510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Syngo.Plaza VB10A

K132532510(k) TraditionalCleared Dec 9, 2013119-day reviewClass II21 CFR 892.2050Recalls 11

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens AG Healthcare · LLZ System, Image Processing, Radiological

Cleared on Dec 9, 2013 as substantially equivalent to K101666 (Siemens AG Medical Solutions, 2010) after a 119-day review. Its family tree reaches back 8 generations to 1997. 8 later devices descend from it. It has been recalled 11 times.

Decided
Dec 9, 2013
Review
119 days
Class
Class II
Product code
LLZ
Recalls
11
Predicates
1

Its longest line of descent runs back 8 generations to K971684 (1997). 25 ancestors in all, the oldest K945188 (1995). 8 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2010: 1 other ancestor2007: 2 other ancestors2008: 2 other ancestors2005: 3 other ancestors2001: 2 other ancestors2000: 2 other ancestors1999: 1 other ancestor1997: 2 other ancestors1996: 1 other ancestor1995: 1 other ancestor2015: 1 descendant cleared2018: 2 descendants cleared2016: 1 descendant cleared2020: 2 descendants cleared2021: 1 descendant cleared2023: 1 descendant clearedK971684 · Magnetom Project 047 System (Numaris 3.5 Software) · Siemens Medical Solutions USA, Inc. · 1997 · recalledK032428 · Magnetom Avanto MR System · Siemens Medical Solutions USA, Inc. · 2003 · recalledK041112 · Magnetom Espree System · Siemens Medical Solutions USA, Inc. · 2004 · recalledK052423 · Magnetom Systems with Syngo Expert - I Option · Siemens Medi Cal Solutions, Inc. · 2006K061964 · Syngo Multimodality Workplace · Siemens AG Medical Solutions · 2006K072728 · Syngo Multimodality Workplace · Siemens AG Medical Solutions · 2008K092519 · Syngo.X · Siemens AG Medical Solutions · 2009 · recalledK101666 · Syngo. Plaza VA20A · Siemens AG Medical Solutions · 2010 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 2 generations back: 2 other ancestors3 generations back: 2 other ancestors4 generations back: 1 other ancestor5 generations back: 2 other ancestors6 generations back: 8 other ancestors7 generations back: 1 other ancestor8 generations back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 2 descendants3 generations forward: 1 descendant4 generations forward: 1 descendant5 generations forward: 1 descendant6 generations forward: 1 descendant K101666 · Syngo. Plaza VA20A · Siemens AG Medical Solutions · 2010 · recalledK092519 · Syngo.X · Siemens AG Medical Solutions · 2009 · recalledK072728 · Syngo Multimodality Workplace · Siemens AG Medical Solutions · 2008K061964 · Syngo Multimodality Workplace · Siemens AG Medical Solutions · 2006K052423 · Magnetom Systems with Syngo Expert - I Option · Siemens Medi Cal Solutions, Inc. · 2006K041112 · Magnetom Espree System · Siemens Medical Solutions USA, Inc. · 2004 · recalledK032428 · Magnetom Avanto MR System · Siemens Medical Solutions USA, Inc. · 2003 · recalledK971684 · Magnetom Project 047 System (Numaris 3.5 Software) · Siemens Medical Solutions USA, Inc. · 1997 · recalled−50+5
  1. Root · 8 generations backK971684 Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc. · 1997 · Recalled
  2. 7 generations backK032428 Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc. · 2003 · Recalled
  3. 6 generations backK041112 Magnetom Espree SystemSiemens Medical Solutions USA, Inc. · 2004 · Recalled
  4. 5 generations backK052423 Magnetom Systems with Syngo Expert - I OptionSiemens Medi Cal Solutions, Inc. · 2006
  5. 4 generations backK061964 Syngo Multimodality WorkplaceSiemens AG Medical Solutions · 2006
  6. 3 generations backK072728 Syngo Multimodality WorkplaceSiemens AG Medical Solutions · 2008
  7. 2 generations backK092519 Syngo.XSiemens AG Medical Solutions · 2009 · Recalled
  8. 1 generation backK101666 Syngo. Plaza VA20ASiemens AG Medical Solutions · 2010 · Recalled
  9. This deviceK132532 Syngo.Plaza VB10ASiemens AG Healthcare · 2013
  10. Descendant · cites it directly · 6 of its ownK142750 JiveXVisus Technology Transfer GmbH · 2015
  11. Descendant · 2 generations on · 4 of its ownK162008 JiveXVisus Technology Transfer GmbH · 2016
  12. Descendant · 3 generations on · 3 of its ownK181964 JiveXVisus Health IT GmbH · 2018
  13. Descendant · 4 generations on · 2 of its ownK200703 JiveXVisus Health IT GmbH · 2020
  14. Descendant · 5 generations on · 1 of its ownK212321 JiveX (Model Number / Release: 5.3)Visus Health IT GmbH · 2021

