510(k) Report

510(k) · Cleared

Cleared

Capiox FX15 and FX25 Hollow Fiber Oxygenator/Reservoir

K130520510(k) SpecialCleared Mar 13, 201313-day reviewClass II21 CFR 870.4350

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ashitaka Factory of Terumo Corp. · DTZ Oxygenator, Cardiopulmonary Bypass

Cleared on Mar 13, 2013 as substantially equivalent to K071494 (Terumo Corp., 2007) after a 13-day review. Its family tree reaches back 6 generations to 1992. 6 later devices descend from it.

Decided
Mar 13, 2013
Review
13 days
Class
Class II
Product code
DTZ
Recalls
None
Predicates
1

Its longest line of descent runs back 6 generations to K915573 (1992). 14 ancestors in all. 6 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2005: 1 other ancestor2004: 1 other ancestor2000: 2 other ancestors1996: 2 other ancestors1994: 1 other ancestor1998: 1 other ancestor2014: 1 descendant cleared2015: 2 descendants cleared2025: 1 descendant cleared2018: 2 descendants clearedK915573 · Venous Reservoir W/Cardiotomy Filter · Baxter Healthcare Corp · 1992K982223 · Capiox SX Hardshell Resevoir · Terumo Medical Corp. · 1998K991973 · Capiox SX10 Hardshell Resevoir · Terumo Medical Corp. · 1999K030449 · Modification to Capiox SX10 Hardshell Reservoir · Terumo Corp. · 2003K051997 · Capiox RX15 Hollow, Fiber Oxygenator with/Without Hardshell Reservoir · Terumo Cardiovascular Systems Corp. · 2005 · recalledK071494 · Capiox FX15 and FX25 Hollow Fiber Oxygenator and Arterial Filter · Terumo Corp. · 2007 · recalled1990200020102020
  1. Root · 6 generations backK915573 Venous Reservoir W/Cardiotomy FilterBaxter Healthcare Corp · 1992
  2. 5 generations backK982223 Capiox SX Hardshell ResevoirTerumo Medical Corp. · 1998
  3. 4 generations backK991973 Capiox SX10 Hardshell ResevoirTerumo Medical Corp. · 1999
  4. 3 generations backK030449 Modification to Capiox SX10 Hardshell ReservoirTerumo Corp. · 2003
  5. 2 generations backK051997 Capiox RX15 Hollow, Fiber Oxygenator with/Without Hardshell ReservoirTerumo Cardiovascular Systems Corp. · 2005 · Recalled
  6. 1 generation backK071494 Capiox FX15 and FX25 Hollow Fiber Oxygenator and Arterial FilterTerumo Corp. · 2007 · Recalled
  7. This deviceK130520 Capiox FX15 and FX25 Hollow Fiber Oxygenator/ReservoirAshitaka Factory of Terumo Corp. · 2013
  8. Descendant · cites it directly · 5 of its ownK140774 Terumo Capiox FX15 and FX25 Hollow Fiber Oxygenator/ReservoirTerumo Cardiovascular Systems Corporation · 2014
  9. Descendant · 2 generations on · 2 of its ownK151791 Terumo Capiox FX15 Advance Oxygenator with Integrated Arterial…Terumo Cardiovascular Systems Corporation · 2015 · Recalled
  10. Descendant · 2 generations on · 1 of its ownK151389 Terumo Capiox FX15 Advance Oxygenator with Integrated Arterial…Terumo Cardiovascular Systems Corporation · 2015
  11. Descendant · 3 generations on · 0 of its ownK242503 Trilly Pediatric Oxygenator with Integrated Arterial Filter (US5232)Eurosets S.R.L · 2025
  12. Descendant · 3 generations on · 0 of its ownK180950 Terumo Capiox NX19 Oxygenator with Reservoir (east Orientation)…Terumo Cardiovascular Systems Corporation · 2018 · Recalled

Showing 5 of 6 descendants. See all descendants

Descendants cleared per year

2014: 12015: 22018: 22025: 1
2014high 2 (2015) · low 0 (2016)2026

Most influential descendants

Showing 5 of 6 · See all
  • K140774Terumo Capiox FX15 and FX25 Hollow Fiber Oxygenator/ReservoirTerumo Cardiovascular Systems Corporation · 2014 · cites it directly5descendants
  • K151791Terumo Capiox FX15 Advance Oxygenator with Integrated Arterial Filter and Reservoir…Terumo Cardiovascular Systems Corporation · 2015 · 2 generations on2descendants
  • K151389Terumo Capiox FX15 Advance Oxygenator with Integrated Arterial Filter and Hardshell…Terumo Cardiovascular Systems Corporation · 2015 · 2 generations on1descendant
  • K242503Trilly Pediatric Oxygenator with Integrated Arterial Filter (US5232)Eurosets S.R.L · 2025 · 3 generations on0descendants
  • K180950Terumo Capiox NX19 Oxygenator with Reservoir (east Orientation), Terumo Capiox NX19…Terumo Cardiovascular Systems Corporation · 2018 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K915573Venous Reservoir W/Cardiotomy FilterBaxter Healthcare Corp1992—
−5K982223Capiox SX Hardshell ResevoirTerumo Medical Corp.1998—
−4K991973Capiox SX10 Hardshell ResevoirTerumo Medical Corp.1999—
−3K030449Modification to Capiox SX10 Hardshell ReservoirTerumo Corp.2003—
−2K051997Capiox RX15 Hollow, Fiber Oxygenator with/Without Hardshell ReservoirTerumo Cardiovascular Systems Corp.20051
−1K071494Capiox FX15 and FX25 Hollow Fiber Oxygenator and Arterial FilterTerumo Corp.200712
0 (this device)K130520Capiox FX15 and FX25 Hollow Fiber Oxygenator/ReservoirAshitaka Factory of Terumo Corp.2013—
+1K140774Terumo Capiox FX15 and FX25 Hollow Fiber Oxygenator/ReservoirTerumo Cardiovascular Systems Corporation2014—
+2K151791Terumo Capiox FX15 Advance Oxygenator with Integrated Arterial Filter and…Terumo Cardiovascular Systems Corporation20152
+2K151389Terumo Capiox FX15 Advance Oxygenator with Integrated Arterial Filter and…Terumo Cardiovascular Systems Corporation2015—
+3K242503Trilly Pediatric Oxygenator with Integrated Arterial Filter (US5232)Eurosets S.R.L2025—
+3K180950Terumo Capiox NX19 Oxygenator with Reservoir (east Orientation), Terumo Capiox…Terumo Cardiovascular Systems Corporation20185

