510(k) Report

510(k) · Cleared

Cleared

Magstim Double 70-2 Coil

K130403510(k) SpecialCleared Aug 23, 2013185-day reviewClass II21 CFR 882.1870

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

The Magstim Company, Ltd. · GWF Stimulator, Electrical, Evoked Response

Cleared on Aug 23, 2013 as substantially equivalent to K060847 (Magstim Company Us, LLC, 2006) and 1 other predicate after a 185-day review. Its family tree reaches back 2 generations to 1991.

Decided
Aug 23, 2013
Review
185 days
Class
Class II
Product code
GWF
Recalls
None
Predicates
2

Its longest line of descent runs back 2 generations to K905059 (1991). 5 ancestors in all. No later device cites it as a predicate yet.

Other ancestors, by year cleared 2006: 1 other ancestor2000: 1 other ancestor1993: 1 other ancestorK905059 · High Speed Magnetic Stimulator · Cadwell Laboratories, Inc. · 1991K051864 · Magstim RAPID2, Magstim Super RAPID2, Models 3004-000, 3005-00 · The Magstim Company, Ltd. · 2005 · recalled199019952000200520102015
Other ancestors, by generation 1 generation back: 1 other ancestor2 generations back: 2 other ancestors K051864 · Magstim RAPID2, Magstim Super RAPID2, Models 3004-000, 3005-00 · The Magstim Company, Ltd. · 2005 · recalledK905059 · High Speed Magnetic Stimulator · Cadwell Laboratories, Inc. · 19910
  1. Root · 2 generations backK905059 High Speed Magnetic StimulatorCadwell Laboratories, Inc. · 1991
  2. 1 generation backK051864 Magstim RAPID2, Magstim Super RAPID2, Models 3004-000, 3005-00The Magstim Company, Ltd. · 2005 · Recalled
  3. This deviceK130403 Magstim Double 70-2 CoilThe Magstim Company, Ltd. · 2013
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K905059High Speed Magnetic StimulatorCadwell Laboratories, Inc.1991—
−1K051864Magstim RAPID2, Magstim Super RAPID2, Models 3004-000, 3005-00The Magstim Company, Ltd.20052
0 (this device)K130403Magstim Double 70-2 CoilThe Magstim Company, Ltd.2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
200010
200610
FIG. 1 — Lineage of K130403, 1991–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GWF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GWF device

  1. #1 K263010 ISIS Headboxes, ISIS Neurostimulator, ISIS…score 17.5 · 19 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K060847Magstim 200-2
Summary page 1
…eed 24 hours. Predicate Devices This device is substantially equivalent to the current Magstim Model 2002 with Double 70mmn Remote Coil (K060847) and the Magstim Model Rapid2 with Double 70mm Remote Coil (KO51I864). Modifications…
Magstim Company Us, LLC2006—
*K051864Magstim RAPID2, Magstim Super RAPID2, Models 3004-000, 3005-00
Summary page 3
…redicate Deviee Magstim 2™ Gerieration Double 70mm-Coil P/N 3191-00 (Reference KOGO847) ‘Predicate Device Magstim 2™ Generation Double 70mm Coil P/N 3191-00 (Reference K051864) Intended Use The Magstim D707 coil is a stimulating coil intended for use with the Magstim 200° Stimulating Unit or the Magsti…
The Magstim Company, Ltd.20052

Also cleared against K060847: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 22 years and passes through 2 companies; a new generation was cleared every 14.3 years on median.
  • 1 of the 5 ancestors has a recall of any class.
  • 2 other devices also cite its predicate K060847.

How does its category compare?

All GWF devices

GWF Stimulator, Electrical, Evoked Response: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · GWF

1in 2025

2007: 32008: 22009: 52010: 32011: 22012: 42013: 32014: 12015: 22016: 12017: 32019: 12020: 12021: 22022: 22023: 22024: 42025: 12026 (partial year): 3
2007high 5 (2009) · low 0 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GWF

92in 2024

2007: 69 days median2009: 178 days median2010: 269 days median2012: 230 days median2013: 174 days median2017: 169 days median2024: 92 days median2026 (partial year): 125 days median
2007high 269 (2010) · low 69 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GWF

2so far in 2026, the most since 2007

2012: 12015: 12026 (partial year): 2
2007high 2 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · GWF

1,034in 2025

2007: 22008: 12009: 12010: 42011: 102012: 442013: 892014: 252015: 222016: 202017: 602018: 582019: 942020: 652021: 552022: 452023: 1502024: 8542025: 1,034
2007high 1,034 (2025) · low 1 (2008)2025

MAUDE reports naming product code GWF, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200736902
20082—01
2009517801
2010326904
20112—010
20124230144
20133174089
20141—025
20152—122
20161—020
20173169060
20180—058
20191—094
20201—065
20212—055
20222—045
20232—0150
20244920854
20251—01,034
202631252—

What is it indicated for?

The Magstim 0702 Coil is a stimulating coil intended for use with the MagstimO 2002 Stimulating Unit or the Magstim® Rapid 2 Stimulating Unit for the purpose of peripheral nerve stimulation for diagnostic purposes. The coil is an accessory to the Magstim @2002 Stimulating Unit, and the coil is an accessory to the Magstim® Rapid 2 Stimulating Unit.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K130403: Magstim Double 70-2 Coil.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K130403

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026