510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Mega 8FR. 50CC and Sensation Plus 8FR. 50CC Intra-Aortic Balloon Catheters and Accessories

K133074510(k) SpecialCleared Dec 12, 201373-day reviewClass II21 CFR 870.3535Recalls 9

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Datascope Corp. · DSP System, Balloon, Intra-Aortic And Control

Cleared on Dec 12, 2013 as substantially equivalent to K091449 (Datascope Corp., Cardiac Assist Division, 2009) and 1 other predicate after a 73-day review. Its family tree reaches back 10 generations to 1990. 8 of the 67 devices in its family, counting itself, have a Class I recall. It has been recalled 9 times.

Decided
Dec 12, 2013
Review
73 days
Class
Class II
Product code
DSP
Recalls
9
Predicates
2

Its longest line of descent runs back 10 generations to K892222 (1990). 66 ancestors in all, the oldest K790775 (1979). No later device cites it as a predicate yet. 8 of the 67 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2009: 1 other ancestor1990: 3 other ancestors2006: 1 other ancestor, 1 with a Class I recall12004: 1 other ancestor2003: 1 other ancestor, 1 with a Class I recall12002: 1 other ancestor, 1 with a Class I recall12000: 5 other ancestors, 2 with a Class I recall21998: 4 other ancestors, 1 with a Class I recall11995: 9 other ancestors2001: 1 other ancestor1997: 6 other ancestors, 1 with a Class I recall11996: 8 other ancestors1994: 3 other ancestors1993: 4 other ancestors1992: 1 other ancestor1991: 4 other ancestors1989: 1 other ancestor1982: 1 other ancestor1979: 1 other ancestorK892222 · 9.5F Percu Double Lumen Intra-Aortic Balloon · Kontron Instruments, Inc. · 1990K963187 · Modified Labeling of Cardiac Assist Intra-Aortic Balloon Catheters: Model 930 - 940, 30 - 40 CC Sensation and Nicath 30C · Boston Scientific Corp · 1996K972113 · 30 CC and 40 CC Sub-9 Nitinol and 30 and 40 CC Sub-9 Stainless Steel · Boston Scientific Corp · 1997K980385 · Datascope 8FR. CO-Lumen 34 and 40CC Intra-Aortic Balloon Catheters for Optional Sheathless Insertion with Accessories · Datascope Corp. · 1998K994157 · Datascope's 8FR. Polymide 25CC,34CC and 40CC Iab and Accessories, Models 0684-00-0327, 06844-00-0328, 0684-00-0329 · Datascope Corp. · 2000K013326 · Modification to: Datascope 8FR. Polyimide Alt B Iab, 25CC, 34CC, 40CC, Datascope Insertion Kit for 8 FR, Polyimide Alt B · Datascope Corp. · 2001K031569 · Datascope CA 40 8FR. Iabs, Model 0684-00-04XX · Datascope Corp. · 2003K041281 · Datascope 7.5FR. Iab Catheter and Accessories · Datascope Corp. · 2004 · recalledK063525 · Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter and CS300 Intra-Aortic Balloon Pump Control System · Datascope Corp. · 2007 · Class I recallK112327 · Sensation (TM) Plus 8FR. 50CC Intra-Aortic Balloon Catheter and Accessories · Cardiac Assist, Maquet Cardiovascular, LLC · 2011 · recalled1980199020002010
