Hemoglobin A1C, Hemoglobin A1C Calibrators, Hemoglobin A1C Controls

K130255510(k) TraditionalCleared Feb 28, 2014392-day reviewClass II21 CFR 862.1373Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Abbott Laboratories · PDJ Hemoglobin A1c Test System

Cleared on Feb 28, 2014 as substantially equivalent to DEN130002 (Roche Diagnostics, 2013) and 2 other predicates after a 392-day review. Its family tree reaches back 2 generations to 1994. 4 later devices descend from it. It has been recalled 3 times.

Decided
Feb 28, 2014
Review
392 days
Class
Class II
Product code
PDJ
Recalls
3
Predicates
3

Its longest line of descent runs back 2 generations to K934070 (1994). 4 ancestors in all. 4 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2013: 1 other ancestor2010: 1 other ancestor2015: 1 descendant cleared2017: 1 descendant cleared2018: 1 descendant cleared2021: 1 descendant clearedK934070 · Tina-Quant Hemoglobin A1C · Boehringer Mannheim Corp. · 1994K052101 · C.F.a.S. (Calibrator for Automated Systems) HBA1C · Roche Diagnostics · 20051990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors 1 generation forward: 3 descendants2 generations forward: 1 descendant K052101 · C.F.a.S. (Calibrator for Automated Systems) HBA1C · Roche Diagnostics · 2005K934070 · Tina-Quant Hemoglobin A1C · Boehringer Mannheim Corp. · 19940
  1. Root · 2 generations backK934070 Tina-Quant Hemoglobin A1CBoehringer Mannheim Corp. · 1994
  2. 1 generation backK052101 C.F.a.S. (Calibrator for Automated Systems) HBA1CRoche Diagnostics · 2005
  3. This deviceK130255 Hemoglobin A1C, Hemoglobin A1C Calibrators, Hemoglobin A1C ControlsAbbott Laboratories · 2014
  4. Descendant · cites it directly · 1 of its ownK171771 ADVIA Chemistry Enzymatic Hemoglobin A1c (A1c_E) AssaySiemens Healthcare Diagnostics, Inc. · 2017 · Recalled
  5. Descendant · 2 generations on · 0 of its ownK200256 ADVIA Chemistry Enzymatic Hemoglobin Ale (Alc_E) AssaySiemens Healthcare Diagnostics, Inc. · 2021 · Recalled
  6. Descendant · cites it directly · 0 of its ownK173206 SEKURE HbA1c AssaySekisui Diagnostics P.E.I., Inc. · 2018
  7. Descendant · cites it directly · 0 of its ownK140654 Hemoglobin A1c Assay, Hemoglobin A1c Calibrators, Hemoglobin A1c…Abbott Laboratories · 2015 · Recalled

Most influential descendants

Showing 4 of 4 · See all
  • K171771ADVIA Chemistry Enzymatic Hemoglobin A1c (A1c_E) AssaySiemens Healthcare Diagnostics, Inc. · 2017 · cites it directly1descendant
  • K200256ADVIA Chemistry Enzymatic Hemoglobin Ale (Alc_E) AssaySiemens Healthcare Diagnostics, Inc. · 2021 · 2 generations on0descendants
  • K173206SEKURE HbA1c AssaySekisui Diagnostics P.E.I., Inc. · 2018 · cites it directly0descendants
  • K140654Hemoglobin A1c Assay, Hemoglobin A1c Calibrators, Hemoglobin A1c ControlsAbbott Laboratories · 2015 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K934070Tina-Quant Hemoglobin A1CBoehringer Mannheim Corp.1994—
−1K052101C.F.a.S. (Calibrator for Automated Systems) HBA1CRoche Diagnostics2005—
0 (this device)K130255Hemoglobin A1C, Hemoglobin A1C Calibrators, Hemoglobin A1C ControlsAbbott Laboratories2014—
+1K171771ADVIA Chemistry Enzymatic Hemoglobin A1c (A1c_E) AssaySiemens Healthcare Diagnostics, Inc.20171
+2K200256ADVIA Chemistry Enzymatic Hemoglobin Ale (Alc_E) AssaySiemens Healthcare Diagnostics, Inc.20212
+1K173206SEKURE HbA1c AssaySekisui Diagnostics P.E.I., Inc.2018—
+1K140654Hemoglobin A1c Assay, Hemoglobin A1c Calibrators, Hemoglobin A1c ControlsAbbott Laboratories20152

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
201010
201310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
201710
201810
202110
FIG. 1 — Lineage of K130255, 1994–2021Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in PDJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new PDJ device