Showing 5 of 8 descendants. See all descendants

Descendants cleared per year

2015: 12016: 12018: 22020: 22021: 12023: 1
2015high 2 (2018) · low 0 (2017)2026

Most influential descendants

Showing 5 of 8 · See all
  • K142750JiveXVisus Technology Transfer GmbH · 2015 · cites it directly6descendants
  • K162008JiveXVisus Technology Transfer GmbH · 2016 · 2 generations on4descendants
  • K181964JiveXVisus Health IT GmbH · 2018 · 3 generations on3descendants
  • K200703JiveXVisus Health IT GmbH · 2020 · 4 generations on2descendants
  • K212321JiveX (Model Number / Release: 5.3)Visus Health IT GmbH · 2021 · 5 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K971684Magnetom Project 047 System (Numaris 3.5 Software)Siemens Medical Solutions USA, Inc.19974
−7K032428Magnetom Avanto MR SystemSiemens Medical Solutions USA, Inc.20034
−6K041112Magnetom Espree SystemSiemens Medical Solutions USA, Inc.20044
−5K052423Magnetom Systems with Syngo Expert - I OptionSiemens Medi Cal Solutions, Inc.2006—
−4K061964Syngo Multimodality WorkplaceSiemens AG Medical Solutions2006—
−3K072728Syngo Multimodality WorkplaceSiemens AG Medical Solutions2008—
−2K092519Syngo.XSiemens AG Medical Solutions20094
−1K101666Syngo. Plaza VA20ASiemens AG Medical Solutions20103
0 (this device)K132532Syngo.Plaza VB10ASiemens AG Healthcare2013—
+1K142750JiveXVisus Technology Transfer GmbH2015—
+2K162008JiveXVisus Technology Transfer GmbH2016—
+3K181964JiveXVisus Health IT GmbH2018—
+4K200703JiveXVisus Health IT GmbH2020—
+5K212321JiveX (Model Number / Release: 5.3)Visus Health IT GmbH2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510
199610
199720
199910
200020
200120
200530
200720
200820
201010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
201610
201820
202020
202110
202310
FIG. 1 — Lineage of K132532, 1997–2023Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K101666Syngo. Plaza VA20A
Summary page 2
…dates / upgrades are offered based on service contracts and fulfilled by trained service technicians. This is also applicable for the predicate device syngo.plaza VA2OA (K 101666). 510(k) for svngodW.plaza July 30. 2013 Page K 3 v ) 28eg £10T 0€ AIn…
Siemens AG Medical Solutions20103

What descends from it? (8)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K142750JiveXVisus Technology Transfer GmbH20156
K180563syngo.plazaSiemens Healthcare GmbH20180

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 20, 2018Class IIZ-1700-2018Software designSoftware upgrade to correct format of study dates and issues with Legacy Presentation States (annotations) in order to prevent potential patient misdiagnosis.
Feb 20, 2018Class IIZ-1699-2018Software designSoftware upgrade to correct format of study dates and issues with Legacy Presentation States (annotations) in order to prevent potential patient misdiagnosis.
Aug 23, 2017Class IIZ-0524-2018Software designWhen a prior study is being replaced in the workflow step, in certain scenarios (based on the Display Protocol configuration) the prior study will only be replaced in the active Workflow Step and not show in all other workflow steps. The…
Aug 18, 2017Class IIZ-0495-2018Software designSoftware update to correct several issues that include (1)Potential data loss, (2) Study mix-up, (3) Incorrect measurements on multi-frame images, (4) Dearchiving issue, and (5) Unauthorized access of data due to inadequate permissions…
May 22, 2017Class IIZ-0168-2018Software designSiemens is releasing a non-medical software application LTA Incomplete Archive Check Tool V1.0. It is intended to eliminate an issue that occurs during syngo.plaza de-archiving from Dicom LTA. It has been defined that the number of…
Sep 6, 2016Class IIZ-2892-2016Software designSoftware upgrade to eliminate several issues
Aug 2, 2016Class IIZ-2794-2016Error in labelingSiemens is releasing a Customer Advisory Notice to explain meaning of a suffix o.p. that shows on some DICOM images on area, distance or perimeter values. This suffix stands for on projection and indicates that a shown measurement…
Dec 22, 2015Class IIZ-0862-2016Software designSiemens is releasing an updated software version to address several software issues including RGB images will show "?" since calculation of HU is not possible; save as option enabled; changes in access for loading studies; breast region…
Feb 18, 2015Class IIZ-1354-2015Software designPossibly incomplete archived studies during pre-fetch. In a server farm setup, when pre-fetch/retrieve operation is performed for partially archived studies, the series that have not yet been archived, will remain unarchived.
Jan 13, 2015Class IIZ-1027-2015Software Manufacturing/Software DeploymentIn case of a system crash, images may not be written to the hard disk and this may result in inconsistencies in the database. In case of a system crash (e.g. blue screen, power outage) images may not be written from cache to the hard disk…
Nov 11, 2014Class IIZ-0576-2015Component design/selectionPrintouts may be printed in incorrect anatomical size when using syngo.plaza or syngo Imaging XS filming application in conjunction with a printer not released for anatomical print usage. This may happen due to an invalid combination of…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 16 years and passes through 4 companies; a new generation was cleared every 1.5 years on median.
  • 14 of the 25 ancestors have a recall of any class.
  • Of its descendants, K142750 (Visus Technology Transfer GmbH, 2015) has gone furthest, with 6 descendants of its own.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

syngo.plaza is a Picture Archiving and Communication System intended to display, process, read, report, communicate, distribute, store, and archive digital medical images, including maninographic images. It supports the physician in diagnosis and treatment planning. For primary image diagnosis in Mammography only uncompressed or non-lossy compressed images and only preprocessed DICOM "For Presentation" images must be used. Also monitors (displays) and…

Read the full statement (114 words)

syngo.plaza is a Picture Archiving and Communication System intended to display, process, read, report, communicate, distribute, store, and archive digital medical images, including maninographic images.

It supports the physician in diagnosis and treatment planning.

For primary image diagnosis in Mammography only uncompressed or non-lossy compressed images and only preprocessed DICOM "For Presentation" images must be used. Also monitors (displays) and printers which received FDA clearance for Mammography must be used.

syngo.plaza also supports DICOM Structured Reports.

In a comprehensive imaging suite, syngo.plaza integrates Hospital / Radiology Information Systems (HIS / RIS) to enable customer specific workflows.

s)7igo.plaza optionally uses a variety of advanced postprocessing applications.

Note:

Web-based image distribution is not intended for reporting.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K132532: Syngo.Plaza VB10A.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K132532

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026