BD Phoenix Automated Microbiology System

K023301510(k) TraditionalCleared Nov 12, 200240-day reviewClass II21 CFR 866.1645Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Becton, Dickinson & CO · LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation

Cleared on Nov 12, 2002 as substantially equivalent to K020321 (Becton, Dickinson & CO, 2002) and 2 other predicates after a 40-day review. Its family tree reaches back 1 generation to 2002. 10 later devices descend from it. It has been recalled 2 times.

Decided
Nov 12, 2002
Review
40 days
Class
Class II
Product code
LON
Recalls
2
Predicates
4

Its longest line of descent runs back 1 generation to K020322 (2002). 3 ancestors in all. 10 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2002: 2 other ancestors2008: 2 descendants cleared2007: 1 descendant cleared2013: 2 descendants cleared2011: 1 descendant cleared2024: 1 descendant cleared2010: 1 descendant cleared2012: 1 descendant cleared2019: 1 descendant clearedK020322 · BD Phoenix Automated Microbiology System · Becton, Dickinson & CO · 2002200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors 1 generation forward: 3 descendants2 generations forward: 5 descendants3 generations forward: 2 descendants K020322 · BD Phoenix Automated Microbiology System · Becton, Dickinson & CO · 20020
  1. Root · 1 generation backK020322 BD Phoenix Automated Microbiology SystemBecton, Dickinson & CO · 2002
  2. This deviceK023301 BD Phoenix Automated Microbiology SystemBecton, Dickinson & CO · 2002
  3. Descendant · cites it directly · 9 of its ownK071026 BD Geneohm Staphsr AssayBd Diagnostics (Geneohm Sciences Canada, Inc) · 2007
  4. Descendant · cites it directly · 3 of its ownK082140 Xpert Mrsa/SA Blood Culture AssayCepheid · 2008 · Recalled
  5. Descendant · 2 generations on · 2 of its ownK101879 Xpert Mrsa/SA BC (Blood Culture) Assay Model Gxmrsa/SA-BC-10Cepheid · 2010 · Recalled
  6. Descendant · 2 generations on · 1 of its ownK130894 Xpert Mrsa/SA Blood Culture Assay; Genexpert DX Systems (GX-I…Cepheid · 2013
  7. Descendant · 2 generations on · 1 of its ownK102342 Keypath (TM) Mrsa/Mssa Blood Culture Test- BTMicrophage, Inc. · 2011

Showing 5 of 10 descendants. See all descendants

Descendants cleared per year

2007: 12008: 22010: 12011: 12012: 12013: 22019: 12024: 1
2007high 2 (2008) · low 0 (2009)2026

Most influential descendants

Showing 5 of 10 · See all
  • K071026BD Geneohm Staphsr AssayBd Diagnostics (Geneohm Sciences Canada, Inc) · 2007 · cites it directly9descendants
  • K082140Xpert Mrsa/SA Blood Culture AssayCepheid · 2008 · cites it directly3descendants
  • K101879Xpert Mrsa/SA BC (Blood Culture) Assay Model Gxmrsa/SA-BC-10Cepheid · 2010 · 2 generations on2descendants
  • K130894Xpert Mrsa/SA Blood Culture Assay; Genexpert DX Systems (GX-I, GX-II, GX-IV, GX-Xvi)…Cepheid · 2013 · 2 generations on1descendant
  • K102342Keypath (TM) Mrsa/Mssa Blood Culture Test- BTMicrophage, Inc. · 2011 · 2 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K020322BD Phoenix Automated Microbiology SystemBecton, Dickinson & CO2002—
0 (this device)K023301BD Phoenix Automated Microbiology SystemBecton, Dickinson & CO2002—
+1K071026BD Geneohm Staphsr AssayBd Diagnostics (Geneohm Sciences Canada, Inc)2007—
+1K082140Xpert Mrsa/SA Blood Culture AssayCepheid20082
+2K101879Xpert Mrsa/SA BC (Blood Culture) Assay Model Gxmrsa/SA-BC-10Cepheid20101
+2K130894Xpert Mrsa/SA Blood Culture Assay; Genexpert DX Systems (GX-I, GX-II, GX-IV…Cepheid2013—
+2K102342Keypath (TM) Mrsa/Mssa Blood Culture Test- BTMicrophage, Inc.2011—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200220

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200710
200820
201010
201110
201210
201320
201910
202410
FIG. 1 — Lineage of K023301, 2002–2024Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LON? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LON device

  1. #1 K260619 VITEK 2 AST-Streptococcus Cefepime (<=0.06 –…score 13.3 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.

