510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Nmi Port II Catheter

K131694510(k) AbbreviatedCleared Aug 8, 201359-day reviewClass II21 CFR 880.5965Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Navilyst Medical, Inc. · LJT Port & Catheter, Implanted, Subcutaneous, Intravascular

Cleared on Aug 8, 2013 as substantially equivalent to K122767 (Navilyst Medical, Inc., 2013) and 1 other predicate after a 59-day review. Its family tree reaches back 11 generations to 2000. 2 of the 70 devices in its family, counting itself, have a Class I recall. 2 later devices descend from it. It has been recalled 2 times.

Decided
Aug 8, 2013
Review
59 days
Class
Class II
Product code
LJT
Recalls
2
Predicates
2

Its longest line of descent runs back 11 generations to K001901 (2000). 67 ancestors in all, the oldest K830256 (1983). 2 later devices descend from it, 2 citing it directly. 2 of the 70 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2012: 1 other ancestor2010: 1 other ancestor2009: 3 other ancestors2008: 2 other ancestors2007: 2 other ancestors2003: 6 other ancestors2006: 4 other ancestors2005: 6 other ancestors2004: 5 other ancestors2002: 1 other ancestor2001: 5 other ancestors2000: 4 other ancestors, 1 with a Class I recall11987: 2 other ancestors1999: 2 other ancestors1997: 2 other ancestors1996: 1 other ancestor1991: 2 other ancestors1983: 1 other ancestor1986: 2 other ancestors, 1 with a Class I recall11989: 4 other ancestors2020: 1 descendant cleared2015: 1 descendant clearedK001901 · Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath Midline Catheter, Model 3134135 · C.R. Bard, Inc. · 2000 · recalledK012902 · Poly Per-Q-Cath Single Lumen Picc Catheter, Model 3143100; Poly Per-Q-Cath Dual Lumen Picc Catheter, Model 3246100 · C.R. Bard, Inc. · 2001K031129 · Poly Per-Q-Cath Picc Catheters; Poly Radpicc · C.R. Bard, Inc. · 2003K034019 · Poly Per-Q-Cath Picc Catheters · C.R. Bard, Inc. · 2004 · recalled×2: 2 generations cleared close together, K050931 (2005), K051672 (2005) K060812 · Powerport (TM)Implanted Titanium Port W/8 FR. Chronoflex Catheter-Multiple · C.R. Bard, Inc. · 2006 · recalled×3: 3 generations cleared close together, K070003 (2007), K071993 (2007), K073210 (2007) K122767 · Nmi Port · Navilyst Medical, Inc. · 2013 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 11 other ancestors3 generations back: 9 other ancestors4 generations back: 10 other ancestors5 generations back: 13 other ancestors, 1 with a Class I recall16 generations back: 5 other ancestors, 1 with a Class I recall17 generations back: 2 other ancestors8 generations back: 5 other ancestors 1 generation forward: 2 descendants K122767 · Nmi Port · Navilyst Medical, Inc. · 2013 · recalledK073210 · Modification to: T-Port HP Infusion Port · Pfm Medical, Inc. · 2007K071993 · T-Port HP Infusion Port · Pfm Medical, Inc. · 2007K070003 · Power Injectable Implantable Infusion Port with Silicone Lumen (Attachable and Pre-Attached) · Medcomp · 2007 · recalledK060812 · Powerport (TM)Implanted Titanium Port W/8 FR. Chronoflex Catheter-Multiple · C.R. Bard, Inc. · 2006 · recalledK051672 · Powerpicc · Bard Access Systems, Inc. · 2005 · recalledK050931 · 6 FR DL Powerpicc Catheter · C.R. Bard, Inc. · 2005K034019 · Poly Per-Q-Cath Picc Catheters · C.R. Bard, Inc. · 2004 · recalledK031129 · Poly Per-Q-Cath Picc Catheters; Poly Radpicc · C.R. Bard, Inc. · 2003K012902 · Poly Per-Q-Cath Single Lumen Picc Catheter, Model 3143100; Poly Per-Q-Cath Dual Lumen Picc Catheter, Model 3246100 · C.R. Bard, Inc. · 2001K001901 · Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath Midline Catheter, Model 3134135 · C.R. Bard, Inc. · 2000 · recalled−10−50
  1. Root · 11 generations backK001901 Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath…C.R. Bard, Inc. · 2000 · Recalled
  2. 7 generations between
  3. 3 generations backK071993 T-Port HP Infusion PortPfm Medical, Inc. · 2007
  4. 2 generations backK073210 Modification to: T-Port HP Infusion PortPfm Medical, Inc. · 2007
  5. 1 generation backK122767 Nmi PortNavilyst Medical, Inc. · 2013 · Recalled
  6. This deviceK131694 Nmi Port II CatheterNavilyst Medical, Inc. · 2013
  7. Descendant · cites it directly · 0 of its ownK190559 SmartPort+ Implantable Ports, SmartPort Plastic Implantable PortsAngioDynamics, Inc. · 2020
  8. Descendant · cites it directly · 0 of its ownK153228 NMI Port, NMI Port IINavilyst Medical, Inc. · 2015
Show all 11 generations
  1. Root · 11 generations backK001901 Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath…C.R. Bard, Inc. · 2000 · Recalled
  2. 10 generations backK012902 Poly Per-Q-Cath Single Lumen Picc Catheter, Model 3143100; Poly…C.R. Bard, Inc. · 2001
  3. 9 generations backK031129 Poly Per-Q-Cath Picc Catheters; Poly RadpiccC.R. Bard, Inc. · 2003
  4. 8 generations backK034019 Poly Per-Q-Cath Picc CathetersC.R. Bard, Inc. · 2004 · Recalled
  5. 7 generations backK050931 6 FR DL Powerpicc CatheterC.R. Bard, Inc. · 2005
  6. 6 generations backK051672 PowerpiccBard Access Systems, Inc. · 2005 · Recalled
  7. 5 generations backK060812 Powerport (TM)Implanted Titanium Port W/8 FR. Chronoflex…C.R. Bard, Inc. · 2006 · Recalled
  8. 4 generations backK070003 Power Injectable Implantable Infusion Port with Silicone Lumen…Medcomp · 2007 · Recalled
  9. 3 generations backK071993 T-Port HP Infusion PortPfm Medical, Inc. · 2007
  10. 2 generations backK073210 Modification to: T-Port HP Infusion PortPfm Medical, Inc. · 2007
  11. 1 generation backK122767 Nmi PortNavilyst Medical, Inc. · 2013 · Recalled
  12. This deviceK131694 Nmi Port II CatheterNavilyst Medical, Inc. · 2013

