K131591510(k) TraditionalCleared Nov 27, 2013180-day reviewClass II21 CFR 888.3040Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Pega Medical, Inc. · HWC Screw, Fixation, Bone

Cleared on Nov 27, 2013 as substantially equivalent to K021932 (Synthes (Usa), 2002) and 4 other predicates after a 180-day review. Its family tree reaches back 5 generations to 1990. It has been recalled 2 times.

Decided
Nov 27, 2013
Review
180 days
Class
Class II
Product code
HWC
Recalls
2
Predicates
5

Its longest line of descent runs back 5 generations to K903810 (1990). 17 ancestors in all, the oldest K760454 (1976). No later device cites it as a predicate yet.

Other ancestors, by year cleared 2004: 1 other ancestor2002: 2 other ancestors1997: 1 other ancestor2007: 1 other ancestor1976: 1 other ancestor2000: 1 other ancestor1999: 3 other ancestors1998: 1 other ancestor1996: 1 other ancestorK903810 · Ace Cannulated Self Tapping Cancellous Bone Screw · Ace Medical Co. · 1990K051296 · Sijf Cannulated Screw System · Depuy Spine, Inc. · 2005K060736 · Smith & Nephew 6.5MM and 8.0MM Cannulated Screws · Smith & Nephew, Inc. · 2006 · recalledK080649 · Pogo Screw · Fxdevices · 2008K092189 · Pogo Screw · Fxdevices · 20091980199020002010
  1. Root · 5 generations backK903810 Ace Cannulated Self Tapping Cancellous Bone ScrewAce Medical Co. · 1990
  2. 4 generations backK051296 Sijf Cannulated Screw SystemDepuy Spine, Inc. · 2005
  3. 3 generations backK060736 Smith & Nephew 6.5MM and 8.0MM Cannulated ScrewsSmith & Nephew, Inc. · 2006 · Recalled
  4. 2 generations backK080649 Pogo ScrewFxdevices · 2008
  5. 1 generation backK092189 Pogo ScrewFxdevices · 2009
  6. This deviceK131591 Free Gliding Slipped Capital Femoral Epiphysis Screw (Scfe)Pega Medical, Inc. · 2013
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−5K903810Ace Cannulated Self Tapping Cancellous Bone ScrewAce Medical Co.1990—
−4K051296Sijf Cannulated Screw SystemDepuy Spine, Inc.2005—
−3K060736Smith & Nephew 6.5MM and 8.0MM Cannulated ScrewsSmith & Nephew, Inc.20061
−2K080649Pogo ScrewFxdevices2008—
−1K092189Pogo ScrewFxdevices2009—
0 (this device)K131591Free Gliding Slipped Capital Femoral Epiphysis Screw (Scfe)Pega Medical, Inc.2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197610
199610
199710
199810
199930
200010
200220
200410
200710
FIG. 1 — Lineage of K131591, 1990–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HWC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HWC device

  1. #1 K203011 Monster Screw Systemscore 22.2 · 14 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K021932Synthes 6.5 MM Cannulated Screw
Summary page 2
…K 13 1591 1. Synthes 6.5 mm Cannulated Screw (Stainless Steel screws), Synthes USA. (K021 932) 2. Hansson Pin System (Ti alloy nail and Stainless Steel screws & end caps), Howmedica Osteonics Corporation. (K96…
Synthes (Usa)20027
*K964893Hansson Pin System
Summary page 2
…eel screws), Synthes USA. (K021 932) 2. Hansson Pin System (Ti alloy nail and Stainless Steel screws & end caps), Howmedica Osteonics Corporation. (K964893 & K033968) 3. POGO Screw (Stainless Steel screws), FxDevices. (K092189 & K080649) The intended uses of these devic…
Osteonics Corp.1997—
*K033968Hansson Pin System
Summary page 2
…), Synthes USA. (K021 932) 2. Hansson Pin System (Ti alloy nail and Stainless Steel screws & end caps), Howmedica Osteonics Corporation. (K964893 & K033968) 3. POGO Screw (Stainless Steel screws), FxDevices. (K092189 & K080649) The intended uses of these devices are the…
Howmedica Osteonics Corp.2004—
*K092189Pogo Screw
Summary page 2
…alloy nail and Stainless Steel screws & end caps), Howmedica Osteonics Corporation. (K964893 & K033968) 3. POGO Screw (Stainless Steel screws), FxDevices. (K092189 & K080649) The intended uses of these devices are the same as the Free Gliding SOFE Screw System Summary of Technologie…
Fxdevices2009—
*K080649Pogo Screw
Summary page 2
…inless Steel screws & end caps), Howmedica Osteonics Corporation. (K964893 & K033968) 3. POGO Screw (Stainless Steel screws), FxDevices. (K092189 & K080649) The intended uses of these devices are the same as the Free Gliding SOFE Screw System Summary of Technologies: The technologic…
Fxdevices2008—

