View Plating System,Reduce Fracture Plating System,Bone Screw

K140876510(k) TraditionalCleared Jul 22, 2014106-day reviewClass II21 CFR 888.3030Recalls 15

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Icon Orthopedic Concepts, LLC Dba Edge Orthopaedic · HRS Plate, Fixation, Bone

Cleared on Jul 22, 2014 as substantially equivalent to K061808 (Darco International, Inc., 2006) and 1 other predicate after a 106-day review. Its family tree reaches back 2 generations to 2002. 2 later devices descend from it. It has been recalled 15 times.

Decided
Jul 22, 2014
Review
106 days
Class
Class II
Product code
HRS
Recalls
15
Predicates
2

Its longest line of descent runs back 2 generations to K022325 (2002). 3 ancestors in all. 2 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2009: 1 other ancestor2019: 1 descendant cleared2014: 1 descendant clearedK022325 · Normed Titanium Osteotomy Plating System · Osteomedics, Inc. · 2002 · recalledK061808 · Darco Locking Bone Plate System · Darco International, Inc. · 2006 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 2 descendants K061808 · Darco Locking Bone Plate System · Darco International, Inc. · 2006 · recalledK022325 · Normed Titanium Osteotomy Plating System · Osteomedics, Inc. · 2002 · recalled0
  1. Root · 2 generations backK022325 Normed Titanium Osteotomy Plating SystemOsteomedics, Inc. · 2002 · Recalled
  2. 1 generation backK061808 Darco Locking Bone Plate SystemDarco International, Inc. · 2006 · Recalled
  3. This deviceK140876 View Plating System,Reduce Fracture Plating System,Bone ScrewIcon Orthopedic Concepts, LLC Dba Edge Orthopaedic · 2014
  4. Descendant · cites it directly · 0 of its ownK190388 RIVAL View Plating Systems and Reduce Fracture Plating SystemsOrthofix Srl · 2019
  5. Descendant · cites it directly · 0 of its ownK142135 Reduce Fracture Plating System Line ExtensionIcon Orthopedic Concepts, LLC Dba Edge Orthopaedic · 2014 · Recalled

Most influential descendants

Showing 2 of 2 · See all
  • K190388RIVAL View Plating Systems and Reduce Fracture Plating SystemsOrthofix Srl · 2019 · cites it directly0descendants
  • K142135Reduce Fracture Plating System Line ExtensionIcon Orthopedic Concepts, LLC Dba Edge Orthopaedic · 2014 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K022325Normed Titanium Osteotomy Plating SystemOsteomedics, Inc.20021
−1K061808Darco Locking Bone Plate SystemDarco International, Inc.20063
0 (this device)K140876View Plating System,Reduce Fracture Plating System,Bone ScrewIcon Orthopedic Concepts, LLC Dba Edge Orthopaedic2014—
+1K190388RIVAL View Plating Systems and Reduce Fracture Plating SystemsOrthofix Srl2019—
+1K142135Reduce Fracture Plating System Line ExtensionIcon Orthopedic Concepts, LLC Dba Edge Orthopaedic20145

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201410
201910
FIG. 1 — Lineage of K140876, 2002–2019Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in HRS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new HRS device

  1. #1 K222465 Tyber Medical Anatomical Plating Systemscore 27.5 · 16 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K061808Darco Locking Bone Plate System
Summary page 1
…Product Code: HRS/HWC Panel Code: Orthopedics Predicate Device [21 CFR 807.92(a)(3)] DARCO Locking Plating System - K061808 (Darco International) ORTHOLOC® 2.0/2.4 Plate System - K090692 (Wright Medical Technology) Descriptio…
Darco International, Inc.20063
*K090692Ortholoc 2.0/2.4 Plate System
Summary page 1
…1 CFR 807.92(a)(3)] DARCO Locking Plating System - K061808 (Darco International) ORTHOLOC® 2.0/2.4 Plate System - K090692 (Wright Medical Technology) Description of the Device [21 CFR 807.92(a)(4)] The EDGE Orthopaedics' VIEWTM Plating System has bee…
Wrightmedicaltechnologyinc20091

