510(k) Report

510(k) · Cleared

Recalled · Class II · 2015

Arthroscopic Energy System

K140578510(k) TraditionalCleared Jul 23, 2014139-day reviewClass II21 CFR 878.4400Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Conmed Corporation · GEI Electrosurgical, Cutting & Coagulation & Accessories

Cleared on Jul 23, 2014 as substantially equivalent to K082666 (Arthrocare Corp., 2008) and 1 other predicate after a 139-day review. Its family tree reaches back 7 generations to 2001. 1 later device descends from it. It has been recalled once.

Decided
Jul 23, 2014
Review
139 days
Class
Class II
Product code
GEI
Recalls
1
Predicates
2

Its longest line of descent runs back 7 generations to K011083 (2001). 13 ancestors in all, the oldest K905654 (1991). 1 later device descends from it, 1 citing it directly.

  1. Root · 7 generations backK011083 Arthrocare System 2000 Controller, Cable, Footswitch, Power Cord &…Arthrocare Corp. · 2001
  2. 6 generations backK013463 Arthrocare Arthro Wands, Enetec Evac Plasma WandArthrocare Corp. · 2001
  3. 5 generations backK020557 Arthrocare Controller; Arthrocare Cable; Foot Control; Power Cord…Arthrocare Corp. · 2002
  4. 4 generations backK030551 Arthrocare Controller (System 2000 and 8000); Arthrocare Patient…Arthrocare Corp. · 2003 · Recalled
  5. 3 generations backK032504 Arthrocare SystemArthrocare Corp. · 2003
  6. 2 generations backK071709 Arthrocare System 12000Arthrocare Corp. · 2007
  7. 1 generation backK082666 Arthrocare Controller (System 12000), Arthrocare Patient Cable, Foot…Arthrocare Corp. · 2008
  8. This deviceK140578 Arthroscopic Energy SystemConmed Corporation · 2014
  9. Descendant · cites it directly · 0 of its ownK153499 EDGE Arthroscopic Energy System, Arthroscopic Energy System Probes…Conmed Corporation · 2015

Most influential descendants

Showing 1 of 1 · See all
  • K153499EDGE Arthroscopic Energy System, Arthroscopic Energy System Probes, Arthroscopic Energy…Conmed Corporation · 2015 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−7K011083Arthrocare System 2000 Controller, Cable, Footswitch, Power Cord & Arthrocare…Arthrocare Corp.2001—
−6K013463Arthrocare Arthro Wands, Enetec Evac Plasma WandArthrocare Corp.2001—
−5K020557Arthrocare Controller; Arthrocare Cable; Foot Control; Power Cord; WandsArthrocare Corp.2002—
−4K030551Arthrocare Controller (System 2000 and 8000); Arthrocare Patient Cable; Foot…Arthrocare Corp.20031
−3K032504Arthrocare SystemArthrocare Corp.2003—
−2K071709Arthrocare System 12000Arthrocare Corp.2007—
−1K082666Arthrocare Controller (System 12000), Arthrocare Patient Cable, Foot Control…Arthrocare Corp.2008—
0 (this device)K140578Arthroscopic Energy SystemConmed Corporation2014—
+1K153499EDGE Arthroscopic Energy System, Arthroscopic Energy System Probes…Conmed Corporation2015—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199110
199510
200110
200210
200510
201110

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201510
FIG. 1 — Lineage of K140578, 2001–2015Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in GEI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new GEI device

  1. #1 K231790 The InMode System with the Morpheus8 Applicatorsscore 32.7 · 21 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K082666Arthrocare Controller (System 12000), Arthrocare Patient Cable, Foot Control, Power…
Summary page 1
…edicate/Legally Marketed Devices Device Name: ArthroCareg® System 12000 Company Name: ArthroCare Corporation 5 10O(k) #: K082666 Device Name: ConMed Sabre Genesis Electrosurgical Generator Company Name: ConMed Electrosurgery…
Arthrocare Corp.2008—
*K103665Conmed R Sabre Genesis TM Electrosurgical Unit; Coned R Sabre Genesis TM Electrosurgical…
Summary page 1
…K082666 Device Name: ConMed Sabre Genesis Electrosurgical Generator Company Name: ConMed Electrosurgery 5 10(k)# K103665 Page 1 of 3 - CoNMED…
Conmed Corporation2011—

Also cleared against K082666: 4 other devices →

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K153499EDGE Arthroscopic Energy System, Arthroscopic Energy System Probes, Arthroscopic Energy…Conmed Corporation20150

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jun 17, 2015Class IIZ-0205-2016Software design (manufacturing process)Probe programming error. The incorrect program can generate more heat at the surgical site, resulting in an elevated probability of tissue damage.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 2 companies; a new generation was cleared every 12 months on median.
  • 3 of the 13 ancestors have a recall of any class.
  • Of its descendants, K153499 (Conmed Corporation, 2015) has gone furthest, with 0 descendants of its own.
  • 4 other devices also cite its predicate K082666.

How does its category compare?

All GEI devices

GEI Electrosurgical, Cutting & Coagulation & Accessories: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · GEI

52in 2025

2007: 502008: 612009: 422010: 462011: 352012: 392013: 482014: 492015: 472016: 642017: 522018: 472019: 482020: 592021: 462022: 552023: 652024: 592025: 522026 (partial year): 43
2007high 65 (2023) · low 35 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · GEI

113in 2025

2007: 51 days median2008: 71 days median2009: 130 days median2010: 109 days median2011: 135 days median2012: 133 days median2013: 136 days median2014: 103 days median2015: 123 days median2016: 108 days median2017: 72 days median2018: 93 days median2019: 93 days median2020: 105 days median2021: 146 days median2022: 150 days median2023: 104 days median2024: 125 days median2025: 113 days median2026 (partial year): 183 days median
2007high 183 (2026) · low 51 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · GEI

4in 2025

2007: 212008: 192009: 272010: 512011: 322012: 302013: 132014: 32015: 42016: 32017: 22018: 62019: 52020: 12021: 62022: 492023: 27, 4 Class I2024: 22, 10 Class I2025: 42026 (partial year): 8
2007high 51 (2010) · low 1 (2020)2026

Class I (14 since 2007) Class II/III not classified

Adverse event reports per year · GEI

7,592in 2025

2007: 2,7952008: 2,5562009: 3,5912010: 3,3772011: 4,4632012: 5,2162013: 4,8972014: 4,9082015: 4,9362016: 5,6162017: 6,0732018: 6,5452019: 7,0672020: 6,4902021: 6,3542022: 6,1032023: 7,1762024: 7,8972025: 7,592
2007high 7,897 (2024) · low 2,556 (2008)2025

MAUDE reports naming product code GEI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20075051212,795
20086171192,556
200942130273,591
201046109513,377
201135135324,463
201239133305,216
201348136134,897
20144910334,908
20154712344,936
20166410835,616
2017527226,073
2018479366,545
2019489357,067
20205910516,490
20214614666,354
202255150496,103
20236510427 (4)7,176
20245912522 (10)7,897
20255211347,592
2026431838—

What is it indicated for?

The Arthroscopic Energy™ System is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic surgical procedures.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K140578: Arthroscopic Energy System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K140578

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026