510(k) Report

510(k) · Cleared

Cleared

Xpert Mrsa/SA Blood Culture Assay; Genexpert DX Systems (GX-I, GX-II, GX-IV, GX-Xvi); Genexpert Infinity-48 System; Gene

K130894510(k) TraditionalCleared Jun 20, 201380-day reviewClass II21 CFR 866.1640

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Cepheid · NQX System, Nucleic Acid Amplification Test, Dna, Methicillin Resistant Staphylococcus Aureus, Direct Specimen

Cleared on Jun 20, 2013 as substantially equivalent to K071026 (Bd Diagnostics (Geneohm Sciences Canada, Inc), 2007) and 1 other predicate after an 80-day review. Its family tree reaches back 5 generations to 2002. 1 later device descends from it.

Decided
Jun 20, 2013
Review
80 days
Class
Class II
Product code
NQX
Recalls
None
Predicates
2

Its longest line of descent runs back 5 generations to K020322 (2002). 11 ancestors in all, the oldest K851949 (1985). 1 later device descends from it, 1 citing it directly.

  1. Root · 5 generations backK020322 BD Phoenix Automated Microbiology SystemBecton, Dickinson & CO · 2002
  2. 4 generations backK023301 BD Phoenix Automated Microbiology SystemBecton, Dickinson & CO · 2002 · Recalled
  3. 3 generations backK071026 BD Geneohm Staphsr AssayBd Diagnostics (Geneohm Sciences Canada, Inc) · 2007
  4. 2 generations backK082140 Xpert Mrsa/SA Blood Culture AssayCepheid · 2008 · Recalled
  5. 1 generation backK101879 Xpert Mrsa/SA BC (Blood Culture) Assay Model Gxmrsa/SA-BC-10Cepheid · 2010 · Recalled
  6. This deviceK130894 Xpert Mrsa/SA Blood Culture Assay; Genexpert DX Systems (GX-I…Cepheid · 2013
  7. Descendant · cites it directly · 0 of its ownK190771 Xpert MRSA/SA Blood Culture, GeneXpert Dx System, GeneXpert…Cepheid · 2019

Most influential descendants

Showing 1 of 1 · See all
  • K190771Xpert MRSA/SA Blood Culture, GeneXpert Dx System, GeneXpert Infinity-48s System…Cepheid · 2019 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K020322BD Phoenix Automated Microbiology SystemBecton, Dickinson & CO2002—
−4K023301BD Phoenix Automated Microbiology SystemBecton, Dickinson & CO20022
−3K071026BD Geneohm Staphsr AssayBd Diagnostics (Geneohm Sciences Canada, Inc)2007—
−2K082140Xpert Mrsa/SA Blood Culture AssayCepheid20082
−1K101879Xpert Mrsa/SA BC (Blood Culture) Assay Model Gxmrsa/SA-BC-10Cepheid20101
0 (this device)K130894Xpert Mrsa/SA Blood Culture Assay; Genexpert DX Systems (GX-I, GX-II, GX-IV…Cepheid2013—
+1K190771Xpert MRSA/SA Blood Culture, GeneXpert Dx System, GeneXpert Infinity-48s…Cepheid2019—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198510
198610
200220
200420

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201910
FIG. 1 — Lineage of K130894, 2002–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NQX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NQX device

  1. #1 K243625 Xpert MRSA/SA SSTIscore 9.4 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K071026BD Geneohm Staphsr Assay
Summary page 1
…e: NQX Classification Microbiology (83) Advisory Panel Predicate Devices Xpert MRSA/SA Blood Culture Assay (510(k) #K101879) Name(s): BD GeneOhm™ StaphSR Assay (510(k) #K071026) Device Description: _ The Cepheid Xpert® MRSA/SA Blood Culture Assay (Xpert MRSA/SA Blood Culture Assay) is a rapid, automated…
Bd Diagnostics (Geneohm Sciences Canada, Inc)2007—
*K101879Xpert Mrsa/SA BC (Blood Culture) Assay Model Gxmrsa/SA-BC-10
Summary page 1
…Antimicrobial susceptibility test powder Product code: NQX Classification Microbiology (83) Advisory Panel Predicate Devices Xpert MRSA/SA Blood Culture Assay (510(k) #K101879) Name(s): BD GeneOhm™ StaphSR Assay (510(k) #K071026) Device Description: _ The Cepheid Xpert® MRSA/SA Blood Culture Assay (Xpe…
Cepheid20101

