510(k) Report

510(k) · Cleared

Recalled · Class II · 2022

Nordicbrainex

K133910510(k) TraditionalCleared Apr 4, 2014102-day reviewClass II21 CFR 892.2050Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Nordicneurolab · LLZ System, Image Processing, Radiological

Cleared on Apr 4, 2014 as substantially equivalent to K090546 (Nordicimaginglab AS, 2009) after a 102-day review. Its family tree reaches back 8 generations to 1990. 5 later devices descend from it. It has been recalled 2 times.

Decided
Apr 4, 2014
Review
102 days
Class
Class II
Product code
LLZ
Recalls
2
Predicates
1

Its longest line of descent runs back 8 generations to K896095 (1990). 9 ancestors in all. 5 later devices descend from it, 1 citing it directly.

  1. Root · 8 generations backK896095 Images-on-Call Teleradiology SystemDevices & Services Co. · 1990
  2. 7 generations backK962699 Radworks Medical Imaging SoftwareApplicare Medical Imaging, B.V. · 1996
  3. 6 generations backK982862 Radworks Medical Imaging Software with Quality Control ModuleApplicare Medical Imaging, B.V. · 1998
  4. 5 generations backK012211 Efilm WorkstationEfilm Medical, Inc. · 2001 · Recalled
  5. 4 generations backK020995 Efilm Workstation with ModulesEfilm Medical, Inc. · 2002 · Recalled
  6. 3 generations backK063539 Nordic Image Control and Evaluation (Nice) Software, Version 2.1Nordicicemedical AS · 2007
  7. 2 generations backK082441 Nordicice SoftwareNordicimaginglab AS · 2008 · Recalled
  8. 1 generation backK090546 Modification to Nordicice SoftwareNordicimaginglab AS · 2009
  9. This deviceK133910 NordicbrainexNordicneurolab · 2014
  10. Descendant · cites it directly · 4 of its ownK171754 NovaPACSNovarad Corporation · 2017 · Recalled
  11. Descendant · 2 generations on · 1 of its ownK231662 Aid-UIaid · 2024
  12. Descendant · 2 generations on · 1 of its ownK221796 Cyware 4H and Cysart 4HAbys Medical · 2022
  13. Descendant · 3 generations on · 0 of its ownK250039 HPACSHealthhub · 2025
  14. Descendant · 3 generations on · 0 of its ownK243069 Lumys (V1)Intradys · 2025

Most influential descendants

Showing 5 of 5 · See all
  • K171754NovaPACSNovarad Corporation · 2017 · cites it directly4descendants
  • K231662Aid-UIaid · 2024 · 2 generations on1descendant
  • K221796Cyware 4H and Cysart 4HAbys Medical · 2022 · 2 generations on1descendant
  • K250039HPACSHealthhub · 2025 · 3 generations on0descendants
  • K243069Lumys (V1)Intradys · 2025 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K896095Images-on-Call Teleradiology SystemDevices & Services Co.1990—
−7K962699Radworks Medical Imaging SoftwareApplicare Medical Imaging, B.V.1996—
−6K982862Radworks Medical Imaging Software with Quality Control ModuleApplicare Medical Imaging, B.V.1998—
−5K012211Efilm WorkstationEfilm Medical, Inc.20011
−4K020995Efilm Workstation with ModulesEfilm Medical, Inc.20022
−3K063539Nordic Image Control and Evaluation (Nice) Software, Version 2.1Nordicicemedical AS2007—
−2K082441Nordicice SoftwareNordicimaginglab AS20082
−1K090546Modification to Nordicice SoftwareNordicimaginglab AS2009—
0 (this device)K133910NordicbrainexNordicneurolab2014—
+1K171754NovaPACSNovarad Corporation20171
+2K231662Aid-UIaid2024—
+2K221796Cyware 4H and Cysart 4HAbys Medical2022—
+3K250039HPACSHealthhub2025—
+3K243069Lumys (V1)Intradys2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201710
202210
202410
202520
FIG. 1 — Lineage of K133910, 1990–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in LLZ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new LLZ device

  1. #1 K230955 syngo Application Softwarescore 21.5 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K090546Modification to Nordicice Software
Summary page 1
…1 Panel: Radiology Product Code: LLZ Predicate device name: Nordic Image Control and Evaluation (nordicCE) Software K090546 Device Description The nordidBrainEx Software is a post-processing application for dynamic MRI data developed with focus on eas…
Nordicimaginglab AS2009—

Also cleared against K090546: 8 other devices →

What descends from it? (5)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K171754NovaPACSNovarad Corporation20174

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 16, 2022Class IIZ-1462-2022Software designWhen exporting merged results from the BOLD, DSC, and DCE modules of nordicBrainEx, where both the underlay and overlay originated from a multiframe dataset, the resulting output may have left/right sides flipped.
Sep 17, 2014Class IIZ-2050-2018Software designAn error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 24 years and passes through 6 companies; a new generation was cleared every 2.8 years on median.
  • 3 of the 9 ancestors have a recall of any class.
  • Of its descendants, K171754 (Novarad Corporation, 2017) has gone furthest, with 4 descendants of its own.
  • 8 other devices also cite its predicate K090546.

How does its category compare?

All LLZ devices

LLZ System, Image Processing, Radiological: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · LLZ

58in 2025

2007: 762008: 942009: 702010: 512011: 1062012: 822013: 952014: 782015: 642016: 822017: 782018: 1042019: 742020: 812021: 832022: 652023: 872024: 622025: 582026 (partial year): 45
2007high 106 (2011) · low 51 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · LLZ

173in 2025

2007: 36 days median2008: 52 days median2009: 62 days median2010: 98 days median2011: 79 days median2012: 87 days median2013: 84 days median2014: 89 days median2015: 90 days median2016: 90 days median2017: 107 days median2018: 105 days median2019: 107 days median2020: 115 days median2021: 116 days median2022: 137 days median2023: 157 days median2024: 131 days median2025: 173 days median2026 (partial year): 110 days median
2007high 173 (2025) · low 36 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · LLZ

11in 2025

2007: 202008: 322009: 212010: 172011: 302012: 122013: 642014: 62015: 162016: 122017: 52018: 72019: 22020: 62021: 32022: 42023: 82024: 62025: 112026 (partial year): 14
2007high 64 (2013) · low 2 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · LLZ

116in 2025

2007: 1522008: 1542009: 1452010: 1452011: 4542012: 2662013: 1,1302014: 15,6312015: 2112016: 2902017: 1042018: 722019: 882020: 822021: 702022: 282023: 262024: 422025: 116
2007high 15,631 (2014) · low 26 (2023)2025

MAUDE reports naming product code LLZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007763620152
2008945232154
2009706221145
2010519817145
20111067930454
2012828712266
20139584641,130
20147889615,631
2015649016211
2016829012290
2017781075104
2018104105772
201974107288
202081115682
202183116370
202265137428
202387157826
202462131642
20255817311116
20264511014—

What is it indicated for?

nordicBrainEx provides analysis and visualization capabilities of dynamic MRI data of the brain,

presenting the derived properties and parameters in a clinically useful context.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K133910: Nordicbrainex.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K133910

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026