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K190559SmartPort+ Implantable Ports, SmartPort Plastic Implantable PortsAngioDynamics, Inc. · 2020 · cites it directly0descendants
  • K153228NMI Port, NMI Port IINavilyst Medical, Inc. · 2015 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−11K001901Poly Per-Q-Cath Picc Catheter,Model 3236100 and Poly Per-Q-Cath Midline…C.R. Bard, Inc.20002
−10K012902Poly Per-Q-Cath Single Lumen Picc Catheter, Model 3143100; Poly Per-Q-Cath…C.R. Bard, Inc.2001—
−9K031129Poly Per-Q-Cath Picc Catheters; Poly RadpiccC.R. Bard, Inc.2003—
−8K034019Poly Per-Q-Cath Picc CathetersC.R. Bard, Inc.20041
−7K0509316 FR DL Powerpicc CatheterC.R. Bard, Inc.2005—
−6K051672PowerpiccBard Access Systems, Inc.20051
−5K060812Powerport (TM)Implanted Titanium Port W/8 FR. Chronoflex Catheter-MultipleC.R. Bard, Inc.20061
−4K070003Power Injectable Implantable Infusion Port with Silicone Lumen (Attachable and…Medcomp20073
−3K071993T-Port HP Infusion PortPfm Medical, Inc.2007—
−2K073210Modification to: T-Port HP Infusion PortPfm Medical, Inc.2007—
−1K122767Nmi PortNavilyst Medical, Inc.20136
0 (this device)K131694Nmi Port II CatheterNavilyst Medical, Inc.2013—
+1K190559SmartPort+ Implantable Ports, SmartPort Plastic Implantable PortsAngioDynamics, Inc.2020—
+1K153228NMI Port, NMI Port IINavilyst Medical, Inc.2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198310
198621
198720
198940
199120
199610
199720
199920
200041
200150
200210
200360
200450
200560
200640
200720
200820
200930
201010
201210