Other ancestors, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor3 generations back: 8 other ancestors, 4 with a Class I recall44 generations back: 9 other ancestors, 3 with a Class I recall35 generations back: 29 other ancestors6 generations back: 8 other ancestors K112327 · Sensation (TM) Plus 8FR. 50CC Intra-Aortic Balloon Catheter and Accessories · Cardiac Assist, Maquet Cardiovascular, LLC · 2011 · recalledK063525 · Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter and CS300 Intra-Aortic Balloon Pump Control System · Datascope Corp. · 2007 · Class I recallK041281 · Datascope 7.5FR. Iab Catheter and Accessories · Datascope Corp. · 2004 · recalledK031569 · Datascope CA 40 8FR. Iabs, Model 0684-00-04XX · Datascope Corp. · 2003K013326 · Modification to: Datascope 8FR. Polyimide Alt B Iab, 25CC, 34CC, 40CC, Datascope Insertion Kit for 8 FR, Polyimide Alt B · Datascope Corp. · 2001K994157 · Datascope's 8FR. Polymide 25CC,34CC and 40CC Iab and Accessories, Models 0684-00-0327, 06844-00-0328, 0684-00-0329 · Datascope Corp. · 2000K980385 · Datascope 8FR. CO-Lumen 34 and 40CC Intra-Aortic Balloon Catheters for Optional Sheathless Insertion with Accessories · Datascope Corp. · 1998K972113 · 30 CC and 40 CC Sub-9 Nitinol and 30 and 40 CC Sub-9 Stainless Steel · Boston Scientific Corp · 1997K963187 · Modified Labeling of Cardiac Assist Intra-Aortic Balloon Catheters: Model 930 - 940, 30 - 40 CC Sensation and Nicath 30C · Boston Scientific Corp · 1996K892222 · 9.5F Percu Double Lumen Intra-Aortic Balloon · Kontron Instruments, Inc. · 1990−10−50
  1. Root · 10 generations backK892222 9.5F Percu Double Lumen Intra-Aortic BalloonKontron Instruments, Inc. · 1990
  2. 6 generations between
  3. 3 generations backK041281 Datascope 7.5FR. Iab Catheter and AccessoriesDatascope Corp. · 2004 · Recalled
  4. 2 generations backK063525 Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter…Datascope Corp. · 2007 · Recalled (Class I)
  5. 1 generation backK112327 Sensation (TM) Plus 8FR. 50CC Intra-Aortic Balloon Catheter and…Cardiac Assist, Maquet Cardiovascular, LLC · 2011 · Recalled
  6. This deviceK133074 Mega 8FR. 50CC and Sensation Plus 8FR. 50CC Intra-Aortic Balloon…Datascope Corp. · 2013
Show all 10 generations
  1. Root · 10 generations backK892222 9.5F Percu Double Lumen Intra-Aortic BalloonKontron Instruments, Inc. · 1990
  2. 9 generations backK963187 Modified Labeling of Cardiac Assist Intra-Aortic Balloon Catheters…Boston Scientific Corp · 1996
  3. 8 generations backK972113 30 CC and 40 CC Sub-9 Nitinol and 30 and 40 CC Sub-9 Stainless SteelBoston Scientific Corp · 1997
  4. 7 generations backK980385 Datascope 8FR. CO-Lumen 34 and 40CC Intra-Aortic Balloon Catheters…Datascope Corp. · 1998
  5. 6 generations backK994157 Datascope's 8FR. Polymide 25CC,34CC and 40CC Iab and Accessories…Datascope Corp. · 2000
  6. 5 generations backK013326 Modification to: Datascope 8FR. Polyimide Alt B Iab, 25CC, 34CC…Datascope Corp. · 2001
  7. 4 generations backK031569 Datascope CA 40 8FR. Iabs, Model 0684-00-04XXDatascope Corp. · 2003
  8. 3 generations backK041281 Datascope 7.5FR. Iab Catheter and AccessoriesDatascope Corp. · 2004 · Recalled
  9. 2 generations backK063525 Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter…Datascope Corp. · 2007 · Recalled (Class I)
  10. 1 generation backK112327 Sensation (TM) Plus 8FR. 50CC Intra-Aortic Balloon Catheter and…Cardiac Assist, Maquet Cardiovascular, LLC · 2011 · Recalled
  11. This deviceK133074 Mega 8FR. 50CC and Sensation Plus 8FR. 50CC Intra-Aortic Balloon…Datascope Corp. · 2013