TypeNumberDeviceApplicantDecidedRecalls
Kept off the family treeP050510VITEK System
Not in FDA’s 510(k), De Novo or PMA lists as we hold them (a misreading, a typo, or a number FDA does not list), so it is kept off the family tree
*K020321BD Phoenix Automated Microbiology System
Summary page 3
…quivalent to those used in the VITEK@system, which received approval by the FDA under PMA number N505 10 and BD PhoenixTMAutomated Microbiology System with Gatifloxacin (K020321, May 23,2002), Ofloxacin (K020323, April 14,2002), and Levofloxacin (K020322, March 27,2002). BD Diagnostic Systems Becton, D…
Becton, Dickinson & CO20024
*K020323BD Phoenix Automated Microbiology System
Summary page 1
…M Automated Microbiology System with Gatifloxacin (KO2032 1, May 23,2002), Ofloxacin (K020323, April 14,2002), and Levofloxacin (K020322, March 27,2002). INTENDED USE:…
Becton, Dickinson & CO2002—
*K020322BD Phoenix Automated Microbiology System
Summary page 1
…oxacin (KO2032 1, May 23,2002), Ofloxacin (K020323, April 14,2002), and Levofloxacin (K020322, March 27,2002). INTENDED USE: The BD PhoenixTMAutomated Microbiology System is…
Becton, Dickinson & CO2002—

Also cleared against K020321: 102 other devices →

What descends from it? (10)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K071026BD Geneohm Staphsr AssayBd Diagnostics (Geneohm Sciences Canada, Inc)20079
K082140Xpert Mrsa/SA Blood Culture AssayCepheid20083
K080837Xpert Mrsa/SA Ssti AssayCepheid20081

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 1, 2016Class IIIZ-0312-2017Process controlBD has determined that the PhoenixSpec AP Calibrator referenced in this recall from the following lot (5327782) are labeled with an incorrect expiration date (year). The expiration date printed on the individual calibration tubes within…
Jun 1, 2016Class IIIZ-0311-2017Process controlBD has determined that the BBL PhoenixSpec Calibrator kits referenced in this recall from the following lots (5364776, 5345796 ) are labeled with an incorrect expiration date (year). The expiration date printed on the individual…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 1 of the 3 ancestors has a recall of any class.
  • Of its descendants, K071026 (Bd Diagnostics (Geneohm Sciences Canada, Inc), 2007) has gone furthest, with 9 descendants of its own.
  • 102 other devices also cite its predicate K020321.

How does its category compare?

All LON devices

LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · LON

5in 2025

2007: 192008: 72009: 82010: 12011: 82012: 102013: 72014: 42016: 32017: 102018: 42019: 82020: 52021: 52022: 82023: 112024: 22025: 52026 (partial year): 4
2007high 19 (2007) · low 0 (2015)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LON

90in 2025

2007: 50 days median2008: 66 days median2009: 76 days median2011: 85 days median2012: 200 days median2013: 115 days median2014: 90 days median2016: 242 days median2017: 89 days median2018: 88 days median2019: 84 days median2020: 81 days median2021: 268 days median2022: 189 days median2023: 210 days median2025: 90 days median2026 (partial year): 146 days median
2007high 268 (2021) · low 50 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LON

1in 2025

2008: 12010: 1162011: 32012: 32014: 2, 2 Class I2016: 22017: 12018: 62019: 12020: 82022: 152024: 12025: 12026 (partial year): 6
2007high 116 (2010) · low 0 (2007)2026

Class I (2 since 2007) Class II/III not classified

Adverse event reports per year · LON

690in 2025

2007: 32008: 12010: 32011: 42012: 92013: 32014: 102015: 412016: 942017: 1972018: 5122019: 3152020: 2182021: 2322022: 1802023: 4532024: 4182025: 690
2007high 690 (2025) · low 0 (2009)2025

MAUDE reports naming product code LON, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007195003
200876611
200987600
20101—1163
201188534
20121020039
2013711503
20144902 (2)10
20150—041
20163242294
201710891197
20184886512
20198841315
20205818218
202152680232
2022818915180
2023112100453
20242—1418
20255901690
202641466—

What is it indicated for?

The BD PhoenixTh' Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enrerobacteriaceae and most gram-positive bacteria isolates from pure culture belonging to the genera Sraphylococcus and Enterococcus. This premarket…

Read the full statement (118 words)

The BD PhoenixTh' Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enrerobacteriaceae and most gram-positive bacteria isolates from pure culture belonging to the genera Sraphylococcus and Enterococcus.

This premarket notification is for the addition of the antimicrobial agent Oxacillin at concentrations of 0.0625-4 pg/mL to Gram Positive ID/AST or AST only Phoenix panels. Oxacillin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Active In Vitro and in Clinical Infections Against:

Aerobic Gram-positive microorganisms

Staphylococcus spp.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K023301: BD Phoenix Automated Microbiology System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K023301

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026