Also cleared against K071026: 4 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K190771Xpert MRSA/SA Blood Culture, GeneXpert Dx System, GeneXpert Infinity-48s System…Cepheid20190

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 11 years and passes through 3 companies; a new generation was cleared every 1.8 years on median.
  • 6 of the 11 ancestors have a recall of any class.
  • Of its descendants, K190771 (Cepheid, 2019) has gone furthest, with 0 descendants of its own.
  • 4 other devices also cite its predicate K071026.

How does its category compare?

All NQX devices

NQX System, Nucleic Acid Amplification Test, Dna, Methicillin Resistant Staphylococcus Aureus, Direct Specimen: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · NQX

2in 2024

2007: 22008: 22009: 12010: 32011: 22012: 22013: 42014: 12016: 12019: 22024: 2
2007high 4 (2013) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NQX

75in 2013

2010: 70 days median2013: 75 days median
2007high 75 (2013) · low 70 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NQX

1in 2025

2008: 32010: 12011: 62025: 1
2007high 6 (2011) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · NQX

15in 2025

2010: 12012: 12015: 32017: 12018: 22021: 32022: 52023: 122024: 82025: 15
2007high 15 (2025) · low 0 (2007)2025

MAUDE reports naming product code NQX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
20082—30
20091—00
201037011
20112—60
20122—01
201347500
20141—00
20150—03
20161—00
20170—01
20180—02
20192—00
20200—00
20210—03
20220—05
20230—012
20242—08
20250—115
20260—0—

What is it indicated for?

The Cepheid Xpert® MRSA/SA Blood Culture Assay, performed on the GeneXpert® Instrument Systems, is a qualitative in vitro diagnostic test intended for the detection of Staphylococcus aureus (SA) and methicillin-resistant Staphylococcus aureus (MRSA) DNA directly from positive blood cultures. The assay utilizes automated real-time polymerase chain reaction (PCR) for the amplification of MRSA/SA specific DNA targets and fluorogenic target-specific…

Read the full statement (188 words)

The Cepheid Xpert® MRSA/SA Blood Culture Assay, performed on the GeneXpert® Instrument Systems, is a qualitative in vitro diagnostic test intended for the detection of Staphylococcus aureus (SA) and methicillin-resistant Staphylococcus aureus (MRSA) DNA directly from positive blood cultures. The assay utilizes automated real-time polymerase chain reaction (PCR) for the amplification of MRSA/SA specific DNA targets and fluorogenic target-specific hybridization probes for the real-time detection of the amplified DNA. The assay is performed directly on positive blood culture samples from BD BACTEC™ Plus Aerobic/F, BacT/ALERT® SA (Standard Aerobic) or VersaTREK REDOX 1® (aerobic) blood culture bottles that are determined by Gram stain as Gram Positive Cocci in Clusters (GPCC) or as Gram Positive Cocci in singles (GPC). The Xpert MRSA/SA Blood Culture Assay is indicated for use in conjunction with other laboratory tests, such as culture, and clinical data available to the clinician as an aid in the detection of MRSA/SA from positive blood cultures. Subculturing of positive blood cultures is necessary to recover organisms for susceptibility testing or for epidemiological typing. The Cepheid Xpert MRSA/SA Blood Culture Assay is not intended to monitor treatment for MRSA/SA infections.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K130894: Xpert Mrsa/SA Blood Culture Assay; Genexpert DX Systems (GX-I, GX-II, GX-IV, GX-Xvi); Genexpert Infinity-48 System; Gene.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K130894

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026