
200610

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
201210
201320
201440
201590
201690
201720
201890
201960
202040
202130
202230
202370
202490
202570
202640
FIG. 1 — Lineage of K061718, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DKZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DKZ device

  1. #1 K233019 AllTest Multi-Drug Rapid Test Cup; AllTest…score 14.6 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (7)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K031759Acon Spectrum Multi-Drug Multi-Line Drug Screen Test Card and Acon Spectrum Multi-Drug…
Summary page 5
…Number Code Number Code ACON Spectrum Multi-Drug Multi-Line Drug Screen Test K031759 LDJ DIO DKZ Card and Test Card with Integrated Split E-Z Key Cup DJG LCM JXM…
ACON Laboratories, Inc.20031
*K032903Acon Coc-150 One Step Cocaine Test Strip & Acon Coc-150 One Step Cocaine Test Device
Summary page 5
…Cup DJG LCM JXM DJR DIS LFG ACON COC-150 One Step Cocaine Test Strip/Test Device K032903 DIO ACON mAMP-500 One Step Methamphetamine Test K033299 LAF Strip/Test Device ACON PPX One Step Pr…
ACON Laboratories, Inc.2004—
*K033299Acon Mamp One Step Methamphetamine Test Strip and Test Device
Summary page 5
…DJR DIS LFG ACON COC-150 One Step Cocaine Test Strip/Test Device K032903 DIO ACON mAMP-500 One Step Methamphetamine Test K033299 LAF Strip/Test Device ACON PPX One Step Propoxyphene Test Strip/Test Device K040445 JXN ACON AMP 300 One Step A…
ACON Laboratories, Inc.2003—
*K040445Acon PPX One Step Propoxyphene Test Strip and Acon PPX Step Propoxyphene Test Device
Summary page 5
…903 DIO ACON mAMP-500 One Step Methamphetamine Test K033299 LAF Strip/Test Device ACON PPX One Step Propoxyphene Test Strip/Test Device K040445 JXN ACON AMP 300 One Step Amphetamine Test Strip/Test K041822 DKZ Device ACON OXY |I One Step Oxycodone Test S…
ACON Laboratories, Inc.2004—
*K041822Acon Amp 300 One Step Amphetamine Test Strip and Device
Summary page 5
…033299 LAF Strip/Test Device ACON PPX One Step Propoxyphene Test Strip/Test Device K040445 JXN ACON AMP 300 One Step Amphetamine Test Strip/Test K041822 DKZ Device ACON OXY |I One Step Oxycodone Test Strip/Test Device K043507 DJG ACON BUP One Step Buprenorphine Test…
ACON Laboratories, Inc.2004—
*K043507Acon Oxy II One Step Oxycodone Test Strip; Test Device
Summary page 5
…st Device K040445 JXN ACON AMP 300 One Step Amphetamine Test Strip/Test K041822 DKZ Device ACON OXY |I One Step Oxycodone Test Strip/Test Device K043507 DJG ACON BUP One Step Buprenorphine Test Strip/Test Device K060466 DJG…
ACON Laboratories, Inc.2005—
*K060466Acon Bup One Step Buprenorphine Test Strip and Acron Bup One Step Buprenorphine Test…
Summary page 5
…Test K041822 DKZ Device ACON OXY |I One Step Oxycodone Test Strip/Test Device K043507 DJG ACON BUP One Step Buprenorphine Test Strip/Test Device K060466 DJG 5 DEPARTMENT OF HEALTH & HUMAN SERVICES…
ACON Laboratories, Inc.2006—

Also cleared against K031759: 2 other devices →

What descends from it? (80)

See all 80

7 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K122809Advin Multi-Drug Screen Test Cassette, Dip Card and CupAdvin Biotech201355
K142396Chemtrue BUP/TCA Single/Multi-Panel Drug Screen Dip Card/Cup/Cassette TestsChemtron Biotech, Inc.201410
K111322Chemtrue Single/Multi-Panel Drug Screen Cassette/DipcardChemtron Biotech, Inc.20116
K153192Chemtrue Multi-Panel Drug Screen Dip Card/Cup Tests, Chemtrue Multi-Panel Drug Screen…Chemtron Biotech, Inc.20162
K152590ACCU NEWS Drug Screening Test Card/Urine CupCoretests, Inc.20160
K143599Chemtrue Drug Screen Cup Tests and Chemtrue Drug Screen Cup Tests with OPI 2000Chemtron Biotech, Inc.20150
K130213Quickprofile Single Drug of Abuse Device and Quick Profile Multi-Drugs of Abuse Screen…Lumiquick Diagnostics, Inc.20130

