510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

Bard Mission Disposable Core Biopsy Instrument

K171953510(k) TraditionalCleared Sep 14, 201777-day reviewClass II21 CFR 876.1075Recalls 3

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bard Peripheral Vascular, Inc. · KNW Instrument, Biopsy

Cleared on Sep 14, 2017 as substantially equivalent to K161409 (Mermaid Medical A/S, 2016) and 1 other predicate after a 77-day review. Its family tree reaches back 3 generations to 1995. 5 later devices descend from it. It has been recalled 3 times.

Decided
Sep 14, 2017
Review
77 days
Class
Class II
Product code
KNW
Recalls
3
Predicates
2

Its longest line of descent runs back 3 generations to K950732 (1995). 5 ancestors in all, the oldest K922939 (1993). 5 later devices descend from it, 4 citing it directly.

Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor 1 generation forward: 4 descendants2 generations forward: 1 descendant K161409 · M-Biopsy Semi-Automatic Biopsy Instrument M-Biopsy Coaxial Introducer Needle for Sem · Mermaid Medical A/S · 2016K974814 · Manan (TM) Super-Core Biopsy Needles · Medical Device Technologies, Inc. · 1998 · recalledK950732 · Manan Super-Core Biopsy Needle · Manan Medical Products, Inc. · 19950
  1. Root · 3 generations backK950732 Manan Super-Core Biopsy NeedleManan Medical Products, Inc. · 1995
  2. 2 generations backK974814 Manan (TM) Super-Core Biopsy NeedlesMedical Device Technologies, Inc. · 1998 · Recalled
  3. 1 generation backK161409 M-Biopsy Semi-Automatic Biopsy Instrument M-Biopsy Coaxial…Mermaid Medical A/S · 2016
  4. This deviceK171953 Bard Mission Disposable Core Biopsy InstrumentBard Peripheral Vascular, Inc. · 2017
  5. Descendant · cites it directly · 1 of its ownK222865 M Biopsy /SureCore Automatic Disposable Biopsy Needle, M Biopsy…Canyon Medical, Inc. · 2023
  6. Descendant · cites it directly · 0 of its ownK250032 Marquee™ Disposable Core Biopsy Instrument and Instrument KitBard Peripheral Vascular, Inc. · 2025
  7. Descendant · cites it directly · 0 of its ownK241793 Automatic Core Biopsy Instrument/ Semi-Automatic Core Biopsy…Zhejiang Curaway Medical Technology Co., Ltd. · 2025
  8. Descendant · 2 generations on · 0 of its ownK233031 M·Biopsy /SureCore Automatic Disposable Biopsy Needle, M·Biopsy…Canyon Medical, Inc. · 2024
  9. Descendant · cites it directly · 0 of its ownK201166 Temno Elite Biopsy SystemMerit Medical Systems, Inc. · 2020

Most influential descendants

Showing 5 of 5 · See all
  • K222865M Biopsy /SureCore Automatic Disposable Biopsy Needle, M Biopsy /SureCore Semi-Automatic…Canyon Medical, Inc. · 2023 · cites it directly1descendant
  • K250032Marquee™ Disposable Core Biopsy Instrument and Instrument KitBard Peripheral Vascular, Inc. · 2025 · cites it directly0descendants
  • K241793Automatic Core Biopsy Instrument/ Semi-Automatic Core Biopsy Instrument/ Disposable…Zhejiang Curaway Medical Technology Co., Ltd. · 2025 · cites it directly0descendants
  • K233031M·Biopsy /SureCore Automatic Disposable Biopsy Needle, M·Biopsy /SureCore Semi-automatic…Canyon Medical, Inc. · 2024 · 2 generations on0descendants
  • K201166Temno Elite Biopsy SystemMerit Medical Systems, Inc. · 2020 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K950732Manan Super-Core Biopsy NeedleManan Medical Products, Inc.1995—
−2K974814Manan (TM) Super-Core Biopsy NeedlesMedical Device Technologies, Inc.19981
−1K161409M-Biopsy Semi-Automatic Biopsy Instrument M-Biopsy Coaxial Introducer Needle…Mermaid Medical A/S2016—
0 (this device)K171953Bard Mission Disposable Core Biopsy InstrumentBard Peripheral Vascular, Inc.2017—
+1K222865M Biopsy /SureCore Automatic Disposable Biopsy Needle, M Biopsy /SureCore…Canyon Medical, Inc.2023—
+1K250032Marquee™ Disposable Core Biopsy Instrument and Instrument KitBard Peripheral Vascular, Inc.2025—
+1K241793Automatic Core Biopsy Instrument/ Semi-Automatic Core Biopsy Instrument/…Zhejiang Curaway Medical Technology Co., Ltd.2025—
+2K233031M·Biopsy /SureCore Automatic Disposable Biopsy Needle, M·Biopsy /SureCore…Canyon Medical, Inc.2024—
+1K201166Temno Elite Biopsy SystemMerit Medical Systems, Inc.2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199310
201410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202010
202310
202410
202520
FIG. 1 — Lineage of K171953, 1995–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KNW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KNW device