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−10K8922229.5F Percu Double Lumen Intra-Aortic BalloonKontron Instruments, Inc.1990—
−9K963187Modified Labeling of Cardiac Assist Intra-Aortic Balloon Catheters: Model 930…Boston Scientific Corp1996—
−8K97211330 CC and 40 CC Sub-9 Nitinol and 30 and 40 CC Sub-9 Stainless SteelBoston Scientific Corp1997—
−7K980385Datascope 8FR. CO-Lumen 34 and 40CC Intra-Aortic Balloon Catheters for…Datascope Corp.1998—
−6K994157Datascope's 8FR. Polymide 25CC,34CC and 40CC Iab and Accessories, Models…Datascope Corp.2000—
−5K013326Modification to: Datascope 8FR. Polyimide Alt B Iab, 25CC, 34CC, 40CC…Datascope Corp.2001—
−4K031569Datascope CA 40 8FR. Iabs, Model 0684-00-04XXDatascope Corp.2003—
−3K041281Datascope 7.5FR. Iab Catheter and AccessoriesDatascope Corp.20047
−2K063525Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter and CS300…Datascope Corp.2007Class I
−1K112327Sensation (TM) Plus 8FR. 50CC Intra-Aortic Balloon Catheter and AccessoriesCardiac Assist, Maquet Cardiovascular, LLC20111
0 (this device)K133074Mega 8FR. 50CC and Sensation Plus 8FR. 50CC Intra-Aortic Balloon Catheters and…Datascope Corp.2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197910
198210
198910
199030
199140
199210
199340
199430
199590
199680
199761
199841
200052
200110
200211
200311
200410
200611
200910
FIG. 1 — Lineage of K133074, 1990–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

8 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DSP device

  1. #1 K250542 AC3™ Range™ Intra-Aortic Balloon Pumpscore 11.2 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K091449Datascope's 8FR. 50CC Intra-Aortic Balloon Catheter and Accessories, Models…
Summary page 1
…ct Code: DSP PREDICATE DEVICE INFORMATION: The enhanced 8Fr. Introducer Set is substantially equivalent to the BFr. Introducer Set packaged in the MEGA (K091449) and SENSATION PLUS (K1 12327) 8Fr. 50cc lntra-Aortic Balloon (lAB) Catheters and Accessories predicate devices. DEVICE DESCRIPTI…
Datascope Corp., Cardiac Assist Division2009—
*K112327Sensation (TM) Plus 8FR. 50CC Intra-Aortic Balloon Catheter and Accessories
Summary page 1
…PREDICATE DEVICE INFORMATION: The enhanced 8Fr. Introducer Set is substantially equivalent to the BFr. Introducer Set packaged in the MEGA (K091449) and SENSATION PLUS (K1 12327) 8Fr. 50cc lntra-Aortic Balloon (lAB) Catheters and Accessories predicate devices. DEVICE DESCRIPTION AND INTENDED USE: The 8Fr.…
Cardiac Assist, Maquet Cardiovascular, LLC20111

Also cleared against K091449: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (8 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 17, 2024Class IIZ-0311-2025Error in labelingThere is discrepant labeling on the inner and outer packaging of the device.
Jun 23, 2023Class IIZ-2222-2023Under Investigation by firmDuring IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization…
Jun 23, 2023Class IIZ-2220-2023Under Investigation by firmDuring IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization…
Aug 5, 2022Class IIZ-1687-2022Vendor change controlCertain lots containing undersized dilator.
Aug 5, 2022Class IIZ-1686-2022Vendor change controlCertain lots containing undersized dilator.
Aug 5, 2022Class IIZ-1685-2022Vendor change controlCertain lots containing undersized dilator.
Jul 27, 2020Class IIZ-2837-2020Process change controlPotential Endotoxin Contamination
Feb 21, 2018Class IIZ-1385-2018Error in labelingThe U.S. Instructions For Use for the Mega Intra-Aortic Balloon (IAB) Catheter was missing the Contraindications section.
Apr 12, 2017Class IIZ-2164-2017Device DesignMaquet/Getinge is initiating a voluntary product removal involving four lot numbers of SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK since during internal pressure (leak) testing of the SENSATION…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K063525Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter and CS300 Intra-Aortic Balloon Pump…Datascope Corp. · 2007 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K060309Autocat Intra-Aortic Balloon Pump SeriesArrow Intl., Inc. · 2006 · 3 generations back · ◆ in the 2006 ancestors bar of FIG. 128 Class I recalls
  • K031636CS100 Intra-Aortic Balloon Pump, Model 0998-00-3013-XXDatascope Corp. · 2003 · 3 generations back · ◆ in the 2003 ancestors bar of FIG. 15 Class I recalls
  • K021462Arrow Intra-Aortic Balloon (Iab) Catheter with a Fiber Optic Sensor and a Measurement System, Models…Arrow Intl., Inc. · 2002 · 3 generations back · ◆ in the 2002 ancestors bar of FIG. 12 Class I recalls
  • K981660Bard 8 Fr., 40cc. & 7Fr., 30cc Intra-Aortic Balloon CathetersC.R. Bard, Inc. · 1998 · 3 generations back · ◆ in the 1998 ancestors bar of FIG. 12 Class I recalls
  • K000729Arrow Ultra 8 Intra-Aortic Balloon Catheters (Iabs) 8FR 30CC and 40CC UniversalArrow Intl., Inc. · 2000 · 4 generations back · ◆ in the 2000 ancestors bar of FIG. 17 Class I recalls
  • K9939668FR. Narrowflex Universal Intra-Aortic Balloon CatheterArrow Intl., Inc. · 2000 · 4 generations back · ◆ in the 2000 ancestors bar of FIG. 11 Class I recall
  • K965236System 96 Intra-Aortic Balloon Pump (Iabp)Datascope Corp. · 1997 · 4 generations back · ◆ in the 1997 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 23 years and passes through 4 companies; a new generation was cleared every 1.9 years on median.
  • 16 of the 66 ancestors have a recall of any class; 8 have a Class I recall.
  • 1 other device also cites its predicate K091449.

