510(k) · Cleared
Recalled · Class II · 2024Bard (R) Monopty (R) Disposable Core Biopsy Instrument, Bard (R) Max-Core (R) Disposable Core Instrument
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Bard Peripheral Vascular, Inc. · KNW Instrument, Biopsy
Cleared on Feb 21, 2014 as substantially equivalent to K922939 (C.R. Bard, Inc., 1993) after a 60-day review. Its family tree reaches back 1 generation to 1993. 27 later devices descend from it. It has been recalled 7 times.
- Decided
- Feb 21, 2014
- Review
- 60 days
- Class
- Class II
- Product code
- KNW
- Recalls
- 7
- Predicates
- 1
Its longest line of descent runs back 1 generation to K922939 (1993). 27 later devices descend from it, 13 citing it directly.
- Root · 1 generation backK922939 Bard Monopty Disposable Core Biopsy Isnt. W/ Cent.C.R. Bard, Inc. · 1993
- This deviceK133948 Bard (R) Monopty (R) Disposable Core Biopsy Instrument, Bard (R)…Bard Peripheral Vascular, Inc. · 2014
- Descendant · cites it directly · 11 of its ownK141552 Achieve Programmable Auomated Biopsy SystemsCare Fusion · 2014
- Descendant · cites it directly · 5 of its ownK171953 Bard Mission Disposable Core Biopsy InstrumentBard Peripheral Vascular, Inc. · 2017 · Recalled
- Descendant · 2 generations on · 4 of its ownK160316 SemiCut Semi-automatic Biopsy Needle, Themy Automatic Disposable…M.D.L. S.R.L. · 2016 · Recalled
- Descendant · 2 generations on · 2 of its ownK192101 Medone Ultra, MedextraMedax Srl Unipersonale · 2020
- Descendant · cites it directly · 2 of its ownK153337 Corvocet Biopsy SystemMerit Medical Systems, Inc. · 2016
Showing 5 of 27 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 27 · See all- K141552Achieve Programmable Auomated Biopsy SystemsCare Fusion · 2014 · cites it directly11descendants
- K171953Bard Mission Disposable Core Biopsy InstrumentBard Peripheral Vascular, Inc. · 2017 · cites it directly5descendants
- K160316SemiCut Semi-automatic Biopsy Needle, Themy Automatic Disposable Biopsy Device, Palium…M.D.L. S.R.L. · 2016 · 2 generations on4descendants
- K192101Medone Ultra, MedextraMedax Srl Unipersonale · 2020 · 2 generations on2descendants
- K153337Corvocet Biopsy SystemMerit Medical Systems, Inc. · 2016 · cites it directly2descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K922939 | Bard Monopty Disposable Core Biopsy Isnt. W/ Cent. | C.R. Bard, Inc. | 1993 | — |
| 0 (this device) | K133948 | Bard (R) Monopty (R) Disposable Core Biopsy Instrument, Bard (R) Max-Core (R)… | Bard Peripheral Vascular, Inc. | 2014 | — |
| +1 | K141552 | Achieve Programmable Auomated Biopsy Systems | Care Fusion | 2014 | — |
| +1 | K171953 | Bard Mission Disposable Core Biopsy Instrument | Bard Peripheral Vascular, Inc. | 2017 | 3 |
| +2 | K160316 | SemiCut Semi-automatic Biopsy Needle, Themy Automatic Disposable Biopsy… | M.D.L. S.R.L. | 2016 | 1 |
| +2 | K192101 | Medone Ultra, Medextra | Medax Srl Unipersonale | 2020 | — |
| +1 | K153337 | Corvocet Biopsy System | Merit Medical Systems, Inc. | 2016 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2014 | 1 | 0 |
| 2016 | 2 | 0 |
| 2017 | 2 | 0 |
| 2018 | 1 | 0 |
| 2020 | 4 | 0 |
| 2021 | 3 | 0 |
| 2022 | 6 | 0 |
| 2023 | 1 | 0 |
| 2024 | 2 | 0 |
| 2025 | 3 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in KNW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new KNW device
- #1 K260572 Marrow Marxmanscore 14.5 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K922939 | Bard Monopty Disposable Core Biopsy Isnt. W/ Cent.Summary page 1…21 CFR 676.1075 (Gastroenterology-urology biopsy instrument) 3. Predicate Device: The predicate device is the Bard® Monopty® Disposable Core Biopsy Instrument, K922939, cleared February 16, 1993. 4. Summary of Change: This Special 510(k) provides an updated file to FDA including several changes… | 1993 |
What descends from it? (27)
See all 2713 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K141552 | Achieve Programmable Auomated Biopsy Systems | 2014 | 11 |
| K171953 | Bard Mission Disposable Core Biopsy Instrument | 2017 | 5 |
| K153337 | Corvocet Biopsy System | 2016 | 2 |