  1. #1 K260572 Marrow Marxmanscore 14.5 · 10 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K161409M-Biopsy Semi-Automatic Biopsy Instrument M-Biopsy Coaxial Introducer Needle for Sem
Summary page 4
…stroenterology / Urology Regulation Number: 21 CFR 876.1075 Predicate Devices: x M-Biopsy™ Semi-Automatic Biopsy Instrument by Mermaid Medical (K161409; cleared 2 August 2016) x BARD® MONOPTY® and BARD® MAX-CORE® Disposable Core Biopsy Instruments (K133948/S001; cleared…
Mermaid Medical A/S2016—
*K133948Bard (R) Monopty (R) Disposable Core Biopsy Instrument, Bard (R) Max-Core (R) Disposable…
Summary page 4
…Automatic Biopsy Instrument by Mermaid Medical (K161409; cleared 2 August 2016) x BARD® MONOPTY® and BARD® MAX-CORE® Disposable Core Biopsy Instruments (K133948/S001; cleared 21 February 2014) Reference Devices: y POWERGLIDE® Midline Catheter (K121073; cleared 1 June 2012) y ULTRAC…
Bard Peripheral Vascular, Inc.20147
†K121073Powerglide Midline CatheterC.R. Bard, Inc.20121
†K042341Ultraclip II USInrad2004—
†K051158Encor 7 GaugeSenorx, Inc.200510
†K131493Sidekick and Usher Support CathetersC.R. Bard, Inc.2013—

What descends from it? (5)

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K222865M Biopsy /SureCore Automatic Disposable Biopsy Needle, M Biopsy /SureCore Semi-Automatic…Canyon Medical, Inc.20231
K250032Marquee™ Disposable Core Biopsy Instrument and Instrument KitBard Peripheral Vascular, Inc.20250
K241793Automatic Core Biopsy Instrument/ Semi-Automatic Core Biopsy Instrument/ Disposable…Zhejiang Curaway Medical Technology Co., Ltd.20250
K201166Temno Elite Biopsy SystemMerit Medical Systems, Inc.20200

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 8, 2023Class IIZ-0643-2024Process controlThe internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy needle may not fit…
Apr 11, 2023Class IIZ-1564-2023Mixed-up of materials/componentsCoaxial cannula internal diameter and length of biopsy instruments exceed the external diameter and length of the biopsy needle, prohibiting the biopsy needle from properly fitting into the coaxial cannula and accessing the target tissue…
Feb 28, 2023Class IIZ-1260-2023Mixed-up of materials/componentsThe external diameter of the biopsy instrument is larger than the internal diameter of the coaxial needle, which could lead to: 1) Procedural delay, 2) Requirement of an additional device, 3), Needle tip breakage, if used despite…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 22 years and passes through 4 companies; a new generation was cleared every 2.9 years on median.
  • 2 of the 5 ancestors have a recall of any class.
  • Of its descendants, K222865 (Canyon Medical, Inc., 2023) has gone furthest, with 1 descendant of its own.

How does its category compare?

All KNW devices

KNW Instrument, Biopsy: FDA decisions, review times, recalls and adverse event reports over 20 years. 2017, this device’s year, is highlighted.

FDA decisions per year · KNW

12in 2025

2007: 62008: 42009: 62010: 32011: 102012: 72013: 32014: 82015: 102016: 92017: 82018: 72019: 52020: 82021: 92022: 82023: 72024: 52025: 122026 (partial year): 7
2007high 12 (2025) · low 3 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KNW

178in 2025

2007: 94 days median2008: 53 days median2009: 35 days median2010: 57 days median2011: 64 days median2012: 94 days median2013: 100 days median2014: 95 days median2015: 106 days median2016: 127 days median2017: 90 days median2018: 71 days median2019: 94 days median2020: 92 days median2021: 54 days median2022: 203 days median2023: 164 days median2024: 114 days median2025: 178 days median2026 (partial year): 29 days median
2007high 203 (2022) · low 35 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KNW

1in 2024

2007: 12008: 72009: 32010: 22011: 52012: 12013: 1, 1 Class I2014: 32016: 52017: 12019: 42021: 32022: 32023: 22024: 12026 (partial year): 4, 3 Class I
2007high 7 (2008) · low 0 (2015)2026

Class I (4 since 2007) Class II/III not classified

Adverse event reports per year · KNW

1,667in 2025

2007: 7662008: 3632009: 3512010: 1622011: 672012: 932013: 982014: 1932015: 5912016: 2972017: 7602018: 1,0122019: 4,1952020: 5,1202021: 1,0202022: 5352023: 9312024: 1,2832025: 1,667
2007high 5,120 (2020) · low 67 (2011)2025

MAUDE reports naming product code KNW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20076941766
20084537363
20096353351
20103572162
20111064567
2012794193
201331001 (1)98
20148953193
2015101060591
201691275297
20178901760
201877101,012
201959444,195
202089205,120
202195431,020
202282033535
202371642931
2024511411,283
20251217801,667
20267294 (3)—

What is it indicated for?

The Bard® Mission® Disposable Core Biopsy Instrument is intended for use in obtaining biopsy samples from soft tissues such as lung, liver, spleen, kidney, prostate, lymph nodes, breast, thyroid, and various soft tissue tumors.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K171953: Bard Mission Disposable Core Biopsy Instrument.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K171953

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026