510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

Terumo Capiox FX15 Advance Oxygenator with Integrated Arterial Filter and Reservoir, Terumo Capiox FX25 Advance Oxygenator with Integrated Arterial Filter and Reservoir

K151791510(k) SpecialCleared Nov 19, 2015141-day reviewClass II21 CFR 870.4350Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Terumo Cardiovascular Systems Corporation · DTZ Oxygenator, Cardiopulmonary Bypass

Cleared on Nov 19, 2015 as substantially equivalent to K140774 (Terumo Cardiovascular Systems Corporation, 2014) after a 141-day review. Its family tree reaches back 8 generations to 1992. 2 later devices descend from it. It has been recalled 2 times.

Decided
Nov 19, 2015
Review
141 days
Class
Class II
Product code
DTZ
Recalls
2
Predicates
1

Its longest line of descent runs back 8 generations to K915573 (1992). 20 ancestors in all, the oldest K842908 (1985). 2 later devices descend from it, 2 citing it directly.

  1. Root · 8 generations backK915573 Venous Reservoir W/Cardiotomy FilterBaxter Healthcare Corp · 1992
  2. 7 generations backK982223 Capiox SX Hardshell ResevoirTerumo Medical Corp. · 1998
  3. 6 generations backK991973 Capiox SX10 Hardshell ResevoirTerumo Medical Corp. · 1999
  4. 5 generations backK030449 Modification to Capiox SX10 Hardshell ReservoirTerumo Corp. · 2003
  5. 4 generations backK051997 Capiox RX15 Hollow, Fiber Oxygenator with/Without Hardshell ReservoirTerumo Cardiovascular Systems Corp. · 2005 · Recalled
  6. 3 generations backK071494 Capiox FX15 and FX25 Hollow Fiber Oxygenator and Arterial FilterTerumo Corp. · 2007 · Recalled
  7. 2 generations backK130520 Capiox FX15 and FX25 Hollow Fiber Oxygenator/ReservoirAshitaka Factory of Terumo Corp. · 2013
  8. 1 generation backK140774 Terumo Capiox FX15 and FX25 Hollow Fiber Oxygenator/ReservoirTerumo Cardiovascular Systems Corporation · 2014
  9. This deviceK151791 Terumo Capiox FX15 Advance Oxygenator with Integrated Arterial…Terumo Cardiovascular Systems Corporation · 2015
  10. Descendant · cites it directly · 0 of its ownK180950 Terumo Capiox NX19 Oxygenator with Reservoir (east Orientation)…Terumo Cardiovascular Systems Corporation · 2018 · Recalled
  11. Descendant · cites it directly · 0 of its ownK172071 Terumo Capiox NX19 Oxygenator with Reservoir (east orientation )…Terumo Cardiovascular Systems Corporation · 2018

Most influential descendants

Showing 2 of 2 · See all
  • K180950Terumo Capiox NX19 Oxygenator with Reservoir (east Orientation), Terumo Capiox NX19…Terumo Cardiovascular Systems Corporation · 2018 · cites it directly0descendants
  • K172071Terumo Capiox NX19 Oxygenator with Reservoir (east orientation ), Terumo Capiox NX19…Terumo Cardiovascular Systems Corporation · 2018 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K915573Venous Reservoir W/Cardiotomy FilterBaxter Healthcare Corp1992—
−7K982223Capiox SX Hardshell ResevoirTerumo Medical Corp.1998—
−6K991973Capiox SX10 Hardshell ResevoirTerumo Medical Corp.1999—
−5K030449Modification to Capiox SX10 Hardshell ReservoirTerumo Corp.2003—
−4K051997Capiox RX15 Hollow, Fiber Oxygenator with/Without Hardshell ReservoirTerumo Cardiovascular Systems Corp.20051
−3K071494Capiox FX15 and FX25 Hollow Fiber Oxygenator and Arterial FilterTerumo Corp.200712
−2K130520Capiox FX15 and FX25 Hollow Fiber Oxygenator/ReservoirAshitaka Factory of Terumo Corp.2013—
−1K140774Terumo Capiox FX15 and FX25 Hollow Fiber Oxygenator/ReservoirTerumo Cardiovascular Systems Corporation2014—
0 (this device)K151791Terumo Capiox FX15 Advance Oxygenator with Integrated Arterial Filter and…Terumo Cardiovascular Systems Corporation2015—
+1K180950Terumo Capiox NX19 Oxygenator with Reservoir (east Orientation), Terumo Capiox…Terumo Cardiovascular Systems Corporation20185
+1K172071Terumo Capiox NX19 Oxygenator with Reservoir (east orientation ), Terumo…Terumo Cardiovascular Systems Corporation2018—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198510
199410
199620
199810
200020
200210
200410
200510
200610
201310

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201820
FIG. 1 — Lineage of K151791, 1992–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in DTZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DTZ device