How does its category compare?

All DSP devices

DSP System, Balloon, Intra-Aortic And Control: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · DSP

1in 2025

2007: 22009: 22011: 22012: 22013: 12015: 12017: 32018: 12019: 32020: 22023: 12025: 1
2007high 3 (2017) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DSP

140in 2019

2017: 73 days median2019: 140 days median
2007high 140 (2019) · low 73 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DSP

4in 2025

2007: 32009: 142010: 82012: 32014: 12016: 9, 9 Class I2018: 2, 2 Class I2019: 32021: 5, 3 Class I2022: 33, 33 Class I2024: 12025: 4
2007high 33 (2022) · low 0 (2008)2026

Class I (47 since 2007) Class II/III not classified

Adverse event reports per year · DSP

7,422in 2025

2007: 4612008: 6172009: 6562010: 1,0192011: 5902012: 4902013: 4952014: 1,7382015: 4042016: 4192017: 2,1642018: 3,3312019: 3,5522020: 3,3932021: 4,3452022: 5,1132023: 6,9802024: 7,3172025: 7,422
2007high 7,422 (2025) · low 404 (2015)2025

MAUDE reports naming product code DSP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—3461
20080—0617
20092—14656
20100—81,019
20112—0590
20122—3490
20131—0495
20140—11,738
20151—0404
20160—9 (9)419
201737302,164
20181—2 (2)3,331
2019314033,552
20202—03,393
20210—5 (3)4,345
20220—33 (33)5,113
20231—06,980
20240—17,317
20251—47,422
20260—0—

What is it indicated for?

The MESA and SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheters and Accessories have the following Indications for use: * Refractory Unstable Angina. * Impending Infarction. * Acute Myocardial Infarction. * Refractory Ventricular Failure. * Complications of Acute MI (ie.Acute MR or VSD or papillary muscle rupture) * Cardiogenic Shock. * Support for diagnostic, percutaneous revascularization and Interventionall procedures. * Isohemia misated…

Read the full statement (116 words)

The MESA and SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheters and Accessories have the following Indications for use:

* Refractory Unstable Angina.

* Impending Infarction.

* Acute Myocardial Infarction.

* Refractory Ventricular Failure.

* Complications of Acute MI (ie.Acute MR or VSD or papillary muscle rupture)

* Cardiogenic Shock.

* Support for diagnostic, percutaneous revascularization and Interventionall procedures.

* Isohemia misated Intractable ventricular arrhythmlas.

* Septic Shock.

* Inireoperatie pulsatile flow generation.

* Weaning from cardiopulmonary bypass.

* Cardiac support for non-cardiac surgery.

* Prophylactic support in preparation for cardiac surgery.

* Post-surgical myocardial dysfunctionlow cardiac output syndrome.

* Myocardial Contusion.

* Mechanical bridge to other assist devices.

* Cardiac support following correction of anatomical defects

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K133074: Mega 8FR. 50CC and Sensation Plus 8FR. 50CC Intra-Aortic Balloon Catheters and Accessories.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K133074

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026