510(k) Report

510(k) · Cleared

Recalled · Class II · 2021

Prismaflex System 7.10

K131516510(k) TraditionalCleared Jan 3, 2014220-day reviewClass II21 CFR 876.5860Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Gambro Renal Products, Inc. · KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Cleared on Jan 3, 2014 after a 220-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 1 of the 8 devices in its family, counting itself, has a Class I recall. 7 later devices descend from it. It has been recalled 4 times.

Decided
Jan 3, 2014
Review
220 days
Class
Class II
Product code
KDI
Recalls
4
Predicates
0

7 later devices descend from it, 2 citing it directly. 1 of the 8 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2017: 1 descendant cleared2018: 1 descendant cleared2019: 1 descendant cleared2022: 2 descendants cleared2020: 2 descendants cleared, 1 with a Class I recall12010201520202025
  1. This deviceK131516 Prismaflex System 7.10Gambro Renal Products, Inc. · 2014
  2. Descendant · cites it directly · 4 of its ownK171671 Prismaflex System 8.10Baxter Healthcare Corporation · 2018
  3. Descendant · cites it directly · 4 of its ownK163530 PrisMax Control UnitBaxter Healthcare Corp · 2017
  4. Descendant · 2 generations on · 2 of its ownK190910 PrisMax System Version 2Baxter Healthcare Corp · 2019 · Recalled
  5. Descendant · 2 generations on · 0 of its ownK220281 multiFiltratePRO SystemFresenius Medical Care Renal Therapies Group, LLC · 2022
  6. Descendant · 3 generations on · 0 of its ownK212216 Prismaflex ST60 Set, Prismaflex ST100 Set, Prismaflex ST150 setBaxter Healthcare Corporation · 2022 · Recalled

Showing 5 of 7 descendants. See all descendants

Descendants cleared per year

2017: 12018: 12019: 12020: 22022: 2
2017high 2 (2020) · low 0 (2021)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 7 · See all
  • K171671Prismaflex System 8.10Baxter Healthcare Corporation · 2018 · cites it directly4descendants
  • K163530PrisMax Control UnitBaxter Healthcare Corp · 2017 · cites it directly4descendants
  • K190910PrisMax System Version 2Baxter Healthcare Corp · 2019 · 2 generations on2descendants
  • K220281multiFiltratePRO SystemFresenius Medical Care Renal Therapies Group, LLC · 2022 · 2 generations on0descendants
  • K212216Prismaflex ST60 Set, Prismaflex ST100 Set, Prismaflex ST150 setBaxter Healthcare Corporation · 2022 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K131516Prismaflex System 7.10Gambro Renal Products, Inc.2014—
+1K171671Prismaflex System 8.10Baxter Healthcare Corporation2018—
+1K163530PrisMax Control UnitBaxter Healthcare Corp2017—
+2K190910PrisMax System Version 2Baxter Healthcare Corp20195
+2K220281multiFiltratePRO SystemFresenius Medical Care Renal Therapies Group, LLC2022—
+3K212216Prismaflex ST60 Set, Prismaflex ST100 Set, Prismaflex ST150 setBaxter Healthcare Corporation20223

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
201810
201910
202021
202220
FIG. 1 — Lineage of K131516, 2014–2022Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KDI device

  1. #1 K243874 FX CorAL 60 (F00012969); FX CorAL 80…score 18.2 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it? (7)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K171671Prismaflex System 8.10Baxter Healthcare Corporation20184
K163530PrisMax Control UnitBaxter Healthcare Corp20174

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 28, 2021Class IIZ-1279-2021Process controlVariability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test.
Apr 7, 2020Class IIZ-2066-2020Under Investigation by firmPrismaflex product code 114870 does not have FDA regulatory clearance and was inadvertently distributed.
Apr 24, 2018Class IIZ-1925-2018Software designBaxter Healthcare will be installing new firmware on all Prismaflex control units to address the potential for a small number of these units to exhibit a failure mode with the pump module electronics. The failure mode may result in a…
Apr 24, 2018Class IIZ-1926-2018Software designBaxter Healthcare will be installing new firmware on all Prismaflex control units to address the potential for a small number of these units to exhibit a failure mode with the pump module electronics. The failure mode may result in a…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K192756Aquadex FlexFlow System 2.0Chf Solutions, Inc. · 2020 · 2 generations on · ◆ in the 2020 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K171671 (Baxter Healthcare Corporation, 2018) has gone furthest, with 4 descendants of its own.

How does its category compare?

All KDI devices

KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · KDI

9in 2025

2007: 112008: 62009: 42010: 92011: 52012: 62013: 112014: 82015: 52016: 52017: 82018: 22019: 42020: 92021: 22022: 72023: 72024: 72025: 92026 (partial year): 6
2007high 11 (2007) · low 2 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KDI

199in 2025

2007: 91 days median2008: 148 days median2009: 236 days median2010: 186 days median2011: 85 days median2012: 153 days median2013: 247 days median2014: 164 days median2015: 69 days median2016: 230 days median2017: 188 days median2019: 263 days median2020: 266 days median2022: 259 days median2023: 157 days median2024: 235 days median2025: 199 days median2026 (partial year): 154 days median
2007high 266 (2020) · low 69 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KDI

2in 2025

2007: 42009: 32010: 12011: 42012: 12013: 22014: 32015: 52016: 42017: 12018: 12019: 62021: 22022: 172023: 8, 8 Class I2024: 4, 2 Class I2025: 22026 (partial year): 2
2007high 17 (2022) · low 0 (2008)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · KDI

2,786in 2025

2007: 9162008: 1,0422009: 5222010: 8232011: 2,2702012: 1,7652013: 2,3602014: 6,1542015: 4,1822016: 7662017: 9632018: 1,7162019: 1,5872020: 1,6332021: 1,9442022: 2,1802023: 2,6922024: 2,4412025: 2,786
2007high 6,154 (2014) · low 522 (2009)2025

MAUDE reports naming product code KDI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711914916
2008614801,042
200942363522
201091861823
201158542,270
2012615311,765
20131124722,360
2014816436,154
201556954,182
201652304766
201781881963
20182—11,716
2019426361,587
2020926601,633
20212—21,944
20227259172,180
202371578 (8)2,692
202472354 (2)2,441
2025919922,786
202661542—

What is it indicated for?

The Prismaflex control unit is intended for:

* Continuous Renal Replacement Therapy (CRRT) for patients weighing 20 kilograms or more with acute renal failure and/or fluid overload.

* Therapeutic Plasma Exchange (TPE) therapy for patients weighing 20 kilograms or more with diseases where removal of plasma components is indicated.

All treatments administered via the Prismaflex control unit must be prescribed by a physician.

Indications for use as stated in the FDA 510(k) statement.

Cite this page

510(k) Report. “K131516: Prismaflex System 7.10.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K131516

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026