510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Specular Microscope CEM- 530

K151706510(k) TraditionalCleared Mar 17, 2016267-day reviewClass II21 CFR 886.1850Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Nidek Co., Ltd. · NQE Microscope, Specular

Cleared on Mar 17, 2016 as substantially equivalent to K130565 (Nidek Co., Ltd., 2013) and 1 other predicate after a 267-day review. Its family tree reaches back 5 generations to 1992. 1 later device descends from it. It has been recalled 2 times.

Decided
Mar 17, 2016
Review
267 days
Class
Class II
Product code
NQE
Recalls
2
Predicates
2

Its longest line of descent runs back 5 generations to K920906 (1992). 6 ancestors in all. 1 later device descends from it, 1 citing it directly.

  1. Root · 5 generations backK920906 DGH-500 PachymeterDgh Technology, Inc. · 1992
  2. 4 generations backK980357 Konan Noncon Robo Pachy Specular Microscope and Opticl PachymeterKonan, Inc. · 1998
  3. 3 generations backK062763 Noncon Robo Pachy F&aKonan Medical, Inc. · 2008
  4. 2 generations backK120264 Konan Specular Microscope XivKonan Medical, Inc. · 2012
  5. 1 generation backK130565 Specular Microscope Cem-530Nidek Co., Ltd. · 2013 · Recalled
  6. This deviceK151706 Specular Microscope CEM- 530Nidek Co., Ltd. · 2016
  7. Descendant · cites it directly · 0 of its ownK173980 Specular Microscope CEM-530Nidek Co., Ltd. · 2018 · Recalled

Most influential descendants

Showing 1 of 1 · See all
  • K173980Specular Microscope CEM-530Nidek Co., Ltd. · 2018 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K920906DGH-500 PachymeterDgh Technology, Inc.1992—
−4K980357Konan Noncon Robo Pachy Specular Microscope and Opticl PachymeterKonan, Inc.1998—
−3K062763Noncon Robo Pachy F&aKonan Medical, Inc.2008—
−2K120264Konan Specular Microscope XivKonan Medical, Inc.2012—
−1K130565Specular Microscope Cem-530Nidek Co., Ltd.20131
0 (this device)K151706Specular Microscope CEM- 530Nidek Co., Ltd.2016—
+1K173980Specular Microscope CEM-530Nidek Co., Ltd.20181

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201810
FIG. 1 — Lineage of K151706, 1992–2018Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NQE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NQE device

  1. #1 K191558 Konan Specular Microscope XVIIscore -0.0 · 4 generations · no Class I recall
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K130565Specular Microscope Cem-530
Summary page 4
…s Class II devices pursuant to 21 C.F.R. §886.1850. PRODUCT CODE: CLASSIFICATION / CFR TITLE NQE, 21 CFR 886.1850 PREDICATE DEVICES Nidek Specular Microscope CEM-530 (K130565) Konan Medical, Inc. Cellchek Plus (K120264) INDICATIONS FOR USE The Nidek Specular Microscope CEM-530 is a non-c…
Nidek Co., Ltd.20131
*K120264Konan Specular Microscope Xiv
Summary page 4
…6.1850. PRODUCT CODE: CLASSIFICATION / CFR TITLE NQE, 21 CFR 886.1850 PREDICATE DEVICES Nidek Specular Microscope CEM-530 (K130565) Konan Medical, Inc. Cellchek Plus (K120264) INDICATIONS FOR USE The Nidek Specular Microscope CEM-530 is a non-contact ophthalmic microscope, optical pachym…
Konan Medical, Inc.2012—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K173980Specular Microscope CEM-530Nidek Co., Ltd.20180

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 25, 2019Class IIIZ-0771-2020Device DesignThe electromagnetic noise exceeded the upper limit of the Electromagnetic Compatibility Standard (EMC: IEC 60601-1-2)
Aug 8, 2016Class IIZ-2711-2016Software change controlSoftware version 1.08 and 1.09 for the Specular Microscope CEM 530 included a change of analysis results feature that was not reviewed and approved by the FDA.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 24 years and passes through 4 companies; a new generation was cleared every 4.1 years on median.
  • 1 of the 6 ancestors has a recall of any class.
  • Of its descendants, K173980 (Nidek Co., Ltd., 2018) has gone furthest, with 0 descendants of its own.

How does its category compare?

All NQE devices

NQE Microscope, Specular: FDA decisions, review times, recalls and adverse event reports over 20 years. 2016, this device’s year, is highlighted.

FDA decisions per year · NQE

1in 2020

2008: 12012: 12013: 12015: 12016: 12018: 22020: 1
2007high 2 (2018) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NQE

Too few NQE 510(k)s in any year since 2007 to give a median.

Recalls per year · NQE

1in 2019

2019: 1
2007high 1 (2019) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · NQE

No MAUDE report names NQE since 2007.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20081—00
20090—00
20100—00
20110—00
20121—00
20131—00
20140—00
20151—00
20161—00
20170—00
20182—00
20190—10
20201—00
20210—00
20220—00
20230—00
20240—00
20250—00
20260—0—

What is it indicated for?

The Specular Microscope CEM-530 is a non-contact ophthalmic microscope, optical pachymeter, and camera intended for examination of the corneal endothelium and for measurement of the thickness of the cornea.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K151706: Specular Microscope CEM- 530.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K151706

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026