510(k) Report

510(k) · Cleared

Cleared

Kimguard Smart-Fold Sterilization Wrap

K140963510(k) TraditionalCleared Nov 10, 2014209-day reviewClass II21 CFR 880.6850

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Kimberly-Clark Corp. · FRG Wrap, Sterilization

Cleared on Nov 10, 2014 as substantially equivalent to K112300 (Kimberly-Clark Corp., 2012) and 1 other predicate after a 209-day review. Its family tree reaches back 4 generations to 1988.

Decided
Nov 10, 2014
Review
209 days
Class
Class II
Product code
FRG
Recalls
None
Predicates
2

Its longest line of descent runs back 4 generations to K881471 (1988). 5 ancestors in all. No later device cites it as a predicate yet.

  1. Root · 4 generations backK881471 Kimguard (Regular & Heavy Duty) Sterile WrapKimberly-Clark Corp. · 1988 · Recalled
  2. 3 generations backK082177 Kimguard One-Step Sterilization WrapKimberly-Clark Corp. · 2009
  3. 2 generations backK092167 Kimguard One-Step Sterilization Wrap, Models KC100, KC200, KC300…Kimberly-Clark Corp. · 2009
  4. 1 generation backK113806 Kimguard One Step Sterilization WrapKimberly-Clark Health Care · 2013
  5. This deviceK140963 Kimguard Smart-Fold Sterilization WrapKimberly-Clark Corp. · 2014
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K881471Kimguard (Regular & Heavy Duty) Sterile WrapKimberly-Clark Corp.19881
−3K082177Kimguard One-Step Sterilization WrapKimberly-Clark Corp.2009—
−2K092167Kimguard One-Step Sterilization Wrap, Models KC100, KC200, KC300, KC400…Kimberly-Clark Corp.2009—
−1K113806Kimguard One Step Sterilization WrapKimberly-Clark Health Care2013—
0 (this device)K140963Kimguard Smart-Fold Sterilization WrapKimberly-Clark Corp.2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
201210
FIG. 1 — Lineage of K140963, 1988–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FRG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRG device

  1. #1 K234050 HALYARD® ONE-STEP® Sterilization Wrap…score 15.8 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K112300Kimguard Smart-Fold* Sterilization Wrap
Summary page 6
…KimGuard* Smart-Fold* Sterilization Wrap (Models KC250, KC450, KC550, and KC650) cleared January 19, 2013 (K112300). KimGuard* ONE-STEP Sterilization Wrap (Models KC100, KC200, KC30…
Kimberly-Clark Corp.2012—
*K113806Kimguard One Step Sterilization Wrap
Summary page 6
…* ONE-STEP Sterilization Wrap (Models KC100, KC200, KC300, KC400, KC500, and KC600 cleared August 20, 2013 (K113806) INDICATIONS FOR USE KIMGUARD* Smart Fold* Sterilization Wrap (KC450 and KC650) is intended to be used to enclose another…
Kimberly-Clark Health Care2013—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 26 years and passes through 2 companies; a new generation was cleared every 3.8 years on median.
  • 1 of the 5 ancestors has a recall of any class.

How does its category compare?

All FRG devices

FRG Wrap, Sterilization: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · FRG

9in 2025

2007: 32008: 22009: 42010: 42011: 62012: 102013: 62014: 82015: 122016: 92017: 92018: 52019: 92020: 52021: 42022: 92023: 92024: 112025: 92026 (partial year): 4
2007high 12 (2015) · low 2 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRG

151in 2025

2007: 182 days median2009: 171 days median2010: 213 days median2011: 110 days median2012: 133 days median2013: 146 days median2014: 177 days median2015: 196 days median2016: 131 days median2017: 260 days median2018: 150 days median2019: 258 days median2020: 122 days median2021: 170 days median2022: 217 days median2023: 230 days median2024: 182 days median2025: 151 days median2026 (partial year): 247 days median
2007high 260 (2017) · low 110 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRG

2in 2022

2013: 22017: 292020: 142022: 2
2007high 29 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FRG

18in 2025

2007: 12008: 22009: 22011: 122012: 42013: 52014: 132015: 92016: 202017: 72018: 222019: 102020: 212021: 92022: 42023: 132024: 212025: 18
2007high 22 (2018) · low 0 (2010)2025

MAUDE reports naming product code FRG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007318201
20082—02
2009417102
2010421300
20116110012
20121013304
2013614625
20148177013
20151219609
20169131020
20179260297
20185150022
20199258010
202051221421
2021417009
2022921724
20239230013
202411182021
20259151018
202642470—

What is it indicated for?

KIMGUARD* Smart Fold* Sterilization Wrap (KC450 and KC650) is intended to be used to enclose another medical device that is to be sterilized by a healthcare provider using: Advanced Sterilization Products’ STERRAD® Sterilization Systems that include: STERRAD® 100S STERRAD® NX® [Standard Cycle, Advanced Cycle] STERRAD® 100NX®). [Standard Cycle, Flex Cycle, EXPRESS cycle, DUO Cycle] KIMGUARD* Smart Fold* Sterilization Wraps (KC450 and KC650) are intended…

Read the full statement (838 words)

KIMGUARD* Smart Fold* Sterilization Wrap (KC450 and KC650) is intended to be used to enclose another medical device that is to be sterilized by a healthcare provider using:

Advanced Sterilization Products’ STERRAD® Sterilization Systems that include:

STERRAD® 100S

STERRAD® NX® [Standard Cycle, Advanced Cycle]

STERRAD® 100NX®). [Standard Cycle, Flex Cycle, EXPRESS cycle, DUO Cycle]

KIMGUARD* Smart Fold* Sterilization Wraps (KC450 and KC650) are intended to allow sterilization of the enclosed medical device(s) and also maintain sterility of the enclosed device(s) until used.

