510(k) Report

510(k) · Cleared

Recalled · Class III · 2019

Specular Microscope Cem-530

K130565510(k) TraditionalCleared Nov 27, 2013268-day reviewClass II21 CFR 886.1850Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Nidek Co., Ltd. · NQE Microscope, Specular

Cleared on Nov 27, 2013 as substantially equivalent to K120264 (Konan Medical, Inc., 2012) after a 268-day review. Its family tree reaches back 4 generations to 1992. 2 later devices descend from it. It has been recalled once.

Decided
Nov 27, 2013
Review
268 days
Class
Class II
Product code
NQE
Recalls
1
Predicates
1

Its longest line of descent runs back 4 generations to K920906 (1992). 5 ancestors in all. 2 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1995: 1 other ancestor2016: 1 descendant cleared2018: 1 descendant clearedK920906 · DGH-500 Pachymeter · Dgh Technology, Inc. · 1992K980357 · Konan Noncon Robo Pachy Specular Microscope and Opticl Pachymeter · Konan, Inc. · 1998K062763 · Noncon Robo Pachy F&a · Konan Medical, Inc. · 2008K120264 · Konan Specular Microscope Xiv · Konan Medical, Inc. · 20121990200020102020
Other ancestors, by generationDescendants, by generation 4 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant K120264 · Konan Specular Microscope Xiv · Konan Medical, Inc. · 2012K062763 · Noncon Robo Pachy F&a · Konan Medical, Inc. · 2008K980357 · Konan Noncon Robo Pachy Specular Microscope and Opticl Pachymeter · Konan, Inc. · 1998K920906 · DGH-500 Pachymeter · Dgh Technology, Inc. · 19920
  1. Root · 4 generations backK920906 DGH-500 PachymeterDgh Technology, Inc. · 1992
  2. 3 generations backK980357 Konan Noncon Robo Pachy Specular Microscope and Opticl PachymeterKonan, Inc. · 1998
  3. 2 generations backK062763 Noncon Robo Pachy F&aKonan Medical, Inc. · 2008
  4. 1 generation backK120264 Konan Specular Microscope XivKonan Medical, Inc. · 2012
  5. This deviceK130565 Specular Microscope Cem-530Nidek Co., Ltd. · 2013
  6. Descendant · cites it directly · 1 of its ownK151706 Specular Microscope CEM- 530Nidek Co., Ltd. · 2016 · Recalled
  7. Descendant · 2 generations on · 0 of its ownK173980 Specular Microscope CEM-530Nidek Co., Ltd. · 2018 · Recalled

Most influential descendants

Showing 2 of 2 · See all
  • K151706Specular Microscope CEM- 530Nidek Co., Ltd. · 2016 · cites it directly1descendant
  • K173980Specular Microscope CEM-530Nidek Co., Ltd. · 2018 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K920906DGH-500 PachymeterDgh Technology, Inc.1992—
−3K980357Konan Noncon Robo Pachy Specular Microscope and Opticl PachymeterKonan, Inc.1998—
−2K062763Noncon Robo Pachy F&aKonan Medical, Inc.2008—
−1K120264Konan Specular Microscope XivKonan Medical, Inc.2012—
0 (this device)K130565Specular Microscope Cem-530Nidek Co., Ltd.2013—
+1K151706Specular Microscope CEM- 530Nidek Co., Ltd.20162
+2K173980Specular Microscope CEM-530Nidek Co., Ltd.20181

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201610
201810
FIG. 1 — Lineage of K130565, 1992–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NQE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NQE device

  1. #1 K191558 Konan Specular Microscope XVIIscore -0.0 · 4 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K120264Konan Specular Microscope Xiv
Summary page 2
…d to a filing system via LAN connection. SUBSTANTIAL EQUIVALENCE The Specular Microscope CEM-530 is substantially equivalent to the Konan NMedical, Inc. Cellchek Plus (K120264). The Specular Microscope CEM-530 has the same intended use and indications for use, technological characteristics, and principle…
Konan Medical, Inc.2012—

Also cleared against K120264: 5 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K151706Specular Microscope CEM- 530Nidek Co., Ltd.20161

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 25, 2019Class IIIZ-0771-2020Device DesignThe electromagnetic noise exceeded the upper limit of the Electromagnetic Compatibility Standard (EMC: IEC 60601-1-2)

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 4 companies; a new generation was cleared every 5.5 years on median.
  • None of the 5 ancestors has a recall.
  • Of its descendants, K151706 (Nidek Co., Ltd., 2016) has gone furthest, with 1 descendant of its own.
  • 5 other devices also cite its predicate K120264.

How does its category compare?

All NQE devices

NQE Microscope, Specular: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · NQE

1in 2020

2008: 12012: 12013: 12015: 12016: 12018: 22020: 1
2007high 2 (2018) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NQE

Too few NQE 510(k)s in any year since 2007 to give a median.

Recalls per year · NQE

1in 2019

2019: 1
2007high 1 (2019) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · NQE

No MAUDE report names NQE since 2007.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20081—00
20090—00
20100—00
20110—00
20121—00
20131—00
20140—00
20151—00
20161—00
20170—00
20182—00
20190—10
20201—00
20210—00
20220—00
20230—00
20240—00
20250—00
20260—0—

What is it indicated for?

The Nidek Specular Microscope CEM-530 isa non-contact ophthalmic microscope, optical

pachymeter, and camera intended for examination of the comeal endothelium and for

measurement of the thickness of the cornea.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K130565: Specular Microscope Cem-530.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K130565

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026