| K222865 | M Biopsy /SureCore Automatic Disposable Biopsy Needle, M Biopsy /SureCore Semi-Automatic… | 2023 | 1 |
| K172344 | Medax Biopsy Systems II (MEDONE, MEDCUT, MEDEM, MEDBONE, MEDLOCK) | 2017 | 1 |
| K243296 | Disposable Biopsy Needle | 2025 | 0 |
| K250032 | Marquee™ Disposable Core Biopsy Instrument and Instrument Kit | 2025 | 0 |
| K241793 | Automatic Core Biopsy Instrument/ Semi-Automatic Core Biopsy Instrument/ Disposable… | 2025 | 0 |
| K220611 | SUREcore Plus Biopsy Instrument | 2022 | 0 |
| K214063 | Model 121 Biopsy Needle, Model 111 Biopsy Needle, Model 112 Biopsy Needle, and Model 131… | 2022 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Sep 30, 2024 | Class II | Z-0467-2025 | Process change control | Disposable biopsy instruments were potentially manufactured missing the backend notch retention feature on the needle that results in inadequate securing of the needle inside the device, which could lead to needle detachment and… |
| Aug 10, 2023 | Class II | Z-0027-2024 | Device Design | Diameter mismatch between coaxial and biopsy needle (cutting cannula) in disposable core biopsy instrument kits, where the coaxial cannula inner diameter prohibits the biopsy needle (cutting cannula) from properly fitting into the coaxial… |
| Jan 14, 2021 | Class II | Z-1289-2021 | Process control | Due to the Penetration Depth Switch used to select the desired penetration depth of the biopsy needle for soft tissue biopsies, detaches from the device during use. |
| Oct 25, 2019 | Class II | Z-0724-2020 | Equipment maintenance | During the firm's review of May 2019 complaint data, a significant increase in Disposable Core Biopsy Instrument complaints were observed for Failure to Fire, Firing Problem, and Failure to Obtain Samples. |
| Sep 22, 2017 | Class II | Z-0044-2018 | Device Design | During the review of May 2017 complaint data, a significant increase in MAX-CORE Disposable Core Biopsy Instrument complaints was observed for Failure to Prime, Failure to Fire, Failure to Obtain Sample, and Self-Activation. |
| Apr 9, 2015 | Class II | Z-1586-2015 | Process design | The firm is recalling Bard MAX-CORE Disposable Core Biopsy Instrument due to the risk of having self-activation-related issues. |
| Mar 17, 2015 | Class II | Z-1447-2015 | Nonconforming Material/Component | Bard Peripheral Vascular is recalling the Bard Monopty Disposable Core Biopsy Instruments and Kits because they may be at risk of having activation-related issues. |
The 5 latest of 7 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All KNW devicesKNW Instrument, Biopsy: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.
FDA decisions per year · KNW
12in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · KNW
178in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · KNW
1in 2024
Class I (4 since 2007) Class II/III not classified
Adverse event reports per year · KNW
1,667in 2025
MAUDE reports naming product code KNW, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 6 | 94 | 1 | 766 |
| 2008 | 4 | 53 | 7 | 363 |
| 2009 | 6 | 35 | 3 | 351 |
| 2010 | 3 | 57 | 2 | 162 |
| 2011 | 10 | 64 | 5 | 67 |
| 2012 | 7 | 94 | 1 | 93 |
| 2013 | 3 | 100 | 1 (1) | 98 |
| 2014 | 8 | 95 | 3 | 193 |
| 2015 | 10 | 106 | 0 | 591 |
| 2016 | 9 | 127 | 5 | 297 |
| 2017 | 8 | 90 | 1 | 760 |
| 2018 | 7 | 71 | 0 | 1,012 |
| 2019 | 5 | 94 | 4 | 4,195 |
| 2020 | 8 | 92 | 0 | 5,120 |
| 2021 | 9 | 54 | 3 | 1,020 |
| 2022 | 8 | 203 | 3 | 535 |
| 2023 | 7 | 164 | 2 | 931 |
| 2024 | 5 | 114 | 1 | 1,283 |
| 2025 | 12 | 178 | 0 | 1,667 |
| 2026 | 7 | 29 | 4 (3) | — |
What is it indicated for?
The core needle biopsy device is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors. It is not intended for use in bone.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K133948 · printed Nov 6, 2026
Cite this page
510(k) Report. “K133948: Bard (R) Monopty (R) Disposable Core Biopsy Instrument, Bard (R) Max-Core (R) Disposable Core Instrument.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K133948
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026