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 10, 2011Not classifiedZ-2881-2011Vendor change controlAlere is initiating a voluntary recall on the listed lots of MULTI-DRUG 9 PANEL INTEGRATED CUP manufactured by Alere for sole use by Calloway Labs due to a mistake that occurred while processing the order, these products were incorrectly…
Apr 27, 2009Not classifiedZ-0006-2010Process controlThe cup associated with these lots of product may have a pin-sized hole in the floor of the cup. As a result, there is a potential for affected lots of cups to leak urine after an extended period of time.
Apr 27, 2009Not classifiedZ-0007-2010Process controlThe cup associated with these lots of product may have a pin-sized hole in the floor of the cup. As a result, there is a potential for affected lots of cups to leak urine after an extended period of time.
Apr 27, 2009Not classifiedZ-0008-2010Process controlThe cup associated with these lots of product may have a pin-sized hole in the floor of the cup. As a result, there is a potential for affected lots of cups to leak urine after an extended period of time.
Apr 27, 2009Not classifiedZ-0009-2010Process controlThe cup associated with these lots of product may have a pin-sized hole in the floor of the cup. As a result, there is a potential for affected lots of cups to leak urine after an extended period of time.
Apr 27, 2009Not classifiedZ-0010-2010Process controlThe cup associated with these lots of product may have a pin-sized hole in the floor of the cup. As a result, there is a potential for affected lots of cups to leak urine after an extended period of time.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 4 companies; a new generation was cleared every 3.3 years on median.
  • 1 of the 25 ancestors has a recall of any class.
  • Of its descendants, K122809 (Advin Biotech, 2013) has gone furthest, with 55 descendants of its own.
  • 2 other devices also cite its predicate K031759.

How does its category compare?

All DKZ devices

DKZ Enzyme Immunoassay, Amphetamine: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DKZ

1in 2024

2008: 22009: 22010: 42011: 62012: 52013: 82014: 82015: 52016: 52018: 32019: 32020: 12023: 22024: 1
2007high 8 (2013) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DKZ

180in 2019

2010: 233 days median2011: 123 days median2012: 55 days median2013: 103 days median2014: 128 days median2015: 107 days median2016: 215 days median2018: 128 days median2019: 180 days median
2007high 233 (2010) · low 55 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DKZ

1in 2017

2007: 12008: 32009: 52011: 42013: 12014: 32017: 1
2007high 5 (2009) · low 0 (2010)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · DKZ

2in 2025

2008: 122009: 12010: 12011: 22013: 12014: 92015: 82016: 82021: 32022: 42023: 72024: 12025: 2
2007high 12 (2008) · low 0 (2007)2025

MAUDE reports naming product code DKZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—10
20082—312
20092—51
2010423301
2011612342
201255500
2013810311
2014812839
2015510708
2016521508
20170—10
2018312800
2019318000
20201—00
20210—03
20220—04
20232—07
20241—01
20250—02
20260—0—

What is it indicated for?

The Innovacon Spectrum II Test Card and Innovacon Spectrum II Test Card with Integrated Cups (Innovacon 014 Cup or E-Z Action Multi-Drug Test Cup and Innovacon 022 Cup or E-Z Start Multi-Drug Test Cup) are rapid chromatographic immunoassays for the qualitative and simultaneous detection of Marijuana, Cocaine, Methylenedioxymethamphetamine, Amphetamine, Morphine, Opiates, Methadone, Methamphetamine, Phencyclidine, Benzodiazepine, Oxycodone, Propoxyphene…

Read the full statement (256 words)

The Innovacon Spectrum II Test Card and Innovacon Spectrum II Test Card with Integrated Cups (Innovacon 014 Cup or E-Z Action Multi-Drug Test Cup and Innovacon 022 Cup or E-Z Start Multi-Drug Test Cup) are rapid chromatographic immunoassays for the qualitative and simultaneous detection of Marijuana, Cocaine, Methylenedioxymethamphetamine, Amphetamine, Morphine, Opiates, Methadone, Methamphetamine, Phencyclidine, Benzodiazepine, Oxycodone, Propoxyphene, Tricyclic Antidepressants, Buprenorphine and Barbiturate in human urine at the cutoff concentrations of:

1,000 ng/mL or 300 ng/mL for Amphetamine,

300 ng/mL for Barbiturate,

300 ng/mL for Benzodiazepines,

300 ng/mL or 150 ng/mL for Cocaine,

50 ng/mL for Marijuana,

300 ng/mL for Methadone,

500 ng/mL or 1,000 ng/mL for Methamphetamine,

500 ng/mL for Methylenedioxymethamphetamine,

300 ng/mL for Morphine,

2,000 ng/mL for Opiates,

100 ng/mL for Oxycodone,

25 ng/mL for Phencyclidine,

300 ng/mL for Propoxyphene,

10 ng/mL for Buprenorphine, and

1,000 ng/mL for Tricyclic Antidepressants.

Configurations of the Innovacon Spectrum II Test Card and Innovacon Spectrum II Test Card with Integrated Cups (Innovacon 014 Cup or E-Z Action Multi-Drug Test Cup and Innovacon 022 Cup or E-Z Start Multi-Drug Test Cup) can consist of any combination of the above listed drug analytes. They are intended for healthcare professionals including professionals at point-of-care sites.

This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K061718: Innovacon Spectrum II Test Card.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K061718

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026