Also cleared against K061808: 23 other devices →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K190388RIVAL View Plating Systems and Reduce Fracture Plating SystemsOrthofix Srl20190
K142135Reduce Fracture Plating System Line ExtensionIcon Orthopedic Concepts, LLC Dba Edge Orthopaedic20140

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 6, 2018Class IIZ-2130-2018Device DesignFirm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Apr 6, 2018Class IIZ-2126-2018Device DesignFirm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Apr 6, 2018Class IIZ-2124-2018Device DesignFirm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Apr 6, 2018Class IIZ-2132-2018Device DesignFirm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Apr 6, 2018Class IIZ-2122-2018Device DesignFirm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Apr 6, 2018Class IIZ-2131-2018Device DesignFirm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Apr 6, 2018Class IIZ-2123-2018Device DesignFirm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Apr 6, 2018Class IIZ-2119-2018Device DesignFirm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Apr 6, 2018Class IIZ-2120-2018Device DesignFirm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Apr 6, 2018Class IIZ-2127-2018Device DesignFirm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Apr 6, 2018Class IIZ-2121-2018Device DesignFirm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Apr 6, 2018Class IIZ-2133-2018Device DesignFirm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Apr 6, 2018Class IIZ-2135-2018Device DesignFirm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Apr 6, 2018Class IIZ-2134-2018Device DesignFirm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Apr 6, 2018Class IIZ-2129-2018Device DesignFirm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 12 years and passes through 3 companies; a new generation was cleared every 7.9 years on median.
  • 3 of the 3 ancestors have a recall of any class.
  • Of its descendants, K190388 (Orthofix Srl, 2019) has gone furthest, with 0 descendants of its own.
  • 23 other devices also cite its predicate K061808.

How does its category compare?

All HRS devices

HRS Plate, Fixation, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · HRS

54in 2025

2007: 352008: 352009: 472010: 492011: 432012: 592013: 562014: 762015: 562016: 442017: 572018: 402019: 462020: 482021: 402022: 472023: 402024: 542025: 542026 (partial year): 51
2007high 76 (2014) · low 35 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · HRS

90in 2025

2007: 63 days median2008: 87 days median2009: 85 days median2010: 100 days median2011: 88 days median2012: 106 days median2013: 89 days median2014: 57 days median2015: 97 days median2016: 109 days median2017: 122 days median2018: 116 days median2019: 89 days median2020: 88 days median2021: 87 days median2022: 144 days median2023: 120 days median2024: 96 days median2025: 90 days median2026 (partial year): 82 days median
2007high 144 (2022) · low 57 (2014)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · HRS

1in 2025

2007: 22008: 112009: 32010: 162011: 52012: 42013: 112015: 72016: 52017: 32018: 22019: 252020: 12021: 92022: 32023: 12025: 12026 (partial year): 3
2007high 25 (2019) · low 0 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · HRS

2,306in 2025

2007: 3392008: 3112009: 4352010: 4912011: 7732012: 1,6632013: 2,6432014: 2,4132015: 3,0952016: 2,2762017: 2,2202018: 2,6632019: 3,3592020: 2,7902021: 2,4422022: 1,7002023: 2,0492024: 1,8492025: 2,306
2007high 3,359 (2019) · low 311 (2008)2025

MAUDE reports naming product code HRS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735632339
2008358711311
200947853435
20104910016491
201143885773
20125910641,663
20135689112,643
2014765702,413
2015569773,095
20164410952,276
20175712232,220
20184011622,663
20194689253,359
2020488812,790
2021408792,442
20224714431,700
20234012012,049
2024549601,849
2025549012,306
202651823—

What is it indicated for?

EDGE Orthopaedics' VIEW and REDUCE Plating Systems are intended for use in stabilization and fixation of fresh fractures, revision procedures, joint fusion, and reconstruction of small bones of the hand, feet, wrist, ankles, fingers and toes. The plates are available for use with EDGE Orthopaedics locking and non-locking bone screws.

Plates and screws are intended for single use only. Screws are not intended for use in the spine.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K140876: View Plating System,Reduce Fracture Plating System,Bone Screw.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K140876

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026