Test results validated that KIMGUARD* Smart Fold* Sterilization Wrap (KC450 and KC650) allowed sterilization of the enclosed devices by the Advanced Sterilization Products STERRAD® Sterilization Systems (STERRAD® 100S, NX® [Standard Cycle and Advanced Cycle], and 100NX®. [Standard Cycle, Flex Cycle, EXPRESS cycle, DUO Cycle]). (Refer to the STERRAD® Sterilization System User’s Guide for complete instructions on load(s) and cycle(s), including chamber loading instructions.)

The KC450 and KC650 models of the KIMGUARD* Smart Fold* Sterilization Wrap have been validated for use with the STERRAD® 100S, STERRAD® NX®, and STERRAD® 100 NX® cycles in Table 1.

TABLE 1: Validated Advanced Sterilization Products (ASP) STERRAD® 100S, STERRAD® NX®, and STERRAD® 100NX® Cycles

ASP STERRAD® System and Cycle — Intended Load

STERRAD® 100S: Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load:

An inside diameter of 1 mm or larger and a length of 125 mm or shorter of single-channel stainless steel lumens.

An inside diameter of 2 mm or larger and a length of 250 mm or shorter of single-channel stainless steel lumens.

An inside diameter of 3 mm or larger and a length of 400 mm or shorter of single-channel stainless steel lumens.

An inside diameter of 6 mm or larger and a length of 310 mm or shorter of single-channel Teflon/Polyethylene lumens.

STERRAD® NX® Standard Cycle: Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load:

An inside diameter of 1 mm or larger and a length of 150 mm or shorter of single-channel stainless steel lumens.

An inside diameter of 2 mm or larger and a length of 400 mm or shorter of single-channel stainless steel lumens.

STERRAD® NX® Advanced Cycle: Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load:

An inside diameter of 1 mm or larger and a length of 500 mm or shorter of single-channel stainless steel lumens.

OR

One single-channel Flexible Endoscope with or without a silicone mat and no additional load. The flexible endoscope may contain:

A single-channel Teflon®/Polyethylene lumen with an inside diameter of 1 mm or larger and a length of 850 mm or shorter.

STERRAD® 100NX® Standard Cycle: Reusable metal and non-metal medical devices, including up to 10 lumens of the following dimensions per chamber load:

An inside diameter of 0.7 mm or larger and a length of 500 mm or shorter of single-channel stainless steel lumens. (A maximum of two flexible endoscopes, one per tray per sterilization cycle.)

STERRAD® 100NX® Flex Cycle: One or two single-channel Flexible Endoscope with or without a silicone mat and no additional load. The flexible endoscope may contain:

A single-channel Teflon®/Polyethylene lumen with an inside diameter of 1 mm or larger and a length of 850 mm or shorter. (A maximum of two flexible endoscopes, one per tray per sterilization cycle).

STERRAD® 100NX® EXPRESS Cycle: Non-lumened reusable metal and non-metal devices requiring surface sterilization, and sterilization of diffusion-restricted spaces such as the hinged portions of forceps and scissors, and rigid or semi-rigid endoscopes without lumens.

STERRAD® 100NX® DUO Cycle: One or two single-channel Flexible Endoscope with accessory devices that are normally connected to it, with or without a silicone mat. The flexible endoscope may contain:

A single-channel Teflon®/Polyethylene lumen with an inside diameter of 2 mm or larger and a length of 250 mm or shorter.

Accessory devices that are normally connected to a flexible endoscope during use.

Flexible endoscopes without lumens.

KIMGUARD* Smart Fold* Sterilization Wrap Recommendations for Use with the Advanced Sterilization Products STERRAD® Sterilization Systems are provided in Table 2.

TABLE 2: Recommended Loads for KIMGUARD* Smart-Fold* Sterilization Wrap for use with Advanced Sterilization Products (ASP) STERRAD® Sterilization Systems (STERRAD® 100S, NX® [Standard Cycle, Advanced Cycle], and 100NX® [Standard Cycle, Flex Cycle, EXPRESS Cycle, DUO Cycle])

KIMGUARD* Smart Fold* Sterilization Wrap Models — Intended Loads — Maximum Wrapped Package Content Weights Used in Sterility Maintenance Validation Study — Description of Loads Used in Sterility Maintenance Validation Study

KC450 — Moderate to heavy weight package (e.g., general use medical instruments) — 10.7 lbs. — APTIMAX® Instrument Tray (23 in. x 11 in. x 4in.) with Tray Mat; Metal and non-metal instruments.

KC650 — Very heavy weight package (e.g., general use medical instruments) — 10.7 lbs. — APTIMAX® Instrument Tray (23 in. x 11 in. x 4in.) with Tray Mat; Metal and non-metal instruments.

Intended loads include: Medical Instruments with and without lumens that include telescopes, endoscopes, cameras, light cords, and general use medical instruments.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K140963: Kimguard Smart-Fold Sterilization Wrap.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K140963

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026