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410
199620
199810
200020
200410
200510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201410
201520
201820
202510
FIG. 1 — Lineage of K130520, 1992–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DTZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DTZ device

  1. #1 K240934 Quantum Perfusion Blood Oxygenator ECC…score 14.3 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K071494Capiox FX15 and FX25 Hollow Fiber Oxygenator and Arterial Filter
Summary page 4
…Withi respect to the desig~n of the H~ardshell Reservoir, the reservoir component remains identical to the design of the original reservoir that was cleared by FDA with K07 1494 - except that a positive Pressure relief valve will be included on the lid of the reservoir. The intent of the relief valve is to…
Terumo Corp.200712

Also cleared against K071494: 1 other device →

What descends from it? (6)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K140774Terumo Capiox FX15 and FX25 Hollow Fiber Oxygenator/ReservoirTerumo Cardiovascular Systems Corporation20145

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 5 companies; a new generation was cleared every 3.5 years on median.
  • 3 of the 14 ancestors have a recall of any class.
  • Of its descendants, K140774 (Terumo Cardiovascular Systems Corporation, 2014) has gone furthest, with 5 descendants of its own.
  • 1 other device also cites its predicate K071494.

How does its category compare?

All DTZ devices

DTZ Oxygenator, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · DTZ

2in 2025

2007: 42008: 32009: 92010: 22011: 42012: 92013: 92014: 52015: 92016: 32017: 22018: 32019: 52020: 52021: 52022: 22023: 52024: 32025: 22026 (partial year): 3
2007high 9 (2009) · low 2 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTZ

124in 2024

2007: 55 days median2008: 88 days median2009: 210 days median2011: 160 days median2012: 90 days median2013: 28 days median2014: 28 days median2015: 134 days median2016: 62 days median2018: 72 days median2019: 90 days median2020: 93 days median2021: 336 days median2023: 145 days median2024: 124 days median2026 (partial year): 137 days median
2007high 336 (2021) · low 28 (2013)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DTZ

1in 2025

2007: 72008: 52010: 12011: 132012: 142014: 22016: 12018: 2, 2 Class I2019: 12020: 32021: 1, 1 Class I2022: 62025: 1
2007high 14 (2012) · low 0 (2009)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · DTZ

976in 2025

2007: 1342008: 1052009: 522010: 1192011: 1902012: 2152013: 3462014: 7642015: 6582016: 4432017: 3882018: 3202019: 4892020: 4422021: 5952022: 5592023: 7382024: 5352025: 976
2007high 976 (2025) · low 52 (2009)2025

MAUDE reports naming product code DTZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074557134
20083885105
20099210052
20102—1119
2011416013190
201299014215
20139280346
20145282764
201591340658
20163621443
20172—0388
20183722 (2)320
20195901489
20205933442
202153361 (1)595
20222—6559
202351450738
202431240535
20252—1976
202631370—

What is it indicated for?

The Capiox FX H-ollow Fiber Oxygenator and Arterial Filter is intended to be used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery. The integrated arterial filter is intended to filtrate non-biologic particles and emboli and to facilitate air bubble removal from the blood flowing through the cardiopulmonary bypass circuit. The integrated heat exchanger is used…

Read the full statement (242 words)

The Capiox FX H-ollow Fiber Oxygenator and Arterial Filter is intended to be used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery.

The integrated arterial filter is intended to filtrate non-biologic particles and emboli and to facilitate air bubble removal from the blood flowing through the cardiopulmonary bypass circuit.

The integrated heat exchanger is used to warm or cool blood and/or perfusion fluid as it flows through the device.

The hardshell reservoir is used to stoi-e blood during extra-corpoieal circulation from the venous line and the cardiotorny line. The reservoir contains a venous section that is comprised of a filter and defoamer to facilitate air bubble removal. The cardiotomy section of the reservoir contains a filter to remove particulate matter and a defoarner to facilitate air bubble removal. The 3-liter and 4-liter reservoirs may be used for Vacuum Assisted Drainage procedures and Post Operative Chest Drainage Procedures.

The Capiox FY 15 is for use with patients when the required blood flow rate will not exceed 5.0 L/min. when used with a 4 Liter Reservoir; and when the required blood flowv rate will not exceed 4.0 L/min. when used with a 3 Liter Reservoir.

The Capiox FX25 is for use with patients when the required blood flow rate will not exceed 7.0 L/min.

The Capiox EX Oxygenator/Reservoir/Arterial Filter assemblies can be used in procedures lasting up to 6 hours.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K130520: Capiox FX15 and FX25 Hollow Fiber Oxygenator/Reservoir.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K130520

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026