  1. #1 K252749 Medconn 8K Glycated Hemoglobin Test Systemscore 11.5 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*DEN130002Tina-Quant HBA1C Gen. 2 Test System
Our reading of the summary gave K121291, which is not in FDA’s 510(k) list. The device the summary names is FDA’s De Novo DEN130002, Tina-Quant HBA1C Gen. 2 Test System (Roche Diagnostics), granted May 23, 2013; older De Novo requests were first filed under a 510(k) number, so this links to the grant.
Roche Diagnostics2013—
*K052101C.F.a.S. (Calibrator for Automated Systems) HBA1C
Summary page 2
…. Predicate Device Reagents COBAS INTEGRA 800 Tina-quant HbAlcDx Gen.2 assay (k121291) Calibrators Roche C.f.a.s. (Calibrator for Automated Systems) HbAlc (k052101) Controls Roche PreciControl HbAlc norm and PreciControl HbAlc path (k103099) 4. Description of Device Reagents Th…
Roche Diagnostics2005—
*K103099Precicontrol HBA1C Norm; Precicontrol HBA1C Path
Summary page 2
…(k121291) Calibrators Roche C.f.a.s. (Calibrator for Automated Systems) HbAlc (k052101) Controls Roche PreciControl HbAlc norm and PreciControl HbAlc path (k103099) 4. Description of Device Reagents The Hemoglobin A1 c Reagent Kit contains: Component Number of Bo…
Roche Diagnostics2010—

Also cleared against DEN130002: 6 other devices →

What descends from it? (4)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K171771ADVIA Chemistry Enzymatic Hemoglobin A1c (A1c_E) AssaySiemens Healthcare Diagnostics, Inc.20171
K173206SEKURE HbA1c AssaySekisui Diagnostics P.E.I., Inc.20180
K140654Hemoglobin A1c Assay, Hemoglobin A1c Calibrators, Hemoglobin A1c ControlsAbbott Laboratories20150

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 9, 2022Class IIZ-1455-2022Under Investigation by firmThere is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be generated using the ARCHITECT and Alinity c Hemoglobin A1c assay.
Jun 9, 2022Class IIZ-1454-2022Under Investigation by firmThere is a potential for falsely elevated Hemoglobin %A1c and Hemoglobin A1c results to be generated using the ARCHITECT and Alinity c Hemoglobin A1c assay.
Apr 30, 2014Class IIZ-0135-2015Under Investigation by firmHbA1c Calibrators LN 4P52-02, lot 45063UQ12 may contain a value sheet for a different HbA1c calibrator lot. If calibrator lot 45063UQ12 is used in the calibration of the HbA1c assay using values from an incorrect calibrator value sheet…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 20 years and passes through 3 companies; a new generation was cleared every 10.9 years on median.
  • None of the 4 ancestors has a recall.
  • Of its descendants, K171771 (Siemens Healthcare Diagnostics, Inc., 2017) has gone furthest, with 1 descendant of its own.
  • 6 other devices also cite its predicate DEN130002.

How does its category compare?

All PDJ devices

PDJ Hemoglobin A1c Test System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · PDJ

2in 2025

2013: 22014: 22015: 42016: 22017: 32018: 52019: 12020: 12021: 22024: 12025: 22026 (partial year): 2
2007high 5 (2018) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · PDJ

203in 2018

2015: 234 days median2017: 173 days median2018: 203 days median
2007high 234 (2015) · low 173 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · PDJ

1in 2022

2016: 12022: 12026 (partial year): 1
2007high 1 (2016) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · PDJ

42in 2025

2013: 92014: 102015: 22017: 12020: 52021: 62022: 192023: 162024: 472025: 42
2007high 47 (2024) · low 0 (2007)2025

MAUDE reports naming product code PDJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20100—00
20110—00
20120—00
20132—09
20142—010
2015423402
20162—10
2017317301
2018520300
20191—00
20201—05
20212—06
20220—119
20230—016
20241—047
20252—042
20262—1—

What is it indicated for?

The Hemoglobin A Ic assay is used in clinical laboratories for the quantitative in vitro measurement of percent hemoglobin Ale (NGSP) or HbAlc fraction mmol/mol (]FCC) in human whole blood and hemolysate on the ARCHITECT c 8000 System. Hemoglobin A Ic measurements are used as an aid in the diagnosis of diabetes mellitus, as an aid to identify patients who may be at risk for developing diabetes mellitus, and for the monitoring of long-term blood glucose…

Read the full statement (128 words)

The Hemoglobin A Ic assay is used in clinical laboratories for the quantitative in vitro measurement of percent hemoglobin Ale (NGSP) or HbAlc fraction mmol/mol (]FCC) in human whole blood and hemolysate on the ARCHITECT c 8000 System.

Hemoglobin A Ic measurements are used as an aid in the diagnosis of diabetes mellitus, as an aid to identify patients who may be at risk for developing diabetes mellitus, and for the monitoring of long-term blood glucose control in individuals with diabetes mellitus.

The Hemoglobin Ale Calibrators are for use in the calibration of the Hemoglobin Ale assay on the ARCHITECT c 8000 System.

The Hemoglobin Ale Controls are used for the estimation of test precision and the detection of systematic analytical deviations of the ARCHITECT c 8000 System.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K130255: Hemoglobin A1C, Hemoglobin A1C Calibrators, Hemoglobin A1C Controls.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K130255

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026