Also cleared against K021932: 61 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 20, 2015Class IIZ-0324-2016Unknown/Undetermined by firmSCFE Driver's threaded shaft broke close to the knob when a manual force was applied during the fixation of the screw.
Jun 15, 2015Class IIZ-1848-2015Reprocessing ControlsSCFE Drivers of the Free-Gliding SCFE Screw System are recalled because the handle of the instruments may detach from the shaft if high torques are applied in counter-clockwise rotation.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 5 companies; a new generation was cleared every 2.6 years on median.
  • 3 of the 17 ancestors have a recall of any class.
  • 61 other devices also cite its predicate K021932.

How does its category compare?

All HWC devices

HWC Screw, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · HWC

15in 2025

2007: 352008: 332009: 272010: 232011: 352012: 202013: 252014: 312015: 412016: 262017: 402018: 202019: 302020: 282021: 262022: 262023: 242024: 172025: 152026 (partial year): 21
2007high 41 (2015) · low 15 (2025)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HWC

111in 2025

2007: 77 days median2008: 76 days median2009: 113 days median2010: 110 days median2011: 88 days median2012: 99 days median2013: 90 days median2014: 88 days median2015: 85 days median2016: 131 days median2017: 127 days median2018: 106 days median2019: 108 days median2020: 75 days median2021: 84 days median2022: 167 days median2023: 104 days median2024: 90 days median2025: 111 days median2026 (partial year): 125 days median
2007high 167 (2022) · low 75 (2020)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HWC

1in 2025

2007: 252008: 332009: 292010: 72011: 122012: 302013: 142014: 22015: 52016: 42017: 42018: 12019: 52020: 42021: 22022: 132025: 12026 (partial year): 2
2007high 33 (2008) · low 0 (2023)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HWC

2,893in 2025

2007: 6692008: 6602009: 6512010: 7652011: 1,1042012: 3,3392013: 3,8502014: 2,5492015: 2,7112016: 2,2852017: 2,5612018: 3,6622019: 5,4632020: 4,8202021: 4,4462022: 3,9342023: 3,7182024: 3,5752025: 2,893
2007high 5,463 (2019) · low 651 (2009)2025

MAUDE reports naming product code HWC, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007357725669
2008337633660
20092711329651
2010231107765
20113588121,104
20122099303,339
20132590143,850
2014318822,549
2015418552,711
20162613142,285
20174012742,561
20182010613,662
20193010855,463
2020287544,820
2021268424,446
202226167133,934
20232410403,718
2024179003,575
20251511112,893
2026211252—

What is it indicated for?

The Free-Gliding SCFE Screw System is indicated as a temporary implant for stabilization of pediatric femoral neck fractures and slipped capital femoral epiphysis (SCFE) in all pediatric patients (less than or equal to 21 years old) with the exclusion of newborn and infants under 2 years of age.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K131591: Free Gliding Slipped Capital Femoral Epiphysis Screw (Scfe).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K131591

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026