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
202010
FIG. 1 — Lineage of K131694, 2000–2020Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LJT device

  1. #1 K251253 PowerPort™ isp Implantable Port; PowerPort™…score 14.9 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K122767Nmi Port
Summary page 1
…tion Panel: 2ICFR§880.5965, Class 11 General Hospital Premarket Notification: K 122767 1rade Name: NMI P1CC III Common/Usual name: Peripherally Inserte…
Navilyst Medical, Inc.20136
*K121089Nmi Picc Iii
Summary page 1
…ion Panel: 2ICFR§880.5970, Class 11 General Hospital Premarket Notification: K 121089 5-2 E. Device De…
Navilyst Medical, Inc.2012—

Also cleared against K122767: 1 other device →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K190559SmartPort+ Implantable Ports, SmartPort Plastic Implantable PortsAngioDynamics, Inc.20200
K153228NMI Port, NMI Port IINavilyst Medical, Inc.20150

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 24, 2018Class IIZ-0076-2019Unknown/Undetermined by firmBioFlo Ports was partially manufactured with the incorrect sterile port assembly. The actual sterile port assembly used was, 8F SL Xcela Plus Titanium Valved Non-Filled, when it should have been, 8F SL BioFlo Plastic Valved Non-Filled
Nov 25, 2013Class IIZ-0826-2014Device DesignNavilyst Medical has recalled a Port Kit with Access Infusion Set due to a possible breach of sterilization barrier.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K993691Arrowg +Ard Blue Plus Antimicrobial Multi-Lumen Central Venous CatheterArrow Intl., Inc. · 2000 · 5 generations back · ◆ in the 2000 ancestors bar of FIG. 11 Class I recall
  • K862056[Central Venous][Multi-Lumen] Catheter W/Flex TipArrow Intl., Inc. · 1986 · 6 generations back · ◆ in the 1986 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 13 years and passes through 5 companies; a new generation was cleared every 9 months on median.
  • 30 of the 67 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K190559 (AngioDynamics, Inc., 2020) has gone furthest, with 0 descendants of its own.
  • 1 other device also cites its predicate K122767.

How does its category compare?

All LJT devices

LJT Port & Catheter, Implanted, Subcutaneous, Intravascular: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · LJT

2in 2025

2007: 112008: 72009: 22010: 22011: 52012: 22013: 72014: 22015: 42016: 32019: 12020: 12023: 12024: 12025: 22026 (partial year): 1
2007high 11 (2007) · low 0 (2017)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LJT

182in 2016

2007: 85 days median2008: 119 days median2011: 144 days median2013: 150 days median2015: 164 days median2016: 182 days median
2007high 182 (2016) · low 85 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LJT

1in 2022

2007: 12008: 22009: 22010: 192011: 12012: 22015: 12017: 12018: 12019: 72022: 12026 (partial year): 1, 1 Class I
2007high 19 (2010) · low 0 (2013)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · LJT

4,308in 2025

2007: 2082008: 1612009: 1722010: 2072011: 2022012: 1532013: 1412014: 1432015: 1492016: 1352017: 2622018: 8782019: 7652020: 9592021: 1,2852022: 1,3342023: 1,2212024: 1,8962025: 4,308
2007high 4,308 (2025) · low 135 (2016)2025

MAUDE reports naming product code LJT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711851208
200871192161
20092—2172
20102—19207
201151441202
20122—2153
201371500141
20142—0143
201541641149
201631820135
20170—1262
20180—1878
20191—7765
20201—0959
20210—01,285
20220—11,334
20231—01,221
20241—01,896
20252—04,308
20261—1 (1)—

What is it indicated for?

The NMI Port 11 with and without PASV Valve Technology is indicated for patients who require long-term access to the central venous system for administration of fluids including but not limited to hydration fluids, chemotherapy, analgesics, nutritional therapy and blood products. The device is also indicated for blood specimen withdrawal.

When used with a power injectable needle, the NMI Port 11 is indicated for power injection of contrast media. The maximum recommended infusion rate is 5 mi/sec with a 19G or 20G non-coring power injectable needle or 2 mI/sec with a 220 non-coring power injectable needle.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K131694: Nmi Port II Catheter.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K131694

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026