510(k) Report

510(k) · Cleared

Cleared

Axxent Cervical Applicator

K141343510(k) TraditionalCleared Sep 5, 2014107-day reviewClass II21 CFR 892.5900

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Icad, Inc. · JAD System, Therapeutic, X-Ray

Cleared on Sep 5, 2014 as substantially equivalent to K040704 (Mick Radio-Nuclear Instruments, Inc., 2004) after a 107-day review. Its family tree reaches back 4 generations to 1991.

Decided
Sep 5, 2014
Review
107 days
Class
Class II
Product code
JAD
Recalls
None
Predicates
1

Its longest line of descent runs back 4 generations to K912555 (1991). 6 ancestors in all, the oldest K871217 (1987). No later device cites it as a predicate yet.

  1. Root · 4 generations backK912555 Gammamed 12IT High Dose Rate Remote AfterloaderRts Technology, Inc. · 1991 · Recalled
  2. 3 generations backK983436 Gammamed Plus High Dose Rate Remote Afterloading SystemGammamed USA · 1998 · Recalled
  3. 2 generations backK033371 Varian Intracavitary Brachytherapy ApplicatorsVarian Medical Systems · 2003 · Recalled
  4. 1 generation backK040704 Modification to Henschke HDR Cervix Applicator and Hilaris/Nori HDR…Mick Radio-Nuclear Instruments, Inc. · 2004
  5. This deviceK141343 Axxent Cervical ApplicatorIcad, Inc. · 2014
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K912555Gammamed 12IT High Dose Rate Remote AfterloaderRts Technology, Inc.19912
−3K983436Gammamed Plus High Dose Rate Remote Afterloading SystemGammamed USA199817
−2K033371Varian Intracavitary Brachytherapy ApplicatorsVarian Medical Systems20032
−1K040704Modification to Henschke HDR Cervix Applicator and Hilaris/Nori HDR Cervix…Mick Radio-Nuclear Instruments, Inc.2004—
0 (this device)K141343Axxent Cervical ApplicatorIcad, Inc.2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198710
198910
FIG. 1 — Lineage of K141343, 1991–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAD? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAD device

  1. #1 K241174 INTRABEAM (700); And accessories (INTRABEAM…score 11.7 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K040704Modification to Henschke HDR Cervix Applicator and Hilaris/Nori HDR Cervix Applicator
Summary page 5
…Manufacturer Reference # Date Mick Radio-Nuclear Instruments, Inc. Modified Mick Radio-Nuclear K040704 06/18/2004 Henschke HDR Cervix Instruments, Inc. Applicator Device Description The Axxent Cervical Appl…
Mick Radio-Nuclear Instruments, Inc.2004—

Also cleared against K040704: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 23 years and passes through 5 companies; a new generation was cleared every 7.2 years on median.
  • 3 of the 6 ancestors have a recall of any class.
  • 2 other devices also cite its predicate K040704.

How does its category compare?

All JAD devices

JAD System, Therapeutic, X-Ray: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · JAD

2in 2025

2007: 12008: 22009: 32011: 12012: 22013: 52014: 12015: 12016: 22017: 12018: 12019: 22022: 12023: 12025: 2
2007high 5 (2013) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAD

114in 2013

2009: 106 days median2013: 114 days median
2007high 114 (2013) · low 106 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAD

2in 2021

2009: 22011: 12012: 32017: 12021: 22026 (partial year): 1
2007high 3 (2012) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAD

1in 2025

2007: 22008: 22009: 52011: 32012: 42013: 122014: 42015: 52016: 52017: 32018: 12019: 92022: 52023: 22024: 52025: 1
2007high 12 (2013) · low 0 (2010)2025

MAUDE reports naming product code JAD, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—02
20082—02
2009310625
20100—00
20111—13
20122—34
20135114012
20141—04
20151—05
20162—05
20171—13
20181—01
20192—09
20200—00
20210—20
20221—05
20231—02
20240—05
20252—01
20260—1—

What is it indicated for?

The Axxent Cervical Applicator is indicated for use with the Axxent Electronic Brachytherapy System to deliver high dose rate brachytherapy for intracavitary treatment of cancer of the uterus, cervix, endometrium and vagina.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K141343: Axxent Cervical Applicator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K141343

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026