  1. #1 K240934 Quantum Perfusion Blood Oxygenator ECC…score 14.3 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K140774Terumo Capiox FX15 and FX25 Hollow Fiber Oxygenator/Reservoir
Summary page 4
…tion, materials and specifications to the following devices: Terumo® Capiox“ FX Hollow Fiber Oxygenator with Integrated Arterial Filter and Hardshell Reservoir – K140774. Intended Use The Capiox“ FX Advance Oxygenator with Integrated Arterial Filter and Hardshell Reservoir is intended to be used t…
Terumo Cardiovascular Systems Corporation2014—

Also cleared against K140774: 1 other device →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K180950Terumo Capiox NX19 Oxygenator with Reservoir (east Orientation), Terumo Capiox NX19…Terumo Cardiovascular Systems Corporation20180
K172071Terumo Capiox NX19 Oxygenator with Reservoir (east orientation ), Terumo Capiox NX19…Terumo Cardiovascular Systems Corporation20180

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 1, 2023Class IIZ-1347-2023Process controlSome Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak at the point where the arterial thermistor is improperly bonded to the…
Nov 11, 2020Class IIZ-0565-2021Labeling mix-upsCapiox FX25 Advance oxygenator may be the incorrect configuration (East vs. West)

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 6 companies; a new generation was cleared every 2.5 years on median.
  • 4 of the 20 ancestors have a recall of any class.
  • Of its descendants, K180950 (Terumo Cardiovascular Systems Corporation, 2018) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K140774.

How does its category compare?

All DTZ devices

DTZ Oxygenator, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2015, this device’s year, is highlighted.

FDA decisions per year · DTZ

2in 2025

2007: 42008: 32009: 92010: 22011: 42012: 92013: 92014: 52015: 92016: 32017: 22018: 32019: 52020: 52021: 52022: 22023: 52024: 32025: 22026 (partial year): 3
2007high 9 (2009) · low 2 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTZ

124in 2024

2007: 55 days median2008: 88 days median2009: 210 days median2011: 160 days median2012: 90 days median2013: 28 days median2014: 28 days median2015: 134 days median2016: 62 days median2018: 72 days median2019: 90 days median2020: 93 days median2021: 336 days median2023: 145 days median2024: 124 days median2026 (partial year): 137 days median
2007high 336 (2021) · low 28 (2013)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DTZ

1in 2025

2007: 72008: 52010: 12011: 132012: 142014: 22016: 12018: 2, 2 Class I2019: 12020: 32021: 1, 1 Class I2022: 62025: 1
2007high 14 (2012) · low 0 (2009)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · DTZ

976in 2025

2007: 1342008: 1052009: 522010: 1192011: 1902012: 2152013: 3462014: 7642015: 6582016: 4432017: 3882018: 3202019: 4892020: 4422021: 5952022: 5592023: 7382024: 5352025: 976
2007high 976 (2025) · low 52 (2009)2025

MAUDE reports naming product code DTZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074557134
20083885105
20099210052
20102—1119
2011416013190
201299014215
20139280346
20145282764
201591340658
20163621443
20172—0388
20183722 (2)320
20195901489
20205933442
202153361 (1)595
20222—6559
202351450738
202431240535
20252—1976
202631370—

What is it indicated for?

The Capiox® FX Advance Oxygenator with Integrated Arterial Filter and Hardshell Reservoir is intended to be used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery. The integrated arterial filter is intended to filtrate non-biologic particles and emboli and to facilitate air bubble removal from the blood flowing through the cardiopulmonary bypass circuit. The…

Read the full statement (232 words)

The Capiox® FX Advance Oxygenator with Integrated Arterial Filter and Hardshell Reservoir is intended to be used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during cardiopulmonary bypass surgery. The integrated arterial filter is intended to filtrate non-biologic particles and emboli and to facilitate air bubble removal from the blood flowing through the cardiopulmonary bypass circuit.

The integrated heat exchanger is used to warm or cool blood and/or perfusion fluid as it flows through the device.

The hardshell reservoir is used to store blood during extra-corporeal circulation from the venous line and the cardiotomy line. The reservoir contains a venous section that is comprised of a filter and defoamer to facilitate air bubble removal. The cardiotomy section of the reservoir contains a filter to remove particulate matter and a defoamer to facilitate air bubble removal. The 3-liter and 4-liter reservoirs may be used for Vacuum Assisted Drainage procedures and Post Operative Chest Drainage Procedures.

The Capiox® FX15 Advance Oxygenator with Integrated Arterial Filter and Hardshell Reservoir (both 3-liter and 4-liter) is for use with patients when the required blood flow rate will not exceed 5.0 L/min. The Capiox® FX25 is for use with patients when the required blood flow rate will not exceed 7.0 L/min.

The Capiox® FX Advance Oxygenator/Reservoir/Arterial Filter assemblies can be used in procedures lasting up to 6 hours.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K151791: Terumo Capiox FX15 Advance Oxygenator with Integrated Arterial Filter and Reservoir, Terumo Capiox FX25 Advance Oxygenator with Integrated Arterial Filter and Reservoir.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K151791

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026